← Product Code [MQV](/productcode/MQV) · K241555

# Moldable Bone Void Filler and Moldable Bone Void Filler + CCC (K241555)

_Rti Surgical, Inc. · MQV · Feb 21, 2025 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K241555

## Device Facts

- **Applicant:** Rti Surgical, Inc.
- **Product Code:** [MQV](/productcode/MQV.md)
- **Decision Date:** Feb 21, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3045
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Moldable Bone Void Filler and Moldable Bone Void Filler + CCC is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities and pelvis). These osseous defects are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. Moldable Bone Void Filler and Moldable Bone Void Filler + CCC can be used with autogenous bone marrow. Moldable Bone Void Filler and Moldable Bone Void Filler + CCC resorbs and are replaced with bone during the healing process. Moldable Bone Void Filler and Moldable Bone Void Filler + CCC are for single patient use only.

## Device Story

Bone void filler composed of demineralized bone matrix (DBM), synthetic macromer hydrogel, and cortical cancellous bone chips (CCC) for the + CCC version. Provided with accessory kit containing hydrating solution and spatula. Clinician hydrates components to create a moldable putty; putty is manually placed into osseous defects in extremities or pelvis. Device resorbs over time and is replaced by new bone. Used in surgical settings to manage non-structural bone voids. Benefits patient by facilitating bone healing in traumatic or surgically created defects.

## Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical testing including chemical/physical properties, biocompatibility, sterility, shelf life, and an animal study. Animal study at 12 weeks showed new bone formation averages of 16.21% (Optecure), 13.8% (Optecure + CCC), 15.75% (predicate), and 12.08% (negative control).

## Technological Characteristics

Composition: demineralized bone matrix (DBM), synthetic macromer hydrogel, and cortical cancellous bone chips (CCC). Form factor: moldable putty upon hydration. Single patient use. No electronic or software components.

## Regulatory Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

## Predicate Devices

- Exactech Optecure and Optecure + CCC ([K121989](/device/K121989.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the date February 21, 2025. The text is written in a clear, sans-serif font. The date is written out in full, with the month, day, and year clearly visible. The image is simple and straightforward, with the focus on the date itself.

Image /page/0/Picture/10 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

RTI Surgical, Inc. Ellen Rounds Sr. Regulatory Affairs Specialist 11621 Research Circle Alachua, Florida 32615

Re: K241555

Trade/Device Name: Moldable Bone Void Filler and Moldable Bone Void Filler + CCC Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MOV, MBP Dated: January 24, 2025 Received: January 24, 2025

Dear Ellen Rounds:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

JESSE MUIR -> Digitally signed by JESSE MUR ->
Date: 2025.02.21 11:57:51 -05'00'

Jesse Muir, PhD Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known) K241555

Device Name

Moldable Bone Void Filler and Moldable Bone Void Filler + CCC

Indications for Use (Describe)

Moldable Bone Void Filler and Moldable Bone Void Filler + CCC is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities and pelvis). These osseous defects are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. Moldable Bone Void Filler and Moldable Bone Void Filler + CCC can be used with autogenous bone marrow. Moldable Bone Void Filler and Moldable Bone Void Filler + CCC resorbs and are replaced with bone during the healing process.

Moldable Bone Void Filler and Moldable Bone Void Filler + CCC are for single patient use only.

| Type of Use (Select one or both, as applicable)                                   |                                                                                |
|-----------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span>     X Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>     Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for RTI Surgical. The logo consists of the letters "rti" in a blue rounded square, followed by the word "surgical" in gray. The letters "rti" are white, with a small green accent on the "i".

## 510(k) Summary

| Date Prepared:                            | 02/20/2025                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:                                | RTI Surgical, Inc.<br>11621 Research Circle<br>Alachua, FL 32615 USA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Contact Information:                      | Ellen Rounds<br>Sr. Regulatory Affairs Specialist<br>Email: erounds@rtix.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Name of Device:                           | Moldable Bone Void Filler and Moldable Bone Void Filler + CCC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Common Name:                              | Optecure and Optecure + CCC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Classification Name:                      | Resorbable calcium salt bone void filler device<br>Filler, Bone Void, Calcium Compound                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Regulation Number:                        | 21 CFR 888.3045                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Regulatory Class:                         | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Product Code:                             | MQV, MBP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Panel:                                    | Orthopedic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Legally Marketed Predicate Device(s):     | Exactech Optecure and Optecure + CCC, K121989                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Device Description:                       | Moldable Bone Void Filler and Moldable Bone Void Filler + CCC are<br>bone void fillers composed of processed demineralized bone matrix<br>(DBM), a synthetic macromer hydrogel, and cortical cancellous bone<br>chips (CCC) for Moldable Bone Void Filler + CCC only. Both are<br>provided with an accessory kit containing pre-measured hydrating<br>solution and a spatula to mix the components.<br><br>After the implant is hydrated, the resultant putty can then be handled<br>and placed in the appropriate bone voids or gaps. Moldable Bone Void<br>Filler and Moldable Bone Void Filler + CCC gradually resorb and are<br>replaced with new bone during the healing process. At the 12 week<br>timepoint, animal study data demonstrated new bone formation<br>averages of 16.21% in the Optecure group, 13.8% in the Optecure +<br>CCC group, 15.75% in the Exactech Optecure + CCC predicate group,<br>and 12.08 % in the empty defect negative control group. |
| Indications for Use:                      | Moldable Bone Void Filler and Moldable Bone Void Filler + CCC is<br>an implant intended to fill bony voids or gaps of the skeletal system<br>(i.e., extremities and pelvis). These osseous defects are surgically<br>created or the result of traumatic injury to the bone and are not intrinsic<br>to the stability of the bony structure. Moldable Bone Void Filler and<br>Moldable Bone Void Filler + CCC can be used with autogenous bone<br>marrow. Moldable Bone Void Filler and Moldable Bone Void Filler +<br>CCC resorbs and are replaced with bone during the healing process.<br><br>Moldable Bone Void Filler and Moldable Bone Void Filler + CCC are<br>for single patient use only                                                                                                                                                                                                                                                                           |
| Substantial<br>Equivalence<br>Comparison: | The subject device are demonstrated to be substantially equivalent to<br>the predicate device cited above with respect to indications for use,<br>materials, biocompatibility, storage, and performance.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Non-Clinical<br>Performance Data:         | In accordance with the “Guidance for Industry and FDA Staff - Class<br>II Special Controls Guidance Document: Resorbable Calcium Salt<br>Bone Void Filler Device”, the following testing has been performed to<br>support substantial equivalence and to demonstrate safety and<br>effectiveness  Chemical and physical properties Biocompatibility Sterility Shelf Life Animal Study                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                           | At the 12 week timepoint, animal study data demonstrated new bone<br>formation averages of 16.21% in the Optecure group, 13.8% in the<br>Optecure + CCC group, 15.75% in the Exactech Optecure + CCC<br>predicate group, and 12.08 % in the empty defect negative control<br>group.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Clinical Performance:                     | There was no clinical testing required to support the medical device as<br>the indications for use is equivalent to the predicate device. The non-<br>clinical testing detailed in this submission supports the substantial<br>equivalence of the device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Substantial<br>Equivalence:               | The determination of a device are substantially equivalent to a<br>predicate device when the device has the same intended use and the<br>same technological characteristics as the predicate devices.<br>Additionally, if the device has the same intended use but different<br>technological characteristics and can be demonstrated that the new<br>device does not raise any safety and effectiveness concerns as<br>compared to the predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                           | In the provided 510(k) submission the difference between the predicate<br>device and Moldable Bone Void Filler and Moldable Bone Void Filler<br>+ CCC do not raise safety and/or effectiveness concerns. Therefore,<br>Moldable Bone Void Filler and Moldable Bone Void Filler + CCC is<br>determined to be substantially equivalent to the predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Conclusion:                               | The submitted 510(k) demonstrates Moldable Bone Void Filler and<br>Moldable Bone Void Filler + CCC are substantially equivalent to the<br>predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for RTI Surgical. The logo consists of the letters "rti" in a blue rounded square, followed by the word "surgical" in gray. The "i" in "rti" has a small green leaf on top.

---

**Source:** [https://fda-staging.innolitics.com/device/K241555](https://fda-staging.innolitics.com/device/K241555)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com