Retrospective analysis of clinical outcomes (audiological performance and adverse events) from patients implanted with the PMEI Stapesplasty Prostheses to confirm safety and effectiveness compared to state-of-the-art benchmarks.
Post-Market Clinical Follow-Up (PMCF) investigation; Multicenter, retrospective, single-subjects repeated measures; Follow-up/Duration: Median 49.5 days (audiological); Median 179.0 days (safety)
Patients requiring reconstruction of the ossicular chain (n=168 for audiological outcomes; n=188 for adverse events); Sample Size: 168-188
Not applicable for this study
Postoperative PTA4 ABG (≤20 dB and ≤10 dB), BC-PTA thresholds, adverse events, revision surgeries, device dislocations/extrusions
Indications for Use
The passive middle ear implant – stapesplasty prosthesis is intended to be used for replacement of the stapes arch or stapes arch and incus in case of a fixed stapes footplate. The stapesplasty prosthesis is implanted in the middle ear to restore sound transmission from the tympanic membrane to the oval window by replacing the ossicles partially. The stapesplasty prosthesis is a medical device for single use delivered in sterile condition.
Device Story
Passive middle ear implants (mAXIS, mLOOP, mZAM, mFIX) restore mechanical sound transmission from tympanic membrane to oval window. Distal piston placed in inner ear opening; proximal loop or clip coupled to incus or malleus. Used by ENT surgeons in otological surgery. Devices are sterile, single-use titanium implants. No specific device training required. Output is mechanical sound conduction; clinical benefit is improved conductive hearing. MRI conditional at 1.5T, 3.0T, and 7.0T.
Clinical Evidence
Multicenter, retrospective, single-subject repeated measures PMCF study (n=168). Primary endpoints: postoperative PTA4 air-bone gap (ABG). Results: 91.7% of patients achieved PTA4 ABG ≤20 dB; 66.1% achieved PTA4 ABG ≤10 dB. Mean BC-PTA pre/post-surgery: 26.6 dB HL / 23.9 dB HL. Safety: 0 revision surgeries, 0 device dislocations/extrusions. Results comparable to state-of-the-art literature.
Technological Characteristics
Materials: Medical grade titanium (ASTM F67, ASTM F136). Design: Piston-style partial ossicular replacement prosthesis with various loop/clip coupling structures. Energy: Passive mechanical. Connectivity: None. Sterilization: Gamma radiation (SAL <10^-6). MRI: Conditional at 1.5T, 3.0T, 7.0T.
Indications for Use
Indicated for patients of all ages with congenital or acquired stapes defects (e.g., otosclerosis, congenital fixation, traumatic injury, malformation) or inadequate conductive hearing from previous stapes surgery.
Regulatory Classification
Identification
A partial ossicular replacement prosthesis is a device intended to be implanted for the functional reconstruction of segments of the ossicular chain and facilitates the conduction of sound wave from the tympanic membrane to the inner ear. The device is made of materials such as stainless steel, tantalum, polytetrafluoroethylene, polyethylene, polytetrafluoroethylene with carbon fibers composite, absorbable gelatin material, porous polyethylene, or from a combination of these materials.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 17, 2025
Med-El Elektromedizinische Geräte GmbH Inés Román Santiago Senior Regulatory Affairs Fürstenweg 77a Innsbruck, Tirol 6020 Austria
Re: K241142
Trade/Device Name: mAXIS Stapes Prosthesis, mLOOP Stapes Prosthesis, mZAM Stapes Prosthesis and mFIX Stapes Prosthesis Regulation Number: 21 CFR 874.3450 Regulation Name: Partial Ossicular Replacement Prosthesis Regulatory Class: Class II Product Code: ETB Dated: December 19, 2024 Received: December 19, 2024
Dear Inés Román Santiago:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}------------------------------------------------
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Joyce C. Lin-S
for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241142
### Device Name
mAXIS Stapes Prosthesis, mLOOP Stapes Prosthesis, mZAM Stapes Prosthesis and mFIX Stapes Prosthesis.
### Indications for Use (Describe) INTENDED USE:
INTENDED USE.
The passive middle ear impl
The passive middle ear implant – stapesplasty prosthesis is intended to be used for replacement of the stapes arch or stapes arch and incus in case of a fixed stapes footplate. The stapesplasty prosthesis is implanted in the middle ear to restore sound transmission from the tympanic membrane to the oval window by replacing the ossicles partially. The stapesplasty prosthesis is a medical device for single use delivered in sterile condition.
### INTENDED USER
The stapesplasty prosthesis is intended to be implanted by qualified ENT surgeons only, with adequate skills to perform otological surgeries. Replacement of the ossicular chain is a standard surgical procedure, no additional specific device training is mandatory or required for the safe and effective use.
### TARGET PATIENT POPULATION
The target patient population for the passive middle ear implant are patients of all ages requiring reconstruction of the ossicular chain.
### INDICATIONS:
The stapesplasty prosthesis is indicated to treat patients with:
- congenital or acquired defects of the stapes due to e.g., otosclerosis, congenital fixation of the stapes, traumatic injury, malformation of the ossicular chain or middle ear
- inadequate conductive hearing from previous stapes surgery
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for MED-EL, a company that specializes in hearing implants. The logo is in red and features the company name in bold, sans-serif font. To the right of the letters "D" and "E" is a circular graphic with curved lines, possibly representing an ear or sound waves. The overall design is simple and modern.
# 510(k) SUMMARY of Safety and Effectiveness MED-EL Elektromedizinische Geräte GmbH [As Required by 21 CFR 807.92(c)]
#### Submitter [807.92(a)(1)] 1.0
Manufacturer: MED-EL Elektromedizinische Geräte GmbH (hereafter MED-EL) Fürstenweg 77a, 6020 Innsbruck Austria
### Contact Person:
Ines Roman Santiago MED-EL Regulatory Affairs Senior Specialist Phone: +43 577885786 E-Mail: ines.romansantiago@medel.com
FDA Official Correspondent: Elizabeth Gfoeller MED-EL Corporate Director, Regulatory Affairs Phone: +43 577885614 E-Mail: elizabeth.gfoeller@medel.com
Date the Summary was prepared: 17th January 2025.
#### Device Names [807.92(a)(2)] 2.0
## Table 1 Device/Trade Names of MED-EL PMEI Stapesplasty Prostheses
| Device/Trade<br>Name | Generic category<br>& Medical Specialty | Classification | Product<br>code | Regulation<br>(CFR) |
|----------------------------|-------------------------------------------------------------------------------------------------------|----------------|-----------------|---------------------|
| mAXIS Stapes<br>Prosthesis | Middle ear, Prosthesis,<br><b>Partial</b> Ossicular<br>Replacement Prostheses<br>(Ear, Nose & Throat) | Class II | ETB | 874.3450 |
| mLOOP Stapes<br>Prosthesis | | | | |
| mZAM Stapes<br>Prosthesis | | | | |
| mFIX Stapes<br>Prosthesis | | | | |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the MED-EL logo in red. The logo consists of the word "MED" followed by a stylized image of a cochlea, and then the letters "EL". The cochlea image is a red circle with a white swirl inside.
#### 3.0 Predicate Devices [807.92(a)(3)]
| MED-EL device | Primary Predicate Device A | Predicate Device B |
|-------------------------|----------------------------|-----------------------------------|
| mAXIS Stapes Prosthesis | | N/A |
| mLOOP Stapes Prosthesis | | Kurz K-Piston<br>K002221 |
| mZAM Stapes Prosthesis | Detroit Piston<br>K130512 | Skarzynski<br>Piston<br>K130512 |
| mFIX Stapes Prosthesis | | Clip Piston<br>àWengen<br>K021479 |
# Table 2 Predicate devices manufactured by Heinz Kurz GmbH
#### Description of the Devices [807.92(a)(4)] 4.0
The MED-EL "Passive Middle Ear Implants" (PMEls) Stapesplasty Prostheses are partial ossicular replacement prostheses which restore the mechanical sound transmission to the oval window. A partial (stapedotomy) or total opening of the footplate (stapedectomy) is required. The distal end (=piston) of the device is placed into the opening of the inner ear. The other end of the prosthesis (=Loop or Clip) is coupled to the long process of the incus or to the handle of the malleus.
They are offered in different designs, functional lengths and/or piston diameters to allow the surgeon to have more flexibility in terms of selecting the correct prosthesis according to the individual patient's anatomical needs, and the surgeon's surgical preferences and proficiency.
| Device/Trade<br>Name | mAXIS | mLOOP | mZAM | mFIX |
|-------------------------|------------------------------------------------------------|----------------------------|-----------------------------|--------------------------------|
| | Stapesplasty Prosthesis | | | |
| Design | Image: MAXIS design | Image: mLOOP design | Image: mZAM design | Image: mFIX design |
| | Spiral-shaped loop for easier application | | | Clip for standardized coupling |
| Coupling<br>structure | Loop band width =<br>0.5mm<br>(broad &<br>perforated loop) | Loop band width<br>= 0.3mm | Loop band width<br>= 0.20mm | Clip band width<br>=0.2mm |
| Material | Medical grade titanium | | | |
| Method of<br>Attachment | Manually, with crimping | | | Manually, with<br>clipping |
| Sterile /<br>Single-Use | Yes | | | |
# Table 3 Description of the MED-EL PMEI Stapesplasty
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for MED-EL, a company that specializes in hearing implants. The logo is in red and features the company name in bold letters. To the right of the letters "MED" is a circular graphic with curved lines inside.
#### 5.0 Statement of the Intended use [807.92(a)(5)]
The intended use and indications of the MED-EL PMEI Stapesplasty Prostheses and of the predicate device(s) manufactured by Heinz Kurz GmbH are substantially equivalent. Comparing the subject devices and the respective predicate devices, the same medical conditions or underlying pathologies can be treated, and the principal functionality of the devices is the same, hence no different questions of safety and effectiveness arise.
# PMEI Stapesplasty Prostheses
## Intended Use
The passive middle ear implant - stapesplasty prosthesis is intended to be used for replacement of the stapes arch or stapes arch and incus in case of a fixed stapes footplate. The stapesplasty prosthesis is implanted in the middle ear to restore sound transmission from the tympanic membrane to the oval window by replacing the ossicles partially. The stapesplasty prosthesis is a medical device for single use delivered in sterile condition.
# Intended User
The stapesplasty prosthesis is intended to be implanted by qualified ENT surgeons only, with adequate skills to perform otological surgeries. Replacement of the ossicular chain is a standard surgical procedure, no additional specific device training is mandatory or required for the safe and effective use.
## Target Patient Population
The target patient population for the passive middle ear implant are patients of all ages requiring reconstruction of the ossicular chain.
## Indications:
The stapesplasty prosthesis is indicated to treat patients with:
- congenital or acquired defects of the stapes due to e.g., otosclerosis, congenital fixation of the stapes, traumatic injury, malformation of the ossicular chain or middle ear.
- inadequate conductive hearing from previous stapes surgery.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for MED-EL, a company that specializes in hearing implants. The logo is in red and features the company name in bold letters. The "O" in MED-EL is replaced with a stylized ear symbol. The logo is simple and modern, and it effectively communicates the company's focus on hearing solutions.
#### 6.0 Comparison on Technological Characteristics with predicate devices [807.92(a)(6)]
The MED-EL PMEIs have the same technological characteristics (i.e. design, material, chemical composition, etc.) as the predicate devices identified in paragraph (a)|3). Hence, no difference in safety and effectiveness of the MED-EL PME! Stapesplasty Prostheses as compared to the predicate device(s) from Heinz Kurz GmbH is expected. A summary of the technological characteristics is provided in the following table.
#### 6.1 PMEI Stapesplasty Prostheses – Comparison with predicate devices
| MED-EL Device /<br>Predicate Device | "mAXIS Stapes<br>Prosthesis" | Predicate<br>Device [A]<br>"Detroit<br>Piston"<br>(K130512) | "mLOOP<br>Stapes<br>Prosthesis" | Predicate<br>Device [B]<br>"K-Piston"<br>(K002221) | "mZAM Stapes<br>Prosthesis" | Predicate<br>Device [B]<br>"Skarzynski<br>Piston"<br>(K130512) | "mFIX Stapes<br>Prosthesis" | Predicate<br>Device [B]<br>"Clip Piston<br>àWengen"<br>(K021479) |
|-------------------------------------|----------------------------------------------------------------------------------------------|-------------------------------------------------------------|---------------------------------------------------------------------------------|----------------------------------------------------|-------------------------------------------------------------------------------------------------|----------------------------------------------------------------|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------|
| Design | Image: mAXIS Stapes Prosthesis | Image: Detroit Piston | Image: mLOOP Stapes Prosthesis | Image: K-Piston | Image: mZAM Stapes Prosthesis | Image: Skarzynski Piston | Image: mFIX Stapes Prosthesis | Image: Clip Piston àWengen |
| | Broad, perforated & spiral-<br>shaped loop | | Spiral-shaped loop | | Narrow band & spiral-shaped<br>loop | | Clip-type coupling structure | |
| Attachment | Manually, with crimping | | | | | Manually, without crimping | | |
| # of sizes | 27 | 24 | 28 | | 27 | 16 | 18 | 16 |
| Piston Ø [mm] | 0.4 / 0.5 / 0.6 | | 0.4 / 0.6 | | 0.4 / 0.5 / 0.6 | 0.4 / 0.6 | 0.4 / 0.6 | |
| Functional lengths<br>[mm] | 3.5, 3.75, 4.0,<br>4.25, 4.5, 4.75,<br>5.0, <span style="font-weight:bold">5.25</span> , 5.5 | 3.5, 3.75, 4.0,<br>4.25, 4.5,<br>4.75, 5.0, 5.5 | 3.5, 3.75, 4.0, 4.25, 4.5, 4.75,<br>5.0, 5.25, 5.5, 6.0, 7.0, 8.0, 9.0,<br>10.0 | | 3.5, 3.75, 4.0,<br>4.25, 4.5,<br>4.75, 5.0,<br><span style="font-weight:bold">5.25</span> , 5.5 | 3.5, 3.75, 4.0,<br>4.25, 4.5,<br>4.75, 5.0, 5.5 | 3.5, 3.75, 4.0,<br>4.25, 4.5,<br>4.75, 5.0,<br><span style="font-weight:bold">5.25</span> , 5.5 | 3.5, 3.75, 4.0,<br>4.25, 4.5, 4.75,<br>5.0, 5.5 |
| Loop width [mm] | 0.50 | | | 0.30 | 0.20 | 0.25 | 0.20 | 0.25 |
## Table 4: PMEI Stapesplasty Prostheses - Comparison of Technical Characteristics
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for MED-EL, a company that specializes in hearing implants. The logo is in red and features the company name in a bold, sans-serif font. The "O" in MED-EL is replaced with a stylized ear symbol.
| Materials | Medical grade titanium ASTM F67 | Medical grade<br>titanium ASTM<br>F136 (clip) and<br>ASTM F67<br>(shaft, piston) | Medical grade<br>titanium ASTM<br>F67 |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|---------------------------------------|
| Biocompatible | Yes (EN ISO 10993) | | |
| Surgical Tools<br>(mandatory) | N/A | | |
| Surgical tools<br>(optional) | MED-EL does not offer any optional surgical tool for the PMEI Stapesplasty Prosthesis.<br>The manufacturer of the predicate device(s) offers the "KURZ-Meter" as an optional surgical tool. | | |
| Packaging | One length/diameter-variant per package; Single-Use; Sterile | | |
| MRI | MRI Conditional at 1.5, 3.0 and 7.0 T | | |
Comparing the MED-EL PMEI Stapesplasty Prostheses and their identified predicated by Heinz Kurz GmbH, the method of attachment, biocompatibility, packaging configuration and MRI compatibility are IDENTICAL.
The following features are considered EQUIALENT: device design, number and dimensions of length/diameter variants, materials in body contact. As there is no difference in technological characteristics, there is no difference in safety and/or effectiveness.
There is just one difference: the surgical tools. For intraoperative measuring of the required prosthesis length, Heinz Kurz GmbH provides the KURZ-Meter however MED-EL does not provide such a tool due to other optional surgical tools are available on the market that can be used to determine the required prosthesis length to be used during a well-established stapes replacement surgical procedure. Thus, as there is no difference in technological characteristics, there is no difference in safety and effectiveness of the MED-EL PMEI Stapesplasty Prostheses compared to the predicate devices from Heinz Kurz GmbH is expected.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for MED-EL, a company that specializes in hearing implants. The logo is in red and features the company name in bold, sans-serif font. To the right of the name is a stylized graphic of an ear, also in red. The logo is simple and modern, and the use of red conveys a sense of energy and innovation.
#### 7.0 Non-clinical Test [807.92(b)(1)]
Non-clinical testing was conducted to assess the performance data for a determination of substantial equivalence.
The following tests were conducted:
- . MRI environment according to ASTM F2119, F2052, F2182 с MRI Conditional at 1.5T, 3.0T and 7.0T
- . Biocompatibility according to EN ISO 10993
- The performed evaluation provides objective evidence to support the o conclusion that the devices can be considered biocompatible for their intended use.
- Shelf-Life testing according to EN ISO 11607
- Sterilization Validation according to EN ISO 11137-1, EN ISO 11137-2. .
- Gamma Sterilization with a confidence sterility assurance level of <10 o
- . Packaging Validation according EN ISO 11607
- . Mechanical & other testing
Safety and effectiveness have been demonstrated with non-clinical testing, concluding that the design and performance requirements are met.
#### 8.0 Clinical Test [807.92(b)(2)]
The benefit for patients of the PMEI Stapesplasty Prostheses has been successfully demonstrated and the clinical outcomes of the MED-EL PMEI Stapesplasty Prostheses are comparable to the predicate devices as per the state of the art for these types of devices.
The goal of a stapes prosthesis is to close the postoperative pure tone average (PTA4) air-bone gap (ABG) to within 20 dB, ideally to within 10 dB. According to the state of the art, as assessed in several systematic literature reviews and meta-analysis, more than 90 % of patients implanted with a stapes prosthesis achieve a postoperative PTA4 ABG ≤20 dB. A variable fraction of stapes prosthesis implanted patients (41.0-87.2 %) even achieves a postoperative PTA4 ABG to within 10 dB.
MED-EL conducted a large multicenter, retrospective, single-subjects repeated measures (each subject served as his or her own control) Post-Market Clinical Follow-Up (PMCF) investigation with the PMEI Stapesplasty Prosthesis in Europe:
- Audiological outcomes: .
- o Number of patients: 168 (115 female, 53 male)
- Age: mean 46.4 ± 12.9 years (range 12-70) O
- Implanted ear: 86 right, 82 left o
- Follow-up: median 49.5 days (mean 64.6 ± 61.4; range 7-470 days) o
- Prosthesis type: 88 mAXIS, 50 mLOOP, 11 mZAM, 19 mFIX O
- O % patients with a postoperative PTA4 ABG ≤20 dB: 91.7 %
- % patients with a postoperative PTA4 ABG ≤10 dB: 66.1 % O
- Mean BC-PTA« pre/post-surgery: 26.6 ± 10.3 dB HL / 23.9 ± 10.8 dB HL o
- o BC-PTA deterioration >10 dB HL: 9/168 patients = 5.4 % (range 11.3-21.3 dB HL)
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for MED-EL, a medical device company specializing in hearing implants. The logo is in red and features the company name in bold, sans-serif font. To the right of "MED" is a stylized graphic of a cochlea, also in red, which is a key component of hearing implants.
- Adverse events: ●
- Number of analyzed patients: 188 (127 female, 61 male) o
- Age: mean 47 ± 13 years (range 12-82) റ
- Implanted ear: 98 right, 90 left o
- Follow-up: median 179.0 days (mean 246.4 ± 212.8; range 7-709 days) o
- Prosthesis type: 93 mAXIS, 64 mLOOP, 11 mZAM, 20 mFIX O
- Number of revision surgeries: 0 (2 recommended) o
- o Number of device dislocations/extrusions: 0
- Total 16 AEs in 13 patients: 7 Device unrelated AEs, 2 Medical related O AEs, 2 Clinical/surgical related AEs, 5 AEs "unknown", 0 Follow-up AE
In the above mentioned PMCF investigation the safety and effectiveness of the PMEI Stapesplasty Prostheses was confirmed as demonstrated by postoperative PTA4 ABG <20 dB rates and postoperative PTA ABG ≤10 dB rates comparable to the state of the art, low rates of device dislocations/extrusions and revision surgeries, and overall stable bone conduction thresholds, an indicator that residual inner ear hearing is not compromised after implantation with an ossicular replacement prosthesis. Safety and effectiveness of the MED-EL PMEI Stapesplasty Prostheses are comparable to the current state of the art.
#### 9.0 Conclusion [807.92(b)(3)]
The collective result of the non-clinical testing demonstrates that:
- the MED-EL PMEI Stapesplasty Prostheses meet the established specifications, . the design and performance requirements to ensure safety and effectiveness according to the intended use.
- . there is no difference in safety and/or effectiveness between MED-EL PMEI Stapesplasty Prostheses and the predicate devices manufactured by Heinz Kurz GmbH.
It can be concluded that the MED-EL PMEI Stapesplasty Prostheses are substantially equivalent to the predicate devices manufactured by Heinz Kurz GmbH.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.