Q-FIX Knotless All-Suture Anchor, 1.8 mm, with one ULTRABRAID suture, blue; Q-FIX Knotless All-Suture Anchor, 1.8 mm, with one ULTRABRAID suture, co-braid-blue; Q-FIX Knotless All-Suture Anchor, 1.8 mm, with one MINITAPE suture, blue; Q-FIX Knotless All-Suture Anchor, 1.8 mm, with one MINITAPE suture, co-braid-blue
K241435 · Smith & Nephew, Inc. · MBI · Jun 20, 2024 · Orthopedic
Device Facts
Record ID
K241435
Device Name
Q-FIX Knotless All-Suture Anchor, 1.8 mm, with one ULTRABRAID suture, blue; Q-FIX Knotless All-Suture Anchor, 1.8 mm, with one ULTRABRAID suture, co-braid-blue; Q-FIX Knotless All-Suture Anchor, 1.8 mm, with one MINITAPE suture, blue; Q-FIX Knotless All-Suture Anchor, 1.8 mm, with one MINITAPE suture, co-braid-blue
Applicant
Smith & Nephew, Inc.
Product Code
MBI · Orthopedic
Decision Date
Jun 20, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Q-FIX® Knotless All-Suture Anchor is intended for use during surgical procedures for the fixation of soft tissue to bone.
Device Story
The Q-FIX Knotless All-Suture Anchor is a sterile, single-use orthopedic fixation device used in arthroscopic or mini-open surgical procedures. It consists of a braided polyester all-suture anchor preloaded with a transfer suture and either ULTRABRAID or MINITAPE (UHMWPE) repair sutures, mounted on an insertion device. The device is inserted into a predrilled bone hole; the anchor utilizes an expandable compression fit mechanism for bone fixation. A knotless suture locking mechanism allows for adjustable one-way tensioning of the soft tissue against the bone. Operated by orthopedic surgeons, the device provides secure fixation for various ligament and tendon repairs. The output is the mechanical stabilization of soft tissue to bone, which facilitates healing and restores joint function. The device is MR safe.
Clinical Evidence
Bench testing only. Testing included insertion testing, cyclic loading, load to pass, and knot tensile strength testing. All tests met acceptance criteria. Biocompatibility was evaluated per ISO 10993-1 and deemed safe.
Indicated for soft tissue to bone fixation in shoulder (Bankart, SLAP, AC repair, capsular shift/reconstruction, deltoid repair, rotator cuff repair, biceps tenodesis), foot & ankle (medial/lateral repair, midfoot/forefoot repair, hallux valgus, metatarsal ligament/tendon repair, Achilles repair), elbow (ulnar/radial collateral ligament reconstruction, lateral epicondylitis, biceps reattachment), knee (extra-capsular repair of MCL/LCL/posterior oblique ligament, IBT, patellar tendon repair, VMO advancement, joint capsule closure), and hip (acetabular labral repair).
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
June 20, 2024
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Smith & Nephew, Inc. Rebecca Peterkins Regulatory Affairs Specialist II 150 Minuteman Road Andover, Massachusetts 01810
## Re: K241435
Trade/Device Name: Q-FIX◊ Knotless All-Suture Anchor, 1.8 mm, with one ULTRABRAID0 suture, blue; Q-FIX Knotless All-Suture Anchor, 1.8 mm, with one ULTRABRAID0 suture, co-braid-blue; Q-FIX > Knotless All-Suture Anchor, 1.8 mm, with one MINITAPE suture, blue; Q-FIX Knotless All-Suture Anchor, 1.8 mm, with one MINITAPE◊ suture, co-braid-blue Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: May 21, 2024 Received: May 21, 2024
Dear Ms. Peterkins:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Robert M. Stefani Digitally signed by Robert M. Stefani -S -ટ Date: 2024.06.20 13:47:43 -04'00"
For: Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K241435
Device Name
Q-FIX® Knotless All-Suture Anchor, 1.8 mm, with one ULTRABRAIDV suture, blue;
Q-FIX() Knotless All-Suture Anchor, 1.8 mm, with one ULTRABRAID( suture, co-braid-blue;
Q-FIX() Knotless All-Suture Anchor, 1.8 mm, with one MINITAPE() suture, blue;
Q-FIX() Knotless All-Suture Anchor, 1.8 mm, with one MINITAPE( suture, co-braid-blue
Indications for Use (Describe)
The Q-FIX\ Knotless All-Suture Anchor is intended for use for soft tissue to bone fixation for the following indications:
Shoulder:
- · Bankart lesion repair
- · SLAP lesion repair
- · Acromio-clavicular repair
- · Capsular shift/capsulolabral reconstruction
- · Deltoid repair
- · Rotator cuff tear repair
- · Biceps tenodesis
Foot & Ankle:
- · Medial/Lateral repair and reconstruction
- · Midfoot and forefoot repair
- · Hallux valgus reconstruction
- · Metatarsal ligament/tendon repair or reconstruction
- · Achilles tendon repair
Elbow:
- · Ulnar or radial collateral ligament reconstruction
- · Lateral epicondylitis repair
- · Biceps tendon reattachment
Knee:
· Extra-capsular repair: medial collateral ligament (MCL), lateral collateral ligament (LCL) and posterior oblique ligament
- · Iliotibial band tenodesis (IBT)
- · Patellar tendon repair
- · Vastus medialis obliquus advancement (VMO)
- · Joint capsule closure
Hip:
## · Acetabular labral repair
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
T:+ 1 978 749 1000 T:+ 1 800 343 8386 (USA toll free) www.smith-nephew.com
# Smith Nephew
K241435 Page 1 of 5
# 510(k) Summary
#### Prepared: 20 June 2024
| Submitter Information | Contact Information |
|-----------------------|------------------------------------|
| Smith & Nephew, Inc. | Ms. Rebecca Peterkins |
| Endoscopy Division | Regulatory Affairs Specialist II |
| 150 Minuteman Road | rebecca.peterkins@smith-nephew.com |
| Andover, MA 01810 | (978) 749-1057 |
| Device Name & Classification | |
|------------------------------|----------------------------------------------------------------------------------------|
| Proprietary Name | Q-FIX® Knotless All-Suture Anchor, 1.8mm, with one ULTRABRAID suture,<br>blue |
| | Q-FIX® Knotless All-Suture Anchor, 1.8mm, with one ULTRABRAID suture,<br>co-braid-blue |
| | Q-FIX® Knotless All-Suture Anchor, 1.8mm, with one MINITAPE suture,<br>blue |
| | Q-FIX® Knotless All-Suture Anchor, 1.8mm, with one MINITAPE suture,<br>co-braid-blue |
| Common Name | Soft Tissue Fixation Device |
| Classification Name | Fastener, Fixation, Nondegradable, Soft Tissue |
| Classification Regulation | 21 CFR 888.3040 |
| Class | II |
| Product Code(s) | MBI |
| Panel | Orthopedic |
#### Legally Marketed Predicate Device
The Smith & Nephew Q-FIX® Knotless All-Suture Anchor is substantially equivalent in intended use and fundamental scientific technology to the following legally marketed devices in commercial distribution:
| Description | Submission Number | Clearance Date |
|-----------------------|-------------------|----------------|
| Q-FIX ◊ Suture Anchor | K172165 | 17AUG2017 |
#### Legally Marketed Reference Devices
| Description | Submission Number | Clearance Date |
|---------------------|-------------------|----------------|
| MINITAPE ◊ Suture | K132357 | 30OCT2013 |
| ULTRABRAID ◊ Suture | K101377 | 08APR2011 |
{6}------------------------------------------------
#### Device Description
The Smith & Nephew Q-FIX® Knotless All-Suture Anchor is a fixation device intended to provide fixation of soft tissue to bone. The device consists of an all-suture anchor preloaded with one (1) transfer suture and either one (1) ULTRABRAID® or MINITAPE® suture assembled inside an insertion device. The device is supplied sterile, for single use only.
The 1.8mm implant (all-suture anchor) is made of braided polyester with either one strand of ULTRABRAID® or MINITAPE® suture and one transfer suture pre-loaded. The ULTRABRAID®, MINITAPE®, and transfer sutures are non-absorbable sutures of ultra-high molecular weight polyethylene (UHMWPE). The Q-FIX® Knotless All-Suture Anchor has a knotless suture locking mechanism that provides adjustable one-way tensioning.
#### Intended Use
The Q-FIX® Knotless All-Suture Anchor is intended for use during surgical procedures for the fixation of soft tissue to bone.
#### Indications for Use
The Q-FIX® Knotless All-Suture Anchor is intended for use for soft tissue to bone fixation for the following indications:
#### Shoulder:
- . Bankart lesion repair
- SLAP lesion repair .
- . Acromio-clavicular repair
- Capsular shift/capsulolabral reconstruction .
- Deltoid repair ●
- . Rotator cuff tear repair
- Biceps tenodesis .
#### Foot & Ankle:
- Medial/Lateral repair and reconstruction ●
- Midfoot and forefoot repair .
- Hallux valgus reconstruction ●
- Metatarsal ligament/tendon repair or reconstruction .
- Achilles tendon repair ●
#### Elbow:
- Ulnar or radial collateral ligament reconstruction ●
- Lateral epicondylitis repair
- Biceps tendon reattachment ●
#### Knee:
- Extra-capsular repair: medial collateral ligament (MCL), lateral collateral ligament (LCL) and . posterior oblique ligament
- Iliotibial band tenodesis (IBT) ●
- Patellar tendon repair ●
- Vastus medialis obliquus advancement (VMO) ●
- Joint capsule closure
#### Hip:
- Acetabular labral repair
{7}------------------------------------------------
## Technological Characteristics
| | Q-FIX® Knotless All-Suture Anchor<br>(Subject Device) | Predicate Q-FIX®<br>Suture Anchor<br>(K172165) |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| Intended Use | Soft tissue to bone fixation | Same |
| Indications for Use | Shoulder:<br>• Bankart lesion repair<br>• SLAP lesion repair<br>• Acromio-clavicular repair<br>• Capsular shift/capsulolabral reconstruction<br>• Deltoid repair<br>• Rotator cuff tear repair<br>• Biceps tenodesis<br>Foot & Ankle:<br>• Medial/Lateral repair and reconstruction<br>• Midfoot and forefoot repair<br>• Hallux valgus reconstruction<br>• Metatarsal ligament/tendon repair or reconstruction<br>• Achilles tendon repair<br>Elbow:<br>• Ulnar or radial collateral ligament reconstruction<br>• Lateral epicondylitis repair<br>• Biceps tendon reattachment<br>Knee:<br>• Extra-capsular repair: medial collateral ligament<br>(MCL), lateral collateral ligament (LCL) and posterior<br>oblique ligament<br>• Iliotibial band tenodesis (IBT)<br>• Patellar tendon repair<br>• Vastus medialis obliquus advancement (VMO)<br>• Joint capsule closure<br>Hip:<br>• Acetabular labral repair | Subset of predicate device |
| Delivery Method | Arthroscopic and mini open | Same |
| How Supplied | Packaged in thermoform tray with Tyvek lid, Sterile<br>(EtO), Single Use | Same |
| Shelf Life | 3 years | Same |
| MR Compatibility | MR Safe | Same |
| Anchor Material | Polyester (PET) | Same |
| Suture Material | UHMWPE (ultra-high molecular weight polyethylene) | Same |
| Inserter Shaft<br>Materials | Stainless Steel | Same |
| Inserter Handle<br>Materials | Polycarbonate, ABS, Nylon | Same |
| Method of Anchor<br>Insertion | Inserted into a predrilled hole | Same |
| Inserter Shaft<br>Working Length | 8.60 in | Same as Q-FIX® 1.8mm<br>Suture Anchor |
| Bone Locking<br>Mechanism | Expandable Compression Fit | Same |
| Suture Locking<br>Mechanism | Knotless suture mechanism | Manually tied suture knot |
| Size Offered<br>(Diameter) | 1.8mm | Same as Q-FIX® 1.8mm<br>Suture Anchor |
| Suture Size (USP) | #2 or MINITAPE (#2) | Same |
| Number of Suture<br>Legs | 3 | Within range of predicate<br>device |
| Bone Hole OD | 2.1 mm | Same as Q-FIX® 1.8mm<br>Suture Anchor |
| Bone Hole Depth | 22.3 mm | Same as Q-FIX® 1.8mm<br>Suture Anchor |
| Accessories | Nonsterile, reusable and sterile, disposable drills, drill guides, and obturators | Subset of predicate device |
{8}------------------------------------------------
Q-FIX® Knotless All-Suture Anchor is an iterative design of the legally marketed Q-FIX® Suture Anchor (K172165). The modifications to the legally marketed predicate device Q-FIX® Suture Anchor (K172165) include adding one (1) transfer and either one (1) ULTRABRAID® or MINITAPEº repair suture to facilitate a knotless repair. Q-FIX® Knotless All-Suture Anchor is similar in its intended use, indications for use, sterilization method, materials, packaging configuration, and design. The ULTRABRAID® and MINITAPE® sutures in the subject device are similar to the reference devices cleared under K132357 & K101377. Additionally, many of the subject device components are identical compared to the predicate. The differences between the subject device and predicate device are minor and raise no new questions of safety of effectiveness.
#### Performance Data
Non-clinical bench testing was completed on the subject device, and the device met all required specifications for each test. Testing included insertion testing, cyclic loading testing, load to pass testing, and knot tensile strength testing. A summary of test acceptance criteria and results have been provided. Results for all tests passed.
The biocompatibility of Q-FIX® Knotless All-Suture Anchor was evaluated against the requirements per ISO 10993-1 and was deemed biologically safe.
Based on similarities to the subject device repair sutures, reference devices, ULTRABRAID® and MINITAPE, were included to support product performance and biocompatibility.
{9}------------------------------------------------
#### Conclusion
The substantial equivalence of the Q-FIX® Knotless All-Suture Anchor is based on similarities in indications for use, design features, operational principles, material biocompatibility and composition, and performance to the predicate/reference devices listed above. Modifications in design are minor and do not raise additional questions of safety or effectiveness compared to the predicate device. Based on the similarities, Q-FIX® Knotless All-Suture Anchor is substantially equivalent to its predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.