K240999 · Pollogen, Ltd. · GEI · May 8, 2024 · General, Plastic Surgery
Device Facts
Record ID
K240999
Device Name
Legend X Applicator VO
Applicant
Pollogen, Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
May 8, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Legend X Applicator VO and its accessories are intended for dermatological procedures requiring ablation and resurfacing of the skin when using VoluDerm Energy (tips gen12, gen100, gen 36, gen36L, H7x7). The Legend X Applicator VO with genXL tip is intended for use in dermatological procedures for electrocoagulation and hemostasis. At higher energy levels greater than 62 mJ/pin, use of the Applicator VO with genXL tip is limited to Skin Types I-IV.
Device Story
Legend X Applicator VO is a handheld electrosurgical handpiece connecting to the Legend X console; delivers fractional radiofrequency (RF) energy to skin via micro-electrode pin arrays on disposable tips. Used for ablation, resurfacing, electrocoagulation, and hemostasis. Operated by clinicians in clinical settings; requires continuous direct user control and visualization. Console software manages energy delivery based on user-selected parameters; provides real-time status display. Benefits include non-invasive treatment of face and body skin conditions. Device is an extension of the cleared Legend X platform, adding the genXL tip for deeper tissue effects (up to 4mm).
Clinical Evidence
Bench testing only. Performance verified through biocompatibility testing (ISO 10993-1, -5, -10, -11, -23), electrical safety/EMC compliance (IEC 60601-1), and software verification and validation per FDA guidance. No clinical data presented.
Technological Characteristics
Handheld electrosurgical applicator; stainless steel and gold-plated micro-electrode pin arrays. Energy source: RF generator (Legend X console). Connectivity: Wired to console. Sterilization: EO (disposable tips). Software: Embedded control system. Standards: IEC 60601-1 (electrical/EMC), ISO 10993 (biocompatibility).
Indications for Use
Indicated for dermatological procedures requiring skin ablation, resurfacing, electrocoagulation, and hemostasis. Use of genXL tip at >62 mJ/pin restricted to Fitzpatrick Skin Types I-IV.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
InMode System with the Morpheus8 Applicators (K200947)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the word "ADMINISTRATION".
May 8, 2024
Pollogen Ltd. % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
RE: K240999
Trade/Device Name: Legend X Applicator VO Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: April 11, 2024 Received: April 11, 2024
Dear Prithul Bom:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed by Long H. Chen -S Lona H. Chen-Date: 2024.05.08 13:39:12 -04'00'
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices
{2}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k)Number (if known) K240999
DeviceName Legend X Applicator VO
#### Indications for Use (Describe)
The Legend X Applicator VO and its accessories are intended for dermatological procedures requiring ablation and resurfacing of the skin when using VoluDerm Energy (tips gen12, gen361, H7x7). The Legend X Applicator VO with genXL tip is intended for use in dermatological procedures for electrocoagulation and hemostass. At higher energy levels greater than 62 m.J/pin, use of the Applicator VO with genXL tip is limited to Skin Types I-IV.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
|×| Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary K240999
## 1. General information
| 510(k) Submitter | Pollogen Ltd<br>6 Kaufman Street<br>Tel Aviv, IL 6801298 |
|-------------------------|--------------------------------------------------------------------------------------------------------|
| FDA Registration Number | 3008753275 |
| Primary Correspondent | Karen Smith<br>Vice President, Regulatory & Quality<br>Lumenis Be, Inc., a Pollogen Ltd Family Company |
| Contact Information | Email: karen.smith@lumenis.com |
| Secondary Correspondent | Lola Fajinmi<br>Regulatory Affairs Consultant<br>MedTeCheck |
| Contact Information | Email: lola@MedTeCheck.com |
| Date Prepared | April 11, 2024 |
#### 2. Device Identification
The Legend X Applicator VO consists of an applicator accessories (disposable tips) needed to perform some of Legend X dermatological procedures. The Legend X Applicator VO can be defined as follows:
| Proprietary Name | Legend X Applicator VO |
|-----------------------------|-------------------------------------------------------------------|
| Device Classification Name: | Electrosurgical, Cutting & Coagulation & Accessories |
| Regulation Description: | Electrosurgical Cutting and Coagulation Device and<br>Accessories |
| Regulation Number: | 878.4400 |
| Device Class: | Class II |
| Product Code: | GEI |
#### 3. Predicate Devices
Primary Predicate:
| Proprietary Name | Legend X Platform |
|-----------------------------|-------------------------------------------------------------------|
| Device Classification Name: | Electrosurgical, Cutting & Coagulation & Accessories |
| Regulation Description: | Electrosurgical Cutting and Coagulation Device and<br>Accessories |
| 510(k) Number: | K232903 |
| Regulation Number: | 878.4400 |
{5}------------------------------------------------
| Device Class: | Class II |
|---------------|----------|
| Product Code: | GEI, NGX |
Secondary Predicate:
| Proprietary Name | InMode System with the Morpheus8 Applicators |
|-----------------------------|-------------------------------------------------------------------|
| Device Classification Name: | Electrosurgical, Cutting & Coagulation & Accessories |
| Regulation Description: | Electrosurgical Cutting and Coagulation Device and<br>Accessories |
| 510(k) Number: | K200947 |
| Regulation Number: | 878.4400 |
| Device Class: | Class II |
| Product Code: | GEI |
#### 4. Device Description
The Legend X Applicator VO ("Proposed Device") is connected to the Legend X softwarecontrolled capital equipment platform to enable application of radiofrequency onto the skin for ablation and resurfacing or electrocoagulation and hemostasis. The proposed device applicator and applicator accessories are identical to the recently cleared Legend X Platform (K232903), with the exception of the new optional Applicator VO genXL disposable tip. The application of radiofrequency with the Applicator VO Disposable Tip only occurs under the continuous and direct control of a user with direct visualization, via the user inputs into the Legend X Console and Applicator indicators.
The proposed device consists of applicators and applicator accessories (disposable tips) needed to perform some of Legend X dermatological procedures. The summary descriptions of the proposed device are outlined below.
Legend X Applicator VO: The Legend X Applicator VO is a handheld handpiece which delivers energy to the treatment area. Depending on the selected user inputs, the user can utilize the applicator to deliver fractional RF energy for ablation and resurfacing, or electrocoagulation and hemostasis, as a means for noninvasive skin treatment throughout the face and body. The Applicator VO is connected to the cleared Legend X Console via an Applicator Connector cord.
Legend X Software: The Legend X Software provides the user and patient with the ability to safely commence, drive, and stop the operation of the Applicator on the skin area of interest throughout the face and body. It receives user input from the Legend X Console and computes the appropriate output to the Applicator VO connected to the Console. It provides a graphical user interface where the treatment parameters are adjusted as well as status of operation is shown in real time and displays important system status information.
{6}------------------------------------------------
Legend X Applicator VO Disposable Tips: The Legend X Applicator VO Disposable Tips set is part of the Legend X Applicator VO kit. The Legend X Applicator VO Disposable Tips Set includes:
- 1. gen12 disposable tip
- 2. gen36 disposable tip
- 3. gen36L disposable tip
- 4. gen100 disposable tip
- 5. H7X7 disposable tip
- 6. genXL disposable tip
The patient-contacting portions of the disposable tips are constructed of stainless steel and gold plating. The disposable tips are connected to handheld Applicator VO for delivery of radiofrequency electrical current via an array of micro-electrode pins onto the skin surface.
## 5. Intended Use/Indications for Use
The Legend X Applicator VO and its accessories are intended for dermatological procedures requiring ablation and resurfacing of the skin when using VoluDerm Energy (tips gen12, gen100, gen 36, gen36L, H7x7). The Legend X Applicator VO with genXL tip is intended for use in dermatological procedures for electrocoagulation and hemostasis. At higher energy levels greater than 62 mJ/pin, use of the Applicator VO with genXL tip is limited to Skin Types I-IV.
## 6. Comparison of Technological Characteristics with the Predicate Devices
Overall, the proposed and predicate devices are based on the following similar basic technological elements:
- 1. Device contains a central console that connects each of the main components to facilitate dermatologic procedures and facilitates the selection of fractional RF treatment modes via the graphic user interface
- 2. Device contains an applicator to allow energy delivery and application on to the area of interest on the skin
- 3. Device contains similar components between the subject and predicate devices (i.e., Applicator VO and disposable tips)
- 4. Device requires continuous direct control by the user to operate the Applicator
- 5. Device mode operation is only at the command of the user
Comparison of the technological characteristics with the predicate devices for each of the Legend X Applicator VO disposable tip components are described below:
{7}------------------------------------------------
| Proposed Device | Primary Predicate | Secondary Predicate | Technological Characteristics<br>Comparison |
|-------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Legend X<br>Applicator VO | Legend X Platform<br>(K232903) | InMode System<br>(K200947) | |
| Components:<br>- Applicator VO<br>(connected to a<br>cleared Legend X<br>Console) with<br>accessories | System Components:<br>- Console<br>- Patient-controlled<br>manual switch<br>- Applicators (1, 2, 3,<br>and VO)<br>- Footswitch | System Components:<br>- Console<br>- Applicator<br>- Footswitch | Identical Characteristics:<br>- The Legend X Applicator<br>VO is identical to the<br>primary predicate Legend<br>X Platform in<br>functionality, design, and<br>specifications. |
| | | | Differences<br>- The subject Legend X<br>Applicator VO is<br>compatible with six<br>disposable tip accessories,<br>whereas the primary<br>predicate is compatible<br>with five disposable tip<br>accessories. |
| Applicator VO<br>Disposable Tips:<br>- gen12<br>- gen36<br>- gen36L<br>- gen100<br>- H7x7<br>- genXL | Applicator VO<br>Disposable Tips:<br>- gen12<br>- gen36<br>- gen36L<br>- H7x7<br>- gen100 | Accessories<br>- T tip head<br>- 12-pin tip head<br>- 24-pin tip head<br>- 40-pin tip head | Identical Characteristics:<br>- Functionality: To provide<br>delivery of fractional RF<br>energy<br>- Similar materials,<br>specifications, and intended<br>functionality<br>Differences<br>- The Applicator VO genXL<br>disposable tip is an<br>additional disposable tip<br>with insertion depth<br>options of up to 4 mm,<br>which is within the range<br>of the secondary<br>predicate's disposable tips<br>that are up to 7 mm. |
# 7. Performance Data
{8}------------------------------------------------
The Legend X Applicator VO was assessed for performance in accordance with internal design specification with the applicable performance standards to demonstrate safety and effectiveness.
The testing identified no new issues of safety or effectiveness. The testing performed are summarized below:
### Summary of Performance Testing:
The reprocessing, sterilization, and shelf life for the subject device are identical to those evaluated for the primary predicate. The Legend X. Applicator VO disposable tips are EO sterilized for single use, and the Legend X Applicator VO is reusable.
Biocompatibility: The final finished form of the subject device has been used for the biocompatibility evaluation. Biocompatibility for patient-contacting components has been evaluated and validated in accordance with the provisions of the following guidance document and standards:
- . FDA Guidance for Industry and Food and Drug Administration Staff – Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process."
- ISO 10993-1 Biological evaluation of medical devices Part 1: Evaluation and testing ● within a risk management process
- ISO 10993-5 Biological evaluation of medical devices Part 5: Tests for in vitro ● cvtotoxicitv
- ISO 10993-10 Biological evaluation of medical devices Part 10: Tests for irritation . and skin sensitization
- . ISO 10993-11 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
- ISO 10993-23 Biological evaluation of medical devices Part 23: Tests for irritation ●
Electrical Safety and Electromagnetic Compatibility: The subject device has been fully evaluated for electrical safety and EMC compliance and is compliant with IEC 60601-1 standards.
Software: Software was developed, tested, and verified per Guidance for Industry and FDA Staff: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
Results of verification and validation (V&V) testing confirm that the proposed device conforms to design specifications and meets the needs of the intended users.
V & V Testing: Performance testing was executed to verify the overall functionality of the proposed device to operate as specified by the design input requirements, including applicator and controls, various functional safety features, and other general functionality. Requirements
{9}------------------------------------------------
for safety and effectiveness of the device, including the adherence to regulatory standards, were verified. Results of verification testing confirm that the subject device conforms to design specifications and requirements and meets the needs of the intended users.
#### 8. Conclusion
Based on the same Indications for Use and similar technological characteristics as the secondary predicate and performance testing, the Legend X Applicator VO with genXL tip raises no new questions of safety and effectiveness and is substantially equivalent to the secondary predicate device in terms of safety, effectiveness, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.