Horizon® 3.0 TMS Therapy System (Horizon 3.0 Inspire); Horizon® 3.0 TMS Therapy System (Horizon 3.0 with StimGuide Pro); Horizon® 3.0 TMS Therapy System (Horizon 3.0)
K241518 · The Magstim Company Limited · OBP · Aug 30, 2024 · Neurology
Device Facts
Record ID
K241518
Device Name
Horizon® 3.0 TMS Therapy System (Horizon 3.0 Inspire); Horizon® 3.0 TMS Therapy System (Horizon 3.0 with StimGuide Pro); Horizon® 3.0 TMS Therapy System (Horizon 3.0)
Applicant
The Magstim Company Limited
Product Code
OBP · Neurology
Decision Date
Aug 30, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5805
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Horizon 3.0 TMS Therapy System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode, as well as an adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder (OCD).
Device Story
Computerized, electromechanical system delivering non-invasive, repetitive transcranial magnetic stimulation (rTMS) to cerebral cortex; uses brief, rapidly alternating magnetic pulses to induce electrical currents. System components: stimulating unit, power supply, user interface, applicating coils, cart, and holding arm. Used in inpatient/outpatient settings (clinics, hospitals) by licensed physicians. Physician determines motor threshold (MT) via visual monitoring or EMG; treatment targets specific cortical regions (e.g., 4cm from MT hotspot for OCD, 5.5cm for MDD). Output affects neuronal activity to treat MDD and OCD. Horizon 3.0 Inspire configuration reorganizes existing components (PC moved to touchscreen, EMG amplifier to MEP pod) to reduce footprint and cost while maintaining identical principles of operation, safety, and effectiveness as predicate configurations.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing: electromagnetic compatibility (IEC 60601-1-2), electrical/mechanical/thermal safety (IEC 60601-1), software verification/validation (IEC 62304), human factors/usability (IEC 62366-1), and magnetic/electric field output measurements in saline-filled head phantom models. Acoustic testing confirmed noise levels within safe limits.
Indicated for adult patients with Major Depressive Disorder (MDD) who failed prior antidepressant medication in current episode, and as an adjunct for adult patients with Obsessive-Compulsive Disorder (OCD).
Regulatory Classification
Identification
A repetitive transcranial magnetic stimulation system is an external device that delivers transcranial repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex to treat the symptoms of major depressive disorder without inducing seizure in patients who have failed at least one antidepressant medication and are currently not on any antidepressant therapy.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's “Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation System.” See § 882.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 30, 2024
The Magstim Company Limited Daniel Gregory Principal Systems Engineer Spring Gardens Whitland Carmarthenshire, SA34 0HR United Kingdom
Re: K241518
Trade/Device Name: Horizon® 3.0 TMS Therapy System (Horizon 3.0 Inspire); Horizon® 3.0 TMS Therapy System (Horizon 3.0 with StimGuide Pro); Horizon® 3.0 TMS Therapy System (Horizon 3.0) Regulation Number: 21 CFR 882.5805 Regulation Name: Repetitive Transcranial Magnetic Stimulation System Regulatory Class: Class II Product Code: OBP; QCI Dated: May 29, 2024 Received: May 29, 2024
Dear Daniel Gregory:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Vivek J. Pinto -S
Vivek Pinto, PhD Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K241518
Device Name
Horizon® 3.0 TMS Therapy System (Horizon 3.0 Inspire);
Indications for Use (Describe)
Horizon 3.0 TMS Therapy System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode, as well as an adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder (OCD).
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
Prepared on: 2024-05-29
| Contact Details | 21 CFR 807.92(a)(1) | | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|---------------------|
| Applicant Name | The Magstim Company Limited | | | |
| Applicant Address | Spring Gardens Whitland Carmarthenshire SA34 0HR United Kingdom | | | |
| Applicant Contact Telephone | +441994240798 | | | |
| Applicant Contact | Mr. Daniel Gregory | | | |
| Applicant Contact Email | daniel.gregory@magstim.com | | | |
| Device Name | 21 CFR 807.92(a)(2) | | | |
| Device Trade Name | Horizon® 3.0 TMS Therapy System (Horizon 3.0 Inspire);<br>Horizon® 3.0 TMS Therapy System (Horizon 3.0 with StimGuide Pro);<br>Horizon® 3.0 TMS Therapy System (Horizon 3.0) | | | |
| Common Name | Repetitive transcranial magnetic stimulation system | | | |
| Classification Name | Transcranial Magnetic Stimulator | | | |
| Regulation Number | 882.5805 | | | |
| Product Code(s) | OBP, QCI - 21 C.F.R. § 882.5802 | | | |
| Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) | | | |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | | | Product Code |
| K232235 | Horizon 3.0 TMS Therapy System | | | OBP |
| K182853 | Horizon TMS Therapy System | | | OBP |
| Device Description Summary | | | | 21 CFR 807.92(a)(4) |
| The Horizon® 3.0 TMS Therapy System is a computerized, electromechanical medical device that<br>produces and delivers non-invasive, magnetic stimulation using brief duration rapidly alternating, or<br>pulsed, magnetic fields to induce electrical currents directed at spatially discrete regions of the<br>cerebral cortex. This method of cortical stimulation by application of brief magnetic pulses to the<br>head is known as Transcranial Magnetic Stimulation. ("TMS"). | | | | |
| The Horizon® 3.0 TMS Therapy System is a non-invasive tool for the stimulation of cortical neurons<br>for the treatment of Major Depressive Disorder in adult patients who have failed to achieve<br>satisfactory improvement from prior antidepressant medication in the current episode, as well as an<br>adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder (OCD). | | | | |
| The Horizon® 3.0 TMS Therapy System is used for patient treatment by prescription only under the<br>supervision of a licensed physician. It can be used in both inpatient and outpatient settings, including<br>physicians' offices, clinics, and hospitals. | | | | |
| The Horizon® 3.0 TMS Therapy System configurations are an integrated system consisting of a combination of hardware, software, and | | | | |
The Horizon 3.0 TMS Therapy System Conngurations are an integrated system configurations:
{5}------------------------------------------------
- Horizon 3.0 Inspire (Subject of this submission).
- Horizon 3.0 (Previously cleared under K232235, K223154, K222171 and K211389).
- Horizon 3.0 with StimGuide Pro (Cleared under K232235).
The three tiers of system offer equivalent safety and effectiveness with the main purpose allowing for physician offices, clinics and hospitals to choose a configuration that suits the organizational needs and provide different entry levels to promote the accessibility of TMS Therapy Treatments to health delivery organizations.
All three configurations, including the subject Horizon 3.0 Inspire configuration, have identical intended use/indications for use, common specifications, equivalent performance and equivalent composition. All three devices share equivalent technological characteristics and principles of operation.
All configurations are composed from the following main components:
### - Stimulating Unit & Power Supply
- User Interface
- Applicating Coil for Motor Threshold
- Applicating Coil for Treatment Delivery
- System and Applicating Coil Cart and Holding Arm
The Horizon 3.0 with StimGuide Pro specifically includes a stereotactic infrared tracking system for aiding coil positioning.
## Intended Use/Indications for Use
Horizon 3.0 TMS Therapy System is indicated for the treatment of Major Depressive Disorder in adult patieve to achieve satisfactory improvement from prior antidepressant medication in the current episode, as well as an adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder (OCD).
## Indications for Use Comparison
The indications for use for the subject device and the primary predicate device (K232235) are identical.
The reference predicate, K182853, where some of the subject device components have been re-used from was previously cleared for the treatment of Major Depressive Disorder in adult patieve satisfactory improvement from prior antidepressant medication in the current episode. The performance data within this submission, such as the applicating coil comparison to the primary predicate, demonstrate why these components do not raise any differing questions of safety or effectiveness when also used as an adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder (OCD).
## Technological Comparison
## Introduction:
The Horizon® 3.0 TMS Therapy System is offered in three system configurations, the third of which is the subject of this submission:
Horizon 3.0, Horizon 3.0 with StimGuide Pro and Horizon 3.0 Inspire.
All system configurations, including the subject Horizon 3.0 Inspire configuration, feature the same core components and operating principles that drive the device. All three configurations in subject Horizon 3.0 Inspire configuration, have identical intended use/indications for use, common specifications, equivalent composition. All three devices share equivalent technological characteristics and principles of operation, substantiated below.
### Principles of Operation:
Horizon 3.0 Inspire is identical to the primary principles of operation. Horizon 3.0 Inspire is capable of delivering and provides the same recommended treatment protocols (OCD, TTMS and iTBS) as the primary predicate device.
The method for coll positioning is also identical to the primary predicate, with the Horizon 3.0 "Inspire" following the same principle of the statistical distance from the Motor Threshold (MT) Hotspot to the chosen treatment (4cm from MT Hotspot, Dorsomedial Prefrontal Cortex, for OCD and 5.5cm from MT Hotspot, Left Dorsolateral Prefrontal Cortex, for MDD).
The principles of operation with respect to MT Determination remain identical to the primary predicate device - either by using visual qualitative monitoring for APB response or quantitative data provided by EMG.
In summary, the principles of operation of the subject Horizon 3.0 Inspire configuration remain identical to the primary predicate device configurations - Horizon 3.0 and Horizon 3.0 with StimGuide Pro.
## 21 CFR 807.92(a)(5)
21 CFR 807.92(a)(6)
{6}------------------------------------------------
#### Technological Characteristics:
The differences between Horizon 3.0 Inspire and the previously cleared device configurations from the primary predicate device are listed below. The intent of which is to reduce the cost and footprint for the end user whilst also providing an equivalent level of safety and effectiveness as the predicate device configurations. The three tiers of system offectiveness with the main purpose allowing for physician offices, clinics and hospitals to choose a configuration that suits the organizational needs and provide different entry levels to promote the accessibility of Therapy Treatments to health delivery organizations.
- The Horizon 3.0 Inspire Mainframe features a different "interface" panel on the mainframe. This is to support the connection to the Horizon 3.0 Inspire Touchscreen. The mainframe supplies power and communications via this interface. The fundamental principles of the Horizon 3.0 Inspire Main identical to the predicate Horizon 3.0 Mainframe. The maximum output voltage, frequency, biphasic waveform, charging circuity and protocol performance remain identical across all configurations. The only change was interface related to mount a Ul touchscreen on the Mainframe.
- The Horizon 3.0 Interface Unit from the primary predicate device configurations is not required for the Horizon 3.0 Inspire configuration. It's functions have been into other components. The PC has been moved from the Horizon 3.0 Interface Unit into the Horizon 3.0 Inspire Touchscreen, the EMG Amplifier has been moved into the Horizon 3.0 MEP Pod and some device control mechanisms previously performed via Horizon 3.0 Interface Unit now reside in the Horizon 3.0 Inspire Mainframe. In summary, the functions provided remain identical to the primary predicate and the technological changes were purely a reorganization of these functions to reduce the footprint of the system. Components such as the EMG Amplifier within the new Horizon 3.0 MEP Pod is the same module that is used on the primary predicate configurations.
- The Horizon 3.0 Inspire Touchscreen as described above now contains the PC that previously resided in the Horizon 3.0 Interface Unit. The technical specification is equivalent to the Horizon 3.0 Interface Unit. Further, the Operating System and U Software executing on the touchscreen are the same as that which executed on the Horizon 3.0 Interface Unit. The software was modified sightly to detect a Horizon 3.0 Inspire configuration and set up device communications accordingly. In summary, the functions provided remain identical to the primary predicate and the technological changes were purely a reorganization of the footprint of the system.
- The Horizon 3.0 Inspire Mains Filters slightly from the standard Horizon 3.0 Mains Filter of the primary predicate device to ensure the Horizon 3.0 Inspire system as a whole remains compliant to IEC 60601-1-2 with respect to leakage currents, emissions and immunity.
- The treatment coil is different between the subject Horizon 3.0 Inspire compared to that used on the primary predicate device. The Horizon Air Film Coil is the designated treatment coil for the Horizon 3.0 Inspire configuration. The Horizon Air Film Coil has received prior clearance under K182853, K18097 and K171051 and was previously a predicate for introducing the E-Z Cool Coil range to market. The coil magnetic and electric field output properties were compared to the primary predicate Horizon 3.0 E-z Cool Coil in accordance to the FDA's guidance "Class II Special Control Guidance Document: Repetitive Transcranial Magnetic Stimulation (TMS)" and special controls 21 C.F.R. § 882.5802 and 21 C.F.R. § 882.5805 to reaffirm it's suitability for treatment in accordance with the indications for use and were found to be substantially equivalent.
- The remaining minor technological changes are for the Cart, Coil Support and Accessory Cables. The Cart and Coil Support mechanism for the Horizon 3.0 Inspire configuration are the Magstim Coll Stand respectively. Both components are of known provenance for supporting the weight of the system components and applicating coil as remaining stable during treatment delivery as these components have been previously cleared under K182853 and K180907. Most of the same cables have been re-used with the exception of the Mainframe communications and power USB leads. These cables have been testing as part of IEC 60601-1 type testing.
As can be seen from the summary above - technologically the Horizon 3.0 Inspire configuration which is the submission is very similar, and in some cases identical, to the primary predicate device configurations cleared under K232235. The technological changes can be considered more of a re-organizational exercise to provide a system of reduced footprint and cost to the encluser whilst still maintaining the same core principles, technology, safety and effectiveness of the primary predicate device.
This demonstrates all three configurations including the subject Horizon 3.0 Inspire configuration, have identical intended use/ indications for use, common specifications, equivalent performance and equivalent composition.
All changes have been appropriately verified and conforms to the same safety standards as the primary predicate device such as IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 62366-1, AAM/ANSI HE75, ISO 10993-1, ISO 10993-10, ISO 14971 for electrical safety, thermal safety, mechanical safety, alarm systems and biocompatibility. Including IEC 62304, AAM TRS7 and AAM TIR97 for software and security testing of the Horizon 3.0 Inspire configuration such as Security Requirement, Threat Mitigation and Vulnerability testing was performed and documented in accordance to demonstrate the Horizon 3.0 Inspire configuration is as secure as the primary predicate device and demonstrate the continued security of the Horizon® 3.0 TMS Therapy System family.
{7}------------------------------------------------
In conclusion - the Horizon® 3.0 TMS Therapy System with the addition of the Horizon 3.0 Inspire configuration and its technological differences from the previously cleared predicate device configurations do not raise any new or differing questions of safety or effectiveness and are substantially equivalent.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Performance testing of the Horizon® 3.0 TMS Therapy System with the addition of the Horizon 3.0 Inspire configuration demonstrates that the device performs as intended for the safe and effective treatment of MDD and as an adjunct for the treatment of OCD and is therefore substantially equivalent to the primary predicate device configurations - Horizon 3.0 with StimGuide Pro. The following nonclinical testing within this submission is relied upon for a determination of substantial equivalence:
Electromagnetic Compatibility in accordance with IEC 60601-1-2 - The Horizon 3.0 "Inspire" configuration has been tested and found compliant to the requirements of IEC 60601-1-2 to demonstrate an equivalent level of Electromagnetic Compatibility as the primary predicate device.
Electrical & Mechanical Safety in accordance with IEC 60601-1 - The Horizon 3.0 "Inspire" configuration has been tested and found compliant to the requirements of IEC 60601-1 (including IEC 60601-1-8) to demonstrate an equivalent level of Electrical and Mechanical safety as the primary predicate device.
Thermal Safety in accordance with IEC 60601-1 - The Horizon 3.0 "Inspire" configuration has been tested and found compliant to the requirements of EC 60601-1 (including IEC 60601-1-6 and IEC 60601-1-8) to demonstrate an equivalent level of Thermal safety as the primary predicate device. Additional testing was performed to ensure that the Horizon Air Film Coil can execute OCD, iTBS and rTMS protocols at worst-case ambient conditions without resulting in excessive temperatures.
Software Verification & Validation (IEC 62304, FDA Software has been developed in a structured manner following IEC 62304. Software lifecycle documentation demonstrates appropriate function of software and demonstrates the software cannot contribute to any unacceptable risk. Changes made to inspire" configuration have been verified for correctness and also do not negatively impact the Horizon 3.0 with StimGuide Pro configurations.
Usability/Human Factors Engineering (IEC 60601-1-6, IEC 62366-1, ANSI/AAMI HE75) - A human factors evaluation demonstrates appropriate human factors and usability of the Horizon® 3.0 TMS Therapy System device configurations for its intended use and demonstrates that the device is free from any unacceptable use related risks.
Magnetic Pulse Output Testing and Magnetic and Electrical Field Testing (Special Controls: 21 C.F.R. § 882.5805 and FDA Guidance "Repetitive Transcranial Magnetic Stimulation (rTMS) Systems – Class II Special Controls Guidance for Industry and FDA Staff" – July 26, 2011) - The electric field distribution produced by the subject coil in a human head phantom model filled with physiologic saline solution were measured for comparison. Values of the Magnetic and Electric field were obtained by measuring induced voltage into a measuring probe inside a phantom head model filled with a physiologic saline solution. The results demonstrated equivalent Power Outputs, Electric Fields between the subject device configuration applicating coll and the primary predicate device configuration applicating coil.
Safety Feature Testing (IEC 60601-1) - IEC 60601-1 testing exercises many fault scenarios to ensure appropriate safety features are maintained. System level testing documented as part of the software verification also exercises the various interlocks with respect to safety features. Testing introducing failure modes to test multiple fault conditions, such as disabling a software check and ensuring a hardware backup interlock is active.
Acoustic Testing (IEC 60601-1) - Acoustic testing of the system under a simulated use scenario demonstrates the Horizon 3.0 "Inspire" configuration with the Horizon Air Film Coil does not reach excessive/unacceptable noise levels.
## Clinical Testing - Not Applicable
Based on the above nonclinical testing – it can be concluded that the subject Horizon® 3.0 TMS Therapy System device introducing the Horizon 3.0 "Inspire" configuration is substantially equivalent to the primary predicate device cleared under K232235. The Horizon 3.0 "Inspire" offers an equivalent level of safety and effectiveness as the primary predicate Horizon 3.0 with StimGuide Pro configurations, whilst meeting its goal of offering an alternative system to meet such as footprint and cost to ensure that as many patients as possible can gain access to safe and effective TMS Therapy Treatments.
Electromagnetic Compatibility was demonstrated to be the primary predicate device through compliance to IEC 60601-1-2. The test plan executed to evaluate the subject Horizon 3.0 "Inspire" configuration utilized the same performance criteria as that used for the primary predicate device configurations. The test plan was composed of the same device specific criteria that was used to evaluate the primary predicate device configurations (Horizon 3.0 with StimGuide Pro) and therefore directly demonstrates a substantial equivalence of the subject Horizon® 3.0 TMS Therapy System introducing the Horizon 3.0 "Inspire" configuration.
{8}------------------------------------------------
Electrical, Mechanical and Thermal safety was demonstrated through the application of IEC 60601-1. The subject Horizon 3.0 "Inspire" configuration was found to be compliant to the standard. Further, in-house device specific protocol testing determined that for TMS, iTBS and OCD treatment protocols under reasonable worst-case simulated ambient conditions (80% - 100% of machine output, at 30°C ambient room temperature) - the Horizon 3.0 "Inspire" configuration with the Horizon Air Film coil could perform the recommended protocols safely and effectively - equivalent to the primary predicate device.
Software Verification and Validation was performed on the subject Horizon 3.0 "Inspire" device configuration. The test procedure used to verify the subject Horizon 3.0 "Inspire" configuration was comprised of test cases used to verify the primary predicate device and its configurations: Horizon 3.0 and Horizon 3.0 with StimGuide Pro. This supports substantial equivale behavior is consistent and compatible across all configurations of the Therapy System. The testing also included cybersecurity testing to ensure the subject Horizon 3.0 "Inspire" device configuration is as secure as the primary predicate device configurations.
Human Factors and Usability Engineering processes demonstrate the Horizon is substantially equivalent to the primary predicate device configurations for its interms of human factors and usability, and demonstrates that the device is free from any unacceptable use related risks. The consistency of workflows, User Interfaces and components of known provenance between all configurations of the device promote a healthy usability profile for all system configurations and the introduction of the Horizon 3.0 "Inspire" configuration of safety or effectiveness with respect to usability and human factors.
Magnetic and Electric Field Testing demonstrated that the subject device configuration - the Horizon Air Film Col, is substantially equivalent to the applicating coil of the primary predicate device configuration. Values of the Magnetic and Electric feld were obtained by measuring induced voltage into a measuring probe inside a phantom head model filled with a physiologic saline solution, with a total volume of 4cm x 3cm x 3cm and a resolution of 5mm.
Key characteristics of the fleld output were equivalent between both coils; such as the required stimulator outputy of age, field spatial distribution, E-Field decay, output waveforms and rate of change. This is especially apparent at the clinically relevant depths of stimulation - where the values of the characteristics of both coils were virtually identical.
The Horizon Air Film Coil has been on the market since K143531 and is used by reference predicate device cleared under K182853. The Horizon Air Film Coil was previously used as the predicate for introducing the Horizon E-z Cool Coil range to market. The Horizon Air Film Coll has a known safety and effectiveness profile, especially for MDD treatments, and the magnetic and electric fields of the Horizon Air Film Coil performed as part of this submission also demonstrates an equivalent level of safety and effectiveness for use as an adjunct for Obsessive Compulsive Disorder. As a result this demonstrates substantial equivalence between the subject Horizon 3.0 "Inspire" device configuration coil and the primary predicate device configurations/applicating coil.
Safety Feature Testing ensured all fault scenarios and safety features tested utilized similar (in the vast majority of cases, the same) procedures that were used to evaluate the primary predicate device, including compliance to IEC 60601-1. The testing provided within this submission demonstrates that there is substantial equivalence of safety features amongst all device configurations – importantly, including the subject Horizon 3.0 "Inspire" configuration.
Acoustic Testing in simulated use conditions (OCD Protocol at maximum machine output) for the subject Horizon 3.0 "Inspire" device configuration demonstrated equivalent acoustic properties to the values recorded for the subject device configuration fall below safe exposure limits demonstrating that Horizon 3.0 "Inspire" is as safe as the primary predicate device.
In conclusion - the subject Horizon® 3.0 TMS Therapy System with the addition of the Horizon 3.0 Inspire configuration is as safe and effective and is substantially equivalent to the primary predicate device cleared under K232235.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.