Borvo EVAC System (Ergo); Borvo EVAC System (Classic)
K241458 · Borvo Medical, Inc. · JXG · Aug 20, 2024 · Neurology
Device Facts
Record ID
K241458
Device Name
Borvo EVAC System (Ergo); Borvo EVAC System (Classic)
Applicant
Borvo Medical, Inc.
Product Code
JXG · Neurology
Decision Date
Aug 20, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5550
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EVAC is indicated to facilitate the evacuation of a chronic or subacute hematoma or hygroma in the cranium. The EVAC is intended for draining of fluid accumulated in the subdural space, including chronic or subacute hematomas and hygromas. The EVAC is also intended for removing air and fluid from the subdural space following open surgical procedures to remove subdural collections.
Device Story
Borvo EVAC System facilitates drainage of subdural fluid accumulations (chronic/subacute hematomas, hygromas) or post-surgical air/fluid. System comprises titanium port (Ergo or Classic configuration), tubing, and vacuum bulb reservoir. Operates by creating low negative pressure in subdural space to gradually drain fluid into external reservoir. Used in clinical settings by physicians; device provides passive drainage mechanism. Benefits include controlled evacuation of intracranial collections. Device is single-use and sterile.
Clinical Evidence
Bench testing only. Testing included dimensional analysis, tensile strength, pull-out force, simulated use, sterile barrier/packaging integrity, shelf life, and biocompatibility per ISO 10993-1.
Indicated for patients requiring evacuation of chronic or subacute cranial hematomas or hygromas, or removal of air/fluid from the subdural space post-craniotomy.
Regulatory Classification
Identification
A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.
{0}------------------------------------------------
August 20, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
Borvo Medical, Inc. Ming Cheng Chew Regulatory Consultant 2500 Old Middlefield Way, Suite E Mountain View. California 94043
Re: K241458
Trade/Device Name: Borvo EVAC System (Ergo); Borvo EVAC System (Classic) Regulation Number: 21 CFR 882.5550 Regulation Name: Central Nervous System Fluid Shunt And Components Regulatory Class: Class II Product Code: JXG Dated: May 22, 2024 Received: May 23, 2024
Dear Ming Cheng Chew:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam D Pierce -S Digitally signed by Adam D. Pierce -S Date: 2024.08.20 14:37:29 -04'00'
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality
{2}------------------------------------------------
Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241458
Device Name Borvo EVAC System (Ergo); Borvo EVAC System (Classic)
### Indications for Use (Describe)
The EVAC is indicated to facilitate the evacuation of a chronic or subacute hematoma or hygroma in the cranium. The EVAC is intended for draining of fluid accumulated in the subdural space, including chronic or subacute hematomas and hygromas. The EVAC is also intended for removing air and fluid from the subdural space following open surgical procedures to remove subdural collections.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg class="bi bi-check-square-fill" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2zm10.03 4.97a.75.75 0 0 0-1.08.022L7.477 9.417 5.384 7.323a.75.75 0 0 0-1.06 1.06L6.97 11.03a.75.75 0 0 0 1.079-.02l3.992-4.99a.75.75 0 0 0-.01-1.05z" fill-rule="evenodd"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M0 2a2 2 0 0 1 2-2h12a2 2 0 0 1 2 2v12a2 2 0 0 1-2 2H2a2 2 0 0 1-2-2V2zm1.5 0a.5.5 0 0 0-.5.5v11a.5.5 0 0 0 .5.5h11a.5.5 0 0 0 .5-.5V2.5a.5.5 0 0 0-.5-.5h-11z" fill-rule="evenodd"></path> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary: K241458 1
#### 1.1 Contact Information
## Sponsor/Manufacturer
Borvo Medical Inc. 2500 Old Middlefield Way, Suite E Mountain View, CA 94043 Phone: 551-697-8927 Email: sthompson@borvomedical.com
## Correspondent
Ming Cheng Chew Regulatory Consultant, ROM+ Phone: 763-232-3701 Email: mchew@rqmplus.com
Date: August 20, 2024
#### 1.2 Device Information
| Name of Device: | Borvo EVAC System |
|----------------------|---------------------------------------------------|
| Classification Name: | Central Nervous System Fluid Shunt and Components |
| Regulation Number: | 21 CFR 882.5550 |
| Regulatory Class: | II |
| Product Code: | JXG |
#### 1.3 Predicate Device
Medtronic SEPS Cranial Access Kit (K042359)
#### 1.4 Device Description
The Borvo EVAC System is intended for the drainage of subdural fluid accumulations. The EVAC System is designed to gradually drain fluid to an external reservoir by creating a low negative pressure in the subdural space. It consists of 3 components:
- 1. EVAC Port (Ergo or Classic)
- 2. Tubing
- 3. Vacuum bulb
An optional Cranial Access Kit can be purchased with the EVAC System, or a commercially available Cranial Access Kit with a 5.31mm drill bit can be used.
{5}------------------------------------------------
#### INDICATIONS FOR USE 1.5
The EVAC is indicated to facilitate the evacuation of a chronic or subacute hematoma or hygroma in the cranium. The EVAC is intended for draining of fluid accumulated in the subdural space, including chronic or subacute hematomas and hygromas. The EVAC is also intended for removing air and fluid from the subdural space following open surgical procedures to remove subdural collections.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE 1.6 PREDICATE DEVICE
| Characteristic | Borvo EVAC System | Medtronic SEPS Cranial<br>Access Kit (K042349) |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Intended for the for<br>drainage of subdural fluid<br>accumulations. | Same |
| Indication for Use | The EVAC is indicated to<br>facilitate the evacuation<br>of a chronic or subacute<br>hematoma or hygroma in<br>the cranium. The EVAC.<br>is intended for draining<br>of fluid accumulated in<br>the subdural space,<br>including chronic or<br>subacute hematomas and<br>hygromas. The EVAC is<br>also intended for<br>removing air and fluid<br>from the subdural space<br>following open surgical<br>procedures to remove<br>subdural collections | Use of the SEPS Cranial<br>Access Kit is indicated<br>when access to and<br>evacuation of a cranial<br>subacute or chronic<br>hematoma or hygroma is<br>necessary. The SEPS<br>Cranial Access Kit is<br>intended for drainage of<br>subdural fluid<br>accumulations such as<br>hygromas and chronic or<br>subacute hematomas to an<br>external suction reservoir.<br>The SEPS Cranial Access<br>Kit is also intended for<br>draining air and fluids from<br>the subdural space<br>immediately following<br>craniotomy procedures<br>performed to remove a<br>chronic or subacute<br>subdural hematoma. |
| Product Code | JXG | Same |
| Regulation | 21 CFR 882.5550 | Same |
| Components | Port | Same |
| Characteristic | Borvo EVAC System | Medtronic SEPS Cranial<br>Access Kit (K042349) |
| | Tubing<br>Vacuum Bulb<br>Cranial Access Kit | |
| Drill Bit<br>Compatibility | 5.31 mm | Same |
| Port Material | Titanium | Stainless Steel |
| Port ID Distal | 0.150 inches | 0.149 inches |
| Port ID Proximal | 0.160 inches | 0.149 inches |
| Port OD at threads | 0.241 inches | 0.236 inches |
| Port OD at barbs | 0.250 inches | 0.236 inches |
| Thread pitch | 0.75 mm | 0.75 mm |
| Port Wing<br>Configuration | Angular (Classic)<br>or Curvilinear (Ergo) | Angular |
| Tubing Material | Thermoplastic Elastomer | Silicone |
| Bulb Configuration | Jackson-Pratt Reservoir | Same |
| Sterile | Yes | Same |
| Single-use only | Yes | Same |
| Biocompatibility | ISO 10993-1 | Same |
Table 1-1: Device Comparison to the Predicate Device
{6}------------------------------------------------
# K241458 Page 3 of 4
#### 1.7 PERFORMANCE DATA
A high-level summary of the testing conducted to confirm the safety and efficacy of the Borvo EVAC System is as follows:
- Dimensional .
- Tensile
- Pull-Out Force .
- . Simulated Use
- Sterile Barrier/Packaging Testing .
- . Shelf Life
- Biocompatibility .
{7}------------------------------------------------
K241458 Page 4 of 4
#### 1.8 SUBSTANTIAL EQUIVALENCE
Based on the same intended use, similar technological characteristics, and the summary of data submitted, the Borvo EVAC System is found to be substantially equivalent to the predicate device. Testing demonstrated the device safety and effectiveness as intended without raising new and different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.