The sagAlign Lumbar Cage System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). sagAlign Lumbar Cage implants are to be used with allogenic bone graft comprised of cancellous and/or corticocancellous bone graft and implanted via an open posterior approach. saqAlian Lumbar Cage implant is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment with an intervertebral cage.
Device Story
Expandable interbody lumbar cage; implanted via open posterior approach; requires supplemental posterior fixation. Surgeon expands cage in situ to restore disc height and lordosis. Features graft windows and ridges/teeth to resist migration. Used in clinical settings by orthopedic or neurosurgeons. Provides structural support for intervertebral fusion. Benefits patient by restoring disc height and facilitating spinal fusion in DDD cases.
Clinical Evidence
Bench testing only. Mechanical verification included static compression, static compression-shear, dynamic compression-shear (ASTM F2077), and subsidence testing (ASTM F2267) on worst-case devices. No clinical data presented.
Technological Characteristics
Expandable interbody cage; titanium alloy (ASTM F136/ISO 5832-2); optional Promimic HAnano Surface (hydroxyapatite) coating. Mechanical expansion mechanism. Sterilization via gamma or steam. No software or electronic components.
Indications for Use
Indicated for skeletally mature patients with lumbar degenerative disc disease (DDD) at one or two contiguous levels (L2-S1) with or without Grade I spondylolisthesis. Requires 6 months of failed non-operative treatment. Must be used with supplemental fixation and allogenic bone graft.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the word "ADMINISTRATION".
July 17, 2024
Acuity Surgical Devices, LLC Chuck Forton Director of Engineering and Regulatory Affairs 8710 N. Royal Lane Irving, Texas 75063
Re: K241413
Trade/Device Name: sagAlign Lumbar Cage System (Various PNs) Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: May 17, 2024 Received: May 17, 2024
Dear Chuck Forton:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Aakash Jain -S
for Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
Submission Number (if known)
K241413
Device Name
sagAlign Lumbar Cage System (Various PNs)
### Indications for Use (Describe)
The sagAlign Lumbar Cage System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s).
sagAlign Lumbar Cage implants are to be used with allogenic bone graft comprised of cancellous and/or corticocancellous bone graft and implanted via an open posterior approach. saqAlian Lumbar Cage implant is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment with an intervertebral cage.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 1 of 1
{3}------------------------------------------------
K241413 Page 1 of 2
Prepared on: 2024-07-11
# Contact Details
21 CFR 807.92(a)(1)
21 CFR 807.92(a)(5)
| Applicant Name | | Acuity Surgical Devices LLC | | |
|------------------------------------|--|----------------------------------------------------------|--|---------------------|
| Applicant Address | | 8710 N. Royal Lane Irving TX 75063 United States | | |
| Applicant Contact Telephone | | 512-585-3537 | | |
| Applicant Contact | | Mr. Chuck Forton | | |
| Applicant Contact Email | | cforton@acuitysurgical.com | | |
| Device Name | | | | 21 CFR 807.92(a)(2) |
| Device Trade Name | | sagAlign Lumbar Cage System (Various PNs) | | |
| Common Name | | Intervertebral body fusion device | | |
| Classification Name | | Intervertebral Fusion Device With Bone Graft, Lumbar | | |
| Regulation Number | | 888.3080 | | |
| Product Code(s) | | MAX | | |
| Legally Marketed Predicate Devices | | | | 21 CFR 807.92(a)(3) |
| Predicate # | | Predicate Trade Name (Primary Predicate is listed first) | | Product Code |
| K080537 | | L-Varlock Lumbar Cages | | MAX |
| K222561 | | Align | | MAX |
| K221535 | | Align Lumbar Interbody Fusion System | | MAX |
| Device Description Summary | | | | 21 CFR 807.92(a)(4) |
umbar Cage System is an interbody expandable lumbar cage. It is implanted by a posterior approach and supp by posterior fixation which is required to achieve fusion. The lumbar cages can be gradually expanded to reach thosen by the surgeon, allowing restoration of disc height and the ideal lordosis.
The sagAlign Lumbar Cages are delivered non-sterile packaged. The cages are available in several sizes in different heights, widths, and lengths. One or two cages may be implanted for each treated level. The sagAlign Lumbar Cages will be composed of titanium alloy and may have a surface treating of Promimic HAnano Surface (hydroxyapatite (HA) coating). The Promimic HAnano Surface coated cages are only available sterile.
Reusable instruments to support implantation of the subject device are provided with non-sterilization trays.
# Intended Use/Indications for Use
The sagAlign Lumbar Cage System is indicated for intervertebral body fusion procedures in skeletally mature pative disc disease (DDD) of the lumbar spine at one or two
contiguous levels from L2-S1. DDD is defined as discogenic pain with degeneration of the disc
confirmed by history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s).
sagAlign Lumbar Cage implants are to be used with allogenic bone graft comprised of cancellous bone graft
{4}------------------------------------------------
and/or autograft and implanted via an open posterior approach. sagAlign Lumbar Cage implant is to be used with supplementa fixation. Patients should have at least six (6) months of non-operative treatment with an intervertel
# Indications for Use Comparison
The sagAlign Lumbar Cage System has the similar indications for use as the devices cleared in K080537, K222561. The difference in indications for use between the subject device (K080537) is the description of the bone graft material.
## Technological Comparison
Acuity Surgical Devices LLC believes that the sagAlign Lumbar Cage System is substantially equivalent to the primary predicate and additional predicate devices based on information summarized here:
For the sagAlign Lumbar Cage System, all devices have similar indications for use as the devices cleared in K080537.
The sagAlign Lumbar Cage System has the same surgical approach, implanted via an open posterior surgical approach in the lumbar spine at one or two contiguous levels from L2-S1, as the primary predicate device cleared in K080537. Both the subject device and primary predicate (K080537) also require supplement fixation. The subject device, sagAlign Lumbar design, rectangular shape, and dimensions as the primary predicate device (K080537) and additional predicate devices (K22256). The subject device is the same design as the primary predicate device (K080537) with expanded range of sizes. The subject device and primary predicate device (K080537) have the same expansion mechanism and instrument. The subject device has graft windows and same ridges / teeth to resist migration and increase surface contact as the primary predicate device cleared in K080537. The subject device and primary predicate devices (K080537) implants are material, titanium alloy per ASTM F136/ISO 5832-2, except the subject device's implants may have an optional implant coating for the subject device and additional predicate device (K222561) are the same material, hydroxapatite coating (Promimic HAnano Surface). The subject device and additional predicate devices (K222561 and K221535) may be sterilized using either gamma or steam sterilization.
In order to ensure that the different technological characteristics do not affectiveness of the subject device, a performance evaluation was performed to demonstrate that the mechanical strength and fatigue endurance of the sagAlign Lumbar Cage is comparable to the primary predicate and additional predicate devices.
The sagAlign Lumbar Cage System is substantially equivalent to and is as safe and as effective as, the legally marketed predicate device, L-Varlock Lumbar Cages (K080537) under regulation 21CFR 888,3080, product code MAX.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
The subject device underwent verification (static compression, static compression, static compression-shear, dynamic compression-shear per ASTM F2077 Test Methods for Intervertebral Body Fusion Devices, and subsidence testing per ASTM F2267 Standard Test Method for Measuring Load Induced Subsidence of Intervertebral Body Fusion Device Under Static Axial Compression) to ensure that the design features met the required mechanical strength criteria for their intended use.
It was verified that the new modifications of the subject device were demonstrated not to the predicate device.
The results on the non-clinical testing demonstrated that the substantially equivalent to the predicate devices.
Based on the results of mechanical testing conducted on the worst-case devices according to FDA Class II Special Controls Guidance, the subject device is considered substantially equivalent to the predicate devices cleared in K080537 and does not safety or effectiveness compared to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.