The intended use of the Tyber Medical Anatomical Plating System is to bridge or otherwise stabilize bone fragments to facilitate healing. It is composed of the Mini-Fraq System and Ankle Fracture/ Fusion System with the following indications: Indications for Use: Mini-fraq System The Tyber Medical Mini-Fraq System is indicated for fixation of fractures, osteotomies, nonunions, malunions, replantations, and fusions of short bones and small fragments of bone including, but not limited to, the hand, wrist, foot, and ankle. The mini-frag system is also intended for reduction and stabilization of non-load bearing long bone fragments. The Tyber Medical Mini-frag System is not for Spinal Use. Ankle Fracture/Fusion System Tyber Medical Ankle Fracture/Fusion System is indicated for Use in: Fixation of fractures of the distal tibia included, but not limited to, ankle fractures, periarticular fractures, corrective osteotomies, non-unions, intra- and extra- articular and distal tibia fractures with a shaft extension, and malleolar fractures; - In intra- and extra articular fractures, osteotomies, medial malleolar fractures and nonunions of the metaphyseal and diaphyseal region of the distal fibula, and calcaneus; - In distal tibia/fibula long bones which include the metaphyseal regions of the tibia and fibula in the ankle. The Tyber Medical Ankle Fracture/Fusion Plating System is not for Spinal Use.
Device Story
System consists of bone plates, locking/non-locking screws, and optional bone screw washers; used for bone fragment stabilization and healing. Operates via mechanical fixation of bone fragments; locking screws allow on-axis or 15-degree off-axis placement. Washers aid load distribution at screw head/bone interface. Used in clinical settings by orthopedic surgeons for trauma and reconstruction. Output is physical stabilization of bone; healthcare providers use intraoperative imaging to verify placement. Benefits include restored bone alignment and facilitated healing.
Clinical Evidence
Bench testing only. No clinical data provided. Design control activities and verification testing demonstrate substantial equivalence.
Technological Characteristics
Materials: Ti-6Al-4V ELI (ASTM F136) or Stainless Steel (316L). Surface treatment: Anodized Type II. Components: Bone plates, locking/non-locking screws (2.0mm-4.0mm), and washers. Locking mechanism: On-axis or 15-degree off-axis. Sterilization: Offered sterile and non-sterile.
Indications for Use
Indicated for fixation of fractures, osteotomies, nonunions, malunions, replantations, and fusions of short bones/small fragments (hand, wrist, foot, ankle) and reduction/stabilization of non-load bearing long bone fragments (Mini-Fraq). Also indicated for distal tibia/fibula fractures, periarticular fractures, corrective osteotomies, nonunions, and malleolar fractures (Ankle Fracture/Fusion). Contraindicated for spinal use.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
May 31, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Tyber Medical, LLC Nicole Merlini Regulatory Affairs Specialist 83 S Commerce Way Suite 310 Bethlehem, Pennsylvania 18017
#### Re: K241218
Trade/Device Name: Tyber Medical Anatomical Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC, HTN Dated: April 30, 2024 Received: May 1, 2024
#### Dear Nicole Merlini:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Tejen D. Soni -S
For
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
### Indications for Use
Submission Number (if known)
K241218
Device Name
Tyber Medical Anatomical Plating System
#### Indications for Use (Describe)
Intended Use:
The intended use of the Tyber Medical Anatomical Plating System is to bridge or otherwise stabilize bone fragments to facilitate healing. It is composed of the Mini-Fraq System and Ankle Fracture/ Fusion System with the following indications:
Indications for Use:
Mini-fraq System
The Tyber Medical Mini-Fraq System is indicated for fixation of fractures, osteotomies, nonunions, malunions, replantations, and fusions of short bones and small fragments of bone including, but not limited to, the hand, wrist, foot, and ankle. The mini-frag system is also intended for reduction and stabilization of non-load bearing long bone fragments. The Tyber Medical Mini-frag System is not for Spinal Use.
Ankle Fracture/Fusion System
Tyber Medical Ankle Fracture/Fusion System is indicated for Use in:
Fixation of fractures of the distal tibia included, but not limited to, ankle fractures, periarticular fractures, corrective osteotomies, non-unions, intra- and extra- articular and distal tibia fractures with a shaft extension, and malleolar fractures;
- In intra- and extra articular fractures, osteotomies, medial malleolar fractures and nonunions of the metaphyseal and diaphyseal region of the distal fibula, and calcaneus;
- In distal tibia/fibula long bones which include the metaphyseal regions of the tibia and fibula in the ankle.
The Tyber Medical Ankle Fracture/Fusion Plating System is not for Spinal Use.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary Prepared on: 05/29/2024 Contact Details 21 CFR 807.92(a)(1) Tyber Medical, LLC Applicant Name 83 S Commerce Way Suite 310 Bethlehem PA 18017 United States Applicant Address 484-274-4471 Applicant Contact Telephone Mrs. Nicole Merlini Applicant Contact nmerlini@tybermed.com Applicant Contact Email Device Name 21 CFR 807.92(a)(2) Tyber Medical Anatomical Plating System Device Trade Name Single/multiple component metallic bone fixation appliances and Common Name accessories Plate, Fixation, Bone Classification Name 888.3030, 888.3040 Regulation Number HRS, HTN, HWC Product Code(s)
# Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
21 CFR 807.92(a)(4
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|-------------|----------------------------------------------------------|--------------|
| K153575 | Tyber Medical Trauma Screw System | HWC |
| K232652 | Tyber Medical Anatomical Plating System | HRS |
# Device Description Summary
Note: The only subject devices of this Special 510(k) submission are the optional bone screw Washers are only to be used with the non-locking screws in the Mini-fraq and Ankle Fracture/Fusion Systems.
Tyber Medical Anatomical Plating System Information
The systems presented in this Special 510(k) submission consist of various bone plates, locking screws, washers, and instruments. The Tyber Medical Anatomical Plating System consists of the following categories:
1.Mini-frag System
2.Ankle Fracture/Fusion System
A description of each system is provided below:
Mini-frag/Small Bone System
The Tyber Medical Mini-frag System is designed to address a variety of indications for short bones and sincluding, but not limited to, the hand, wrist, foot and ankle.
The System will incorporate both Locking Screws and standard non-locking Screws in 2.0, 2.5, and 4.0mm sizes in various lengths. Optional bone screw Washers are available for the non-locking Screws in 2.0mm sizes. All locking screws can be placed on-axis with the internal plate threads or up to 15 degrees off axis in any direction.
{4}------------------------------------------------
All plates are composed of Ti-6Al-4V ELI (ASTM F136) or Stainless Steel (316L) with an Anodized Type II surface treatment. All screws and optional bone screw washers are composed of Ti-6Al-4V ELI (ASTM F136) or Stainless Steel (316L) with color anodized for sizing.
A full set of Ancillary Instrumentation is available with the system. The instruments are not specific to the system. The Tyber Medical Mini-frag/Small Bone System is offered both sterile and non-sterile.
Ankle Fracture/Fusion System
The Tyber Medical Ankle Fracture/Fusion System is designed to address a variety of indications in ankle reconstruction mid-shaft and distal tibia/fibula fixation surgery.
The System will incorporate both locking Screws and standard non-locking screws in 2.7, 3.0, 3.5, and 4.0mm sizes in various lengths. Optional bone screw Washers are available for the non-locking Screws in 2.7, 3.5, and 4.0mm sizes. All locking screws can be placed onaxis with the internal plate threads or up to 15 degrees off axis in any direction.
All plates are composed of Ti-6Al-4V ELI (ASTM F136) or Stainless Steel (316L) with an Anodized Type II surface treatment. All screws and optional screw washers are composed of Ti-6AI-4V ELI (ASTM F136) or Stainless Steel (316L) with color anodized for sizing.
A full set of Ancillary Instrumentation) is available with the system. The instruments are not specific to the system.
The Tyber Medical Ankle Fracture/Fusion System is offered both sterile and non-sterile.
# Intended Use/Indications for Use
## 21 CFR 807.92(a)(5)
#### Intended Use:
The intended use of the Tyber Medical Plating System is to bridge or otherwise stabilize bone fragments to facilitate healing. lt is composed of the Mini-Frag System and Ankle Fracture/ Fusion System with the following indications:
Indications for Use:
Mini-fraq System
The Tyber Medical Mini-Fraq System is indicated for fixation of fractures, nonunions, malunions, replantations, and fusions of short bones and small fragments of bone including, but not limited to, the hand, wrist, foot, and ankle. The mini-frag system is also intended for reduction and stabilization of non-load bearing long bone fragments. The Tyber Mini-frag System is not for Spinal Use.
Ankle Fracture/Fusion System
Tyber Medical Ankle Fracture/Fusion System is indicated for Use in:
-Fixation of fractures of the distal tibia included, but not limited to, ankle fractures, corrective osteotomies, nonunions, intra- and extra- articular and distal tibia fractures with a shaft extension, and malleolar fractures;
-In intra- and extra articular fractures medial malleolar fractures and nonunions of the metaphyseal reqion of the distal fibula, and calcaneus;
•In distal tibia/fibula long bones which include the metaphyseal reqions of the tibia and fibula in the ankle. The Tyber Medical Ankle Fracture/Fusion Plating System is not for Spinal Use.
{5}------------------------------------------------
# Indications for Use Comparison
The subject indications for use are within those of the primary predicate and identical to the additional predicate indications.
# Technological Comparison
The subject plates and screws are identical to those cleared in the additional predicate system. The se in the primary predicate indications (aid in load distribution at the screw head/bone interface), design, dimensions, material, and operating principle.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
All design contol activities, including verification, and other testing, demonstrates that adding optional screw Washers to the Tyber Medical Anatomical Plating System does not raise any questions of safety or performance and is substantially equivalent to the predicate device(s).
21 CFR 807.92(a)(6)
21 CFR 807.92(a)(5)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.