K241144 · Wontech Co., Ltd. · GEX · Jun 24, 2024 · General, Plastic Surgery
Device Facts
Record ID
K241144
Device Name
Picosecond Nd:YAG Laser (PICOCAREMAJESTY)
Applicant
Wontech Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 24, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PICOCAREMAJESTY is indicated for the following at the specified wavelength: · 1064 nm: Removal of tattoos for all skin types (Fitzpatrick I-VI) to treat the following tattoo colors: black, brown, green, blue and purple. · 532 nm: Removal of tattoos for Fitzpatrick skin types I-III to treat the following tattoo colors: red, yellow and orange. · Treatment of benign pigmented lesions for Fitzpatrick skin types I-IV.
Device Story
Solid-state Nd:YAG laser system delivering 1064 nm or 532 nm wavelengths in picosecond pulses. System comprises main unit, articulated arm, handpiece, and foot switch. User selects fluence and frequency via LCD touch panel on main unit; energy output adjusted automatically based on selected fluence. Laser energy delivered to skin via handpiece for tattoo removal and treatment of benign pigmented lesions. Used in clinical settings by trained professionals. Output provides therapeutic laser energy to target chromophores in skin; benefits include removal of tattoos and pigmented lesions.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by bench testing, including electrical safety (AAMI ES60601-1), electromagnetic compatibility (IEC 60601-1-2), usability (IEC 60601-1-6), and laser safety (IEC 60825-1).
Technological Characteristics
Solid-state Nd:YAG laser; 1064 nm and 532 nm wavelengths; 275-300 ps pulse width; max 10 Hz repetition rate; 2-10 mm spot size. Handpiece tip material: Aluminum powder (CAS 7429-90-5). Delivery via articulated arm. Software: Moderate level of concern. Standards: AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-22, IEC 60825-1.
Indications for Use
Indicated for tattoo removal (1064 nm for Fitzpatrick I-VI, black/brown/green/blue/purple; 532 nm for Fitzpatrick I-III, red/yellow/orange) and treatment of benign pigmented lesions (Fitzpatrick I-IV).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 24, 2024
Won Tech Co., Ltd. Hyun Sik Yoon Team Leader 64 Techno8-ro 64, Yuseong-gu Daejeon, 34028 Korea, South
### Re: K241144
Trade/Device Name: Picosecond Nd:YAG Laser (PICOCAREMAJESTY) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: April 15, 2024 Received: April 25, 2024
Dear Hyun Sik Yoon:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Tanisha L. Digi Hithe -5 Date: 2024.06.24
15:13:13 -04'00' Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K241144
Device Name
Picosecond Nd:YAG Laser (PICOCAREMAJESTY)
Indications for Use (Describe)
The PICOCAREMAJESTY is indicated for the following at the specified wavelength:
· 1064 nm: Removal of tattoos for all skin types (Fitzpatrick I - VI) to treat the following tattoo colors: black,brown, green, blue and purple.
· 532 nm: Removal of tattoos for Fitzpatrick skin types I · III to treat the following tattoo colors: red, yellow and orange.
· Treatment of benign pigmented lesions for Fitzpatrick skin types I-IV.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circle with a stylized white "W" inside of it. To the right of the circle is the text "WONTECH", with "WON" in orange and "TECH" in gray. The logo is simple and modern, and the use of orange and gray gives it a professional look.
# 510(k) Summary K241144
[As required by 21 CFR 807.92]
#### 1. Date Prepared [21 CFR 807.92(a)(a)]
June 6th, 2024
#### Submitter's Information & Contact Person [21 CFR 807.92(a)(1)] 2.
- Name of Manufacturer: WON TECH Co., Ltd.
- Address: 64 Techno 8-ro, Yuseong-gu, Daejeon, 34028, Republic of Korea Hyun Sik Yoon - Contact Name: - Telephone No.: +82-10-6750-5346 - Fax No.: +82-70-7836-0110 - Email Address: yoonhs21@wtlaser.com
-
#### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)|
Common name: Nd:YAG Laser System Trade name: PICOCAREMAJESTY
| Classification Description | 21 CFR Section | Product Code |
|-----------------------------------|----------------|--------------|
| Powered Laser Surgical Instrument | 878.4810 | GEX |
As stated in 21 CFR, parts 878.4810, this generic types of devices has been classified as Class II.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circle with a stylized "W" inside, followed by the text "WONTECH". The "WON" part of the text is in orange, while the "TECH" part is in gray.
#### 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)|
The identified predicate devices within this submission are shown as follow:
#### #1 Predicate device
| • 510(k) Number: | K234104 |
|------------------------|-----------------------------------|
| • Applicant: | WON TECH Co., Ltd. |
| • Classification Name: | Powered Laser Surgical Instrument |
| • Trade Name: | PICOANDY |
#### #2 Predicate device
| • 510(k) Number: | K201773 |
|------------------------|-----------------------------------|
| • Applicant: | WON TECH Co., Ltd. |
| • Classification Name: | Powered Laser Surgical Instrument |
| • Trade Name: | PICOHI |
#### ર. Description of the Device [21 CFR 807.92(a)(4)]
The PICOCAREMAJESTY is the solid state laser capable of delivering energy at wavelengths of 1064mm or 532nm at short durations. The device system consists of a main unit, an articulated arm, a handpiece and a foot switch. The laser output is delivered to the skin through the articulated arm delivery system terminated by the handpiece. The fluence (energy density) and frequency are controlled from the LCD display/Touch Pad located on the front of the main unit. The LCD display is used to obtain feedback from the system, such as the number of pulses delivered or spot size selected.
For treatment, the user can select the appropriate fluence value. (The treatmented is provided in the user Manual) The energy is changed automatically in accordance with the selected fluence (down) button and then, the fluence is increased / decreased by 0.1 J/cm2.
The selectable fluence values are 0.2 to 10.0 J/cm2 at 1064nm and 0.1 to 2.5 J/cm2 at 532nm
#### 6. Indications for Use [21 CFR 807.92(a)(5)]
The PICOCAREMAJESTYis indicated for the following at the specified wavelength:
· 1064 nm: Removal of tattoos for all skin types (Fitzpatrick I-VI) to treat the following tattoo colors: black, brown, green, blue and purple.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circle with a stylized "W" inside, followed by the text "WONTECH" in a sans-serif font. The "WON" part of the text is in orange, while the "TECH" part is in gray.
- · 532 nm: Removal of tattoos for Fitzpatrick skin types I-III to treat the following tattoo colors: red, yellow and orange.
- · Treatment of benign pigmented lesions for Fitzpatrick skin types I-IV.
#### Determination of Substantial Equivalence [21 CFR 807.92(a)(6) and 21 CFR 807.92(b)] 7.
There are no significant differences between the PICOCAREMAJESTY and the predicate devices that would adversely affect the use of the product. It is substantially equivalent to this device in design, function, and technical characteristics.
| | Proposed Device | Predicate Device #1 | Predicate Device #2 | SE decision |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| K Number | - | K234104 | K201773 | - |
| Manufacturer | WON TECH Co., Ltd. | WON TECH Co., Ltd. | Hironic Co., Ltd | - |
| Model | PICOCAREMAJESTY | PICOANDY | PICOHI | - |
| Indications<br>for Use | The<br>PICOCAREMAJESTY is<br>indicated for the<br>following at the specified<br>wavelength:<br><br>• 1064 nm: Removal of<br>tattoos for all skin types<br>(Fitzpatrick I-VI) to treat<br>the following tattoo<br>colors: black, brown,<br>green, blue and purple.<br><br>• 532 nm: Removal of<br>tattoos for Fitzpatrick skin<br>types I-III to treat the<br>following tattoo colors:<br>red, yellow and orange.<br><br>• Treatment of benign<br>pigmented lesions for<br>Fitzpatrick skin types I-<br>IV. | The PICOANDY is<br>indicated for the<br>following at the specified<br>wavelength:<br><br>• 1064 nm: Removal of<br>tattoos for all skin types<br>(Fitzpatrick I-VI) to treat<br>the following tattoo<br>colors: black, brown,<br>green, blue and purple.<br><br>• 532 nm: Removal of<br>tattoos for Fitzpatrick<br>skin types I-III to treat<br>the following tattoo<br>colors: red, yellow and<br>orange.<br><br>• Treatment of benign<br>pigmented lesions for<br>Fitzpatrick skin types I-<br>IV. | The PICOHI is indicated<br>for the following at the<br>specified wavelength:<br><br>• 1064 nm: Removal of<br>tattoos for all skin types<br>(Fitzpatrick I-VI) to treat<br>the following tattoo<br>colors: black, brown,<br>green, blue and purple.<br><br>• 532 nm: Removal of<br>tattoos for Fitzpatrick<br>skin types I-III to treat<br>the following tattoo<br>colors: red, yellow and<br>orange.<br><br>• Treatment of benign<br>pigmented lesions for<br>Fitzpatrick skin types I-<br>IV. | Same |
| Anatomical<br>site | Skin and subcutaneous<br>tissue | Skin and subcutaneous<br>tissue | Skin and subcutaneous<br>tissue | Same |
| Wavelength | 1064 & 532nm | 1064 & 532nm | 1064 & 532nm | Same |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circular icon with a stylized "W" inside, followed by the text "WONTECH". The "WON" part of the text is in orange, while the "TECH" part is in gray.
| | Proposed Device | Predicate Device #1 | Predicate Device #2 | SE decision |
|--------------------------------|-----------------------------------------|-----------------------------------------|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Pulse Width | 275~300ps | 350~550 ps | 275~300ps | Same with Predicate<br>Device #2 |
| Pulse Energy | 1. Zoom 1064nm : 30 ~<br>500mJ ±10% | 1. Zoom 1064nm : 30 ~<br>550mJ ±10% | 1. Zoom 1064nm : 10~<br>508mJ ±20% | Diffenence<br>However, the pulse<br>energy of<br>PICOCAREMAJESTY<br>is included in those of<br>the Predicate Device<br>#2 |
| | 2. Zoom 532nm : 10 ~<br>250mJ ±10% | 2. Zoom 532nm : 10 ~<br>200mJ ±10% | 2. Zoom 532nm : 4 ~<br>255mJ ±20% | |
| | 3. Collimation Mode: 38<br>~ 500mJ ±10% | 3. Collimation Mode: 38<br>~ 500mJ ±10% | 3. Collimation Mode: 7 ~<br>471mJ ±20% | |
| | 4. MLA (1064nm): 30 ~<br>500mJ ±10% | 4. MLA (1064nm): 30 ~<br>550mJ ±10% | 4. ZMLA (1064nm): 5 ~<br>508mJ ±20% | |
| | 5. MLA (532nm): 10 ~<br>250mJ ±10% | 5. MLA (532nm): 10 ~<br>200mJ ±10% | 5. ZMLA (532nm): 2 ~<br>255mJ ±20% | |
| | 6. 1064nm DOE : 30 ~<br>450mJ ±10% | 6. 1064nm DOE : 30 ~<br>550mJ ±10% | 6. 1064nm DOE : 10 ~<br>500mJ ±20% | |
| | 7. 532nm DOE : 20 ~<br>200mJ ±10% | 7. 532nm DOE : 20 ~<br>200mJ ±10% | 7. 532nm DOE : 10~<br>250mJ ±20% | |
| Spot Size | 2 to 10 mm | 2 to 10 mm | 1.5 to 13 mm | Same |
| Pulse<br>Repetition | Max.10Hz | Max.10Hz | Max.10Hz | Same |
| Laser<br>Delivery<br>Type | Articulated Arm with<br>Handpiece | Articulated Arm with<br>Handpiece | Articulated Arm with<br>Handpiece | Same |
| Handpiece | Zoom(1064nm, 532nm) | Zoom(1064nm, 532nm) | Zoom(1064nm, 532nm) | Same |
| | Collimated(1064nm) | Collimated(1064nm) | Collimated(1064nm) | |
| | MLA (1064nm, 532nm) | MLA (1064nm, 532nm) | ZMLA (1064nm, 532nm) | |
| | DOE (1064nm, 532nm) | DOE (1064nm, 532nm) | DOE (1064nm, 532nm) | |
| Patient<br>Contact<br>Material | Aluminum powder<br>(Handpiece) | Aluminum powder<br>(Handpiece) | Aluminum powder<br>(Handpiece) | Same |
### Non-Clinical Test Summary |21 CFR 807.92(b)(1)]
1) Electrical Safety, Electromagnetic Compatibility Testing
Bench tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
| Standard<br>(Edition) | Standard Title |
|-----------------------|----------------|
|-----------------------|----------------|
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circle with a stylized white "W" inside of it. To the right of the circle is the text "WONTECH" in a sans-serif font. The "WON" part of the text is orange, while the "TECH" part is gray.
| AASI AAMI ES60601-<br>1:2005/(R)2012 and A1:2012 | Medical electrical equipment – Part 1: General requirements for<br>basic safety and essential performance |
|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ANSI AAMI IEC 60601-1-<br>2:2014 | Medical electrical equipment - Part 1-2: General requirements<br>for safety and essential performance - Collateral standard:<br>Electromagnetic compatibility - Requirements and tests |
| IEC 60601-1-6 Edition 3.1<br>2013 | Medical electrical equipment - Part 1-6: General requirements<br>for basic safety and essential performance - Collateral standard:<br>Usability |
| IEC 60601-2-22 Edition 3.1<br>2012-10 | Medical electrical equipment - Part 2-22: Particular<br>requirements for basic safety and essential performance of<br>surgical, cosmetic, therapeutic and diagnostic laser equipment |
| IEC 60825-1:2014 (Third<br>Edition | Safety of laser products - Part 1: Equipment classification and<br>requirements |
### 2) Software Validation
The PICOCAREMAJESTY contains MODERATE level of concern software was designed and developed according to a software development process and was verified and validated.
The software information is provided in accordance with FDA guidance: The content of premarket submissions for software contained in medical devices, on May 11, 2005.
### 3) Biocompatibility
| Part | Material | Patient<br>Contact | Duration of Contact<br>by ISO 10993-1 | Bio-<br>compatibility |
|------------------------------|-----------------------------------------|--------------------|---------------------------------------|-----------------------|
| Handpiece Tip of<br>PICOANDY | Aluminium Powder<br>(Cas No. 7429-90-5) | Intact Skin | Limited<br>(< 24 hours) | Yes |
- The material of handpiece for the PICOCAREMAJESTY is same to previous registered material of own ● product (PICOCARE Family <K181272>, ICOCAREMAJESTY <K212127>, and PICOANDY <K234104)) in formulation, processing, and sterilization, and no other chemicals have been added. Thus, biocompatibility is not necessary for this device, as this material has been deemed biocompatible in others previously cleared submissions.
- 4) Performance Testing
performance as The of the PICOCAREMAJESTY has been defined follows. -Laser wavelength: 1064nm±10%, 532nm±10% - Laser output power: 1064nm mode: (30 to 500)mJ, 532nm mode: (20 to 250)mJ - Pulse Width: 1064mm mode: 300-400 ps, 532mm mode : 300-400 ps - Pulse repetition rate: 1064nm
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circle with a stylized "W" inside it, followed by the text "WONTECH". The "WON" part of the text is in orange, while the "TECH" part is in gray. The logo is simple and modern, and the use of orange and gray gives it a professional look.
64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea TEL: +82 (42) 934 6800 FAX: +82 (42) 934 9491 mode:1-10Hz, 532nm mode : 1-10Hz - Radiation diameter: 1064mm mode: (2 to 10mm) step: 1mm, 532nm mode: (2 to 10mm) step: 1mm
5) Cleaning Validation
Cleaning validation is not needed, because PICOCAREMAJESTY is not in contact with patients
### Clinical Test Summary [21 CFR 807.92(b)(2)]
No clinical studies were considered necessary and performed.
### Conclusion [21 CFR 807.92(b)(3)]
In according with the Federal Food & drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification WON TECH Co., Ltd. concludes that the PICOCAREMAIESTY is substantially equivalent to predicate devices as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.