K241488 · Trainfes · IPF · Feb 5, 2025 · Physical Medicine
Device Facts
Record ID
K241488
Device Name
TrainFES Advance
Applicant
Trainfes
Product Code
IPF · Physical Medicine
Decision Date
Feb 5, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
For Prescription and Home Use by prescription from a medical professional: The TrainFES advanced is a neuromuscular electrical stimulator indicated for use under medical supervision for adjunctive therapy in the treatment of medical diseases and conditions. As a powered muscle stimulator, TrainFES advanced is indicated for the following conditions: - Relaxation of muscle spasms - Prevention or retardation of disuse atrophy - increasing local blood circulation - immediate post-surgical stimulation of calf muscles to prevent venous thrombosis - Maintaining or increasing range of motion - Muscle re-education As an external functional neuromuscular stimulator (FES), TrainFES Advanced is indicated for the following conditions: - Helps to relearn voluntary motor functions of the extremities. TrainFES' Advanced intended population is anybody aged 22 or over. Environments of use: TrainFES Advanced devices can be used by both therapists and patients, in the clinic, hospitals or at home. Platform: TrainFES is a battery-powered, wireless device, accessible through software.
Device Story
TrainFES Advanced is a 6-channel, battery-powered, wireless functional electrostimulator. Used in clinics, hospitals, or home environments by patients and medical professionals. Device generates electrical impulses to stimulate paralyzed segments; facilitates movement relearning and neuromodulation of tone. Configured via smartphone/tablet app; session settings retrievable from PC or cloud. Output parameters (e.g., current, frequency, pulse width) are adjusted by the user/therapist via software. Output affects muscle contraction, aiding in physical therapy and rehabilitation. Benefits include improved motor function, circulation, and muscle maintenance.
Clinical Evidence
No clinical testing was performed in support of this submission. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), performance of nerve/muscle stimulators (IEC 60601-2-10), home healthcare environment requirements (IEC 60601-1-11), battery safety (IEC 62133-2), software lifecycle (IEC 62304), usability (IEC 62366), and risk management (ISO 14971).
Indicated for patients aged 22+ for relaxation of muscle spasms, prevention/retardation of disuse atrophy, increasing local blood circulation, post-surgical calf stimulation to prevent venous thrombosis, maintaining/increasing range of motion, muscle re-education, and relearning voluntary motor functions of extremities. Used under medical supervision.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym and name on the right. The FDA acronym is in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 5, 2025
Biomedical Devices Spa % Juan Tezak Consultant Compliance 4 Devices 118 W Prive Cr. Delray Beach, Florida 33445
Re: K241488
Trade/Device Name: TrainFES Advanced Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF. GZI Dated: May 24, 2024 Received: May 24, 2024
Dear Juan Tezak:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Heather L. Dean -S
for Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
#### K241488
Device Name
#### TrainFES Advanced
Indications for Use (Describe)
For Prescription and Home Use by prescription from a medical professional:
The TrainFES advanced is a neuromuscular electrical stimulator indicated for use under medical supervision for adjunctive therapy in the treatment of medical diseases and conditions.
As a powered muscle stimulator, TrainFES advanced is indicated for the following conditions:
- Relaxation of muscle spasms
- Prevention or retardation of disuse atrophy
- increasing local blood circulation
- immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
- Maintaining or increasing range of motion
- Muscle re-education
As an external functional neuromuscular stimulator (FES), TrainFES Advanced is indicated for the following conditions:
- Helps to relearn voluntary motor functions of the extremities.
TrainFES' Advanced intended population is anybody aged 22 or over.
Environments of use: TrainFES Advanced devices can be used by both therapists and patients, in the clinic, hospitals or at home.
Platform: TrainFES is a battery-powered, wireless device, accessible through software.
Type of Use (Select one or both, as applicable)
| <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
|--------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
X | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary K241488 Prepared July 03, 2024
| I) | Applicant<br>Submitter<br>Address | Biomedical Devices Spa<br>Camino El Alba 8760, of 903 Las<br>Condes RM Chile |
|------|-----------------------------------|------------------------------------------------------------------------------|
| | Contact<br>Telephone<br>E-mail | Mr. Matias Hosiasson<br>+56 998 954326<br>matias@trainfes.com |
| II) | Device<br>Trade Name | TrainFES Advanced |
| | Common Name | External functional neuromuscular stimulator |
| | Classification Name | Stimulator, Neuromuscular, External Functional |
| | Regulation Number | 890.5850 IPF / 882.5810 GZI |
| | Product Code | IPF, GZI |
| III) | Predicate Device<br>510(k) Number | K210002 |
| | Applicant | EGZOTech Sp. Z.o.o |
| | Trade Name | Stella BIO |
Power Muscle Stimulator
876.5320 882.5810
NYN
890.5850 / 882.5890 / 882.5050 /
IPF, GZJ, HCC, GZI, KPI, NGX, NUH,
Classification Name
Regulation Number
Product Code
{5}------------------------------------------------
#### IV) Indication For Use
For Prescription and Home Use by prescription from a medical professional: The TrainFES advanced is a neuromuscular electrical stimulator indicated for use under medical supervision for adjunctive therapy in the treatment of medical diseases and conditions.
As a powered muscle stimulator, TrainFES advanced is indicated for the following conditions:
Relaxation of muscle spasms
Prevention or retardation of disuse atrophy
Increasing local blood circulation
Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis Maintaining or increasing range of motion
Muscle re-education
As an external functional neuromuscular stimulator (FES), TrainFES Advanced is indicated for the following conditions: Helps to relearn voluntary motor functions of the extremities. TrainFES' Advanced intended population is anybody aged 22 or over.
Environments of use: TrainFES Advanced devices can be used by both therapists and patients, in the clinic, hospitals or at home.
Platform: TrainFES is a battery-powered, wireless device, accessible through software.
#### V) Device Description
TrainFES Advanced is a portable functional electrostimulator with 6 channels designed for use in the clinics and hospitals by the medical professionals as well as in the home environment by the patient. This device generates electrical impulses to stimulate the musculature of paralyzed seqments and facilitate both the relearning of movement and neuromodulation of tone.
TrainFES Advanced is a battery-powered, wireless device, configurable from the TRAINFES App, available for Smartphone and Tablets, which allows you to adjust different parameters and follow a training plan from your smartphone. Session settings can be retrieved from the PC or Cloud.
#### VI) Comparison of Technological Characteristics
| Basic Device Characteristics | | | |
|------------------------------------|------------------------|-------------------------------------------|------------|
| Characteristics /<br>Specification | TrainFES Advanced | Stella BIO | Difference |
| 510(k) Number | K241488 | K210002 | N/A |
| FDA product code | IPF, GZI | IPF, GZJ, HCC, GZI, KPI,<br>NGX, NUH, NYN | N/A |
| Manufacturer | Biomedical Devices Spa | EGZO Tech | N/A |
{6}------------------------------------------------
| Classification | Class II | Class II | Same |
|------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| Prescription or OTC | Prescription and Home Use | Prescription and Home Use | Same |
| Environment of Use | Clinics, hospitals and home environments | Clinics, hospitals and home environments | Same |
| Indications for use | For Prescription and Home Use by prescription from a medical professional:<br>The TrainFES advanced is a neuromuscular electrical stimulator indicated for use under medical supervision for adjunctive therapy in the treatment of medical diseases and conditions.<br><br>As a powered muscle stimulator, TrainFES advanced is indicated for the following conditions:<br>• Relaxation of muscle spasms<br>• Prevention or retardation of disuse atrophy<br>• Increasing local blood circulation<br>• Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis<br>• Maintaining or increasing range of motion<br>• Muscle re-education<br><br>As an external functional neuromuscular stimulator (FES), TrainFES Advanced is indicated for the following conditions:<br>• Helps to relearn voluntary motor functions of the extremities. | The Stella BIO is a neuromuscular electrical stimulator indicated for use under medical supervision for adjunctive therapy in the treatment of medical diseases and conditions.<br><br>As a powered muscle stimulator, Stella BIO is indicated for the following conditions:<br>• Relaxation of muscle spasms,<br>• Prevention or retardation of disuse atrophy,<br>• Increasing local blood circulation,<br>• Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis,<br>• Maintaining or increasing range of motion,<br>• Muscle re-education.<br><br>As a transcutaneous electrical nerve stimulator for pain relief, Stella BIO is indicated for the following conditions:<br>• Symptomatic relief and management of chronic (long-term), intractable pain,<br>• Adjunctive treatment in the management of post-surgical pain and post traumatic acute pain.<br><br>As a biofeedback device, Stella BIO is indicated for:<br>• Biofeedback, relaxation, and muscle re- education. | Similar Note 1 |
| | TrainFES' Advanced<br>intended population is<br>anybody aged 22 or over.<br><br>Environments of use:<br>TrainFES Advanced<br>devices can be used by<br>both therapists and<br>patients, in the clinic,<br>hospitals or at home.<br><br>Platform: TrainFES is a<br>battery-powered, wireless<br>device, accessible through<br>software. | As an external functional<br>neuromuscular stimulator,<br>Stella BIO is indicated for<br>the following<br>conditions:<br>• Helps to relearn voluntary<br>motor functions of the<br>extremities.<br><br>As a non-implanted<br>electrical continence<br>device, Stella BIO is<br>indicated for the following<br>conditions:<br>• Acute and ongoing<br>treatment of stress, urge or<br>mixed urinary incontinence<br>and where the following<br>results may improve<br>urinary control: inhibition of<br>the detrusor muscles<br>through reflexive<br>mechanisms<br>and strengthening of pelvic<br>floor muscles.<br>• Incontinence treatment<br>for assessing EMG activity<br>of the pelvic floor and<br>accessory muscles such as<br>abdominal and the gluteus<br>muscles.<br><br>Patient population: Stella<br>BIO Prescription and Home<br>Use by prescription from a<br>medical professional<br>can be used on adults<br>aged 22 yrs and older.<br>Environments of Use:<br>Clinics, hospital and home<br>environments.<br><br>Platform: Stella BIO is a<br>battery-powered, wireless<br>device, accessible through<br>software. | |
| Power Source | Battery: Li-Ion 3,7V<br>(4000mAh) | Battery pack Li-Ion 7,4V | Different<br>Note 2 |
| Method of Line<br>Current Isolation | N/A (Battery) | N/A (Battery) | Same |
| | | | |
| Patient leakage<br>current (Normal<br>condition, μA) | N/A<br>Battery Operated Device<br>(<100μA patient leakage) | N/A<br>Battery Operated Device<br>(<100μA patient leakage) | Same |
| Patient leakage<br>current (Single fault<br>condition, μA) | N/A<br>Battery Operated Device<br>(<100μA patient leakage) | N/A<br>Battery Operated Device<br>(<100μA patient leakage) | Same |
| Number of Output<br>Modes | 2 | 4 | Different<br>Note 3 |
| Software/<br>Microprocessor<br>control | Yes | Yes | Same |
| Automatic Overload<br>Trip | Yes | Yes | Same |
| Automatic No-Load<br>Trip | Yes | Yes | Same |
| Automatic Shut Off | Yes | Yes | Same |
| Patient Override<br>Control | Yes (Button) | Yes | Same |
| Indicator Display<br>On/Off Status | Yes (Led) | Yes | Same |
| Indicator Display<br>Low Battery | Yes | Yes | Same |
| Indicator Display<br>Voltage/Current<br>Level | Yes | Yes | Same |
| Standards | ISO 13485:2016<br>Bluetooth: Complies with<br>FCC 15.2473 (for U.S.)<br>regulations<br>IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-10<br>IEC 60601-1-11<br>IEC 62133<br>Compliance<br>IEC 62304:2015<br>IEC 62366 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-10<br>IEC 60601-1-11<br>IEC 62304<br>IEC 62366<br>IEC 62133<br>ANSI/AAMI NS4 | Different<br>Note 4 |
| Compliance with 21<br>CFR 898 | Yes | Yes | Same |
| Weight | 218 g | 112 g | Similar<br>Note 5 |
| Dimension | 130 x 79 x 33mm | 91.5 x 68.4 x 24 mm | Similar<br>Note 5 |
| Housing material<br>and Construction | ABS Plastic | Plastic (Injection Molded<br>ABS) | Same |
| Apply Part | BF | | |
| Powered Muscle Stimulator | | | |
| Characteristics /<br>Specification | TrainFES Advanced | Stella BIO | Difference |
| 510(k) Number | K241488 | K210002 | N/A |
| Manufacturer | Biomedical Devices Spa | EGZO Tech | N/A |
| Number of channels | 6 Channels | 8 Channels | Different<br>Note 1 PMS |
| Synchronous or<br>alternating | Alternating | Synchronous | Different<br>Note 2 PMS |
| Method of channel<br>isolation | Multiplexer | Multiplexer…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.