The trophon Wireless Ultrasound Probe Holder is an accessory to be added to the trophon EPR and trophon2 to allow wireless ultrasound probes to be disinfected. The trophon EPR is designed to provide High-Level Disinfection of ultrasound transducers. The system uses the trophon Disinfectant which is intended to be used exclusively with the trophon EPR device. The trophon Disinfectant is intended for use as a High-Level Disinfectant to be used exclusively with the trophon EPR for the High-Level Disinfection of ultrasound transducers. The trophon EPR is suitable for use in general hospital and health care facilities by trained personnel. The trophon EPR system consists of a multiple use instrument combined with a single use disinfectant, delivered from a multi-dose cartridge. The trophon Disinfectant should be used with the following contact conditions: Minimum Operational Cycle Time: 7 minutes Minimum Concentration: 31.5% Minimum Disinfectant Dose: 1.0 g Minimum Chamber Temperature: 56℃. The trophon2 is designed to provide High-Level Disinfection (HLD) of validated ultrasound transducers. High-Level Disinfection is achieved by surface exposure to a controlled dose of hydrogen peroxide mist delivered to a disinfection chamber contain the ultrasound probe. The trophon2 system consists of a multiple use instrument combined with a single use disinfectant "trophon Sonex-HL", delivered from a multi-dose cartridge. The trophon2 is suitable for use in general hospital and health care facilities by trained personnel. The trophon Sonex-HL should be used with the following contact conditions: Minimum Operational Cycle Time: 7 minutes Minimum Concentration: 31.5% Minimum Disinfectant Dose: 1.0 g Minimum Chamber Temperature: 56°C
Device Story
Accessory for trophon EPR and trophon2 disinfection systems; enables high-level disinfection (HLD) of wireless ultrasound probes. Device holds wireless probe in place within existing disinfection chamber using cable clamp mechanism. System operates via automated cycle: nebulizer generates hydrogen peroxide mist; mist contacts probe surfaces for specified time, temperature, and dosage; automated aeration removes disinfectant. Used in hospitals/clinics by trained personnel. Output is a disinfected probe ready for immediate clinical use. Benefits include expanded compatibility for wireless transducers while maintaining established HLD efficacy and safety standards.
Clinical Evidence
No clinical data. Bench testing only, including HLD efficacy (Mycobacterium terrae), residual H2O2 assessment, material compatibility (5000 cycles), temperature monitoring, leak testing, cleaning validation, and human factors validation.
Technological Characteristics
Accessory for trophon EPR/trophon2. Operates via hydrogen peroxide mist (35% concentration) generated by nebulizer. Includes integrated catalytic destruct system. Software-controlled automated cycle. Complies with IEC 61010-1, IEC 61010-2-040, IEC 61326, IEC 62304, ISO 62366, ISO 10993-1, and ISO 14971.
Indications for Use
Indicated for high-level disinfection of ultrasound transducers, including wired probes and validated wireless ultrasound probes (e.g., GE Vscan Air CL), in hospital and healthcare facilities by trained personnel.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 27, 2024
Nanosonics Limited Nancy Kaiser Regulatory Affairs Manager 7-11 Talavera Road Macquarie Park, NSW 2113 Australia
Re: K241536
Trade/Device Name: trophon® Wireless Ultrasound Probe Holder Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic Ultrasonic Transducer Regulatory Class: Class II Product Code: OUJ Dated: May 30, 2024 Received: August 29, 2024
Dear Nancy Kaiser:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Image /page/2/Picture/3 description: The image shows the name "Katharine Segars" followed by "-S". The text is in a large, clear font, making it easily readable. The background is plain, which helps the text stand out.
Katharine Segars, Ph.D. Assistant Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
## 510(k) Number (if known) K241536
### Device Name
trophon® Wireless Ultrasound Probe Holder
### Indications for Use (Describe)
The trophon Wireless Ultrasound Probe Holder is an accessory to be added to the trophon EPR and trophon2 to allow wireless ultrasound probes to be disinfected.
trophon EPR:
The trophon EPR is designed to provide High-Level Disinfection of ultrasound transducers. The system uses the trophon Disinfectant which is intended to be used exclusively with the trophon EPR device.
The trophon Disinfectant is intended for use as a High-Level Disinfectant to be used exclusively with the trophon EPR for the High-Level Disinfection of ultrasound transducers.
The trophon EPR is suitable for use in general hospital and health care facilities by trained personnel.
The trophon EPR system consists of a multiple use instrument combined with a single use disinfectant, delivered from a multi-dose cartridge.
The trophon Disinfectant should be used with the following contact conditions:
Minimum Operational Cycle Time: 7 minutes Minimum Concentration: 31.5% Minimum Disinfectant Dose: 1.0 g Minimum Chamber Temperature: 56℃
trophon2:
The trophon2 is designed to provide High-Level Disinfection (HLD) of validated ultrasound transducers. High-Level Disinfection is achieved by surface exposure to a controlled dose of hydrogen peroxide mist delivered to a disinfection chamber contain the ultrasound probe.
The trophon2 system consists of a multiple use instrument combined with a single use disinfectant "trophon Sonex-HL", delivered from a multi-dose cartridge.
The trophon2 is suitable for use in general hospital and health care facilities by trained personnel.
The trophon Sonex-HL should be used with the following contact conditions: Minimum Operational Cycle Time: 7 minutes Minimum Concentration: 31.5% Minimum Disinfectant Dose: 1.0 g Minimum Chamber Temperature: 56°C Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Nanosonics. The word "nanosonics" is written in a sans-serif font, with "nano" in blue and "sonics" in gray. A blue arc is above the word "nano". Below the word "nanosonics" is the phrase "Infection Prevention. For Life." in a smaller font.
# 510(k) SUMMARY - trophon® Wireless Ultrasound Probe Holder
### I. DATE PREPARED September 24, 2024
II. 510(k) NUMBER K241536
#### III. 510(k) SUBMITTER
Nanosonics Limited 7-11 Talavera Road Macquarie Park NSW 2113 Australia
Contact Person:
Nancy Kaiser, Regulatory Affairs Manager Address: 7-11 Talavera Road Macquarie Park, NSW 2113 Australia Email: n.kaiser@nanosonics.com Telephone: (317) 854-7625
#### IV. DEVICE
| Trade Name of Device: | trophon® Wireless Ultrasound Probe Holder. |
|-----------------------|----------------------------------------------------------------------------------|
| Common or Usual Name: | High Level Disinfection Reprocessing Instrument for Ultrasonic Transducers, Mist |
| Classification: | II |
| Regulation Number: | 21 CFR 892.1570 |
| Product Code: | OUJ |
#### V. PREDICATE DEVICE
| Predicate Device | |
|------------------|-----------------|
| Trade Name: | trophon2 |
| 510(k) Number: | K173865 |
| Company Name: | Nanosonics Ltd. |
## Reference Device
| Trade Name: | trophon EPR |
|----------------|-----------------|
| 510(k) Number: | K103059 |
| Company Name: | Nanosonics Ltd. |
#### VI. DEVICE DESCRIPTION
The trophon Wireless Ultrasound Probe Holder is an accessory to be added to the trophon EPR and trophon2 to allow wireless ultrasound probes to be disinfected.
The Nanosonics trophon EPR and trophon2 are software-controlled devices which provide High- Level Disinfection of ultrasound transducers. The device consists of a sealed disinfection chamber and operates in conjunction with a multi-dose cartridge of concentrated hydrogen peroxide disinfectant, supplied as an accessory to the device. Precleaned and dried ultrasound transducers are placed within the trophon EPR/trophon2
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Nanosonics. The word "nanosonics" is written in a combination of blue and gray letters. The word "nano" is in blue, and the word "sonics" is in gray. Below the word "nanosonics" is the phrase "Infection Prevention. For Life." in a smaller font.
chamber and disinfected by means of an automated disinfection and aeration cycle. The disinfected ultrasound transducer is removed from the chamber and is ready for immediate use.
#### VII. INDICATIONS FOR USE
The trophon Wireless Ultrasound Probe Holder is an accessory to be added to the trophon EPR and trophon2 to allow wireless ultrasound probes to be disinfected.
trophon EPR:
The trophon EPR is designed to provide High-Level Disinfection of ultrasound transducers. The system uses the trophon Disinfectant which is intended to be used exclusively with the trophon EPR device.
The trophon Disinfectant is intended for use as a High-Level Disinfectant to be used exclusively with the trophon EPR for the High-Level Disinfection of ultrasound transducers.
The trophon EPR is suitable for use in general hospital and health care facilities by trained personnel.
The trophon EPR system consists of a multiple use instrument combined with a single use disinfectant, delivered from a multi-dose cartridge.
The trophon Disinfectant should be used with the following contact conditions: Minimum Operational Cycle Time: 7 minutes Minimum Concentration: 31.5% Minimum Disinfectant Dose: 1.0 g Minimum Chamber Temperature: 56°C
## trophon2:
The trophon2 is designed to provide High-Level Disinfection (HLD) of validated ultrasound transducers. High-Level Disinfection is achieved by surface exposure to a controlled dose of hydrogen peroxide mist delivered to a disinfection chamber contain the ultrasound probe.
The trophon2 system consists of a multiple use instrument combined with a single use disinfectant "trophon Sonex-HL", delivered from a multi-dose cartridge.
The trophon2 is suitable for use in general hospital and health care facilities by trained personnel.
The trophon Sonex-HL should be used with the following contact conditions: Minimum Operational Cycle Time: 7 minutes Minimum Concentration: 31.5% Minimum Disinfectant Dose: 1.0 g Minimum Chamber Temperature: 56°C
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Nanosonics. The word "nanosonics" is written in a sans-serif font, with "nano" in blue and "sonics" in gray. Below the word "nanosonics" is the phrase "Infection Prevention. For Life." in a smaller, sans-serif font. A blue arc is above the word "nanosonics".
### VIII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The trophon2 and the trophon EPR with the proposed Ultrasound Probe Holder are as safe and as effective in terms of technological characteristics and principles of operation to the predicate device, trophon2 and reference device, trophon EPR without the Ultrasound Probe Holder.
Like the predicate and reference devices, the trophon2 and trophon EPR still achieve HLD by having the probe suspended in their chamber, the Probe Holder is held in place by the cable clamp during high level disinfection to achieve this positioning for a wireless ultrasound probe. During each disinfection cycle, a controlled quantity of hydrogen peroxide mist is generated by a nebulizer and is delivered to the trophon EPR/trophon2 chamber. In the chamber, the mist directly contacts the wired or wireless ultrasound probe covering its surface. Contact occurs for a specified time, temperature, and liquid/mist dosage which determines the germicidal efficacy. The disinfected wired or wireless ultrasound transducer is removed from the chamber and is ready for immediate use. Design verification testing demonstrates that the proposed Wireless Ultrasound Probe Holder meets the product specifications and validation testing confirmed, intended use and user needs continue to be met.
| Feature | Table 1. A Comparison between the Subject and Predicate Device<br>Subject Device:<br>trophon2 (K241536) | Predicate Device:<br>trophon2 (K173865) | Comparison |
|-------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Nanosonics Limited | Nanosonics Limited | Same |
| Regulation Number | 21 CFR 892.1570 | 21 CFR 892.1570 | Same |
| Product Code | OUJ | OUJ | Same |
| Indication for<br>Use/Intended Use | The trophon Wireless Ultrasound<br>Probe Holder is an accessory to be<br>added to the trophon EPR and<br>trophon2 to allow wireless ultrasound<br>probes to be disinfected.<br>The trophon2 is designed to provide<br>High-level Disinfection (HLD) of<br>validated ultrasound probes, High-<br>Level Disinfection is achieved by<br>surface exposure to a controlled dose<br>of hydrogen peroxide mist delivered to<br>a disinfection chamber containing the<br>ultrasound probe.<br>The trophon2 is suitable for use in<br>general hospital and health care<br>facilities by trained personnel.<br>The trophon Sonex-HL should be<br>used with the following contact<br>conditions:<br>Minimum Operational Cycle Time: 7<br>minutes<br>Minimum Concentration: 31.5%<br>Minimum Disinfectant Dose: 1.0 g<br>Minimum Chamber Temperature: | The trophon2 is designed to provide<br>High-Level Disinfection (HLD) of<br>validated ultrasound transducers. High-<br>Level Disinfection is achieved by<br>surface exposure to a controlled dose of<br>hydrogen peroxide mist delivered to a<br>disinfection chamber contain the<br>ultrasound probe.<br>The trophon2 system consists of a<br>multiple use instrument combined with a<br>single use disinfectant "trophon Sonex-<br>HL", delivered from a multi-dose<br>cartridge.<br>The trophon2 is suitable for use in<br>general hospital and health care facilities<br>by trained personnel.<br>The trophon Sonex-HL should be used<br>with the following contact conditions:<br>Minimum Operational Cycle Time: 7<br>minutes<br>Minimum Concentration: 31.5%<br>Minimum Disinfectant Dose: 1.0 g<br>Minimum Chamber Temperature: 56°C | Same |
| Operating Principle | Software controlled systems<br>that deliver measured doses of<br>hydrogen peroxide disinfectant<br>to achieve High -Level<br>Disinfection (HLD) | Software controlled systems that<br>deliver measured doses of<br>hydrogen peroxide disinfectant to<br>achieve High -Level Disinfection<br>(HLD) | Same |
| Disinfectant | trophon Sonex-HL (35%<br>hydrogen peroxide in<br>cartridge) | trophon Sonex-HL (35%<br>hydrogen peroxide in cartridge) | Same |
| Disinfectant Delivery | Liquid Aerosol Mist | Liquid Aerosol Mist | Same |
| Disinfectant Removal<br>Process | Automate Aeration | Automate Aeration | Same |
| Process Monitoring | Automated process monitoring<br>in the device | Automated process monitoring in<br>the device | Same |
| Chemical Indicator<br>Required? | Yes, trophon chemical<br>indicator (cleared in K103126) | Yes, trophon chemical indicator<br>(cleared in K103126) | Same |
| Microbiology/Efficacy<br>AOAC Performance<br>Standards | Meets AOAC Methods | Meets AOAC Methods | Same |
| Device Performance<br>Standards | IEC 601010-1<br>IEC 61010-2-040<br>IEC 61326<br>IEC 62304<br>ISO 62366 -1 and -2<br>ISO 10993-1<br>ISO 14971 | IEC 601010-1<br>IEC 61010-2-040<br>IEC 61326<br>IEC 62304<br>ISO 62366 -1 and -2<br>ISO 10993-1<br>ISO 14971 | Same |
| Residue Testing | Effectively removes residues<br>from disinfected transducers | Effectively removes residues<br>from disinfected transducers | Same |
| Chamber Design | Regular shaped chamber | Irregular shaped chamber | Same |
| Chamber Temperature | Set point 65°C | Set point 75°C | Same |
| Door Lock | Motor and hook assembly | Motor and hook assembly | Same |
| Probe Clamp | Spring loaded cleats | Spring loaded cleats | Same |
| All in One Catalytic<br>Destruct | An integrated catalytic<br>destruct system | An integrated catalytic destruct<br>system | Same |
| Touch Screen | Color touch screen panel | Color touch screen panel | Same |
| Software/Firmware | Multiple software/firmware<br>components which are<br>deployed on two PCBAs | Multiple software/firmware<br>components which are deployed<br>on two PCBAs | Same |
| Traceability | An integrated RFID module<br>allowing automated<br>traceability features in the<br>device.<br>Patient information is not<br>received or recorded by the<br>device; therefore, it cannot be<br>accessed via RFID. | An integrated RFID module<br>allowing automated traceability<br>features in the device.<br>Patient information is not<br>received or recorded by the<br>device; therefore, it cannot be<br>accessed via RFID. | Same |
| Communication Ports | 3 USB ports to connect to<br>external device (i.e. printer)<br>1 Ethernet port<br>The trophon2 will be able to<br>connect externally to a | 3 USB ports to connect to<br>external device (i.e. printer)<br>1 Ethernet port<br>The trophon2 will be able to<br>connect externally to a network | Same |
| | network via the Ethernet port | via the Ethernet port | |
| Key Accessories | trophon AcuTrace Operator<br>Card<br>trophon AcuTrace Medical<br>Instrument Tag<br>trophon AcuTrace PLUS<br>activation Card | trophon AcuTrace Operator Card<br>trophon AcuTrace Medical<br>Instrument Tag<br>trophon AcuTrace PLUS<br>activation Card<br>trophon Sonex-HL | Same<br>An optional<br>Wireless<br>Ultrasound<br>Probe Holder |
| Image: Infection Prevention. For Life. | trophon Sonex-HL | | (provided<br>separately) to<br>provide high<br>level<br>disinfection for<br>wireless GE<br>ultrasound<br>probe (model:<br>Vscan Air CL) |
| Intended to provide high<br>level disinfection to<br>Wired Probes | Yes | Yes | Same<br>Wired Probes<br>and Validated<br>Wireless GE<br>Probe (model:<br>Vscan Air CL) |
# Table 1. A Comparison between the Subject and Predicate Device
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Nanosonics. The word "nanosonics" is written in a combination of blue and gray colors. A blue arc is above the word. Below the word "nanosonics" is the phrase "Infection Prevention. For Life."
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for "nanosonics". The word "nano" is in blue, and the word "sonics" is in gray. There is a blue arc above the word "nano". Below the word "nanosonics" is the phrase "Infection Prevention. For Life."
# Table 2. A Comparison between the Subject and Reference Device
| Feature | Subject Device:<br>trophon EPR (K241536) | Reference Device:<br>trophon EPR (K103059) | Comparison |
|-------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Nanosonics Limited | Nanosonics Limited | Nanosonics Limited | Same |
| Regulation Number | 21 CFR 892.1570 | 21 CFR 892.1570 | Same |
| Product Code | OUJ | OUJ | Same…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.