K241433 · Weero Co., Ltd. · IPF · Jun 4, 2025 · Physical Medicine
Device Facts
Record ID
K241433
Device Name
EVE Synergy (EVE-20M)
Applicant
Weero Co., Ltd.
Product Code
IPF · Physical Medicine
Decision Date
Jun 4, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
EMS is used for: - Relaxation of muscle spasms - Prevention or retardation of disuse atrophy - Increasing local blood circulation - Muscle re-education - Maintaining or increasing range of motion - Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis TENS is used for: - Symptomatic relief and management of chronic, intractable pain - Post-surgical acute pain - Post-trauma acute pain
Device Story
EVE Synergy (EVE-20M) is a powered muscle stimulator and TENS device for clinical use. It accepts user inputs via an LCD touch screen to control electrical stimulation parameters. The device delivers EMS and TENS signals through handpieces (Synergy and Pro models) with various tips (Ball, T, COMB). Synergy mode includes an integrated temperature sensor that automatically terminates output if skin temperature exceeds 43°C. The device operates via an AC/DC power supply. Healthcare providers use the device to manage pain, prevent atrophy, and facilitate muscle re-education. Output affects clinical decision-making by providing non-pharmacological pain management and physical therapy support. Benefits include improved muscle function, pain relief, and prevention of venous thrombosis in post-surgical patients.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and performance testing for nerve/muscle stimulators (IEC 60601-2-10).
Technological Characteristics
Materials: Stainless Steel 304 (electrodes), Polycarbonate (display). Energy: 24V DC power adapter. Output: Symmetrical biphasic (EMS/TENS) and pulsed monophasic (TENS). Connectivity: Standalone console with wired handpieces. Safety: Automatic overload/no-load trip, temperature sensor (43°C limit). Standards: IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2, ISO 14971, IEC 62304.
Indications for Use
Indicated for patients requiring muscle stimulation for spasm relaxation, disuse atrophy prevention, blood circulation improvement, muscle re-education, range of motion maintenance, or post-surgical calf stimulation; also indicated for symptomatic relief of chronic intractable pain, post-surgical acute pain, and post-trauma acute pain.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 4, 2025
WEERO Co., Ltd.
Moon Young Han
Regulatory Affairs
A-605, Venture Valley II, 142-10, Saneop-Ro 156 Beon-Gil,
Gwonseon-Gu Suwon-Si, Gyeonggi-Do
Suwon, Gyeonggi 16648
Korea, South
Re: K241433
Trade/Device Name: EVE Synergy (EVE-20M)
Regulation Number: 21 CFR 890.5850
Regulation Name: Powered Muscle Stimulator
Regulatory Class: Class II
Product Code: IPF, GZJ
Dated: May 17, 2024
Received: May 21, 2024
Dear Moon Young Han:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K241433 - Moon Young Han
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K241433 - Moon Young Han
Page 3
Sincerely,
Jitendra V. Virani -S
CDR Jitendra Virani, MS, MBA
Assistant Director
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K241433
Device Name
EVE Synergy (EVE-20M)
Indications for Use (Describe)
EMS is used for:
- Relaxation of muscle spasms
- Prevention or retardation of disuse atrophy
- Increasing local blood circulation
- Muscle re-education
- Maintaining or increasing range of motion
- Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis
TENS is used for:
- Symptomatic relief and management of chronic, intractable pain
- Post-surgical acute pain
- Post-trauma acute pain
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
{4}
EVE Synergy
510(k) Summary_K241433
# 1. Submitter and US Official Correspondent
Submitter: WEERO Co.,Ltd.
Address: A205 VentureValley II, 142-10, Saneop-ro 156beon-gil, Gwonseon-gu, Suwon-si, Gyeonggi-do, Republic of Korea
Telephone No.: +82-31-427-0787
Official Correspondent: Moon-young Han
Correspondent: WEERO Co.,Ltd.
Address: A205 VentureValley II, 142-10, Saneop-ro 156beon-gil, Gwonseon-gu, Suwon-si, Gyeonggi-do, Republic of Korea
Telephone No.: +82-31-427-0787
Email: ra@weeroweero.com
Date prepared: May 26, 2025
# 2. Device Information
Trade/Device Name: EVE Synergy / EVE-20M
Regulation Name:
- Powered Muscle Stimulator
- Transcutaneous electrical nerve stimulator for pain relief
Classification Name:
- Stimulator, Muscle, Powered
- Stimulator, Nerve, Transcutaneous, For Pain Relief
Product Code: IPF, GZJ
Regulatory Class: Class II
Regulation Number: 21CFR890.5850, and 21CFR882.5890
# 3. Predicate Device(Equivalent Legally Marketed Device)
| Manufacturer | Device | 510(k) No. |
| --- | --- | --- |
| Primary Predicate Device | | |
| INMODE LTD. | EVOLVE System with the T3 Applicator | K210877 |
| Secondary Predicate Device | | |
| INMODE LTD. | EVOLVE System with the Tone Applicator | K201285 |
# 4. Description of the Device
EVE Synergy has Synergy handpiece(1EA), Pro handpiece(2EA), Micro cable, Tip(Ball type, T type, COMB type) and Power adaptor & cable.
EVE Synergy has Synergy mode and Pro mode.
Synergy mode has radio frequency and low frequency functions using a synergy handpiece. The Synergy handpiece has a temperature sensor, so the output is automatically turned off when the skin surface temperature exceeds $43^{\circ}\mathrm{C}$ .
Pro mode has low frequency function that uses the Pro handpiece with tip.
It should be noted that the device has radio frequency (RF Mode) indicated for relief of minor muscle aches and pain, relief of muscle spasm, temporary improvement of local blood circulation. This RF mode was previously reviewed by the FDA under K212253 and
{5}
EVE Synergy
510(k) Summary_K241433
was cleared under regulation 21 CFR 878.4400 (Product code PBX). In the subject submission, there are no changes being proposed related to the RF Mode, and, hence, the review of the RF Mode is not within the scope of the subject 510(k) submission.
## 5. Indications for use (intended use)
EMS is used for:
- Relaxation of muscle spasms
- Prevention or retardation of disuse atrophy
- Increasing local blood circulation
- Muscle re-education
- Maintaining or increasing range of motion
- Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis
TENS is used for:
- Symptomatic relief and management of chronic, intractable pain
- Post-surgical acute pain
- Post-trauma acute pain
Page 2 / 13
{6}
EVE Synergy
510(k) Summary_K241433
## 6. Substantial Equivalence Discussion
### 1) Comparison Information
| Name | Subject device | Primary Predicate Device | Secondary Predicate device | Comparison |
| --- | --- | --- | --- | --- |
| Device Name | EVE Synergy(EVE-20M) | EVOLVE System with the T3 Applicator | EVOLVE System with the Tone Applicator | |
| Manufacturer | WEERO Co.,Ltd. | InMode Ltd. | InMode Ltd. | |
| 510(k) No. | K241433 | K210877 | K201285 | |
| Product Code, Class | IPF, GZJ
Class II | IPF, GZJ
Class II | IPF, GZJ
Class II | Identical |
| Indications for use | EMS is used for:
• Relaxation of muscle spasms
• Prevention or retardation of disuse atrophy
• Increasing local blood circulation
• Muscle re-education
• Maintaining or increasing range of motion
• Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
TENS is used for:
• Symptomatic relief and management of chronic, intractable pain
• Post-surgical acute pain
• Post-trauma acute pain | EMS mode is intended for:
• Relaxation of muscle spasms
• Prevention or retardation of disuse atrophy
• Increasing local blood circulation
• Muscle re-education
• Maintaining or increasing range of motion
• Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
TENS mode is intended for:
• Symptomatic relief and management of chronic, intractable pain
• Post-surgical acute pain
• Post-trauma acute pain | The EVOLVE System with the Tone Applicator is an electro-muscle and transcutaneous nerve stimulation device for the treatment of different body areas.
The EVOLVE System with Tone Applicator is designed to operate in two modes – EMS and TENS.
In EMS mode it is used for:
• Relaxation of muscle spasms
• Prevention or retardation of disuse atrophy
• Increasing local blood circulation
• Muscle re-education
• Maintaining or increasing range of motion
• Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
And in TENS mode is intended for
• Symptomatic relief and management of chronic, intractable pain
• Post-surgical acute pain
• Post-trauma acute pain | Identical |
Page 3 / 13
{7}
EVE Synergy
510(k) Summary_K241433
| Name | Subject device | Primary Predicate Device | Secondary Predicate device | Comparison |
| --- | --- | --- | --- | --- |
| Device Name | EVE Synergy(EVE-20M) | EVOLVE System with the T3 Applicator | EVOLVE System with the Tone Applicator | |
| Manufacturer | WEERO Co.,Ltd. | InMode Ltd. | InMode Ltd. | |
| 510(k) No. | K241433 | K210877 | K201285 | |
| Design | The EVE Synergy consists of an AC/DC power supply unit, controller, and user interface including an LCD touch screen. The delivery of the electrical energy is controlled by a Start/Stop button positioned on the front panel.
The System supports the following components:
• LCD display touch screen
• Audio loudspeaker
• 24V AC/DC power supply
• Controller | The EVOLVE System with T3 Applicator consists of an AC/DC power supply unit, Controller and user interface including an LCD touch screen. The delivery of the RF/electrical energy is controlled by a Start/Stop button positioned on the front panel.
The System supports the following components:
• LCD display touch screen
• Audio loudspeaker
• 48V AC/DC power Supply
• Controller | The EVOLVE System with Tone Applicator consists of an AC/DC power supply unit, controller, and user interface including an LCD touch screen. The delivery of the electrical energy is controlled by a Start/Stop button positioned on the front panel.
The System supports the following components:
• LCD display touch screen
• Audio loudspeaker
• 48V AC/DC power supply
• Controller | Not identical.
The subject device was tested and is compliant with IEC60601-1 and IEC 60601-2-10. Therefore the difference doesn't impact essential performance, basic safety or substantial equivalence. |
| - Components Console | The EVE Synergy consists of the following components:
• Power supply unit, Console, including controller and user interface including an LCD touch screen.
• Handpieces with up to 3 units connected to the console via 3 designated cables and 3 designated connection ports. | The EVOLVE System consists of the following components:
• Console, including a power supply unit, controller and user interface including an LCD touch screen.
• T3 Applicator with up to 6 units connected to the console via 6 designated cables and 6 designated connection ports. | The EVOLVE System consists of the following components:
• Console, including a power supply unit, controller and user interface including an LCD touch screen.
• Tone applicator with up to 4 units connected to the console via 4 designated cables and 4 designated connection ports. | Not identical.
Although the number of connection ports is different from the predicate device, both the subject and predicate devices comply with IEC 60601-1 and IEC 60601-2-10 requirements and the difference doesn't impact essential performance, basic safety or substantial equivalence. |
| Dimension Console [W x H x D]
Applicator unit [L x D] | 22.4cm W x 41.8cm D x 32.5cm H [8.8” W x 16.5” D x 12.8” H]
Synergy handpiece:
167.5mm L x 43 mm D [6.6” L x 1.7” D]
Pro handpiece:
136mm L x 27.5mm D [5.4” L x 1.1” D]
Ball type tip:
86mm L x 18mm D [3.4” L x 0.7” D]
T type tip:
80mm L x 50mm D [3.1” L x 2.0” D]
COMB type tip:
88mm L x 30mm D [3.5” L x 1.2” D] | 46cm W x 46cm D x 100cm H [18.2” W x 18.2” D x 44” H]
T3 Applicator unit:
67.3mm L x 54.3mm D
[2.7” L x2.2” D] | 46cm W x 46cm D x 100cm H [18.2” W x 18.2” D x 44” H]
Tone Applicator:
12cm L x 10cm D
[4.7” L x 4” D] | Not identical.
Although the dimension of the subject device is different from the predicate device, both the subject and predicate devices comply with IEC 60601-1 and IEC 60601-2-10 requirements, which ensures that they meet the necessary safety and performance standards. The difference in dimension does not affect essential performance, basic safety, or substantial equivalence. |
Page 4 / 13
{8}
EVE Synergy
510(k) Summary_K241433
| Name | Subject device | Primary Predicate Device | Secondary Predicate device | Comparison |
| --- | --- | --- | --- | --- |
| Device Name | EVE Synergy(EVE-20M) | EVOLVE System with the T3 Applicator | EVOLVE System with the Tone Applicator | |
| Manufacturer | WEERO Co.,Ltd. | InMode Ltd. | InMode Ltd. | |
| 510(k) No. | K241433 | K210877 | K201285 | |
| Weight Console
Weight applicator | 7 kg [15.4 lbs.]
Synergy handpiece: 0.12 kg [0.26 lbs.]
Pro handpiece: 0.06 kg [0.13 lbs.]
Ball type tip: 0.04 kg [0.08 lbs.]
T type tip: 0.02 kg [0.04 lbs.]
COMB type tip: 0.02 kg [0.04 lbs.] | 33.0 Kg [73 lbs.]
T3: 0.16 Kg [0.4 lbs.] | 33.0 Kg [73 lbs.]
Tone: 0.22 Kg [0.5 lbs.] | Not identical.
Although the weight of the subject device is different from the predicate device, both the subject and predicate devices comply with IEC 60601-1 and IEC 60601-2-10 requirements, which ensures that they meet the necessary safety and performance standards. The difference in weight does not affect essential performance, basic safety, or substantial equivalence. |
| Applicator unit treatment area | Synergy handpiece electrode(2ea): 6 cm²
Ball type tip: 10.17 cm²
T type tip: 7.85 cm²
COMB type tip: 28.83 cm² | 6.46 cm² | 12 cm² | Not identical.
However, the difference doesn't raise new questions of safety and effectiveness. |
| Power Source(s) | Power Supply Adapter
- Input: 100-240Vac, 50/60Hz, 1.4-0.7A
- Output: 24Vdc, 5A | Main Line Frequency (nominal)
50-60Hz
Input Voltage (nominal)
100-240VAC
Input Current (rms)
4A | Main Line Frequency (nominal)
50-60Hz
Input Voltage (nominal)
100-240VAC
Input Current (rms)
4A | Not identical.
The subject device was tested and is compliant with IEC60601-1 and IEC 60601-2-10. Therefore the difference doesn't impact essential performance, basic safety or substantial equivalence. |
| Method of Line Current Isolation | Independent transformer isolated | Independent transformer isolated | Independent transformer isolated | Identical |
| Electrical Type | Type BF | Type BF | Type BF | Identical |
| Patient Leakage Current – Normal Condition (μA) | <100uA patient leakage | <100uA patient leakage | <100uA patient leakage | Identical |
| Patient Leakage Current – Single Fault Condition (μA) | <300uA line leakage | <300uA line leakage | <300uA line leakage | Identical |
| Number of Output Modes | 2 | Unknown | 2 | Identical |
| Number of Output Channels | 3 | 6 | 2 | Not identical.
However, the difference doesn't raise new questions of safety and effectiveness. |
Page 5 / 13
{9}
EVE Synergy
510(k) Summary_K241433
| Name | Subject device | Primary Predicate Device | Secondary Predicate device | Comparison |
| --- | --- | --- | --- | --- |
| Device Name | EVE Synergy(EVE-20M) | EVOLVE System with the T3 Applicator | EVOLVE System with the Tone Applicator | |
| Manufacturer | WEERO Co.,Ltd. | InMode Ltd. | InMode Ltd. | |
| 510(k) No. | K241433 | K210877 | K201285 | |
| Synchronous or Alternating | See Output Specifications Below | See Output Specifications Below | See Output Specifications Below | Identical |
| Method of Channel Isolation | Through transformers and isolators | Through transformers and isolators | Through transformers and isolators | Identical |
| Regulated Current or Regulated Voltage (output signals only) | Regulated voltage on all channels With current limit | Regulated voltage on all channels With current limit | Regulated voltage on all channels With current limit | Identical |
| Software/Firmware / Microprocessor Control | Yes | Yes | Yes | Identical |
| Automatic Overload Trip | Yes | Yes | Yes | Identical |
| Automatic No-Load Trip | Yes | Yes | Yes | Identical |
| Automatic Shut Off | Yes | Yes | Yes | Identical |
| Patient Override Control | No | Yes | Yes | Not identical. However, the difference doesn't raise new questions of safety and effectiveness. |
| - On/Off Status | Yes | Yes | Yes | Identical |
| - Battery | No battery | No battery | No battery | Identical |
| -Voltage/Current level | Yes | Yes | Yes | Identical |
| Timer Range | 0-60 [minutes] | 0-60 [minutes] | 0-60 [minutes] | Identical |
| Compliance with 21 CFR 890.5850 (IPF) | Yes | Yes | Yes | Identical |
| Compliance with 21 CFR 882.5890 (GZJ) | Yes | Yes | Yes | Identical |
Page 6 / 13
{10}
EVE Synergy
510(k) Summary_K241433
| Name | Subject device | Primary Predicate Device | Secondary Predicate device | Comparison |
| --- | --- | --- | --- | --- |
| Device Name | EVE Synergy(EVE-20M) | EVOLVE System with the T3 Applicator | EVOLVE System with the Tone Applicator | |
| Manufacturer | WEERO Co.,Ltd. | InMode Ltd. | InMode Ltd. | |
| 510(k) No. | K241433 | K210877 | K201285 | |
| EMS Output | | | | |
| Output Specifications: Waveform | Symmetrical Biphasic Waveform | Symmetrical Biphasic Waveform | Symmetrical Biphasic Waveform | Identical |
| Pulse Shape | Rectangular | Rectangular | Rectangular | Identical |
| Maximum Output Voltage (± 20%) | 34 V @500Ω
48 V @2kΩ
48 V @10kΩ | 30 V @500Ω
54 V @2kΩ
54 V @10kΩ | 56 V @500Ω
56 V @2kΩ
56 V @10kΩ | Not identical.
Since the maximum output voltage & current of the subject device are similar to the predicate devices, they do not raise new questions about safety and effectiveness. |
| Maximum Output Current (± 20%) | 68 mA @500Ω
24 mA @2kΩ
4.8 mA @10kΩ | 60 mA @500Ω
27 mA@2kΩ
5.4 mA @10kΩ | 92.86 mA @500Ω
26.7 mA@2kΩ
5.4 mA @10kΩ | |
| Pulse Width(μsec.) – The output active positive pulse width | 100 to 155 [μs] | 20 to 400 [μs] | 20 to 400 [μs] | Not identical.
Since the pulse width of the subject device is within the range of the predicate devices, it does not raise new questions about safety and effectiveness. |
| Frequency (Hz) | 3 to 200 [Hz] | 3 to 200 [Hz] | 3 to 200 [Hz] | Identical |
| Net Charge @ 500 Ohms [μC/pulse] | 0 [μC] @ 500Ω | 0 [μC] @ 500Ω | 0 [μC] @ 500Ω | Identical |
| Maximum Phase Charge [μC] | 10.54 [μC] @500Ω | 24 [μC] @500Ω | 43.2 [μC] @ 500Ω | Not identical.
Maximum Phase Charge is related to the maximum output current and pulse width, and because the predicate device has various pulse widths, the range of phase charge depends on the pulse width range. For example, for the primary predicate device, the phase charge ranges from 1.2 μC to 24 μC. Because the phase charge of the subject device is within the range of the predicate device, it does not raise new questions about safety and effectiveness. |
Page 7 / 13
{11}
EVE Synergy
510(k) Summary_K241433
| Name | Subject device | Primary Predicate Device | Secondary Predicate device | Comparison |
| --- | --- | --- | --- | --- |
| Device Name | EVE Synergy(EVE-20M) | EVOLVE System with the T3 Applicator | EVOLVE System with the Tone Applicator | |
| Manufacturer | WEERO Co.,Ltd. | InMode Ltd. | InMode Ltd. | |
| 510(k) No. | K241433 | K210877 | K201285 | |
| Maximum Current (RMS) Density [mA/cm²] | 1.31 [mA/cm²] Surface = 6 cm² | 0.74 [mA/cm²] Surface = 6.46cm² | 0.72 [mA/cm²] Surface = 12 cm² | Not identical. Although the subject device demonstrates a higher maximum current density compared to the predicate devices, this value does not exceed the safety level 2 mA/cm² specified in IEC 60601-2-10. Furthermore, the subject device demonstrates a maximum power density comparable to the predicate devices, indicating that the overall energy delivered to the tissue is similar. Therefore, this difference does not raise new questions of safety or effectiveness. |
| Maximum Power Density [mW/cm²] | 47 [mW/cm²] @ 500Ω | 22.2 [mW/cm²] @ 500Ω | 55 [mW/cm²] @ 500Ω | Not identical. The maximum power density of the subject device are similar to the predicate devices; therefore, they do not raise new questions about safety and effectiveness. |
| Burst Mode (i.e.,pulse trains) a. Pulses per burst b. Bursts per second c. Burst duration (seconds) d. Duty Cycle [Line (b) x Line(c)] | Yes: a. 3 – 200 b. 1 c. 1-60 sec d. Time on / off | Yes: a. 3 – 200 b. 1 c. 1-60 sec d. Time on / off | Yes: a. 3 – 200 b. 1 c. 1-60 sec d. Time on / off | Identical |
| On time (sec.) | 1 – 60 [sec] | 1 – 60 [sec] | 1 – 60 [sec] | Identical |
| Off time (sec.) | 1 – 60 [sec] | 1 – 60 [sec] | 1 – 60 [sec] | Identical |
| Treatment Time (min) - the time limit that will put the system in STOP state Level | Up to 60 min | Up to 60 min | Up to 60 min | Identical |
Page 8 / 13
{12}
EVE Synergy
510(k) Summary_K241433
| Name | Subject device | Primary Predicate Device | Secondary Predicate device | Comparison |
| --- | --- | --- | --- | --- |
| Device Name | EVE Synergy(EVE-20M) | EVOLVE System with the T3 Applicator | EVOLVE System with the Tone Applicator | |
| Manufacturer | WEERO Co.,Ltd. | InMode Ltd. | InMode Ltd. | |
| 510(k) No. | K241433 | K210877 | K201285 | |
| TENS Output | | | | |
| Output Specifications: Waveform | Symmetrical Biphasic Waveform, Pulsed Monophasic | Symmetrical Biphasic Waveform | Symmetrical Biphasic Waveform | The difference between a symmetrical biphasic waveform and a pulsed monophasic waveform is solely in whether the current flows in both positive and negative directions or in a single direction. Both waveforms are present in predicate devices cleared by the FDA for TENS applications. This difference does not raise new questions about the safety and effectiveness. |
| Pulse Shape | Rectangular | Rectangular | Rectangular | Identical |
| Maximum Output Voltage(±20%) | 10 V @500Ω
10 V @2kΩ
10 V @10kΩ | 19 V @500Ω
19 V @2kΩ
19 V @10kΩ | 36 V @500Ω
36 V @2kΩ
36 V @10kΩ | Not identical.
The maximum output voltage and current of the subject device are similar to those of the primary predicate device. In addition, since the maximum power density of both devices is very similar, the total energy delivered to the tissue remains similar in its effect on tissue response. Therefore, these differences do not raise new questions about safety and effectiveness. |
| Maximum Output Current(±20%) | 20 mA @500Ω
5.0 mA @2kΩ
1.0 mA @10kΩ | 38 mA @500Ω
9.5 mA@2kΩ
1.9 mA @10kΩ | 67.8 mA @500Ω
17.7 mA@2kΩ
3.6 mA @10kΩ | |
| Pulse Width(μsec.) – The output active positive pulse width | 100 to 150 [μs] | 20 to 400 [μs] | 20 to 400 [μs] | Not identical.
Since the pulse width of the subject device is within the range of the predicate devices, it does not raise new questions about safety and effectiveness. |
| Frequency(Hz) | 3 to 200 [Hz] | 3 to 200 [Hz] | 3 to 200 [Hz] | Identical |
Page 9 / 13
{13}
EVE Synergy
510(k) Summary_K241433
| Name | Subject device | Primary Predicate Device | Secondary Predicate device | Comparison |
| --- | --- | --- | --- | --- |
| Device Name | EVE Synergy(EVE-20M) | EVOLVE System with the T3 Applicator | EVOLVE System with the Tone Applicator | |
| Manufacturer | WEERO Co.,Ltd. | InMode Ltd. | InMode Ltd. | |
| 510(k) No. | K241433 | K210877 | K201285 | |
| Net Charge @ 500 Ohms [μC/pulse] | 0 [μC] @500Ω | 0 [μC] @ 500Ω | 0 [μC] @ 500Ω | Identical |
| Maximum Phase Charge [μC] | 3.0 [μC] @500Ω | 15.2 [μC] @ 500Ω | 28.8 [μC] @ 500Ω | Not identical.
Maximum Phase Charge is related to the maximum output current and pulse width, and because the predicate device has various pulse widths, the range of phase charge depends on the pulse width range. For example, for the primary predicate device, the phase charge ranges from 0.76 μC to 15.2 μC. Because the phase charge of the subject device is within the range of the predicate device, it does not raise new questions about safety and effectiveness. |
| Maximum Current Density [mA/cm²] | 0.83 [mA/cm²] Surface = 6 cm² | 0.47 [mA/cm²] Surface = 6.46 cm² | 0.48 [mA/cm²] Surface = 12 cm² | Not identical.
Although the subject device demonstrates a higher maximum current density compared to the predicate devices, this value does not exceed the safety level 2 mA/cm² specified in IEC 60601-2-10. Furthermore, the subject device demonstrates a maximum power density comparable to the predicate devices, indicating that the overall energy delivered to the tissue is similar. Therefore, this difference does not raise new questions of safety or effectiveness. |
| Maximum Power Density [mW/cm²] | 8.3 [mW/cm²] @ 500Ω | 8.9 [mW/cm²] @ 500Ω | 17.3 [mW/cm²] @ 500Ω | Not identical.
The maximum power density of the subject device are similar to the primary predicate device; therefore, they do not raise new questions about safety and effectiveness. |
Page 10 / 13
{14}
EVE Synergy
510(k) Summary_K241433
| Name | Subject device | Primary Predicate Device | Secondary Predicate device | Comparison |
| --- | --- | --- | --- | --- |
| Device Name | EVE Synergy(EVE-20M) | EVOLVE System with the T3 Applicator | EVOLVE System with the Tone Applicator | |
| Manufacturer | WEERO Co.,Ltd. | InMode Ltd. | InMode Ltd. | |
| 510(k) No. | K241433 | K210877 | K201285 | |
| Burst Mode (i.e.,pulse trains)
a. Pulses per burst
b. Bursts per second
c. Burst duration (seconds)
d. Duty Cycle[Line (b) x Line(c)] | Yes:
a. 3 – 200
b. 1
c. 1-60 sec
d. Time on / off | Yes:
a. 3 – 200
b. 1
c. 1-60 sec
d. Time on / off | Yes:
a. 3 – 200
b. 1
c. 1-60 sec
d. Time on / off | Identical |
| On time (sec.) | 1 – 60 [sec] | 1 – 60 [sec] | 1 – 60 [sec] | Identical |
| Off time (sec.) | 1 – 60 [sec] | 1 – 60 [sec] | 1 – 60 [sec] | Identical |
| Treatment Time (min) - the time limit that will put the system in STOP state Level | Up to 60 min | Up to 60 min | Up to 60 min | Identical |
## 2) Substantial Equivalence Discussion
There are differences on a few things. However, these differences do not affect the significant equivalence of the device and its predicates.
The subject device and predicate devices utilize the same technology, for the same indication for use, and with almost identical design specifications. The device emits EMS or TENS electrical signals with similar power and current densities, pulse characteristics, and bear almost identical system components to its predicate devices such as; user interface, and hardware components. All of the subject device performance specifications are equal or similar to those of its predicate devices. The minor differences in technical specifications should not alter the device safety and effectiveness.
Furthermore, the subject device had underwent the required performance testing and validation testing and demonstrates its conformance with device design requirements and with applicable standards.
The safety features and compliance with safety standards of the subject device are similar to the safety features and compliance with safety standards of the predicate devices. For biocompatibility, all user-contacting materials (i.e., the Stainless Steel 304, and Polycarbonate) are the same as those previously cleared with K212253.
Page 11 / 13
{15}
EVE Synergy
510(k) Summary_K241433
Furthermore, the design and development phases of the subject device were validated throughout a set of performance tests, including software validation testing, electrical and mechanical safety testing according to IEC 60601-1 standard, electromagnetic compatibility testing according to IEC 60601-1-2 standard, safety and essential performance of nerve and muscle stimulators testing according to IEC 60601-2-10 standard. These performance tests demonstrated that the device specifications meet the system requirements and do not raise new safety or effectiveness concerns.
3) Conclusion
"EVOLVE System with the T3 Applicator" and "EVOLVE System with the Tone Applicator" were chosen as predicate devices for the subject device in consideration of the intended use, indications, performance and principles of operation. Minor differences between the subject and the predicate devices were found and listed within the above comparison table and discussion. Considerable amount of testing, including electrical safety, electromagnetic compatibility, performance, software verification and validation and usability testing were performed to support the claims of appropriately chosen predicate device. The test results show that the specifications and performance of EVE Synergy are as safe and effective as legally marketed predicate devices.
Therefore, it is concluded that the EVE Synergy is substantially equivalent to the legally marketed predicate devices.
7. Non-Clinical (Bench) Performance Data:
As part of demonstrating safety and effectiveness of the EVE Synergy and in showing substantial equivalence to the predicate device that are subject to this 510(k) submission, we completed a number of non-clinical performance tests against applicable standards.
- Basic safety and essential performance of the EVE Synergy was tested and evaluated according to the IEC 60601-1:2005/A2:2020.
- Effect to the device by electromagnetic disturbances was tested and evaluated according to the FDA-recognized consensus standard IEC 60601-1-2:2014/A1:2020.
- Particular requirements for the basic safety and essential performance of nerve and muscle stimulators was tested and evaluated according to the IEC 60601-2-10:2012/A2:2023.
- Risk management was recorded by referring to ISO 14971:2019.
- Usability was documented by referring to IEC 60601-1-6:2010/A2:2020.
- For biocompatibility, all user-contacting materials (i.e., Stainless Steel 304 and Polycarbonate) are the same as those previously cleared under K212253.
- Software was tested and evaluated according to IEC 62304:2015
- Content of Premarket Submissions for Device Software Functions
The EVE Synergy passed all the testing in accordance with internal requirements, national standards, and international standards shown above, to support substantial equivalence of the subject device.
Also, to demonstrate that the EVE Synergy meets all design specifications and performance requirements, and to measure the accuracy of the output parameters
Page 12 / 13
{16}
EVE Synergy
510(k) Summary_K241433
of EVE Synergy and to compare the output parameters with predicate devices, nonclinical bench testing was performed in accordance with the internal process in compliance with the recommendations of the FDA Guidance Document for Powered Muscle Stimulator 510(k)s.
The testing results support that the requirements for performance and electrical safety were met for the acceptance of the device. The EVE Synergy passed all testing and supports the claims of substantial equivalence to the predicate device.
# 8. Sterilization/Disinfection/Cleaning/Shelf Life
The EVE Synergy is intended for multiple use and therefore must be cleaned according to the instructions provided in the device Instructions for Use. There are no sterilized parts or accessories involved with this device.
# 9. Biocompatibility
| Part | Material | Patient Contact | Duration of Contact by ISO 10993-1 | Bio-compatibility |
| --- | --- | --- | --- | --- |
| Electrodes of handpiece | SUS 304 | Intact Skin | Limited (< 24 hours) | Yes |
| LED Display part | PC(SR3108FM) | | | |
# 10. Conclusion
Based on the comparison with the predicate devices and on the non-clinical performance testing results demonstrating that the EVE Synergy is as safe and effective as the predicate devices, it can be concluded that the EVE Synergy is substantially equivalent to the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.