← Product Code [JXG](/productcode/JXG) · K241458

# Borvo EVAC System (Ergo); Borvo EVAC System (Classic) (K241458)

_Borvo Medical, Inc. · JXG · Aug 20, 2024 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K241458

## Device Facts

- **Applicant:** Borvo Medical, Inc.
- **Product Code:** [JXG](/productcode/JXG.md)
- **Decision Date:** Aug 20, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5550
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The EVAC is indicated to facilitate the evacuation of a chronic or subacute hematoma or hygroma in the cranium. The EVAC is intended for draining of fluid accumulated in the subdural space, including chronic or subacute hematomas and hygromas. The EVAC is also intended for removing air and fluid from the subdural space following open surgical procedures to remove subdural collections.

## Device Story

Borvo EVAC System facilitates drainage of subdural fluid accumulations (chronic/subacute hematomas, hygromas) or post-surgical air/fluid. System comprises titanium port (Ergo or Classic configuration), tubing, and vacuum bulb reservoir. Operates by creating low negative pressure in subdural space to gradually drain fluid into external reservoir. Used in clinical settings by physicians; device provides passive drainage mechanism. Benefits include controlled evacuation of intracranial collections. Device is single-use and sterile.

## Clinical Evidence

Bench testing only. Testing included dimensional analysis, tensile strength, pull-out force, simulated use, sterile barrier/packaging integrity, shelf life, and biocompatibility per ISO 10993-1.

## Technological Characteristics

Components: Titanium port, thermoplastic elastomer tubing, Jackson-Pratt vacuum bulb. Dimensions: 5.31mm drill bit compatibility, 0.75mm thread pitch. Sensing/Actuation: Passive negative pressure drainage. Connectivity: None. Sterilization: Sterile, single-use. Biocompatibility: ISO 10993-1 compliant.

## Regulatory Identification

A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.

## Predicate Devices

- Medtronic SEPS Cranial Access Kit ([K042359](/device/K042359.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

August 20, 2024

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.

Borvo Medical, Inc. Ming Cheng Chew Regulatory Consultant 2500 Old Middlefield Way, Suite E Mountain View. California 94043

Re: K241458

Trade/Device Name: Borvo EVAC System (Ergo); Borvo EVAC System (Classic) Regulation Number: 21 CFR 882.5550 Regulation Name: Central Nervous System Fluid Shunt And Components Regulatory Class: Class II Product Code: JXG Dated: May 22, 2024 Received: May 23, 2024

Dear Ming Cheng Chew:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Adam D Pierce -S Digitally signed by Adam D. Pierce -S Date: 2024.08.20 14:37:29 -04'00'

Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality 

{2}------------------------------------------------

Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K241458

Device Name Borvo EVAC System (Ergo); Borvo EVAC System (Classic)

### Indications for Use (Describe)

The EVAC is indicated to facilitate the evacuation of a chronic or subacute hematoma or hygroma in the cranium. The EVAC is intended for draining of fluid accumulated in the subdural space, including chronic or subacute hematomas and hygromas. The EVAC is also intended for removing air and fluid from the subdural space following open surgical procedures to remove subdural collections.

| Type of Use (Select one or both, as applicable)                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                         |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg class="bi bi-check-square-fill" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2zm10.03 4.97a.75.75 0 0 0-1.08.022L7.477 9.417 5.384 7.323a.75.75 0 0 0-1.06 1.06L6.97 11.03a.75.75 0 0 0 1.079-.02l3.992-4.99a.75.75 0 0 0-.01-1.05z" fill-rule="evenodd"></path> </svg> </span>    Prescription Use (Part 21 CFR 801 Subpart D)   </div> | <div> <span> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M0 2a2 2 0 0 1 2-2h12a2 2 0 0 1 2 2v12a2 2 0 0 1-2 2H2a2 2 0 0 1-2-2V2zm1.5 0a.5.5 0 0 0-.5.5v11a.5.5 0 0 0 .5.5h11a.5.5 0 0 0 .5-.5V2.5a.5.5 0 0 0-.5-.5h-11z" fill-rule="evenodd"></path> </svg> </span>    Over-The-Counter Use (21 CFR 801 Subpart C)   </div> |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

#### 510(k) Summary: K241458 1

#### 1.1 Contact Information

## Sponsor/Manufacturer

Borvo Medical Inc. 2500 Old Middlefield Way, Suite E Mountain View, CA 94043 Phone: 551-697-8927 Email: sthompson@borvomedical.com

## Correspondent

Ming Cheng Chew Regulatory Consultant, ROM+ Phone: 763-232-3701 Email: mchew@rqmplus.com

Date: August 20, 2024

#### 1.2 Device Information

| Name of Device:      | Borvo EVAC System                                 |
|----------------------|---------------------------------------------------|
| Classification Name: | Central Nervous System Fluid Shunt and Components |
| Regulation Number:   | 21 CFR 882.5550                                   |
| Regulatory Class:    | II                                                |
| Product Code:        | JXG                                               |

#### 1.3 Predicate Device

Medtronic SEPS Cranial Access Kit (K042359)

#### 1.4 Device Description

The Borvo EVAC System is intended for the drainage of subdural fluid accumulations. The EVAC System is designed to gradually drain fluid to an external reservoir by creating a low negative pressure in the subdural space. It consists of 3 components:

- 1. EVAC Port (Ergo or Classic)
- 2. Tubing
- 3. Vacuum bulb

An optional Cranial Access Kit can be purchased with the EVAC System, or a commercially available Cranial Access Kit with a 5.31mm drill bit can be used.

{5}------------------------------------------------

#### INDICATIONS FOR USE 1.5

The EVAC is indicated to facilitate the evacuation of a chronic or subacute hematoma or hygroma in the cranium. The EVAC is intended for draining of fluid accumulated in the subdural space, including chronic or subacute hematomas and hygromas. The EVAC is also intended for removing air and fluid from the subdural space following open surgical procedures to remove subdural collections.

### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE 1.6 PREDICATE DEVICE

| Characteristic             | Borvo EVAC System                                                                                                                                                                                                                                                                                                                                                                                                                                   | Medtronic SEPS Cranial<br>Access Kit (K042349)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use               | Intended for the for<br>drainage of subdural fluid<br>accumulations.                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Indication for Use         | The EVAC is indicated to<br>facilitate the evacuation<br>of a chronic or subacute<br>hematoma or hygroma in<br>the cranium. The EVAC.<br>is intended for draining<br>of fluid accumulated in<br>the subdural space,<br>including chronic or<br>subacute hematomas and<br>hygromas. The EVAC is<br>also intended for<br>removing air and fluid<br>from the subdural space<br>following open surgical<br>procedures to remove<br>subdural collections | Use of the SEPS Cranial<br>Access Kit is indicated<br>when access to and<br>evacuation of a cranial<br>subacute or chronic<br>hematoma or hygroma is<br>necessary. The SEPS<br>Cranial Access Kit is<br>intended for drainage of<br>subdural fluid<br>accumulations such as<br>hygromas and chronic or<br>subacute hematomas to an<br>external suction reservoir.<br>The SEPS Cranial Access<br>Kit is also intended for<br>draining air and fluids from<br>the subdural space<br>immediately following<br>craniotomy procedures<br>performed to remove a<br>chronic or subacute<br>subdural hematoma. |
| Product Code               | JXG                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Regulation                 | 21 CFR 882.5550                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Components                 | Port                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Characteristic             | Borvo EVAC System                                                                                                                                                                                                                                                                                                                                                                                                                                   | Medtronic SEPS Cranial<br>Access Kit (K042349)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                            | Tubing<br>Vacuum Bulb<br>Cranial Access Kit                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Drill Bit<br>Compatibility | 5.31 mm                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Port Material              | Titanium                                                                                                                                                                                                                                                                                                                                                                                                                                            | Stainless Steel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Port ID Distal             | 0.150 inches                                                                                                                                                                                                                                                                                                                                                                                                                                        | 0.149 inches                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Port ID Proximal           | 0.160 inches                                                                                                                                                                                                                                                                                                                                                                                                                                        | 0.149 inches                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Port OD at threads         | 0.241 inches                                                                                                                                                                                                                                                                                                                                                                                                                                        | 0.236 inches                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Port OD at barbs           | 0.250 inches                                                                                                                                                                                                                                                                                                                                                                                                                                        | 0.236 inches                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Thread pitch               | 0.75 mm                                                                                                                                                                                                                                                                                                                                                                                                                                             | 0.75 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Port Wing<br>Configuration | Angular (Classic)<br>or Curvilinear (Ergo)                                                                                                                                                                                                                                                                                                                                                                                                          | Angular                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Tubing Material            | Thermoplastic Elastomer                                                                                                                                                                                                                                                                                                                                                                                                                             | Silicone                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Bulb Configuration         | Jackson-Pratt Reservoir                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Sterile                    | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Single-use only            | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Biocompatibility           | ISO 10993-1                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |

Table 1-1: Device Comparison to the Predicate Device

{6}------------------------------------------------

# K241458 Page 3 of 4

#### 1.7 PERFORMANCE DATA

A high-level summary of the testing conducted to confirm the safety and efficacy of the Borvo EVAC System is as follows:

- Dimensional .
- Tensile
- Pull-Out Force .
- . Simulated Use
- Sterile Barrier/Packaging Testing .
- . Shelf Life
- Biocompatibility .

{7}------------------------------------------------

K241458 Page 4 of 4

#### 1.8 SUBSTANTIAL EQUIVALENCE

Based on the same intended use, similar technological characteristics, and the summary of data submitted, the Borvo EVAC System is found to be substantially equivalent to the predicate device. Testing demonstrated the device safety and effectiveness as intended without raising new and different questions of safety and effectiveness.

---

**Source:** [https://fda-staging.innolitics.com/device/K241458](https://fda-staging.innolitics.com/device/K241458)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).