iQ 2 Nasal Vented Mask; Phantom 2 Nasal Vented Mask
Applicant
Sleepnet Corporation
Product Code
BZD · Anesthesiology
Decision Date
Jun 15, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The iQ 2 Nasal Vented Mask and Phantom 2 Nasal Vented Mask are intended to be used with positive airway pressure devices, such as CPAP or bi-level, operating at 3 cmH2O to 30 cmH2O. The masks are to be used on adult patients (>30 kg) for whom positive airway pressure therapy has been prescribed. The masks are intended for single-patient multi-use in the home, hospital or institutional environment.
Device Story
Nasal vented masks (iQ 2, Phantom 2) provide interface between positive airway pressure device (CPAP/bi-level) and patient; deliver pressurized air (3-30 cmH2O) to nose; feature exhalation port for CO2 flushing. Used in home, hospital, or institutional settings by adult patients. Device is passive; connects to standard 22mm circuits. Includes magnets in headgear clips (380mT field strength). Healthcare providers prescribe therapy; patients self-administer. Output is pressurized air delivery; mask seal ensures therapy efficacy. Benefits include non-invasive respiratory support. Updated labeling includes contraindications and warnings regarding magnetic field interactions with medical implants.
Clinical Evidence
No clinical data. Bench testing only. Performance data (exhaust flow, dead space, resistance to flow, sound levels) and biocompatibility/reprocessing data from the predicate device were leveraged as the subject device design is identical.
Technological Characteristics
Nasal vented mask with gel bladder, mask shell, and headgear. Standard 22mm connection. Passive device; no energy source. Materials identical to predicate. Magnetic headgear clips (380mT). Non-sterile. Single-patient multi-use. Shelf life 5 years.
Indications for Use
Indicated for adult patients (>30 kg) prescribed positive airway pressure therapy (CPAP or bi-level) at 3-30 cmH2O. Contraindicated for patients or individuals in close physical contact with the mask who have active medical implants or metallic implants that interact with magnets (e.g., pacemakers, ICDs, neurostimulators, aneurysm clips, stents, ocular implants, insulin pumps, CSF shunts, embolic coils, metallic splinters, cochlear implants, flow disruption devices, contact lenses with metal, dental implants, metallic cranial plates, screws, burr hole covers, bone substitutes, or magnetic implants/valves in upper limbs/torso).
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 15, 2024
Sleepnet Corporation % Paul Dryden President ProMedic Consulting LLC 131 Bay Point Dr NE St. Ptersburg, Florida 33704
Re: K241469
Trade/Device Name: iQ 2 Nasal Vented Mask; Phantom 2 Nasal Vented Mask Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: Class II Product Code: BZD Dated: May 23, 2024 Received: May 24, 2024
Dear Paul Dryden:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
### Rachana Visaria -S
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices
{2}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241469
Device Name
iQ 2 Nasal Vented Mask and Phantom 2 Nasal Vented Mask
Indications for Use (Describe)
The iQ 2 Nasal Vented Mask and Phantom 2 Nasal Vented Mask are intended to be used with positive airway pressure devices, such as CPAP or bi-level, operating at 3 cmH2O to 30 cmH2O. The masks are to be used on adult patients (>30 kg) for whom positive airway pressure therapy has been prescribed. The masks are intended for single-patient multi-use in the home, hospital or institutional environment.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| <span style="font-size: 10pt;"> </span> | |
| ☒ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------|-----------------------------------------------|
|------------------------------------------------|-----------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| Sponsor:<br>Sleepnet Corporation<br>5 Merrill Industrial Drive<br>Hampton, NH 03842 | Tel - 603-758-6625 |
|-------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| Sponsor Contact: | Jennifer Kennedy - Director of Regulatory and Quality |
| Submission Correspondent: | Paul Dryden<br>ProMedic, LLC |
| Proprietary or Trade Name: | iQ 2 Nasal Vented Mask and Phantom 2 Nasal Vented Mask |
| Common/Usual Name: | Patient interface for CPAP |
| Regulation Number:<br>Regulation Code:<br>Product Code: | 21CFR 868.5905<br>Non-continuous ventilator (IPPB)<br>BZD |
| Device: | iQ 2 and Phantom 2 Nasal Vented Mask |
| Predicate Device:<br>Regulation Number:<br>Regulation Code:<br>Product Code: | K211274 - Sleepnet iQ 2 and Phantom 2 Nasal Mask<br>21CFR 868.5905<br>Non-continuous ventilator (IPPB)<br>BZD |
#### Device Description:
The Sleepnet iQ 2 and Phantom 2 Nasal Vented masks are nearly identical to the Predicate Sleepnet iQ 2 and Phantom 2 Nasal Mask (K211274).
The subject of this submission is:
· Addition of a contraindication related to magnets on the iQ 2 Nasal Vented Mask and Phantom 2 Nasal Vented Mask
• Update to the warning related to magnets on the iQ 2 Nasal Vented Mask and Phantom 2 Nasal Vented Mask
#### Indications for Use:
The iQ 2 Nasal Vented Mask and Phantom 2 Nasal Vented Mask are intended to be used with positive airway pressure devices, such as CPAP or bi-level, operating at 3 cmH2O to 30 cmH2O. The masks are to be used on adult patients (>30 kg) for whom positive airway pressure therapy has been prescribed. The masks are intended for single-patient multi-use in the home, hospital or institutional environment.
Patient Population: For adults (>30 kg)
Environment of Use: Home or hospital / institutional environments
Substantial Equivalence Discussion:
{5}------------------------------------------------
The Sleepnet iQ 2 and Phantom 2 Nasal Vented masks are viewed as substantially equivalent to the predicate device because:
#### Indications -
- We have updated the device name and removed "or above" in the Indications for Use which are similar.
- Similar to the predicate - K211274.
#### Patient Population -
- The masks are to be used on adult patients (>30 kg) for whom positive airway pressure ● therapy has been prescribed.
- Identical to the predicate K211274. ●
#### Environment of Use -
- The masks are intended for use in the home or hospital/institutional environment.
- Identical to the predicate K211274. ●
#### Technological Characteristics -
- Identical technology to the predicate K211274. ●
#### Non-Clinical Testing Summary -
#### Bench testing -
No new performance testing was conducted on the subject device, since the device designs are identical to the predicates.
Discussion - The performance test results from the predicate device were leveraged for the subject device.
#### Biocompatibility -
There has been no change in the materials – the subject and predicate device utilize the same materials.
Discussion - The biocompatibility test results from the predicate device were leveraged for the subject device.
#### Reprocessing -
There has been no change in the reprocessing - the subject and predicate device utilize the same reprocessing instructions.
Discussion - The reprocessing test results from the predicate device were leveraged for the subject device.
We have demonstrated that the proposed device is equivalent to the predicate, K211274.
{6}------------------------------------------------
# K241469 510(k) Summary Page 3 of 7 15-Jun-24
# Table of Comparison to Predicate
| Attributes | Subject -<br>iQ 2 Nasal Vented Mask and | Predicate -<br>iQ 2/ Phantom 2 Nasal Mask | Explanation of Differences |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| | Phantom 2 Nasal Vented Mask | | |
| 510(k) | K241469 | K211274 | |
| Product<br>Classification<br>CFR | BZD<br>CFR 868.5905 | BZD<br>CFR 868.5905 | Identical |
| Indications for Use | The iQ 2 Nasal Vented Mask and Phantom 2<br>Nasal Vented Mask are intended to be used with<br>positive airway pressure devices, such as CPAP<br>or bi-level, operating at 3 cmH2O to 30 cmH2O.<br>The masks are to be used on adult patients (>30<br>kg) for whom positive airway pressure therapy<br>has been prescribed. The masks are intended for<br>single-patient multi-use in the home, hospital or<br>institutional environment. | The iQ 2/ Phantom 2 Nasal Mask is intended to<br>be used with positive airway pressure devices,<br>such as CPAP or bi-level, operating at or above 3<br>cm H2O. The mask is to be used on adult patients<br>(>30kg) for whom positive airway pressure<br>therapy has been prescribed. The mask is<br>intended for single- patient multi-use in the<br>home, hospital or institutional environment. | Similar<br>Removed "or above" for clarity |
| Patient Population | Adult (>30 kg) | Adult (>30 kg) | Identical |
| Patient type | Patients for whom positive airway pressure therapy<br>has been prescribed | Patients for whom positive airway pressure<br>therapy has been prescribed | Identical |
| Prescriptive | Yes | Yes | Identical |
| Principle of Operation | Device provides a seal over the nose to allow for<br>delivery of pressurized air from a positive airway<br>pressure device. Device has an exhalation port for<br>flushing out exhaled CO2.<br>The device is passive until connected to the<br>positive pressure device. | Device provides a seal over the nose to allow for<br>delivery of pressurized air from a positive airway<br>pressure device. Device has an exhalation port<br>for flushing out exhaled CO2.<br>The device is passive until connected to the<br>positive pressure device. | Identical |
| Contraindication and<br>Warnings | New Contraindication<br>Do not use this mask if you or anyone (example:<br>household members, bed partners, caregivers, etc.)<br>in close physical contact with your mask has an<br>active medical implant or metallic implant that will<br>interact with magnets. Implant examples include,<br>but are not limited to, pacemakers, implantable | There were no contraindications in K211274. | New as per recall event #94169<br>Original K211274 labeling<br>available in Section<br>006_Labeling. |
| | | | |
| | cardioverter defibrillators (ICD), neurostimulators,<br>aneurysm clips, metallic stents, ocular implants,<br>insulin/infusion pumps, cerebral spinal fluid (CSF)<br>shunts, embolic coils, metallic splinter, implants to<br>restore hearing or balance with implanted magnets<br>(such as cochlear implants), flow disruption<br>devices, contact lenses with metal, dental implants,<br>metallic cranial plates, screws, burr hole covers,<br>bone substitute device, magnetic metallic<br>implants/electrodes/valves placed in upper limbs,<br>torso, or higher, etc. If you have any questions<br>regarding the implant, consult your physician or<br>the manufacturer of your implant.<br>Updated Warning<br>Magnets are used in the mask and headgear clips<br>with a field strength of 380mT. With the exception<br>of the devices identified in the contraindication,<br>ensure that the mask is kept at least 6 inches<br>(approx.16 cm) away from any other medical<br>implants or medical devices that can be impacted by<br>the magnetic fields to avoid possible effects from<br>localized magnetic fields. For instance, the<br>functioning of implants may change, or implants<br>may move within the body due to magnetic<br>attraction/repulsion. This applies to you or anyone<br>in close physical contact with your mask. | Magnet Warning<br>Magnets are used in the swivel frame/tube set and<br>the headgear clips. The magnetic field strength is<br>380 mT. Ensure that the magnets are kept at least<br>2.25 inches (6 cm) away from any active medical<br>implant or medical device that can be impacted by<br>the magnetic field (e.g., pacemaker, defibrillators,<br>neurostimulators, cochlear implants, hearing aids)<br>to avoid possible effects from localized magnetic<br>fields | |
| Therapy Pressure<br>Range | 3 cm H2O to 30 cm H2O. | 3 cm H2O to 30 cm H2O. | Identical |
| Anatomical site | Face (seals around nose) | Face (seals around nose) | Identical |
| | | | |
| User Interface<br>to administer<br>therapy | Masks have a standard 22mm connection that<br>connects to 22mm CPAP/bi-level circuits. | Masks have a standard 22mm connection that<br>connects to 22mm CPAP/bi-level circuits. | Identical |
| Environment of Use | The masks are intended for use in the home or<br>hospital/institutional environment. | The masks are intended for use in the home or<br>hospital/institutional environment. | Identical |
| Attributes | Subject -<br>iQ 2/ Phantom 2 nasal vented mask | Predicate -<br>iQ 2/ Phantom 2 nasal mask | Explanation of Differences |
| Useful life | Mask cushion - 1 month<br>Swivel frame/tube set - 3 months Headgear<br>and magnetic clips – 6 months | Mask cushion - 1 month<br>Swivel frame/tube set - 3 months Headgear<br>and magnetic clips - 6 months | Identical |
| Shelf life | 5 years. | 5 years. | Identical |
| Non-sterile | Yes | Yes | Identical |
| Cleaning methods | Mild Soap (such as Ivory) and warm water | Mild Soap (such as Ivory) and warm water | Identical |
| Available sizes | 1 | 1 | Identical |
| Shape | Identical | Identical | Identical |
| Incorporates an Exhaust port | Yes | Yes | Identical |
| Components of the mask | Mask cushion Mask shell Gel bladder Headgear<br>Swivel frame/tube set (tubing assembly) | Mask cushion Mask shell Gel bladder Headgear<br>Swivel frame/tube set (tubing assembly) | Identical |
| Shell design | Soft | Soft | Identical |
{7}------------------------------------------------
# K241469 510(k) Summary
{8}------------------------------------------------
# K241469 510(k) Summary
| Duration of Use | Single patient, multi-use |
|-----------------|---------------------------|
| | Single patient, multi-use |
| | Identical |
{9}------------------------------------------------
# K241469 510(k) Summary Page 6 of 7
15-Jun-24
| Attributes | Subject -<br>iQ 2/ Phantom 2 nasal vented mask | | | Predicate -<br>iQ 2/ Phantom 2 nasal mask | | | Explanation of Differences |
|-----------------------------------------|------------------------------------------------------------------------------------------------------|------------------------|-----------|------------------------------------------------------------------------------------------------------|------------------------|-----------|----------------------------|
| Patient Contact per<br>ISO 10993-1 | Skin contact, Skin and<br>Externally Communicating with tissue (air<br>pathway)<br>Permanent contact | | | Skin contact, Skin and<br>Externally Communicating with tissue (air<br>pathway)<br>Permanent contact | | | Identical |
| Exhaust Flow | Pressure<br>(cm H2O) | Exhaust flow (lpm) | | Pressure<br>(cm H2O) | Exhaust flow (lpm) | | Identical |
| | | iQ 2 | Phantom 2 | | iQ 2 | Phantom 2 | |
| | 4 | 18.2 | 18.9 | 4 | 18.2 | 18.9 | |
| | 8 | 26.3 | 27.5 | 8 | 26.3 | 27.5 | |
| | 12 | 34.3 | 35.2 | 12 | 34.3 | 35.2 | |
| | 16 | 39.6 | 39.2 | 16 | 39.6 | 39.2 | |
| | 20 | 44.0 | 45.3 | 20 | 44.0 | 45.3 | |
| | 24 | 49.8 | 50.6 | 24 | 49.8 | 50.6 | |
| | 28 | 53.4 | 55.3 | 28 | 53.4 | 55.3 | |
| | 30 | 55.6 | 56.8 | 30 | 55.6 | 56.8 | |
| Dead space | iQ 2 –<br>Mask – 47.3 ml<br>Tubing – 68.33 ml | | | iQ 2 –<br>Mask – 47.3 ml<br>Tubing – 68.33 ml | | | Identical |
| | Phantom 2 –<br>Mask – 30.7 ml<br>Tubing – 68.33 ml | | | Phantom 2 –<br>Mask – 30.7 ml<br>Tubing – 68.33 ml | | | |
| Resistance to flow (pressure<br>drop) | Flow rate<br>(lpm) | Pressure drop (cm H2O) | | Flow rate<br>(lpm) | Pressure drop (cm H2O) | | Identical |
| | | iQ 2 | Phantom 2 | | iQ 2 | Phantom 2 | |
| | 50 | 0.45 | 0.45 | 50 | 0.45 | 0.45 | |
| | 100 | 2.05 | 2.06 | 100 | 2.05 | 2.06 | |
| Sound pressure and Sound<br>power level | iQ 2 –<br>Sound pressure - 30.32 dB<br>Sound power - 33.33 dB | | | iQ 2 –<br>Sound pressure - 30.32 dB<br>Sound power - 33.33 dB | | | Identical |
| | Phantom 2 –<br>Sound pressure – 30.49 dB Sound power –<br>33.50 dB | | | Phantom 2 –<br>Sound pressure – 30.49 dB Sound power –<br>33.50 dB | | | |
{10}------------------------------------------------
# K241469 510(k) Summary Page 7 of 7 15-Jun-24
# Substantial Equivalence Conclusion:
The iQ 2 Nasal Vented Mask and Phantom 2 Nasal Ventical indications, technological characteristics and principles of operation and performance to the predicate and performance testing demonstrates that the subject device is use the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.