K240994 · Nextbiomedical Co., Ltd. · QAU · Nov 27, 2024 · General, Plastic Surgery
Device Facts
Record ID
K240994
Device Name
Nexpowder
Applicant
Nextbiomedical Co., Ltd.
Product Code
QAU · General, Plastic Surgery
Decision Date
Nov 27, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4456
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K240994 · Nov 27, 2024
Nexpowder
Nextbiomedical Co., Ltd.
Retrospective clinical studies (Korea); Aggregated clinical data (European Union)
Retrospective clinical data and aggregated EU data were used to support the expansion of the device's indications for use to include lower gastrointestinal bleeding.
Lower GI bleeding; Retrospective study; Indication expansion
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective study
Patients with lower gastrointestinal bleeding; Sample Size: 260 (total cases including prospective and retrospective)
Hemospray (DEN170015)
Hemostasis effectiveness
Indications for Use
Nexpowder™ is used for hemostasis of nonvariceal gastrointestinal bleeding.
Device Story
Nexpowder is a single-use, prescription-only device for hemostasis of nonvariceal gastrointestinal bleeding. It consists of a hemostatic powder (succinic anhydride/ε-poly-L-lysine and aldehyded dextran) pre-packaged in a vial and a delivery system comprising a spray body, connector, and catheter. The device is operated by a clinician during an endoscopic procedure; the catheter is inserted through the endoscope's working channel to the target site. A battery-powered air pump in the spray body generates air pressure to propel the powder through the catheter onto the bleeding site. This low-pressure delivery minimizes particle scattering and maintains visualization. The powder acts to achieve hemostasis. The device is used in clinical settings where endoscopy is performed. It benefits patients by providing a controlled, effective method for managing gastrointestinal bleeding.
Clinical Evidence
Clinical evidence for lower GI indication expansion included a prospective study and retrospective studies conducted in Korea, plus aggregated data from the EU. Total sample size: 260 lower GI cases. Results were compared to the predicate device (Hemospray, DEN170015) and found to be comparable, supporting safety and effectiveness in the lower GI tract.
Technological Characteristics
Single-use endoscopic delivery system. Powder composition: Succinic anhydride (ε-poly-L-lysine) and aldehyded dextran. Energy source: Battery-powered air pump. Delivery method: Air-propelled powder through a catheter inserted via endoscope working channel. Non-mechanical, low-pressure delivery.
Indications for Use
Indicated for hemostasis of nonvariceal gastrointestinal bleeding in patients.
Regulatory Classification
Identification
A hemostatic device for intraluminal gastrointestinal use is a prescription device that is endoscopically applied to the upper and/or lower gastrointestinal tract and is intended to produce hemostasis via absorption of fluid or by other physical means.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The device must be demonstrated to be biocompatible.
(2) Performance data must support the sterility and pyrogenicity of the device.
(3) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
(4) In vivo performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The testing must evaluate the following:
(i) The ability to deliver the hemostatic material to the bleeding site;
(ii) The ability to achieve hemostasis in a clinically relevant model of gastrointestinal bleeding; and
(iii) Safety endpoints, including thromboembolic events, local and systemic toxicity, tissue trauma, gastrointestinal tract obstruction, and bowel distension and perforation.
(5) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be evaluated:
(i) Materials characterization of all components must demonstrate the device meets established specifications, which must include compositional identity and purity, characterization of impurities, physical characteristics, and reactivity with fluids.
(ii) Performance testing must demonstrate the mechanical integrity and functionality of the system used to deliver the device and demonstrate the device meets established specifications, including output pressure for propellant-based systems.
(6) Labeling must include:
(i) Information identifying and explaining how to use the device and its components; and
(ii) A shelf life.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 27, 2024
Nextbiomedical Co., Ltd. % Gali Shrieber Senior Regulatory Affairs Specialist Covidien llc. 15 Hampshire Street Mansfield, Massachusetts 02048
Re: K240994
Trade/Device Name: Nexpowder™ Regulation Number: 21 CFR 878.4456 Regulation Name: Hemostatic Device For Intraluminal Gastrointestinal Use Regulatory Class: Class II Product Code: QAU Dated: April 11, 2024 Received: October 28, 2024
Dear Gali Shrieber:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tek N.
Digitally signed by
Tek N. Lamichhane -S
Lamichhane -S Date: 2024.11.27
10:36:14-05'00'
Tek N. Lamichhane, Ph.D. Assistant Director DHT4B: Division of Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K240994
Device Name Nexpowder™
Indications for Use (Describe) Nexpowder™ is used for hemostasis of nonvariceal gastrointestinal bleeding.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------|--|
| <div> <span style="font-size: 12px;"></span></div> | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY (K240994) NextBiomedical CO., LTD. Nexpowder™
#### I. SUBMITTER
| Submitter Name and Address: | NEXTBIOMEDICAL CO., LTD. |
|-----------------------------|-------------------------------------------------------|
| | 6 Venture-ro 100 beon-gil, Yeonsu-gu |
| | Incheon, 22013 |
| | Republic of Korea |
| Correspondent Information: | Covidien llc. |
| | 15 Hampshire Street |
| | MA, 02048 |
| | USA |
| Contact person: | Gali Shrieber |
| | Senior Regulatory Affairs Specialist |
| | gali.shrieber@medtronic.com |
| Phone: | 972-73-2501577 |
| Date prepared: | November 26, 2024 |
| II. | <b>DEVICE</b> |
| Name of Device: | Nexpowder |
| Device Common Name: | Hemostatic Device For Endoscopic Gastrointestinal Use |
| Classification: | Class II |
| | Regulation 21 CFR 878.4456 Hemostatic device for |
| | intraluminal gastrointestinal use |
| | Product Code: QAU |
#### PREDICATE DEVICE(S) III.
Hemospray® Endoscopic Hemostat (DEN170015) - Predicate device Nexpowder (K202929) - Reference device
{5}------------------------------------------------
#### IV. DEVICE DESCRIPTION
The Nexpowder is used for hemostasis of non-variceal gastrointestinal bleeding. It is a prescription only, single-use device provided with a powder pre-packaged in a vial, and a delivery system, which consists of a spray body, a connector and a delivery catheter is inserted through an endoscope's working channel to deliver the powder to the intended hemostasis target site. Nexpowder is almost identical to the currently marketed device (K202929) and has the same technological characteristics and mechanism of action, but differs in the indications for use, where the subject device is also indicated for lower GI bleeding.
#### INDICATIONS FOR USE v.
Nexpowder is used for hemostasis of nonvariceal gastrointestinal bleeding.
| | Subject Device<br>Nexpowder | Predicate Device<br>Hemospray<br>Endoscopic<br>Hemostat<br>(DEN170015) | Reference Device<br>Nexpowder<br>(K202929) | Comparison | |
|------------------------|--------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|-----------------------------------------------------|-----------|
| Indications<br>for Use | Nexpowder is<br>used for<br>hemostasis of non-<br>variceal<br>gastrointestinal<br>bleeding | The COOK<br>Hemospray<br>Endoscopic<br>Hemostat is used<br>for hemostasis of<br>non-variceal<br>gastrointestinal<br>bleeding | Nexpowder is used<br>for hemostasis of<br>non-variceal, upper<br>gastrointestinal<br>bleeding | Identical to<br>predicate device | |
| | Device main<br>components | Hemostatic<br>powder agent and<br>a delivery system | Hemostatic powder<br>agent and a delivery<br>system | Hemostatic powder<br>agent and a delivery<br>system | Identical |
#### TECHNOLOGICAL CHARACTERISTICS VI.
{6}------------------------------------------------
| | Subject Device<br>Nexpowder | Predicate Device<br>Hemospray<br>Endoscopic<br>Hemostat<br>(DEN170015) | Reference Device<br>Nexpowder<br>(K202929) | Comparison |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|
| Operation<br>principle | The powder is<br>delivered by use<br>of a delivery<br>system and<br>through a catheter<br>inserted through<br>the working<br>channel of an<br>endoscope which<br>provides access to<br>the site of the<br>bleeding | The powder is<br>delivered by use of<br>a delivery system<br>and through a<br>catheter inserted<br>through the<br>working channel of<br>an endoscope<br>which provides<br>access to the site of<br>the bleeding | The powder is<br>delivered by use of<br>a delivery system<br>and through a<br>catheter inserted<br>through the<br>working channel of<br>an endoscope<br>which provides<br>access to the site of<br>the bleeding | Identical |
| Powder main<br>chemical<br>composition | • Succinic<br>anhydride (ε-<br>poly-L-lysine)<br>• Aldehyded<br>dextran | • Sodium<br>bentonite | • Succinic<br>anhydride (ε-<br>poly-L-lysine)<br>• Aldehyded<br>dextran | Different from<br>predicate device |
| Delivery method | Powder propelled<br>by battery power | Powder propelled<br>by CO2 | Powder propelled<br>by battery power | Different from<br>predicate device |
### Nexpowder and the primary predicate
Nexpowder and Hemospray (DEN170015) have similar technological characteristics and mechanism of action. In both devices, the powder is delivered using a delivery system and a catheter inserted through the working channel of an endoscope, allowing access to the targeted site for hemostasis.
The main difference in technological characteristics is the composition of the hemostatic powder. Nexpowder contains mainly Aldehyded Dextran and Poly-L-Lysine (Polysaccharide), while the predicate is composed of proprietary components and bentonite powder. Even though the hemostatic agent of Nexpowder differs from the predicate device in chemical composition, it does not raise different questions of safety and effectiveness.
{7}------------------------------------------------
In terms of Principles of operation. Nexpowder and the predicate device, Hemospray, differ in the delivery method and energy source that provides the physical force to move the hemostatic powder agent into the delivery catheter.
The Hemospray includes a handle with a pressurized CO2 cartridge and the powder is propelled through the catheter by release of CO2 from the cartridge located in the device.
The Nexpowder hemostatic agent is deployed by insoluble air propellant. It utilizes a pre-installed battery as its power source, allowing air pressure generated from the air pump in the delivery system's spray body to provide effective physical force to move the powder into the delivery catheter.
The air-powered mechanism of Nexpowder results in low-pressure, controlled delivery of material with minimal particle scattering or loss of visualization. Other than the air pressure itself, no mechanical force or direct contact is made with the target hemostasis site during the delivery process of Nexpowder.
### Nexpowder and reference device
Nexpowder is technologically identical to the reference device (Nexpowder cleared in K202929). The overall design, materials, chemical composition, and energy source of Nexpowder have not been modified since cleared by the agency.
The only difference from the cleared device is the expansion of indications to the lower GI and several minor device modifications that were assessed by the company and determined to not significantly affect the safety or efficacy of the device.
#### PERFORMANCE DATA VII.
The company collected clinical data outside the United States (OUS) in support of the lower GI indication expansion. Clinical data was gathered from a prospective study and retrospective studies conducted in Korea, in addition to aggregated data collected in the European Union (EU). A total sample size of 260 lower GI cases were collected and analyzed. Results were compared to the predicate device- Hemospray. The results of Nexpowder were found to be comparable with that of Hemospray as presented in its De-Novo authorization (DEN170015), further supporting Nexpowder indication expansion to the lower GI.
{8}------------------------------------------------
All other pre-clinical testing in support of Nexpowder was leveraged from the previously cleared Nexpowder device, cleared under K202929.
#### CONCLUSIONS VIII.
Nexpowder has the same intended use and indications for use as the predicate device- Hemosprav (DEN170015). The two devices differ in the powder chemical characteristics, delivery method and energy source, but those differences do not raise different questions of safety and effectiveness, as the safety and effectiveness of Nexpowder powder and device design were already established in K202929.
The Nexpowder has the same intended use, technological characteristics, and principles of operation as its reference device (K202929). The expanded indication does not alter the intended use of the device. The safety and effectiveness of the device in the lower GI was established by clinical data and is further supported by comparing Nexpowder data to that of the predicate device, Hemospray, already cleared for the lower GI.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.