The GD-P-E Enhanced Handheld Pain Relief Laser Instrument generally indicated: - adjunctive use in providing temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin; - adjunctive use in providing temporary relief of minor chronic pain associated with Carpal Tunnel Syndrome (CTS)
Device Story
Handheld, non-invasive, non-heating infrared cold laser instrument; used for adjunctive pain relief. Device emits dual laser wavelengths (808nm and 650nm) to target tissue; light absorption facilitates cellular reaction promoting pain relief. Operated by clinician or patient (prescription use); features LCD, buttons, and rechargeable lithium-ion battery. Output parameters include pulsed and continuous modulation modes. Healthcare providers use device to manage chronic musculoskeletal pain; patient benefits from temporary pain reduction. Device includes power adapter and charging components.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including electrical safety, electromagnetic compatibility, laser safety, and output parameter verification (wavelength, power, energy fluence, irradiance, and beam focus).
Technological Characteristics
Handheld infrared laser instrument; 650nm (±10nm) and 808nm (±10nm) wavelengths. Powered by 3.7V 1950mAh lithium-ion battery. Features pulsed and continuous modulation. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60825-1 (laser safety), IEC 60601-1-11 (home healthcare environment).
Indications for Use
Indicated for patients requiring adjunctive temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin or minor chronic pain associated with Carpal Tunnel Syndrome (CTS).
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and name on the right. The FDA part of the logo is in blue and reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.
January 24, 2025
Wuhan Guangdun Technology Co.,Ltd. % Boyle Wang Manager Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai. 200120 China
Re: K241057
Trade/Device Name: Enhanced Handheld Pain Relief Laser Instrument (GD-P-E) Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: NHN Dated: December 24, 2024 Received: December 30, 2024
Dear Boyle Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Amber T. Ballard -S
Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241057
Device Name
Enhanced Handheld Pain Relief Laser Instrument (GD-P-E)
Indications for Use (Describe)
The GD-P-E Enhanced Handheld Pain Relief Laser Instrument generally indicated:
- adjunctive use in providing temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin;
- adjunctive use in providing temporary relief of minor chronic pain associated with Carpal Tunnel Syndrome (CTS)
Type of Use (Select one or both, as applicable):
| <div> <span style="font-size: 10pt;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span style="font-size: 10pt;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
|---------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.
#### Submitter's Information
| Name: | Wuhan Guangdun Technology Co.,Ltd. |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant address: | No.1,Floor 3, Building 2-05,Optics Valley Chip Center<br>II, NO.303,Optics Valley Avenue,East Lake New Technology<br>Development Zone,Wuhan,Hubei P.R.China |
| Telephone: | +86-27-59368661 |
| Contact: | Suzy Pan |
# Designated Submission Correspondent
| Contact: | Mr. Boyle Wang |
|----------------------|------------------------------------------------------------------------|
| Name: | Shanghai Truthful Information Technology Co., Ltd. |
| Address: | Room 1801, No. 161 East Lujiazui Rd.,<br>Pudong Shanghai, 200120 China |
| Tel: | +86-21-50313932 |
| Email: | Info@truthful.com.cn |
| Date of Preparation: | January 21, 2025 |
Device Information
| 510(k) Number: | K241057 |
|----------------------|----------------------------------------------------------------------------------------------------------------|
| Trade name: | Enhanced Handheld Pain Relief Laser Instrument |
| Common name: | Infrared lamp |
| Classification name: | Powered Light Based Laser Non-Thermal Instrument With<br>Non-Heating Effect for Adjunctive Use in Pain Therapy |
| Model(s): | GD-P-E |
| Production code: | NHN |
| Regulation number: | 21 CFR 890.5500 |
| Classification: | Class II |
| Panel: | Physical Medicine |
## Predicate Devices
## Primary Predicate Device:
{5}------------------------------------------------
| Manufacturer: | GRT Solutions, Inc. |
|---------------------|-------------------------------|
| Trade/Device Name: | GRT LITE Model PRO-8A |
| 510(k) number: | K050668 |
| Predicate Device 2: | |
| Manufacturer: | TUCO Innovations, Inc. |
| Trade/Device Name: | TUCO Erchonia® PL2000 |
| 510(k) number: | K012580 |
| Predicate Device 3: | |
| Manufacturer: | THØR International Ltd. |
| Trade/Device Name: | THØR DDII 830CL3 Laser System |
| 510(k) number: | K030226 |
| Predicate Device 4: | |
| Manufacturer: | Erchonia Corporation |
| Trade/Device Name: | Erchonia® XLR8 |
| 510(k) number: | K130996 |
#### Device Description
The GD-P-E Enhanced Handheld Pain Relief Laser Instrument consists of a LCD, button, battery, power charger jack. Accessories include a power cords, a power adapter. It is a handheld, non-invasive, non-heating infrared cold laser instrument designed to be used as an adjunct treatment for managing the temporary relief of pain associated with Carpal Tunnel Syndrome. A key feature of the device is its ability to produce multiple laser wavelengths (808nm and 650nm).
A key feature of the device is its ability to emit dual laser wavelengths (808nm and 650nm), providing targeted low-level light therapy to the affected area. This "cold" laser therapy enables body tissue to absorb the light, facilitating a temporary reaction within the cells that promotes minor pain relief.
#### Indication for Use Statement
The GD-P-E Enhanced Handheld Pain Relief Laser Instrument generally indicated:
- adjunctive use in providing temporary relief of minor chronic neck and shoulder pain ● of musculoskeletal origin;
- adjunctive use in providing temporary relief of minor chronic pain associated with ● Carpal Tunnel Syndrome (CTS).
#### Comparison to the Predicate Device
{6}------------------------------------------------
| Item | Subject Device (K241057) | Primary Predicate Device (K050668) | Predicate Device 2 (K012580) | Predicate Device 3 (K030226) | Predicate Device 4 (K130996) |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Enhanced Handheld Pain Relief Laser Instrument | GRT LITE Model PRO-8A | TUCO Erchonia® PL2000 | THØR DDII 830CL3 Laser System | Erchonia® XLR8 |
| Class Product Code Regulation | Class II<br>NHN<br>21 CFR 890.5500 | Class II<br>NHN<br>21 CFR 890.5500 | Class II<br>NHN<br>21 CFR 890.5500 | Class II<br>NHN<br>21 CFR 890.5500 | Class II<br>NHN<br>21 CFR 890.5500 |
| Indications for Use | The GD-P-E Enhanced Handheld Pain Relief Laser Instrument generally indicated:<br>• adjunctive use in providing temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin;<br>• adjunctive use in providing temporary relief of minor chronic pain associated with Carpal Tunnel Syndrome (CTS). | The GRT LITE Model PRO-8A is a non-heating lamp, infrared as described under the provisions of 21 CFR §890.5500 and is clinically indicated for:<br>• Adjunctive use in providing temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin.<br>• Adjunctive use in providing temporary relief of minor chronic pain associated with Carpal Tunnel Syndrome (CTS). | The TUCO Erchonia PL2000 is indicated for adjunctive use in providing temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin. | The THØR DDII 83OCL3 Laser System is indicated for adjunctive use in the temporary relief of hand and wrist pain associated with Carpal Tunnel Syndrome. | The Erchonia® XLR8TM laser is indicated for the following three indications:<br>a. adjunctive use in providing temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin<br>b. as an adjunct to liposuction procedures of the thighs, hips and stomach for reduction of pain associated with the recovery process<br>c. temporary reduction in post-surgery pain at 24 hours after surgery following bilateral breast augmentation surgery |
| Type of Use | Prescription Use Only | Prescription Use Only | Prescription Use Only | Prescription Use Only | Prescription Use Only |
| Wavelength | 650nm (±10nm)<br>808nm (±10nm) | 628nm (±5nm)<br>850nm (±5nm) | Not Publicly Available | 830nm | Not Publicly Available |
| Aiming Beam | Laser Diodes, Class 3R | Visible LEDs (628nm)<br>Infrared LEDs (850nm) | Not Publicly Available | Laser Diodes | Not Publicly Available |
| Terminal Output | 13 x 650nm (Laser Diodes)<br>4 x 808nm (Laser Diodes) | 2 x 628nm (Visible LEDs)<br>2 x 850nm (Infrared LEDs) | Not Publicly Available | 3 x 830nm (Laser Diodes) | Not Publicly Available |
| Timing Range | 5 or 8 minutes | Not Publicly Available | Not Publicly Available | Not Publicly Available | Not Publicly Available |
| Max Power | 650nm x 5mW<br>808nm x 3mW | Not Publicly Available | Not Publicly Available | 3 x 30mW | Not Publicly Available |
| Energy Fluence | 1773.15 mJ/mm² | Not Publicly Available | Not Publicly Available | Not Publicly Available | Not Publicly Available |
| Irradiance | 3.43 mW/mm² to 4.95 mW/mm² | Not Publicly Available | Not Publicly Available | Not Publicly Available | Not Publicly Available |
| Spot Size | 1.16 mm² (Avg.) | Not Publicly Available | Not Publicly Available | Not Publicly Available | Not Publicly Available |
| Power Mode | Pulsed & Continuous Modulation | Pulsed & Continuous Modulation | Not Publicly Available | Not Publicly Available | Not Publicly Available |
| Energy Delivery | Handheld treatment probe | Hand-held, non-invasive, pain therapy system | Not Publicly Available | Handheld, non-invasive, low energy, | Not Publicly Available |
| | | | | non-heating<br>infrared<br>therapeutic<br>medical laser | |
| Power | 3.7V-1950mAh<br>Lithium-ion<br>battery | Not Publicly<br>Available | Not Publicly<br>Available | Not Publicly<br>Available | Not Publicly<br>Available |
| Performance<br>Testing | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60825-1<br>IEC 60601-1-11 | 21 CFR –<br>Subchapter J – Part<br>1010<br>FCC Standard – 47<br>CFR Part 15B<br>UL® Approved | Not Publicly<br>Available | EN 60601-1:1993<br>EN 60601-2-22:1996<br>EN 60601-1-2:1994 | Not Publicly<br>Available |
{7}------------------------------------------------
## Non-Clinical Test Data
Non clinical tests were conducted to verify that the subject devices met all design specifications as were Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- · IEC 60601-1: 2020 Medical Electrical Equipment -- Part 1: General Requirements for Basic Safety and essential performance
· IEC 60601-1-2:2020 Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests
- · IEC 60825-1:2014 Safety of Laser products-Part 1: Equipment classification and requirements
. IEC 60601-1-11:2020 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
In addition, the Output Parameters including Output wavelength(s), Power, Energy fluence, Irradiance and the Beam Focus and Energy Distribution Test of the candidate device have been performed. The results of these tests demonstrated that the device performs as intended and meets its specifications:
Software Information:
Software validation was conducted in accordance with a moderate level of concern designation in accordance with the FDA guidance
"Content of Premarket Submissions for Device Software Functions", issued on June 14, 2023
# Clinical Test Conclusion
No clinical study is included in this submission.
#### Conclusion
Performance testing contained in this submission demonstrates the minor differences in technological characteristics between the subject device and the predicate do not raise different questions of safety and effectiveness. And based on the performance testing and compliance with acceptable voluntary standards, we believe the subject device is substantially equivalent to its predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.