The sponsor utilized retrospective clinical data from a multicenter PMCF investigation to demonstrate that audiological outcomes (PTA4 ABG <= 20 dB) and safety profiles (adverse events, revision rates) are comparable to the state of the art for partial ossicular replacement prostheses.
Post-Market Clinical Follow-Up (PMCF) investigation; Multicenter, retrospective, single-subjects repeated measures; Follow-up/Duration: Median 70.0 days (audiological); Median 275.0 days (safety)
Patients requiring reconstruction of the ossicular chain (ages 5-84); Sample Size: 186 (audiological outcomes); 202 (adverse events)
The passive middle ear implant - tympanoplasty partial prosthesis is intended to be used for replacement of components of the ossicular chain, if at least the stapes footplate is present and functional. The tympanoplasty partial prosthesis is implanted in the middle ear to restore sound transmission from the tympanic membrane to the oval window by replacing the ossicles partially. The tympanoplasty prosthesis is a medical device for single use delivered in sterile condition.
Device Story
Passive middle ear implants (PMEI) restore sound transmission from tympanic membrane to oval window in patients with ossicular chain defects. Devices are implanted by ENT surgeons during standard otological surgery. Input is mechanical vibration from tympanic membrane; device acts as bridge to stapes footplate. Output is restored conductive hearing. Designs include mCLIP (clip-on coupling), mCLIP ARC (ball-joint for membrane movement adaptation), mXACT (bell-shaped coupling), and mXACT PRO (length-adjustable with snap-mechanism). Sizers are used intraoperatively to determine required functional length. Benefits include improved hearing thresholds and closure of air-bone gap. Devices are MRI conditional (1.5T, 3.0T, 7.0T).
Clinical Evidence
Retrospective multicenter PMCF study (n=186 for audiological outcomes, n=202 for safety). Primary endpoint: postoperative PTA4 air-bone gap (ABG) ≤20 dB. Results: 72% of patients achieved PTA4 ABG ≤20 dB. Mean BC-PTA4 improved from 19.8 dB HL to 17.9 dB HL. Adverse events (n=34) included 5 revision surgeries and 3 device dislocations/extrusions. Data confirms safety and effectiveness comparable to state-of-the-art.
Technological Characteristics
Materials: Medical grade titanium (ASTM F67). Design: Partial ossicular replacement prostheses with various coupling structures (clip, bell, ball-joint). Dimensions: Functional lengths 0.75-3.50 mm. Connectivity: None (passive implant). Sterilization: Gamma radiation (SAL <10^-6). MRI compatibility: Conditional at 1.5T, 3.0T, 7.0T.
Indications for Use
Indicated for patients of all ages with congenital or acquired ossicular chain defects (e.g., chronic otitis media, traumatic injury, malformation, cholesteatoma) or inadequate conductive hearing from previous middle ear surgery, provided the stapes head and footplate are present and functional.
Regulatory Classification
Identification
A partial ossicular replacement prosthesis is a device intended to be implanted for the functional reconstruction of segments of the ossicular chain and facilitates the conduction of sound wave from the tympanic membrane to the inner ear. The device is made of materials such as stainless steel, tantalum, polytetrafluoroethylene, polyethylene, polytetrafluoroethylene with carbon fibers composite, absorbable gelatin material, porous polyethylene, or from a combination of these materials.
Predicate Devices
TTP-Tuebingen Type BELL Partial Prosthesis (K972492)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date January 17, 2025. The month is January, the day is the 17th, and the year is 2025. The text is in a simple, sans-serif font. The date is written out in full, with the month spelled out and the day and year written numerically.
Image /page/0/Picture/9 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
MED-EL Elektromedizinische Geräte GmbH Inés Román Santiago Senior Regulatory Affairs Fürstenweg 77a Innsbruck, Tirol 6020 Austria
Re: K241261
Trade/Device Name: mCLIP Partial Prosthesis, mCLIP ARC Partial Prosthesis, mXACT Partial Prosthesis and mXACT PRO Partial Prosthesis Kit Regulation Number: 21 CFR 874.3450 Regulation Name: Partial Ossicular Replacement Prosthesis Regulatory Class: Class II Product Code: ETB Dated: December 18, 2024 Received: December 18, 2024
Dear Inés Román Santiago:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Joyce C. Lin -S
for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K241261
### Device Name
mCLIP Partial Prosthesis, mCLIP ARC Partial Prosthesis, mXACT Partial Prosthesis and mXACT PRO Partial Prosthesis Kit
# Indications for Use (Describe)
## INTENDED USE:
The passive middle ear implant - tympanoplasty partial prosthesis is intended to be used for replacement of components of the ossicular chain, if at least the stapes footplate is present and functional. The tympanted in the middle ear to restore sound transmission from the tympanic membrane to the oval window by replacing the ossicles partially. The tympanoplasty prosthesis is a medical device for single use delivered in sterile condition.
### INTENDED USER
The tympanoplasty prosthesis is intended to be implanted by qualified ENT surgeons only, with adequate skills to perform otological surgeries. Replacement of the ossicular chain is a standard surgical procedure, no additional specific device training is mandatory or required for the safe and effective use.
### TARGET PATIENT POPULATION
The target patient population for the passive middle ear implant are patients of all ages requiring reconstruction of the ossicular chain.
### INDICATIONS:
The tympanoplasty prosthesis is indicated to treat patients with:
- congenital or acquired ossicular chain defects (e.g., chronic otitis media, traumatic injury, malformation, cholesteatoma) - inadequate conductive hearing from previous middle ear surgery
The partial ossicular replacement prosthesis is indicated, if at least the stapes footplate are present and functional.
Tympanosplasty Sizers for partial prosthesis
(Note that the tympanosplasty Sizers are packaged together with the mXACT PRO Partial Prosthesis conforming the mXACT PRO Partial Prosthesis Kit).
### INTENDED USE
The tympanoplasty Sizers are intended to be used during a partial ossicular replacement surgery to determine the required functional length of the MED-EL tympanoplasty prosthesis.
The tympanoplasty Sizers are optional, single-use surgical tools for intraoperative transient use only and delivered in sterile condition.
The Sizers for partial prosthesis are intended to intraoperatively determine the head of the stapes and the tympanic membrane, thereby determining the required functional length of the MED-EL tympanoplasty prosthesis.
### INDICATIONS
The indications of the PMEI tympanoplasty partial prostheses apply.
{4}------------------------------------------------
### Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for MED-EL, a company that specializes in hearing implants. The logo is in red and features the company name in bold, sans-serif font. To the right of the name is a stylized graphic of a cochlea, which is the part of the inner ear that is responsible for hearing. The logo is simple and modern, and it effectively communicates the company's focus on hearing solutions.
# 510(k) SUMMARY of Safety and Effectiveness MED-EL Elektromedizinische Geräte GmbH [As Required by 21 CFR 807.92(c)]
### Submitter [807.92(a)(1)] 1.0
Manufacturer: MED-EL Elektromedizinische Geräte GmbH (hereafter MED-EL) Fürstenweg 77a, 6020 Innsbruck Austria
Contact Person:
Ines Roman Santiago MED-EL Regulatory Affairs Senior Specialist Phone: +43 577885786 E-Mail: ines.romansantiago@medel.com
FDA Official Correspondent: Elizabeth Gfoeller MED-EL Corporate Director, Regulatory Affairs Phone: +43 577885614 E-Mail: elizabeth.gfoeller@medel.com
Date the Summary was prepared: 17th January 2025.
#### 2.0 Device Names [807.92(a)(2)]
## Table 1 Device/Trade Names of MED-EL PMEIs Tympanoplasty Partial Prostheses
| Device/Trade<br>Name | Generic category<br>& Medical Specialty | Classification | Product<br>code | Regulation<br>(CFR) |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------|----------------|-----------------|---------------------|
| mCLIP Partial<br>Prosthesis | Middle ear, Prosthesis,<br>Partial Ossicular<br>Replacement Prostheses<br>(Ear, Nose & Throat) | Class II | ETB | 874.3450 |
| mCLIP ARC<br>Partial<br>Prosthesis | | | | |
| mXACT Partial<br>Prosthesis | | | | |
| mXACT PRO<br>Partial<br>Prosthesis Kit | Middle ear, Prosthesis,<br>Length -Adjustable and<br>Partial Ossicular<br>Replacement Prostheses<br>(Ear, Nose & Throat) | | | |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for MED-EL, a company that specializes in hearing implants. The logo is in red and features the company name in bold, sans-serif font. To the right of the "D" in MED is a circular graphic with curved lines, possibly representing sound waves or the inner ear.
#### Predicate Devices [807.92(a)(3)] 3.0
| | Table 2 Predicate devices manufactured by Heinz Kurz GmbH | | |
|--|-----------------------------------------------------------|--|--|
| | | | |
| MED-EL device | Primary Predicate Device A | Predicate Device B |
|-------------------------------------|---------------------------------------------------------|-------------------------------------------------------|
| mXACT Partial Prosthesis | | N/A |
| mXACT PRO Partial<br>Prosthesis Kit | TTP-<br>Tuebingen<br>Type BELL<br>Partial<br>Prosthesis | TTP- VARIAC<br>System Partial<br>K990923 |
| mCLIP Partial Prosthesis | K972492 | CLIP Partial<br>Prosthesis<br>Dresden Type<br>K013573 |
| mCLIP ARC Partial<br>Prosthesis | | CliP Partial<br>FlexiBAL<br>K013573 |
#### Description of the Devices [807.92(a)(4)] 4.0
The MED-EL "Passive Middle Ear Implants" (PMEI) Tympanoplasty Partial Prostheses are partial ossicular replacement prostheses which restore the mechanical sound transmission to the oval window.
They are offered in different designs and fixed or adjustable functional length to allow the surgeon to have more flexibility in terms of selecting the correct prosthesis according to the individual patient's anatomical needs, and the surgeon's surgical preferences and proficiency.
## Table 3 Description of the MED-EL PMEI Tympanoplasty Partial Prostheses
| Device/Trade Name | mXACT<br>Partial Prosthesis | mXACT PRO<br>Partial Prosthesis Kit | mCLIP<br>Partial Prosthesis | mCLIP ARC<br>Partial Prosthesis |
|-------------------|---------------------------------|-----------------------------------------|---------------------------------|-------------------------------------|
| Design | Image: mXACT Partial Prosthesis | Image: mXACT PRO Partial Prosthesis Kit | Image: mCLIP Partial Prosthesis | Image: mCLIP ARC Partial Prosthesis |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for MED-EL, a company that specializes in hearing implants. The logo is in red and features the company name in bold, sans-serif font. A stylized graphic of an ear is placed between the "D" and "E" in MED-EL.
| System/Kit<br>Components<br>and<br>Accessories | N/A | System/Kit content:<br>1 mXACT PRO Partial<br>Prosthesis + 5 Sizers for<br>partial prostheses | N/A |
|------------------------------------------------|---------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Coupling<br>structure | Round & "bell-shaped" for optimized stability<br>on the stapes head | Universal clip for standardized<br>coupling on the stapes head | |
| Headplate | Oval with optional<br>bending<br>possibility | Oval & length adjustable<br>(through snap-<br>mechanism)<br>The surgeon can move<br>the "shaft" inside the<br>headplate to the desired<br>functional length,<br>afterwards the<br>connection between<br>headplate and shaft is<br>fixed through the snap-<br>mechanism inside the<br>headplate. As a last step<br>the overlapping titanium<br>shaft of the prosthesis is<br>cut-off with a scalpel. | Round with a<br>ball-joint<br>connection to the<br>shaft.<br>The additional<br>ball-joint allows<br>the prosthesis to<br>adapt itself to<br>postoperative<br>tympanic<br>membrane<br>movements (for<br>example: caused<br>by the healing<br>process). |
| Headplate | | Round with<br>optional bending<br>possibility | |
| Material | Medical grade<br>titanium | mXACT PRO Partial<br>Prosthesis:<br>Medical grade titanium<br>Sizer for partial<br>prosthesis: Polypropylene | Medical grade titanium |
| Sterile /<br>Single-Use | | Yes | |
#### Statement of the Intended use [807.92(a)(5)] 5.0
The intended use and indications of the MED-EL PMEI Tympanoplasty Partial Prostheses and of the predicate devices manufactured by Heinz Kurz GmbH are substantially equivalent.
Comparing the subject devices and the respective predicate devices, the same medical conditions or underlying pathologies can be treated, and the principal functionality of the devices is the same, hence no different questions of safety and effectiveness arise.
## PMEI Tympanoplasty Partial Prostheses
## Intended Use
The passive middle ear implant - tympanoplasty partial prosthesis is intended to be used for replacement of components of the ossicular chain, if at least the stapes footplate is present and functional. The tympanoplasty partial prosthesis is implanted in the middle ear to restore sound transmission from the tympanic membrane to the oval window by replacing the ossicles partially. The tympanoplasty prosthesis is a medical device for single use delivered in sterile condition.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for MED-EL. The logo is in red and features the company name in bold, sans-serif font. To the right of the name is a circular graphic with curved lines, also in red.
## Intended User
The tympanoplasty prosthesis is intended to be implanted by qualified ENT surgeons only, with adequate skills to perform otological surgeries. Replacement of the ossicular chain is a standard surgical procedure, no additional specific device training is mandatory or required for the safe and effective use.
## Target Patient Population
The target patient population for the passive middle ear implant are patients of all ages requiring reconstruction of the ossicular chain.
## Indications
The tympanoplasty prosthesis is indicated to treat patients with:
- congenital or acquired ossicular chain defects (e.g., chronic otitis media, traumatic injury, malformation, cholesteatoma).
- inadequate conductive hearing from previous middle ear surgery.
The partial ossicular replacement prosthesis is indicated, if at least the stapes head and its stapes footplate are present and functional.
## Tympanoplasty Sizers for partial prosthesis4
## Intended Use
The tympanoplasty Sizers are intended to be used during a partial ossicular replacement surgery to determine the required functional length of the MED-EL tympanoplasty prosthesis. The Tympanoplasty Sizers are optional, single-use surgical tools for intraoperative transient use only and delivered in sterile condition.
The Sizers for partial prosthesis are intended to intraoperatively determine the distance between the head of the stapes and the tympanic membrane, thereby determining the required functional length of the MED-EL tympanoplasty prosthesis.
## Indications
The indications of the PMEI tympanoplasty partial prostheses apply.
<sup>1</sup> Note that the "italic text" in this section shall be used if the tympanoplasty Sizers are packaged together with the mXACT PRO Partial Prosthesis conforming the mXACT PRO Partial Prosthesis Kit. Traditional 510(k) PMEI TYM Partial
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for MED-EL, a company that specializes in hearing implants. The logo is in red and features the company name in bold letters. To the right of the name is a stylized graphic that resembles an ear.
#### 6.0 Comparison on Technological Characteristics with predicate devices [807.92(a)(6)]
The MED-EL PMEls have the same technological characteristics (i.e. design, material composition, etc.) as the predicate devices identified in paragraph |a||3). Hence, no difference in safety and effectiveness of the MED-EL PMET Tympanoplasty Partial Prostheses as compared to the predicate devices from Heinz CmbH is expected. A summary of the technological characteristics is provided in the following table.
#### 6.1 PMEI Tympanoplasty PARTIAL Prostheses - Comparison with predicate devices
| MED-EL<br>Device /<br>Predicate<br>Device | "mXACT Partial<br>Prosthesis" | Predicate<br>Device [A]<br>"TTP-<br>Tuebingen<br>Type BELL<br>Partial<br>Prosthesis"<br>(K972492) | "mXACT PRO<br>Partial<br>Prosthesis Kit" | Predicate Device<br>[B]<br>"TTP-VARIAC<br>System Partial"<br>(K990923) | "mCLIP Partial<br>Prosthesis" | Predicate<br>Device [B]<br>"CLIP Partial<br>Prosthesis<br>Dresden Type"<br>(K013573) | "mCLIP ARC<br>Partial<br>Prosthesis" | Predicate Device<br>[B]<br>"Clip Partial<br>FlexiBAL"<br>(K013573) | |
|-------------------------------------------|-------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|--------------------------------------------------------------------|--|
| Design | Image: mXACT Partial Prosthesis | Image: TTP-Tuebingen Type BELL Partial Prosthesis | Image: mXACT PRO Partial Prosthesis Kit | Image: TTP-VARIAC System Partial | Image: mCLIP Partial Prosthesis | Image: CLIP Partial Prosthesis Dresden Type | Image: mCLIP ARC Partial Prosthesis | Image: Clip Partial FlexiBAL | |
| | mXACT features a slit-headplate<br>design for optional bending<br>possibility | System/Kit content:<br>1x Partial ossicular replacement<br>prosthesis<br>5 Sizers<br>1 Sizer Disc<br>(including 6 | | mCLIP features a slit-headplate<br>design for optional bending<br>possibility and a clip with 8<br>"legs" instead of 7 | | Both devices have a ball-joint in the<br>headplate.<br>mCLIP ARC has a round and centric<br>headplate | | | |
| | | | | Sizers) | | | | | |
| | | | Design: | | | | | | |
| | | | The mXACT<br>headplate<br>works with a<br>snap-closing<br>mechanism | TTP-VARIAC<br>headplate with<br>plastic<br>deformation | | | | | |
| Attachment | | Bell-shaped coupling structure for the placement onto the stapes head | | | Clip-shaped coupling structure for the placement onto the stapes head | | | | |
| | | | | 1 length adjustable partial<br>prosthesis | | | | | |
| # of sizes | 10 | 8 | | 5 Sizers<br>1 Sizer Disc<br>containing 6<br>Sizers | 10 | 8 | 10 | 8 | |
| Functional<br>lengths [mm] | 0.75, 1.00,<br>1.25, 1.50,<br>1.75, 2.00,<br>2.25, 2.50,<br>3.00, 3.50 | 0.75, 1.00,<br>1.25, 1.50,<br>1.75, 2.00,<br>2.25, 2.50 (*) | | length adjustable partial<br>prosthesis can be adjusted to a<br>functional length from 0.75 to<br>3.50<br>Sizers functional length:<br>1.0, 1.5, 2.0,<br>2.5, 3.5<br>1.0, 1.5, 2.0,<br>2.5, 3.0, 3.5 | 0.75, 1.00,<br>1.25, 1.50,<br>1.75, 2.00,<br>2.25, 2.50,<br>3.00, 3.50 | 0.75, 1.00,<br>1.25, 1.50,<br>1.75, 2.00,<br>2.25, 2.50<br>(*) | 0.75, 1.00, 1.25,<br>1.50, 1.75, 2.00,<br>2.25, 2.50, 3.00,<br>3.50 | 0.75, 1.00, 1.25,<br>1.50, 1.75, 2.00,<br>2.25, 2.50 (*) | |
| Shaft Ø [mm] | 0.20 | | | | | | | | |
| Headplate<br>dimensions<br>[mm] | | | Ø2.60 x 3.60 (oval and off-centric) | | | Ø2.60 (round and centric) | Ø2.60 (round<br>and centric) | Ø2.60 x 3.60<br>(oval and off-<br>centric) | |
| Materials | Medical grade titanium ASTM F67 | Medical grade<br>titanium ASTM<br>F67 (bell and<br>shaft) &<br>ASTMF136<br>(headplate); | | Medical grade titanium ASTM F67 | | Medical grade<br>titanium ASTM<br>F67, except the<br>micro-ball which<br>can be out of<br>ASTMF67 or ISO<br>5832-2; | Medical grade<br>titanium ASTM<br>F67 | | |
| Biocompatible | | | | Yes (EN ISO 10993) | | | | | |
| Surgical Tools | N/A | | | Titanium Micro | N/A | | | | |
| (mandatory) | | | | Closing Forceps<br>&<br>Cutting Forceps | | | | | |
| Surgical Tools<br>(optional) | Tympanoplasty<br>Sizers for<br>partial<br>prostheses | ACSizer Disc<br>Partial.<br>Malleus<br>Handle. Cavity<br>Bending pliers. | Tympanoplasty<br>Sizers for<br>partial<br>prostheses<br>(included in<br>the Kit) | ACSizer Disc Partial<br>(included in the<br>sales package).<br>KURZ Titanium<br>Tweezers.<br>Micro Scissors | Tympanoplasty<br>Sizers for<br>partial<br>prostheses | ACSizer Disc<br>Partial | Tympanoplasty<br>Sizers for partial<br>prostheses | ACSizer Disc Partial | |
| Packaging | One length variant per package;<br>Single-Use; Sterile | | | One length variable partial<br>ossicular replacement prosthesis +<br>5 Sizers for partial prostheses.<br>Single-Use; Sterile | | One length variant per package; Single-Use; Sterile | | | |
| MRI | MRI Conditional at 1.5, 3.0 and 7.0 T | | | | | | | | |
## Table 4: PMEI Tympanoplasty Partial Prostheses - Comparison of Technical Characteristics
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for MED-EL, a company that specializes in hearing implants. The logo is in red and features the company name in a bold, sans-serif font. The "O" in MED-EL is stylized with a circular design, adding a unique touch to the logo.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for MED-EL, a company that specializes in hearing implants. The logo is in red and features the company name in bold, sans-serif font. To the right of the "D" in MED, there is a stylized graphic of a cochlea, which is the part of the inner ear responsible for hearing. The logo is simple and modern, and it effectively communicates the company's focus on hearing solutions.
* The predicate device uses a different type of lent in the finctional length of the device (functional length is the distance between stages head and the tympanic nembrane), whereas for the overal length is specified as part of the mentioned 510k. According to the predicate device (see https://www.kurzmed.com/), the overall length of each variant is the functional length plus 1mm.
Comparing the MED-EL PMEI Tympanoplasty Prostheses and their identified predicate devices manufactured by Heinz Kurz GmbH, the, biocompatibility, packaging configuration and MRI compatibility are IDENTICAL.
The following features are considered EQUIVALENT: device design, number and dimensions of length varials in body contact. As there is no difference in technological characteristics, there is no difference in safety and/or effectiveness.
There is just one difference: the surgical tools.
Sizers that can be used optionally determine the required length of the prosthess to be used are available for both, the subject devices as well as for the identified predicate devices.
The MED-EL tympanoplasty Sizers are either available separately (Class ) or packaged together with the mXACT PRO Partial Proshess conforming the mXACT PRO Partial Prosthess, for the predicate devices, also additional optional tools (Malleus Handle Cavity Bending pliers, KURZ Titanium Tweezers, and Micro Scissors) are prosthesis length, the TTP-VARIAC System Partial (Heinz Kurz GmbH) requires the surgeon to use special tools for closing the headplate of the MED-EL mXACT PRO Partial Prosthesis can be closed with e.g. standard delicate tweezers. As this is considered an improvement, no difference in safety and effectiveness is expected. The safety and effectiveness of the MED-EL mXACT PRO Partial Prosthesis headplate was confirmed in non-clinical testing.
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for MED-EL, a company that specializes in hearing implants. The logo is in red and features the company name in bold, sans-serif font. To the right of the name is a stylized graphic of an ear, also in red. The logo is simple and modern, and the use of red conveys a sense of energy and innovation.
#### 7.0 Non-clinical Test [807.92(b)(1)]
Non-clinical testing was conducted to assess the performance data for a determination of substantial equivalence.
The following tests were conducted:
- . MRI environment according to ASTM F2119, F2052, F2182 с MRI Conditional at 1.5T, 3.0T and 7.0T
- . Biocompatibility according to EN ISO 10993
- The performed evaluation provides objective evidence to support the o conclusion that the devices can be considered biocompatible for their intended use.
- Shelf-Life testing according to EN ISO 11607
- Sterilization Validation according to EN ISO 11137-1, EN ISO 11137-2. ●
- Gamma Sterilization with a confidence sterility assurance level of <10 o
- . Packaging Validation according EN ISO 11607
- . Mechanical & other testing
Safety and effectiveness have been demonstrated with non-clinical testing, concluding that the design and performance requirements are met.
#### 8.0 Clinical Test [807.92(b)(2)]
The benefit for patients of the PMEI Tympanoplasty Partial Prostheses has been successfully demonstrated and the clinical outcomes of the MED-EL PMEL Tympanoplasty Partial Prostheses are comparable to the predicate devices as per the state of the art for these types of devices.
The goal of ossicular replacement prosthesis is to close the postoperative pure tone average (PTA4) air-bone gap (ABG) to within 20 dB. According to the state of the art, as assessed in several systematic literature reviews and meta-analysis, 62.5-71.3 % of patients implanted with a partial prosthesis, achieve a postoperative PTA« ABG ≤20 dB.
MED-EL conducted a large multicenter, retrospective, single-subjects repeated measures (each subject served as his or her own control) Post-Market Clinical Follow-Up (PMCF) investigation with the PMEI Tympanoplasty Partial Prostheses in Europe:
- Audiological outcomes:
- O Number of patients: 186 (89 female, 97 male)
- Age: mean 43.9 ± 18.6 years (range 5-81) o
- Implanted ear: 95 right, 91 left o
- Follow-up: median 70.0 days (mean 78.4 ± 55.3; range 3-302 days) O
- Prosthesis type: 48 mXACT Partial, 15 mXACT PRO Partial, 68 mCLIP o Partial, 55 mCLIP ARC Partial
- % patients with a postoperative PTA4 ABG ≤20 dB: 72 % O
- Mean BC-PTA« pre/post-surgery: 19.8 ± 12.2 dB HL / 17.9 ± 12.2 dB HL
- BC-PTA4 deterioration >10 dB HL: 6/186 patients = 3.2 % (range 11.3- O 23.5 dB HL)
- . Adverse events:
O
- Number of analyzed patients: 202 (98 female, 104 male) o
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for MED-EL, a company that specializes in hearing implants. The logo is in red and features the company name in a bold, sans-serif font. To the right of the name is a circular graphic with curved lines, possibly representing sound waves or the inner ear. The logo is simple, modern, and easily recognizable.
- Age: mean 44.5 ± 18.7 vears (range 5-84) o
- Implanted ear: 107 right, 95 left o
- Follow-up: median 275.0 days (mean 284.0 ± 176.9; range 0-666 days) o
- Prosthesis type: 52 mXACT Partial, 16 mXACT PRO Partial, 72 mCLIP o Partial, 62 mCLIP ARC Partial
- Number of (recommended) revision surgeries: 5 O
- Number of (potential) device dislocations/extrusions: 3 o
- o Total 34 AEs in 21 patients: 7 Device unrelated AEs, 4 Medical related AEs, 15 Clinical/surgical related AEs, 8 AEs "unknown", 2 Follow-up AE
In the above mentioned PMCF investigation the safety and effectiveness of the PMEI Tympanoplasty Partial Prostheses was confirmed as demonstrated by postoperative PTA4 ABG ≤20 dB rates comparable to the state of the art, low rates of device dislocations/extrusions and revision surgeries, and overall stable bone conduction thresholds, an indicator that residual inner ear hearing is not compromised after implantation with an ossicular replacement prosthesis. Safety and effectiveness of the MED-EL PMEI Tympanoplasty Partial Prostheses are comparable to the current state of the art.
#### 9.0 Conclusion [807.92(b)(3)]
The results of the non-clinical testing demonstrate that:
- the MED-EL PMEI Tympanoplasty Partial Prostheses meet the established . specifications, the design and performance requirements to ensure safety and effectiveness according to the intended use.
- . there is no difference in safety and/or effectiveness between MED-EL PMEI Tympanoplasty Partial Prostheses and the predicate devices manufactured by Heinz Kurz GmbH.
It can be concluded that the PMEI Tympanoplasty Partial Prostheses are substantially equivalent to the predicate devices manufactured by Heinz Kurz GmbH.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.