The Manual Wheelchair is to provide mobility to persons limited to a sitting position.
Device Story
Manual wheelchair (models S002, S004, S005, S006, S007, S008, S009) provides mobility for individuals restricted to sitting. Device features foldable main frame, seat, backrest, adjustable footrests, four wheels, hand rims for manual propulsion, and manual brakes on rear tires. Operated by user or assistant (via push handles). Used in indoor/outdoor environments on dry, level surfaces (concrete, blacktop, asphalt). Mechanical design; no electronic or software components. Benefits include improved patient mobility and transportability via folding mechanism.
Clinical Evidence
No clinical data. Bench testing only. Compliance verified with ISO 7176-1, -3, -5, -7, -8, -11, -13, -15, -22, and ISO 16840-10. Biocompatibility testing per ISO 10993-1, -5, -10, and -23 confirmed materials are non-cytotoxic, non-irritating, and non-sensitizing.
Technological Characteristics
Mechanical, manually operated wheelchair. Foldable frame with cross-brace. Components: main frame, backrest, seat cushion, footrests, hand rims, brakes, anti-tipper. Materials tested for biocompatibility per ISO 10993. Performance standards: ISO 7176 series (stability, brakes, dimensions, strength). No energy source, connectivity, or software.
Indications for Use
Indicated for persons limited to a sitting position requiring mobility assistance.
Regulatory Classification
Identification
A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and full name on the right. The Department of Health and Human Services logo is a stylized representation of a human figure, while the FDA acronym and full name are written in blue, with the acronym in a square.
February 28, 2025
Anhui JBH Medical Apparatus Co., Ltd % Jarvis Wu Senior Consultant Shanghai SUNGO Management Consulting Co., Ltd. 14th Floor, Dongfang Building, 1500# Century Ave. Shanghai, Shanghai China
Re: K241374
Trade/Device Name: Manual Wheelchair (Models: S002, S004, S005, S006, S007, S008, S009) Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I, reserved Product Code: IOR Dated: February 26, 2025 Received: February 26, 2025
Dear Jarvis Wu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Tushar Bansal -S
for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K241374
Device Name
Manual Wheelchair (Models: S002, S004, S005, S006, S007, S008, S009)
Indications for Use (Describe)
The Manual Wheelchair is to provide mobility to persons limited to a sitting position.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) Summary
### Document Prepared Date: 2025/2/28
#### A. Applicant:
Name: Anhui JBH Medical Apparatus Co., Ltd Address: No.116 QiCang Road Industrial District MingGuang Chuzhou,Anhui, China
Contact Person: Wei yuan wei Tel: +86- 0550-8108866 Email: outreach@jbhmedical.com
Submission Correspondent: Primary contact: Mr. Jarvis Wu Title: Senior Consultant Shanghai SUNGO Management Consulting Co., Ltd. 14th Floor, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: jiawei.wu@sungoglobal.com
Secondary contact: Mr. Raymond Luo Title: Technical Director Shanghai SUNGO Management Consulting Co., Ltd. 14th Floor, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: zxfda(@sungoglobal.com
#### B. Device:
Trade Name: Manual Wheelchair (Models: S002, S004, S005, S006, S007, S008, S009) Common Name: Manual Wheelchair Models: S002, S004, S005, S006, S007, S008, S009
Regulatory Information
Classification Name: Wheelchair, Mechanical Classification: Class I Product code: IOR Regulation Number: 21 CFR 890.3850
Review Panel: Physical Medicine
C. Predicate device: K232230 YJ-K1/K2 Wheelchair Zhenjiang Assure Medical Equipment Co., Ltd
{5}------------------------------------------------
#### D. Device Description:
The S002, S004, S005, S006, S007, S008, S009 series are mechanical wheelchairs which are a manually operated devices with wheels that are intended for medical purposes to provide mobility to persons restricted to a sitting position. It can be folded for transport by bring the two sides together. The manual wheelchair incorporates a main frame, a seat, two adjustable footrests and four wheels. The larger rear wheels have hand rims of slightly smaller diameter projecting just beyond the tire. These allows the user to manoeuvre the chair by pushing them on without requiring them to grasp the tires. The manual wheelchairs have brakes that bear on the tires of the rear wheels and two push handles at the upper rear of the frame to allow for manual propulsion by an assistant.
Main Components: Main frame, Backrest, Seat cushion, handgrip, front wheel, rear wheel, hand rim, crossbar, footrests, brake, Anti-tipper, Seat belt.
The device can be operated indoors, or outdoors on dry, level surfaces composed of concrete, blacktop, or asphalt under normal driving conditions.
#### E. Indications for Use:
The Manual Wheelchair is to provide mobility to persons limited to a sitting position.
## F. Comparison of Technological Characteristics with the Predicate Device Table 1 General Comparison
| Device | Proposed Device | Predicate Device | Results |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| 510K Number | K241374 | K232230 | --- |
| Manufacturer | Anhui JBH Medical Apparatus<br>Co., Ltd | Zhenjiang Assure Medical<br>Equipment Co., Ltd. | --- |
| Proprietary Name | Manual Wheelchair | YJ-K1/K2 Wheelchair | --- |
| Model | S002, S004, S005, S006, S007,<br>S008, S009 (7 models) | 24 models | --- |
| Classification | I | I | Same |
| Indications for use | The Manual Wheelchair is<br>intended for medical purposes<br>to provide mobility to persons<br>restricted to a sitting position. | The YJ-K1/K2 Manual<br>Wheelchair is to provide<br>mobility to persons limited to<br>a sitting position. | Same |
| Design<br>Characteristic | Manual Operation, Four-<br>Wheels, Foldable, Cross-<br>Brace, Quick release hub,Seat<br>cushion ,Backrest, Footrests,<br>Protective cover, Anti-tipper,<br>brake. | Manual Operation, Four-<br>Wheels, Foldable, Cross-<br>Brace, Pull-to- Lock, Armrest,<br>Backrest, Foot rest, skirt<br>guard , Anti-tipper (selected<br>models), brake extender<br>(selected models) | Similar |
| Brake control | occupant-operated brake only | occupant-operated brake only | Same |
| Operation | For indoor/outdoor use | For indoor/outdoor use | Same |
| Environment | | | |
| Control Mode | Mechanical | Mechanical | Same |
| Size(unfold) | S002: 815*610*765mm<br>S004:810*595*810mm<br>S005:810*595*790mm<br>S006:815*570*765mm<br>S007:815*590*765mm<br>S008:815*630*765mm<br>S009:815*650*765mm | 1220 (L) *660/640/700 (W) *<br>857/908mm (H) | Analysis #1 |
| Stowage<br>length/width/height | S002: 815*350*765mm<br>S004: 810*595*650mm<br>S005: 810*595*620mm<br>S006: 815*350*765mm<br>S007: 815*350*765mm<br>S008: 815*350*765mm<br>S009: 815*350*765mm | 1220(L)*280/285/310(W)*857/<br>90 8mm(H) | Analysis #1 |
| Weight(Total) | 12.5kg/14.0kg/14.5kg | 20.0kg/22.6kg | Analysis #1 |
| Weight Capacity | 100kg | 136Kg(300lbs) | Analysis #1 |
| Seat Width | 380mm | 405/455/505mm | Analysis #1 |
| Seat height | 510/520mm | 490mm/445mm | Analysis #1 |
| Seat depth | 430mm | 16",18",20" | Analysis #1 |
| Back type | Fixed | Fixed | Same |
| Tires | Front: 105mm<br>Rear:595mm | Front: 193mm<br>Rear:610mm | Analysis #1 |
| Foot rest | swing away footrest/ elevating<br>leg rest | swing away footrest/ elevating<br>leg rest | Same |
| Rear Axle Position | Single | Single | Same |
| Frame<br>Construction | Foldable frame Push inward<br>from left and right sides to fold | Foldable frame Push inward<br>from left and right sides to fold | Same |
| Safety Feature | Manual Wheel Lock | Manual Wheel Lock | Same |
| Maximum Safe<br>Operational<br>incline | Vertical forward tilt ≥8°,<br>vertical backward tilt ≥8°,<br>and lateral tilt ≥8° | Vertical forward tilt ≥10°,<br>vertical backward tilt ≥10°,<br>and lateral tilt ≥10° | Analysis #2 |
| | | | |
| braking time | The manual wheelchair brake<br>is mechanical structure and<br>manual braking, and the<br>braking time can be ignored<br>when it is used immediately. | The manual wheelchair<br>brake is mechanical structure<br>and manual braking, and the<br>braking time can be ignored<br>when it is used immediately. | Same |
| braking distance | brake is mechanical structure<br>and manual braking, and the<br>braking distance can be<br>ignored. | brake is mechanical structure<br>and manual braking, and the<br>braking distance can be<br>ignored. | Same |
| Performance | Comply with:<br>ISO7176-1<br>ISO7176-3<br>ISO7176-5<br>ISO7176-7<br>ISO7176-8<br>ISO7176-11<br>ISO7176-13<br>ISO7176-15<br>ISO 16840-10 | Comply with:<br>ISO7176-1<br>ISO7176-3<br>ISO7176-5<br>ISO7176-7<br>ISO7176-8<br>ISO7176-11<br>ISO7176-13<br>ISO7176-15<br>ISO7176-16 | Same |
| Biocompatibility | Comply with:<br>ISO10993-1<br>ISO10993-5<br>ISO10993-10<br>ISO10993-23 | Comply with:<br>ISO10993-1<br>ISO10993-5<br>ISO10993-10 | Same |
{6}------------------------------------------------
{7}------------------------------------------------
#### Difference Analysis:
- 1. Compare the predicate device, the subject device have different value on the unfold size, stowage size, device weight, Capacity, Seat Width, Seat height, Seat depth, Tires size. However the subject has pass the <ISO 7176-7-1998 Part7: Measurement of seating and wheel dimensions > and <ISO 7176-5-2008 Part 5: Determination of dimensions, mass and maneuvering space>, so the above different will not raise any new risk of safety or effectiveness.
- 2. The Safe Operational incline is not identical, the subject device has passed the bench performance testing. So it will not raise any new risk of safety or effectiveness.
#### G. Summary of Non-Clinic Performance Testing
#### Performance Testing
Non-clinical tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- A ISO 7176-1:2014, Wheelchairs - Part 1: Determination of static stability
- A ISO 7176-3:2012 Wheelchairs - Part 3: Determination of effectiveness of brakes
- ISO 7176-5: 2008 Wheelchairs Part 5: Determination of overall dimensions, mass and maneuvering A space
- ISO 7176-7: 1998 Wheelchairs Part 7: Measurement of seating and wheel dimensions ISO A 7176-8: 2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strengths
{8}------------------------------------------------
No.116 QiCang Road Industrial District MingGuang Chuzhou,Anhui, China
- ISO 7176-11: 2012 Wheelchairs Part 11: Test dummies V
- ISO 7176-13: 1989 Wheelchairs Part 13: Determination of coefficient of friction of test surfaces V
- > ISO 7176-15: 1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling
- ISO 16840-10:2021 Wheelchair seating- Part 10: Resistance to ignition of postural support devices > Requirements and test method
- ISO 7176-22:2014 Wheelchairs Part 22: Set-up procedures >
### Biocompatibility
Biocompatibility testing was conducted in accordance with the FDA guidance Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process. " Testing included:
- > Cytotoxicity (ISO 10993-5:2009)
- ﺮ Sensitization (ISO 10993-10:2021)
- > Irritation (ISO 10993-23:2021)
The testing supports the biocompatibility of the patient-contacting device materials that were shown to be non-cytotoxic, non-irritating, and non-sensitizing.
#### H. Clinical Test Conclusion
No clinical study is included in this submission.
#### I. Conclusion
The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission, the Manual Wheelchair is as safe, as effective, and performs as well as the legally marketed predicate device cleared under K232230.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.