← Product Code [BZE](/productcode/BZE) · K241268

# Thermalite Heated CPAP Tube (CTUBH06H15DS-2, CTUBH15-DS) (K241268)

_Exceleron Medical · BZE · Aug 2, 2024 · Anesthesiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K241268

## Device Facts

- **Applicant:** Exceleron Medical
- **Product Code:** [BZE](/productcode/BZE.md)
- **Decision Date:** Aug 2, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5270
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The THERMALITE® CPAP Heated Tubing is a heated wire breathing tube intended to provide warmed and/or humidified breathing gases before they enter a patient's airway. It is indicated for single-patient reuse in the home and in clinical settings, such as hospitals, institutions, sleep laboratories, and sub-acute care facilities. It may be used with noninvasive ventilation for adult patients. It is compatible with the Philips Respironics DreamStation 2 heated humidifiers.

## Device Story

Heated wire breathing tube connecting CPAP machine to patient interface (mask); prevents condensate (rainout) by controlling airflow temperature to maintain water vapor. Available in 2-wire or 3-wire configurations. Compatible with Philips Respironics DreamStation 1 and 2 heated humidifiers. Operated by patient (home) or clinician (clinical settings). Power supplied by humidifier; integrated NTC thermistor senses temperature at mask-end (3-wire) or machine-end (2-wire) to regulate current through heating wires. Warmed air reduces/eliminates water pooling in breathing circuit; improves patient comfort and therapy compliance.

## Clinical Evidence

Bench testing only. Performance validated against ISO 5367, ISO 80601-2-70, ISO 80601-2-74, and IEC 60601 series. Biocompatibility confirmed per ISO 10993 and ISO 18562. Results demonstrate compliance with safety and performance requirements.

## Technological Characteristics

Heated wire breathing tube; 1.83m length; 15mm inner diameter. Materials tested for biocompatibility (ISO 10993, ISO 18562). Sensing: NTC thermistor. Energy: Humidifier-controlled electrical heating. Connectivity: Proprietary connector for DreamStation series. Sterilization: Not specified.

## Regulatory Identification

A breathing system heater is a device that is intended to warm breathing gases before they enter a patient's airway. The device may include a temperature controller.

## Predicate Devices

- Philips Heated CPAP Tube ([K140424](/device/K140424.md))

## Reference Devices

- Philips Respironics DreamStation 2 System ([K200480](/device/K200480.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.

August 2, 2024

Exceleron Medical % Paul Dryden Consultant ProMedic Consulting LLC 131 Bay Point Dr. NE Saint Petersburg, Florida 33704

Re: K241268

Trade/Device Name: THERMALITE CPAP Heated Tubing Regulation Number: 21 CFR 868.5270 Regulation Name: Breathing System Heater Regulatory Class: Class II Product Code: BZE Dated: May 4, 2024 Received: May 6, 2024

Dear Paul Dryden:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

### Rachana Visaria -S

Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices 

{2}------------------------------------------------

OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

### Indications for Use

510(k) Number (if known) K241268

Device Name THERMALITE CPAP Heated Tubing

#### Indications for Use (Describe)

The THERMALITE® CPAP Heated Tubing is a heated wire breathing tube intended to provide warmed and/or humidified breathing gases before they enter a patient's airway. It is indicated for single-patient reuse in the home and in clinical settings, such as hospitals, institutions, sleep laboratories, and sub-acute care facilities. It may be used with noninvasive ventilation for adult patients.

It is compatible with the Philips Respironics DreamStation 2 heated humidifiers.

| Type of Use (Select one or both, as applicable)                                                                                            |
|--------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"> <span style="font-size:1.2em;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> |
| <div style="display:flex; align-items:center;"> <span style="font-size:1.2em;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div>  |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# (in accordance with 21 CFR 807.92)

| Date:                              | 30 July 2024                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor:                           | Exceleron Medical, LLC<br>5000 Township Parkway<br>Saint Paul, MN 55110 USA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                    | Keith Roberts - CEO<br>Phone: +1 (877) 834-9279                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Submission Correspondent:          | Paul Dryden<br>ProMedic, LLC<br>Phone: +1 (239) 307-6061                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Trade Name:                        | THERMALITE® CPAP Heated Tubing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Common and<br>Classification Name: | Breathing System, Heater                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Classification<br>Regulation:      | 21 CFR §868.5270                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Product Code:                      | BZE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Predicate:                         | K140424<br>Philips Heated CPAP Tube                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Device Description:                | The Exceleron Medical, LLC (Exceleron) THERMALITE® CPAP Heated<br>Tubing (THERMALITE) is a heated tube that connects between a continuous<br>positive airway pressure (CPAP) machine and the patient's interface, typically a<br>CPAP mask. The THERMALITE is compatible with the Philips Respironics<br>DreamStation 1 and DreamStation 2 heated humidifiers.<br>The THERMALITE has a unique design that helps prevent condensate from<br>forming within the heated air stream during CPAP therapy by controlling the<br>airflow temperature, creating an environment to maintain water in its vapor<br>form. This condition of condensation is commonly known as "rainout," and is<br>usually the result of temperature and humidity differentials between the device<br>versus the surrounding air.<br>The THERMALITE is available in a three (3) or two (2) wire configuration. |
| Indications for Use:               | The THERMALITE® CPAP Heated Tubing is a heated wire breathing tube<br>intended to provide warmed and/or humidified breathing gases before they<br>enter a patient's airway. It is indicated for single-patient reuse in the home and<br>in clinical settings, such as hospitals, institutions, sleep laboratories, and sub-<br>acute care facilities. It may be used with non-invasive ventilation for adult<br>patients.<br>It is compatible with the Philips Respironics DreamStation 1 and DreamStation<br>2 heated humidifiers.                                                                                                                                                                                                                                                                                                                                                 |

{5}------------------------------------------------

| Manufacturer / Device                                                                      | Cleared | Reason for Inclusion                                        | Regulation ( Product Code)                                             |
|--------------------------------------------------------------------------------------------|---------|-------------------------------------------------------------|------------------------------------------------------------------------|
| Primary Predicate<br>Philips Respironics Reusable<br>Heated Tubing                         | K140424 | Same indications for use and<br>same or similar technology. | 21 CFR §868.5270 (BZE)                                                 |
| Reference<br>Philips Respironics Dream<br>Station 2 System which<br>includes heated tubing | K200480 | Same indications for use and<br>same or similar technology. | 21 CFR 868.5905 (BZD)<br>But contains heated tubing as<br>an accessory |

Table - Primary Predicate and Reference Devices

{6}------------------------------------------------

# (

| Characteristics                                                                                | Primary Predicate<br>Philips Respironics Reusable<br>Heated Tubing (K140424)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Reference<br>Philips Respironics<br>Reusable Heated Tubing<br>(K200480)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Exceleron Medical Heated<br>CPAP Tube (3 wire design) -<br>Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Exceleron Medical Heated<br>CPAP Tube (2 wire<br>design) - Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Discussion of<br>Differences                                                                                                                                                                                       |
|------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use                                                                            | The Philips Respironics<br>Reusable Heated Tubing is a<br>heated wire breathing tube<br>intended to provide warmed<br>and/or humidified breathing<br>gases before they enter a<br>patient's airway. It is indicated<br>for single-patient reuse in the<br>home and multi-patient use in<br>clinical settings, such as<br>hospitals, institutions, sleep<br>laboratories, and sub- acute<br>care facilities. It may be used<br>with non-invasive ventilation<br>for patients weighing over 10<br>kg (22lbs).<br>It is compatible with the<br>Philips Respironics System<br>One Heated Humidifier and<br>Philips Respironics A-Series<br>System One Heated<br>Humidifier. | The heated tube is used, along<br>with the heated humidifier, to<br>control the provided<br>humidification. This is<br>accomplished by controlling<br>the temperature of the air in<br>order to ensure that it does not<br>cool down prior to reaching<br>the mask.                                                                                                                                                                                                                                                                                                              | The Exceleron Heated CPAP<br>Tube is a heated wire<br>breathing tube intended to<br>provide warmed and/or<br>humidified breathing gases<br>before they enter a patient's<br>airway. It is indicated for<br>single-patient reuse in the<br>home and in clinical settings,<br>such as hospitals, institutions,<br>sleep laboratories, and sub-<br>acute care facilities. It may be<br>used with non-invasive<br>ventilation for adult patients.<br>It is compatible with the<br>Philips Dreamstation 1<br>(K140424) and Dreamstation<br>2(K200480) Heated<br>Humidifiers         | The Exceleron Heated<br>CPAP Tube is a heated wire<br>breathing tube intended to<br>provide warmed and/or<br>humidified breathing gases<br>before they enter a patient's<br>airway. It is indicated for<br>single-patient reuse in the<br>home and in clinical<br>settings, such as hospitals,<br>institutions, sleep<br>laboratories, and sub-acute<br>care facilities. It may be<br>used with non-invasive<br>ventilation for adult patients.<br>It is compatible with the<br>Philips Dreamstation 1 and<br>Dreamstation 2 Heated<br>Humidifiers. | Similar.<br>Subject<br>devices<br>restricted to:<br>· Single<br>patient<br>reuse.<br>Adult patients.                                                                                                               |
| Compatibility<br>with Humidifiers,<br>Standard<br>Connectors and<br>Humidification<br>Chambers | The tubing has a proprietary<br>connector with two locking<br>tabs that makes it compatible<br>with the Philips Respironics<br>System One Humidifier<br>(K113068) and Philips<br>Respironics A-Series System<br>One Heated Humidifier<br>(K121623).                                                                                                                                                                                                                                                                                                                                                                                                                     | The tubing has a proprietary<br>connector with two locking<br>tabs that makes it compatible<br>with the Dreamstation 2 machine.                                                                                                                                                                                                                                                                                                                                                                                                                                                  | The tubing has a proprietary<br>connector with two locking<br>tabs that make it compatible<br>with the Respironics<br>Dreamstation Series<br>(Dreamstation 1 and<br>Dreamstation 2 CPAP<br>machine (K200480))                                                                                                                                                                                                                                                                                                                                                                  | The tubing has a proprietary<br>connector with two locking<br>tabs that make it<br>compatible with the<br>Respironics Dreamstation<br>Series (Dreamstation 1 and<br>Dreamstation 2 CPAP<br>machine (K200480))                                                                                                                                                                                                                                                                                                                                       | Similar.<br>Compatibility<br>with a subset of<br>the devices with<br>which the<br>predicate is<br>compatible.                                                                                                      |
| Anatomical Site                                                                                | Non-invasive                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Non-invasive                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Non-invasive                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Non-invasive                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same.                                                                                                                                                                                                              |
| Characteristics                                                                                | Primary Predicate<br>Philips Respironics Reusable<br>Heated Tubing (K140424)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Reference<br>Philips Respironics<br>Reusable Heated Tubing<br>(K200480)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Exceleron Medical Heated<br>CPAP Tube (3 wire design) –<br>Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Exceleron Medical Heated<br>CPAP Tube (2 wire<br>design) – Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Discussion of<br>Differences                                                                                                                                                                                       |
| Patient<br>Population                                                                          | Patients weighing over 10 kg<br>(22lbs).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Patients weighing more than 66<br>lb (30 kg)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Adult patients                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Adult patients                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Similar. Device<br>is used with a<br>subset of the<br>predicate<br>device's                                                                                                                                        |
| Environment of<br>Use                                                                          | Home and hospital                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Home and hospital                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Home and hospital                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Home and hospital                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same.                                                                                                                                                                                                              |
| Operating<br>Principle                                                                         | During use a voltage is applied<br>and a current flows through the<br>heating wires, encapsulated in<br>the tubing. Due to the wire<br>resistance, heat is dissipated<br>through the wall of the tube<br>construction into the air flow in<br>the lumen of the tubing. As a<br>result, the air passing through<br>the tubing is warmed reducing<br>or eliminating water<br>condensation and/or pooling of<br>water in the breathing circuit.                                                                                                                                                                                                                            | Due to the wire resistance, heat<br>is dissipated through the wall<br>of the tube construction into<br>the air flow in the lumen of the<br>tubing. As a result, the air<br>passing through the tubing is<br>warmed to or above the dew<br>point (of the air existing the<br>humidifier) reducing or<br>eliminating water condensation<br>and/or pooling of water in the<br>breathing circuit 41 °C.<br>The raising of the gas<br>temperature does not exceed                                                                                                                     | During use a voltage is applied<br>and a current flows through the<br>heating wires, encapsulated in<br>the tubing. Due to the wire<br>resistance, heat is dissipated<br>through the wall of the tube<br>construction into the air flow in<br>the lumen of the tubing. As a<br>result, the air passing through<br>the tubing is warmed reducing<br>or eliminating water<br>condensation and/or pooling of<br>water in the breathing circuit.                                                                                                                                   | During use a voltage is<br>applied and a current flows<br>through the heating wires,<br>encapsulated in the tubing.<br>Due to the wire resistance,<br>heat is dissipated through the<br>wall of the tube construction<br>into the air flow in the lumen<br>of the tubing. As a result, the<br>air passing through the tubing<br>is warmed reducing or<br>eliminating water<br>condensation and/or pooling<br>of water in the breathing<br>circuit.                                                                                                  | Same.                                                                                                                                                                                                              |
| Characteristics<br>(continued)                                                                 | Philips Respironics Reusable<br>Heated Tubing (K140424)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Philips Respironics<br>Reusable Heated Tubing<br>(K200480)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Exceleron Heated CPAP Tube<br>(3 wire design) – Proposed<br>Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Exceleron Medical Heated<br>CPAP Tube (2 awire design) -<br>Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Discussion of<br>Differences                                                                                                                                                                                       |
| Technology                                                                                     | The power is generated by the<br>humidifier and due to the wire<br>resistance in the device, heat is<br>dissipated through the wall of<br>the tube construction into the air<br>flow in the lumen of the tubing.<br>An in-circuit integrated Negative<br>Temperature Coefficient<br>thermistor (NTC) at the mask-<br>end senses the temperature of the<br>passing air flow. The resistance<br>characteristics of the NTC<br>changes with temperature and<br>regulates the current through the<br>heated wires and thereby<br>regulates the temperature of the<br>breathed air.                                                                                          | The power is generated by the<br>humidifier and due to the wire<br>resistance in the device, heat is<br>dissipated through the wall of<br>the tube construction into the<br>air flow in the lumen of the<br>tubing. An in-circuit integrated<br>Negative Temperature<br>Coefficient thermistor (NTC) at<br>the mask-end senses the<br>temperature of the passing air<br>flow. The resistance<br>characteristics of the NTC<br>changes with temperature and<br>regulates the current through the<br>heated wires and thereby<br>regulates the temperature of the<br>breathed air. | The power is generated by the<br>humidifier and due to the wire<br>resistance in the device, heat is<br>dissipated through the wall of the<br>tube construction into the air flow<br>in the lumen of the tubing. An in-<br>circuit integrated Negative<br>Temperature Coefficient<br>thermistor (NTC) at the mask-end<br>senses the temperature of the<br>passing air flow. The resistance<br>characteristics of the NTC<br>changes with temperature and<br>regulates the current through the<br>heated wires and thereby regulates<br>the temperature of the breathed<br>air. | The power is generated by the<br>humidifier and due to the wire<br>resistance in the device, heat is<br>dissipated through the wall of the<br>tube construction into the air flow<br>in the lumen of the tubing. An in-<br>circuit integrated Negative<br>Temperature Coefficient<br>thermistor (NTC) at the machine<br>end. The machine will sense the<br>resistance of the thermistor and<br>will determine the amount of<br>voltage to apply to the heaters to<br>achieve the tubing temperature.                                                | Same with the<br>only difference on<br>the 2 wire design<br>is the NTC is at<br>the machine end<br>and provides the<br>same voltage<br>regulation through the heated wires as<br>both predicate 3<br>wire designs. |
| Dimensions/specs                                                                               | Length: 1.83 mm<br>Tubing Inner Diameter: 15<br>mm version and 22 mm<br>version<br>Patient Fitting – 22 mm<br>Machine Fitting - Proprietary<br>to the humidifier                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Length: 1.83 mm<br>Tubing Inner Diameter: 15<br>mm version and 22 mm<br>version<br>Patient Fitting – 22 mm<br>Machine Fitting -<br>Proprietary to the humidifier                                                                                                                                                                                                                                                                                                                                                                                                                 | Length: 1.83 meter Inner<br>Tubing Inner Diameter: 15 mm<br>Patient Fitting – 22 mm<br>Machine Fitting – Proprietary to<br>the humidifier                                                                                                                                                                                                                                                                                                                                                                                                                                      | Length: 1.83 meter Inner<br>Tubing Inner Diameter: 15 mm<br>Patient Fitting – 22 mm<br>Machine Fitting – Proprietary<br>to the humidifier                                                                                                                                                                                                                                                                                                                                                                                                           | Proposed<br>device is the<br>similar to the<br>15 mm<br>predicate<br>versions.                                                                                                                                     |
| Characteristics                                                                                | Primary Predicate<br>Philips Respironics  Reusable<br>Heated Tubing (K140424)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Reference<br>Philips Respironics<br>Reusable Heated Tubing<br>(K200480)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Exceleron Medical Heated<br>CPAP Tube (3 wire design) -<br>Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Exceleron Medical Heated<br>CPAP Tube (2 wire<br>design) – Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Discussion of<br>Differences                                                                                                                                                                                       |
| Reusable                                                                                       | Single-patient reuse in the<br>home and multi-patient use in<br>clinical setting                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Single patient reuse.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Single patient reuse.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Single patient reuse.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Similar.<br>Proposed device<br>is labeled the<br>same for home<br>use and for<br>single use  only<br>in a clinical                                                                                                 |
| Power Source                                                                                   | Humidifier controlled                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Humidifier controlled                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Humidifier controlled                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Humidifier controlled                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same.                                                                                                                                                                                                              |
| Standards of<br>Conformity /<br>Performance                                                    | ISO 5367<br>IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>ISO 8185                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | ISO 5367<br>IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>ISO 80601-2-70<br>ISO 80601-2-74                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | ISO 5367<br>IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>ISO 80601-2-70<br>ISO 80601-2-74                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | ISO 5367<br>IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>ISO 80601-2-70<br>ISO 80601-2-74                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Similar.<br>Proposed<br>device has<br>applied<br>relevant<br>standards.                                                                                                                                            |
| Biocompatibility                                                                               | ISO 10993,<br>tests for: Cytotoxicity,<br>sensitization, irritation,<br>genotoxicity, implantation,<br>extractables and leachables.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | ISO 10993-1<br>ISO 18562-1<br>ISO 18562-2<br>ISO 18562-3<br>ISO 18562-4<br>tests for: emissions of particulate<br>matter, emissions of volatile<br>organic compounds (VOCs)                                                                                                                                                                                                                                                                                                                                                                                                      | ISO 10993-1<br>ISO 18562-1<br>ISO 18562-2<br>ISO 18562-3<br>ISO 10993-18<br>tests for: Cytotoxicity,<br>sensitization, irritation,<br>emissions of particulate matter,<br>emissions of volatile  organic<br>compounds (VOCs),<br>extractables and leachables.                                                                                                                                                                                                                                                                                                                  | ISO 10993-1<br>ISO 18562-1<br>ISO 18562-2<br>ISO 18562-3<br>ISO 10993-18<br>tests for: Cytotoxicity,<br>sensitization, irritation,<br>emissions of particulate<br>matter, emissions of volatile<br>organic compounds (VOCs),<br>extractables and leachables.                                                                                                                                                                                                                                                                                        | Similar                                                                                                                                                                                                            |

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

# (

{10}------------------------------------------------

### Summary of Performance Testing:

| Biocompatibility                 | The indirect patient-contact materials in the Exceleron THERMALITE were tested for biocompatibility compliance in accordance with the following Standards and guidance documents:<br>ISO 10993-1: 2018, Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process. ISO 10993-18: 2020, Biological evaluation of medical devices — Part 18: Chemical characterization of medical device materials within a risk management process. ISO 18562-1: 2017, Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 1: Evaluation and testing within a risk management process. ISO 18562-2: 2017, Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 2: Tests for emissions of particulate matter. ISO 18562-3: 2017, Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 3: Tests for emissions of volatile organic compounds. Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process," 04 Sep 20.              Test results indicated that the indirect patient-contact materials in the Exceleron THERMALITE complies with the applicable Standards and guidance documents. |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Electrical Safety                | The Exceleron THERMALITE was tested for patient safety in accordance with the following Standards:  IEC 60601-1: 2005, Am1: 2012, Am2: 2020, Medical electrical equipment – Part 1: General requirements for basic safety and essential performance. IEC 60601-1-11: 2020, Medical electrical equipment – Part 1-11: General requirements for basic safety and essential performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.              Test results indicated that the Exceleron THERMALITE complies with the applicable Standards                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Electromagnetic<br>Compatibility | The Exceleron THERMALITE was tested for EMC in accordance with the<br>following Standards and guidance documents:<br>IEC 60601-1-2: 2020, Medical Electrical Equipment, Part 1: Part 1-2:<br>General Requirements for Safety – Collateral Standard: Electromagnetic<br>Compatibility-Requirements and Tests. Electromagnetic Compatibility (EMC) of Medical Devices, 06 Jun 22.          Test results indicated that the Exceleron THERMALITE complies with the<br>applicable Standards and guidance documents.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Performance Testing –<br>Bench   | The Exceleron THERMALITE was tested for performance in accordance with<br>internal requirements, applicable Standards, and guidance documents.  IEC 60601-1-6: 2020, Medical electrical equipment: General<br>requirements for basic safety and essential performance – collateral<br>standard: Usability. IEC 62366-1: 2020, Medical devices – Application of usability<br>engineering to medical devices. ISO 5356-1: 2015, Anaesthetic and respiratory equipment - Conical<br>connectors – Part 1: Cones and sockets. ISO 5367: 2014, Anaesthetic and respiratory equipment -- Breathing sets<br>and connectors.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Performance Testing –<br>Bench   | The Exceleron THERMALITE was tested for performance in accordance with<br>internal requirements, applicable Standards, and guidance documents.  ISO 80601-2-70: 2020, Medical electrical equipment - Part 2-70:<br>Particular requirements for the basic safety and essential performance of<br>sleep apnoea breathing therapy equipment. ISO 80601-2-74: 2017, Medical electrical equipment - Part 2-74:<br>Particular requirements for basic safety and essential performance of<br>respiratory humidifying equipment. ISTA Procedure 3A – 2018, Parcel Delivery System Shipments 150 lb (70<br>kg) or Less.          Test results indicated that the Exceleron THERMALITE complies with<br>internal requirements, applicable Standards, and the guidance documents.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Conclusion                       | Verification and validation activities were conducted to establish the<br>performance and safety characteristics of the Exceleron THERMALITE. The<br>results of these activities demonstrate that the Exceleron THERMALITE is as<br>safe and as effective as the predicate device when used in accordance with its<br>intended use and labeling.<br>Therefore, the Exceleron THERMALITE is considered substantially<br>equivalent to the predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |

{11}------------------------------------------------

# (

---

**Source:** [https://fda-staging.innolitics.com/device/K241268](https://fda-staging.innolitics.com/device/K241268)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).