← Product Code [DQK](/productcode/DQK) · K241197

# DeepRhythmAI (K241197)

_Medicalgorithmics S.A. · DQK · Dec 4, 2024 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K241197

## Device Facts

- **Applicant:** Medicalgorithmics S.A.
- **Product Code:** [DQK](/productcode/DQK.md)
- **Decision Date:** Dec 4, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1425
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** AI/ML, Software as a Medical Device

## Indications for Use

DeepRhythmAl is a cloud-based software for the assessment of cardiac arrhythmias using two lead ECG data in adult patients. It is intended for use by a healthcare solution integrator to build web, mobile or another types of applications to let qualified healthcare professionals review and confirm the analytic result. The product supports downloading and analyzing data recorded in the compatible formats from dedicated ambulatory ECG devices such as Holter, event recorder, Outpatient Cardiac Telemetry devices or other similar recorders when the assessment of the rhythm is necessary. The product can be electronically interfaced and perform analysis with data transferred from other computer-based ECG systems, such as an ECG management system. DeepRhythmAl can be integrated into medical devices. In this case, the medical device manufacturer will identify the indication for use depending on the application of their device. DeepRhythmAl is not for use in life-supporting or sustaining systems or ECG Alarm devices. Interpretation results are not intended to be the sole means of diagnosis. It is offered to physicians and clinicians on an advisory basis only in conjunction with the physician's knowledge of ECG patterns, patient background, clinical history, symptoms and other diagnostic information.

## Device Story

DeepRhythmAI is cloud-based software for automated ECG analysis; inputs two-lead ECG data from ambulatory recorders (Holter, event, telemetry) or ECG management systems; utilizes proprietary deep-learning algorithm to detect/measure cardiac beats and arrhythmias; outputs analysis results via API for integration into third-party web/mobile applications; intended for use by healthcare professionals to support clinical diagnosis; does not provide primary display/GUI; operates as advisory tool; benefits patient by providing automated rhythm assessment to assist physician interpretation.

## Clinical Evidence

Bench testing only. Performance validated against ANSI/AAMI/IEC 60601-2-47:2012/(R)2016 and AAMI/ANSI/EC57:2012 standards. Algorithm validated using a proprietary database (MDG validation db) containing large-scale recordings from the intended patient population, including validation of MLII and MLIII lead configurations.

## Technological Characteristics

Cloud-based software; API-based integration; no physical hardware or GUI; utilizes deep-learning algorithm for signal analysis; compatible with two-lead ECG data (MLII/MLIII); software development follows ANSI/AAMI/IEC 62304.

## Regulatory Identification

A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.

## Predicate Devices

- DeepRhythmAI ([K210822](/device/K210822.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 4, 2024

Medicalgorithmics S.A. Joanna Dabala Product Compliance Leader Aleje Jerozolimskie 81 Warsaw, 02-001 Poland

Re: K241197

Trade/Device Name: DeepRhythmAI Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DOK, DPS, OYX Dated: November 7, 2024 Received: November 7, 2024

Dear Joanna Dabala:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

# Jennifer W. Shih -S

Jennifer Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

Submission Number (if known)

K241197

Device Name

DeepRhythmAl

Indications for Use (Describe)

DeepRhythmAl is a cloud-based software for the assessment of cardiac arrhythmias using two lead ECG data in adult patients.

It is intended for use by a healthcare solution integrator to build web, mobile or another types of applications to let qualified healthcare professionals review and confirm the analytic result. The product supports downloading and analyzing data recorded in the compatible formats from dedicated ambulatory ECG devices such as Holter, event recorder, Outpatient Cardiac Telemetry devices or other similar recorders when the assessment of the rhythm is necessary.

The product can be electronically interfaced and perform analysis with data transferred from other computer-based ECG systems, such as an ECG management system. DeepRhythmAl can be integrated into medical devices. In this case, the medical device manufacturer will identify the indication for use depending on the application of their device.

DeepRhythmAl is not for use in life-supporting or sustaining systems or ECG Alarm devices. Interpretation results are not intended to be the sole means of diagnosis. It is offered to physicians and clinicians on an advisory basis only in conjunction with the physician's knowledge of ECG patterns, patient background, clinical history, symptoms and other diagnostic information.

| Type of Use (Select one or both, as applicable)                                                               |                                                                                               |
|---------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <div> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)             </div> | <div> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)             </div> |

|X | Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for MEDICALgorithmics. The logo is in blue and features the company name in a bold, sans-serif font. Below the company name is the tagline "INNOVATIVE SOLUTIONS IN MEDICINE" in a smaller, lighter font. The logo is simple and professional, and it conveys the company's focus on medical technology.

## 510(k) Summary

#### . Submitter's name and address:

Medicalgorithmics S.A. Aleje Jerozolimskie 81, 02-001 Warsaw, Poland

Contact Person: Joanna Dabała Phone: (+48) 733 888 448 Email: j.dabala@medicalgorithmics.com

Date Prepared:

2024-11-07

- ll. Device
Trade name:

DeepRhythmAl

870.1425 870.2340 870.1025

Class II

Common name: ECG Analysis System

Classification name: Programmable Diagnostic Computer/

Electrocardiograph/ Outpatient Cardiac Telemetry

Requlation number:

Regulatory Class:

DQK, DPS, QYX Classification Product code:

#### Substantial Equivalence lll.

The selected predicate device is:

- 1. DeepRhythmAI, K210822 (Predicate Device),
#### IV. Device description

The DeepRhythmAl is a cloud-based software for automated analysis of ECG data. It uses a scalable Application Programming Interface (API) to enable easy integration with other medical products. The main component of DeepRhythmAl is an automated proprietary deep-learning algorithm, which measures and analyzes ECG data to provide qualified healthcare professional with supportive information for review.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for MEDICALgorithmics. The word "MEDICALgorithmics" is written in a bold, sans-serif font, with the first part of the word in a dark blue color and the second part in a slightly lighter blue. Below the main logo, there is a tagline that reads "INNOVATIVE SOLUTIONS IN MEDICINE" in a smaller, sans-serif font.

DeepRhythmAl can be integrated into medical devices. The product supports downloading and analyzing data recorded in compatible formats from dedicated ambulatory ECG devices such as Holter, event recorder, Outpatient Cardiac Telemetry devices or other similar recorders used when assessment of the rhythm is necessary. The DRAI can also be electronically interfaced and perform analysis with data transferred from other computer-based ECG systems, such as an ECG management system. DeepRhythmAl doesn't have User Interface therefore it should be integrated with the external visualization software used by the ECG technicians for the ECG visualization and analysis reporting.

DeepRhythmAl is not for use in life supporting or sustaining systems or ECG Alarm devices. Interpretation results are not intended to be the sole means of diagnosis. It is offered to physicians and clinicians on an advisory basis only in conjunction with the physician's knowledge of ECG patterns, patient background, clinical history, symptoms, and other diagnostic information.

#### V. Indications for use

DeepRhythmAl is a cloud-based software for the assessment of cardiac arrhythmias using two lead ECG data in adult patients.

It is intended for use by a healthcare solution integrator to build web, mobile or another types of applications to let qualified healthcare professionals review and confirm the analytic result. The product supports downloading and analyzing data recorded in the compatible formats from dedicated ambulatory ECG devices such as Holter, event recorder, Outpatient Cardiac Telemetry devices or other similar recorders when the assessment of the rhythm is necessary.

The product can be electronically interfaced and perform analysis with data transferred from other computer-based ECG systems, such as an ECG management system. DeepRhythmAl can be integrated into medical devices. In this case, the medical device manufacturer will identify the indication for use depending on the application of their device.

DeepRhythmAl is not for use in life-supporting or sustaining systems or ECG Alarm devices. Interpretation results are not intended to be the sole means of diagnosis. It is offered to physicians and clinicians on an advisory basis only in conjunction with the physician's knowledge of ECG patterns, patient background, clinical history, symptoms and other diagnostic information.

#### VI. Comparison to predicate device

The following tables provide a comparison of the detection features and device comparison of DeepRhythmAI and the predicate device.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for MEDICALgorithmics. The word "MEDICAL" is in large, bold, blue letters, with a blue square to the left of the word. The word "gorithmics" is in a smaller, but still bold, blue font. Below the company name is the phrase "INNOVATIVE SOLUTIONS IN MEDICINE" in a smaller, lighter font.

### Detection Features comparison:

| Device functionality                                                                         | Subject device<br>(K241197) | Predicate device<br>(K182344) |
|----------------------------------------------------------------------------------------------|-----------------------------|-------------------------------|
|                                                                                              | DeepRhythmAl                | DeepRhythmAl                  |
| QRS detection                                                                                | YES                         | YES                           |
| Heart rate determination for<br>non-paced adult                                              | YES                         | YES                           |
| R-R interval detection                                                                       | YES                         | YES                           |
| Non-paced arrhythmias<br>interpretation                                                      | YES                         | YES                           |
| Non-paced ventricular<br>arrhythmias calls                                                   | YES                         | YES                           |
| Atrial fibrillation detection                                                                | YES                         | YES                           |
| Cardiac beats detection<br>(Ventricular ectopic beats,<br>Supraventricular ectopic<br>beats) | YES                         | YES                           |
| Patient populations                                                                          | Adult                       | Adult                         |

Device comparison:

| Device<br>functionality                         | Subject device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Similarities/<br>Differences |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
|                                                 | DeepRhythmAl<br>(DRAI)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | DeepRhythmAl<br>(DRAI)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                              |
| Manufacturer                                    | Medicalgorithmics S.A.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Medicalgorithmics S.A.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | N/A                          |
| 510(k) Number                                   | K241197                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | K182344                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | N/A                          |
| Classification                                  | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Equivalent                   |
| Regulation<br>Number(s)                         | 21 CFR §870.1425<br>21 CFR §870.2340<br>21 CFR §870.1025                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 21 CFR §870.1425<br>21 CFR §870.2340                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Equivalent                   |
| Classification<br>name                          | Programmable<br>Diagnostic Computer,<br>Electrocardiograph.<br>Outpatient Cardiac<br>Telemetry                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Programmable<br>Diagnostic Computer,<br>Electrocardiograph                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Equivalent                   |
| Product Code                                    | DOK, DPS, QYX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | DOK, DPS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Equivalent                   |
| Device<br>functionality                         | Subject device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Similarities/<br>Differences |
|                                                 | DeepRhythmAl<br>(DRAI)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | DeepRhythmAl<br>(DRAI)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                              |
| Indications for<br>Use                          | DeepRhythmAl is a<br>cloud-based software for<br>the assessment of<br>cardiac arrhythmias<br>using two lead ECG data<br>in adult patients.<br>It is intended for use by<br>a healthcare solution<br>integrator to build web,<br>mobile or another types<br>of applications to let<br>qualified healthcare<br>professionals review and<br>confirm the analytic<br>result. The product<br>supports downloading<br>and analyzing data<br>recorded in the<br>compatible formats from<br>dedicated ambulatory<br>ECG devices such as<br>Holter, event recorder,<br>Outpatient Cardiac<br>Telemetry devices or<br>other similar recorders<br>when the assessment of<br>the rhythm is necessary.<br>The product can be<br>electronically interfaced<br>and perform analysis<br>with data transferred<br>from other computer-<br>based ECG systems,<br>such as an ECG<br>management system.<br>DeepRhythmAI can be<br>integrated into medical<br>devices. In this case, the<br>medical device<br>manufacturer will identify<br>the indication for use<br>depending on the<br>application of their<br>device. | DeepRhythmAl is a<br>cloud-based software for<br>the assessment of<br>cardiac arrhythmias<br>using two lead ECG data<br>in adult patients.<br>It is intended for use by<br>a healthcare solution<br>integrator to build web,<br>mobile or another types<br>of applications to let<br>qualified healthcare<br>professionals review and<br>confirm the analytic<br>result. The product<br>supports downloading<br>and analyzing data<br>recorded in the<br>compatible formats from<br>dedicated ambulatory<br>ECG devices such as<br>Holter, event recorder,<br>Mobile Cardiac<br>Telemetry or other<br>similar devices when the<br>assessment of the<br>rhythm is necessary.<br>The product can be<br>electronically interfaced<br>and perform analysis<br>with data transferred<br>from other computer-<br>based ECG systems,<br>such as an ECG<br>management system.<br>DeepRhythmAl can be<br>integrated into medical<br>devices. In this case, the<br>medical device<br>manufacturer will identify<br>the indication for use<br>depending on the<br>application of their<br>device. | Equivalent                   |
| Device<br>functionality                         | Subject device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Similarities/<br>Differences |
|                                                 | DeepRhythmAI<br>(DRAI)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | DeepRhythmAI<br>(DRAI)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                              |
|                                                 | DeepRhythmAI is not for<br>use in life-supporting or<br>sustaining systems or<br>ECG Alarm devices.<br>Interpretation results are<br>not intended to be the<br>sole means of diagnosis.<br>It is offered to physicians<br>and clinicians on an<br>advisory basis only in<br>conjunction with the<br>physician's knowledge of<br>ECG patterns, patient<br>background, clinical<br>history, symptoms and<br>other diagnostic<br>information.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | DeepRhythmAI is not for<br>use in life-supporting or<br>sustaining systems or<br>ECG Alarm devices.<br>Interpretation results are<br>not intended to be the<br>sole means of diagnosis.<br>It is offered to physicians<br>and clinicians on an<br>advisory basis only in<br>conjunction with the<br>physician's knowledge of<br>ECG patterns, patient<br>background, clinical<br>history, symptoms and<br>other diagnostic<br>information.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                              |
| Level of Concern                                | Moderate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Moderate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Equivalent                   |
| Components                                      | Software only:<br>1) A web API<br>2) An automated<br>proprietary algorithm.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Software only:<br>1) A web API<br>2) An automated<br>proprietary algorithm.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Equivalent                   |
| Interface                                       | Web application<br>programming interface<br>(API)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Web application<br>programming interface<br>(API)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Equivalent                   |
| Part responsible<br>for ECG signal<br>analysis  | The automated<br>proprietary deep-<br>learning algorithm, which<br>measures and analyzes<br>ECG data to provide<br>qualified healthcare<br>professional with<br>supportive information<br>for review.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | The automated<br>proprietary deep-<br>learning algorithm, which<br>measures and analyzes<br>ECG data to provide<br>qualified healthcare<br>professional with<br>supportive information<br>for review.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Equivalent                   |
| Display or<br>Graphical User<br>Interface (GUI) | No primary display or<br>GUI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | No primary display or<br>GUI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Equivalent                   |

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for MEDICALgorithmics. The logo is in blue and features the company name in a bold, sans-serif font. Below the company name is the tagline "INNOVATIVE SOLUTIONS IN MEDICINE" in a smaller font. The logo is simple and professional, and it conveys the company's focus on medical technology.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for MEDICALgorithmics. The text "MEDICALgorithmics" is in a bold, blue font. Below the company name is the text "INNOVATIVE SOLUTIONS IN MEDICINE" in a smaller, thinner font, also in blue.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for MEDICALgorithmics. The word "MEDICAL" is in bold, blue letters, and the word "gorithmics" is in a smaller font size and is also blue. Below the logo, the words "INNOVATIVE SOLUTIONS IN MEDICINE" are written in a smaller font size.

Device comparison summary:

The algorithm responsible for arrhythmia detection has been changed. Performance data was evaluated on the same requirements according to the

international standards ANSI/AAMI EC57:2012 and ANSI/AAMI/IEC 60601-2-47:2012/(R)2016 as the predicate device. Patient population and monitoring environments for devices are equivalent.

The subject device is considered substantially equivalent to the predicate devices.

#### VII. Summary of performance data

The DeepRhythmAl software for arrhythmia detection and automated analysis of ECG data has been subjected to performance testing according to the recognized consensus standards, ANSI/AAMI/IEC 60601-2-47:2012/(R)2016 and AAMI/ANSI/EC57:2012, Moreover, to enable robust device validation, the algorithm was tested against the proprietary database (MDG validation db) that includes a large number of recordings captured among the intended patient population. MDG validation database allowed for validation of additional compatible hardware configurations (MLII and MLIII leads)

Medicalgorithmics followed ANSI/AAMI/IEC 62304 and the FDA Guidance Document, "General Principles of Software Validation: Final Guidance for Industry and FDA Staff" (January, 2002) with respect to software development and validation.

All necessary testing was conducted on the DeepRhythmAl to support a determination of substantial equivalence to the predicate and reference devices. Test results confirm that DeepRhythmAl meets its intended use.

#### VIII. Conclusion

In conclusion, DeepRythmAl has the same intended use as the predicate device, and any differences in technological characteristics do not raise different questions of safety or effectiveness. Therefore, DRAI is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K241197](https://fda-staging.innolitics.com/device/K241197)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).