Blueprint Patient-Specific Instrumentation; Shoulder iD Primary Reversed Glenoid
Applicant
Stryker Corporation (Tornier, S.A.S.)
Product Code
PHX · Orthopedic
Decision Date
Oct 10, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3660
Device Class
Class 2
Indications for Use
BLUEPRINT™ Patient Specific Instrumentation IFU: Hardware The BLUEPRINT™ Glenoid Guides are patient-specific drill guides. They have been specially designed to assist in the intraoperative positioning of glenoid components used with total anatomic or reversed shoulder arthroplasty procedures using anatomic landmarks that are identifiable on patient-specific preoperative CT scans. Software Blueprint® is a medical device for surgeons. Blueprint® is intended to be used as a pre-surgical planner for shoulder replacement surgery. Blueprint® requires CT scan images showing the anatomical shoulder structure in a DICOM format. Blueprint® allows surgeons to visualize, measure, reconstruct, and annotate anatomic data. Blueprint® allows surgeons to design patient specific components (patient-specific instruments and Shoulder iD™ Primary Reversed Glenoid*) based on the pre-surgical plan. Blueprint® leads to the generation of a planning report. Blueprint® is to be used for adult men and women patients only whose bone maturity is reached and should not be used for diagnostic purpose. Note: Measures and patient specific guide design are provided depending on the case profiles. *Only if patient-specific instruments or Shoulder iD™ Primary Reversed Glenoid are available in your geography. Shoulder iD™ Primary Reversed Glenoid IFU: The Shoulder iD™ Primary Reversed Glenoid implant is indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle and with massive and non-repairable rotator cuff-tear with pain disabled by: · Rheumatoid arthritis - · Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis) - · Correction of functional deformity - · Fractures of the humeral head - · Traumatic arthritis - · Revision of glenohumeral joint if sufficient native glenoid bone remains All components are single use. The Shoulder iD™ Primary Reversed Glenoid implant is anchored to the bone with screws and is for non-cemented fixation. Note: A CT Scan is used to create the Shoulder iD™ Primary Reversed Glenoid implant
Device Story
Blueprint Patient-Specific Instrumentation and Shoulder iD Primary Reversed Glenoid system; used for preoperative planning and intraoperative guidance in shoulder arthroplasty. Input: patient CT scan images (DICOM). Software allows surgeons to visualize, measure, reconstruct, and annotate anatomy; simulate range of motion; select implants; and design patient-specific glenoid drill guides and implants. Output: planning report and patient-specific hardware. Used by orthopedic surgeons in clinical settings. Guides assist in precise intraoperative positioning of glenoid components. Implant (baseplate, screws, glenosphere) replaces native glenoid surface via non-cemented fixation. Benefits: improved surgical planning and implant positioning based on patient-specific anatomy.
Clinical Evidence
No clinical studies were performed. Substantial equivalence supported by non-clinical bench testing, including dimensional analysis, cadaveric testing, and reverse glenoid loosening testing.
Technological Characteristics
System includes patient-specific drill guides and planning software. Implant: baseplate with press-fit post, peripheral screws, and glenosphere; non-cemented fixation. Software: DICOM-based preoperative planning, visualization, and measurement. Connectivity: Online Management System (OMS).
Indications for Use
Indicated for adult patients with reached bone maturity requiring shoulder joint replacement due to massive, non-repairable rotator cuff tear with pain, rheumatoid arthritis, non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), functional deformity, humeral head fractures, traumatic arthritis, or revision of glenohumeral joint with sufficient native bone. Requires functional deltoid muscle.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
BLUEPRINT™ Patient Specific Instrumentation (K232265)
BLUEPRINT™ Patient Specific Instrumentation + Tornier Perform (K211359)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, with the letters FDA in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 10, 2024
Stryker Corporation (Tornier, S.A.S.) Aymen Azaiez Principal Regulatory Affairs Specialist 161 rue Lavoisier Montbonnot-Saint-Martin, 38330 France
Re: K241491
Trade/Device Name: Blueprint Patient-Specific Instrumentation; Shoulder iD Primary Reversed Glenoid Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX, KWS, QHE Dated: September 6, 2024 Received: September 6, 2024
Dear Aymen Azaiez:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Farzana Digitally signed by Farzana Sharmin -S Date: 2024.10.10 Sharmin -S 12:23:24 -04'00'
Farzana Sharmin, PhD Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
Submission Number (if known)
K241491
Device Name
Blueprint Patient-Specific Instrumentation: Shoulder iD Primary Reversed Glenoid
ndications for Use (Describe)
BLUEPRINT™ Patient Specific Instrumentation IFU:
Hardware
The BLUEPRINT™ Glenoid Guides are patient-specific drill guides. They have been specially designed to assist in the intraoperative positioning of glenoid components used with total anatomic or reversed shoulder arthroplasty procedures using anatomic landmarks that are identifiable on patient-specific preoperative CT scans.
Software
Blueprint® is a medical device for surgeons.
Blueprint® is intended to be used as a pre-surgical planner for shoulder replacement surgery. Blueprint® requires CT scan images showing the anatomical shoulder structure in a DICOM format. Blueprint® allows surgeons to visualize, measure, reconstruct, and annotate anatomic data. Blueprint® allows surgeons to design patient specific components (patient-specific instruments and Shoulder iD™ Primary Reversed Glenoid*) based on the pre-surgical plan.
Blueprint® leads to the generation of a planning report.
Blueprint® is to be used for adult men and women patients only whose bone maturity is reached and should not be used for diagnostic purpose.
Note: Measures and patient specific guide design are provided depending on the case profiles. *Only if patient-specific instruments or Shoulder iD™ Primary Reversed Glenoid are available in your geography.
Shoulder iD™ Primary Reversed Glenoid IFU:
The Shoulder iD™ Primary Reversed Glenoid implant is indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle and with massive and non-repairable rotator cuff-tear with pain disabled by:
· Rheumatoid arthritis
- · Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis)
- · Correction of functional deformity
- · Fractures of the humeral head
- · Traumatic arthritis
- · Revision of glenohumeral joint if sufficient native glenoid bone remains
All components are single use.
The Shoulder iD™ Primary Reversed Glenoid implant is anchored to the bone with screws and is for non-cemented fixation.
Note: A CT Scan is used to create the Shoulder iD™ Primary Reversed Glenoid implant
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
| 510(k) #: K241491 | 510(k) Summary | | | Prepared on: 2024-09-09 |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|----------------------------------------------------------------------------------|--|-------------------------|
| Contact Details | | | | 21 CFR 807.92(a)(1) |
| Applicant Name | | Stryker Corporation (Tornier, S.A.S.) | | |
| Applicant Address | | 161 rue Lavoisier Montbonnot-Saint-Martin 38330 France | | |
| Applicant Contact Telephone | | +33648381036 | | |
| Applicant Contact | | Mr. Aymen Azaiez | | |
| Applicant Contact Email | | aymen.azaiez@stryker.com | | |
| Correspondent Name | | Stryker Corporation (Tornier S.A.S.) | | |
| Correspondent Address | | 161 rue Lavoisier Montbonnot-Saint-Martin 38330 France | | |
| Correspondent Contact Telephone | | +33648381036 | | |
| Correspondent Contact | | Mr. Aymen Azaiez | | |
| Correspondent Contact Email | | aymen.azaiez@stryker.com | | |
| Device Name | | | | 21 CFR 807.92(a)(2) |
| Device Trade Name | | Blueprint Patient-Specific Instrumentation; Shoulder iD Primary Reversed Glenoid | | |
| Common Name | | Shoulder Prosthesis, Reverse Configuration | | |
| Classification Name | | Shoulder joint metal/polymer semi-constrained cemented prosthesis | | |
| Regulation Number | | 888.3660 | | |
| Product Code(s) | | PHX, KWS, QHE | | |
| Legally Marketed Predicate Devices | | | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | | | Product Code |
| K232265 | BLUEPRINT™ Pati et Specific Instrumentation | | | PHX |
| K211359 | BLUEPRINT™ Patient Specific Instrumentation + Tornier Perform | | | PHX |
| K222987 (Reference) | Akunah REFLECT | | | LLZ |
| Device Description Summary | | | | 21 CFR 807.92(a)(4) |
| BLUEPRINT™ Patient Specific Instrumentation<br>Blueprint Patient Specific Instrumentation is composed of two components: Blueprint Glenoid Guides (hardware) and Blueprint Planning Software (software) | | | | |
{6}------------------------------------------------
### Blueprint™ Glenoid Guides
The Blueprint Glenoid Guides are patients specially designed to facilitate the implantation of Wright-Tornier glenoid prostheses.
The Blueprint Glenoid Guides are designed and manufactured based on a pre-operated only by the software Blueprint Planning Software.
Blueprint™ Planning Software
Blueprint Planning Software is a software connected to an Online Management System (OMS). The user interface software is installed on a computer and is intended to be used by orthopedic surgeons, as a preoperative planning software for shoulder arthroplasty surgery (anatomic and reversed).
It is intended to help to plan an operation by allowing surgeons to:
- · Plan for shoulder arthroplasty cases
- · Position and select glenoid and humeral implants,
- · Simulate the prosthetic range of motion,
- · Interact with implants and different computed measurements
- · Generate information required to design a patient-specific glenoid component when appropriate.
Shoulder iD™ Primary Reversed Glenoid
The Shoulder iD Primary Reversed Glenoid implant (Shoulder iD) is intended to replace the native glenoid surface of the scapulohumeral joint as part of a reverse shoulder prosthesis.
The glenoid implant is composed of a baseplate with a press-fit post, peripheral anchoring screws, and a glenosphere. Ancillary instruments are also provided for the implantation of the prosthesis.
## Intended Use/Indications for Use
21 CFR 807.92(a)(5)
BLUEPRINT™ Patient Specific Instrumentation IFU:
Hardware
The BLUEPRINT™ Glenoid Guides are patient-specific drill guides. They have been specially designed to assist in the intraoperative positioning of glenoid components used with total anatomic or reversed shoulder arthroplasty procedures using anatomic landmarks that are identifiable on patient-specific preoperative CT scans.
Software
Blueprint® is a medical device for surgeons.
Blueprint® is intended to be used as a pre-surgical planner for shoulder replacement surgery.
Blueprint® requires CT scan images showing the anatomical shoulder structure in a DICOM format.
Blueprint® allows surgeons to visualize, measure, reconstruct, and annotate anatomic data.
Blueprint® allows surgeons to design patient-specific instruments and Shoulder iD™ Primary Reversed Glenoid*) based on the pre-surgical plan.
Blueprint® leads to the generation of a planning report.
Blueprint® is to be used for adult men and women patients only whose bone maturity is reached and should not be used for diagnostic purpose.
Note: Measures and patient specific guide design are provided depending on the case profiles.
*Only if patient-specific instruments or Shoulder iD™ Primary Reversed Glenoid are available in your geography.
Shoulder iD™ Primary Reversed Glenoid IFU:
The Shoulder iD™ Primary Reversed Glenoid in indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle and with massive and non-repairable rotator cuff-tear with pain disabled by:
· Rheumatoid arthritis
- · Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis)
- · Correction of functional deformity
- · Fractures of the humeral head
- · Traumatic arthritis
- · Revision of glenohumeral joint if sufficient native glenoid bone remains
All components are single use.
The Shoulder iD™ Primary Reversed Glenoid implant is anchored to the bone with screws and is for non-cemented fixation. Note: A CT Scan is used to create the Shoulder iD™ Primary Reversed Glenoid implant
## Indications for Use Comparison
## 21 CFR 807.92(a)(5)
For BLUEPRINT Patient Specific Instrumentations for use of the subject device are identical to the predicate device. For Shoulder iD Primary Reversed Glenoid, the indications for use of the subject device (Shoulder iD Primary Reversed Glenoid) and of the predicate device (Tornier Perform Patient-Matched Primary Reversed Glenoid) are identical except for the name of the device.
{7}------------------------------------------------
# Technological Comparison
## 21 CFR 807.92(a)(6)
### BLUEPRINT Patient Specific Instrumentation
The subject and predicate hardware devices have identical technological features.
Technological differences between the subject and predicate software devices do not raise any different questions of safety and effectiveness and they are addressed though performance testing.
Shoulder iD Primary Reversed Glenoid.
The Shoulder iD Primary Reversed Glenoid and the predicate Tornier Perform Patient-Matched Primary Reversed Glenoid have the same technological features. The only differences between the subject and the brand name and the updated Operative Technique for the subject device. The difference in the operative technique does not raise any different questions of safety and effectiveness and is addressed though performance testing.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
### BLUEPRINT™ Patient Specific Instrumentation:
Technological differences between the subject and predicate hardware devices are supported by the dimensional the predicate device hardware which remain applicable to the subject hardware device, as the subject device do not impact functional dimensions or material, and cadaveric test performed on the subject device.
Technological differences between the subject and predicate software are supported with verification and validation evaluations. The operating principle of the subject device is the same as that of the predicate device.
The differences in design specifications do not raise new questions of safety and effectiveness over the predicate device as demonstrated in validation testing.
Shoulder iD Primary Reversed Glenoid
Non-clinical Reverse Glenoid Loosening testing was confirm substantial equivalence to the predicate device in manual bone preparation conditions.
No clinical studies were performed.
#### BLUEPRINT™ Patient Specific Instrumentation:
The subject device hardware BLUEPRINT™ Patient Specific Instrumentation does not raise any different questions of safety and effectiveness over the predicate device. Technological differences between the subject and predicate are upported by the dimensional and cadaveric tests performed on the predicate device hardware which are still applicable on the subject hardware device as the changes to the subject device do not impact functional dimensions nor material.
#### Shoulder iD Primary Reversed Glenoid
The Shoulder iD Primary Reversed Glenoid does not raise different questions of safety or effectiveness. There are in technological characteristics between the predicate and subject devices. The results of the Shoulder iD Primary Reversed Glenoid support substantial equivalence to the predicate Tornier Patient-Matched Primary Reversed Glenoid (K211359, cleared November 12, 2021).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.