SteriTite rigid reusable sterilization container with MediTray Products
K240898 · Case Medical · KCT · Jul 31, 2024 · General Hospital
Device Facts
Record ID
K240898
Device Name
SteriTite rigid reusable sterilization container with MediTray Products
Applicant
Case Medical
Product Code
KCT · General Hospital
Decision Date
Jul 31, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The SteriTite container and MediTray products are a reusable sterilization container system to be used to enclose other medical devices, which are to be sterilized, transported, and stored by a health care provider. It is intended to allow sterilization of the enclosed devices and maintain sterility of the enclosed device until it is used. The solid and perforated bottom container configurations are compatible for use with the following low temperature sterilizers and cycles identified below: - STERIS V-Pro 60: Flexible and Lumen cycle (Solid and perforated bottom container) - STERIS V-Pro s2: Flexible and Lumen cycle (Solid and perforated bottom container) - STERIS V-Pro maX 2: Lumen, Non-Lumen and Flexible cycle (Solid and perforated bottom container) - STERIS V-Pro maX: Lumen, Non-Lumen and Flexible cycle (Solid and perforated bottom container)
Device Story
Rigid reusable sterilization container system; consists of anodized aluminum lid/base, passivated stainless-steel hardware, and platinum-cured silicone gaskets. Used by healthcare providers in clinical settings to enclose medical instruments during sterilization, transport, and storage. Input: medical devices placed inside container with MediTray inserts. Operation: container utilizes perforated lid vents with single-use polypropylene filters to allow sterilant penetration while maintaining microbial barrier; solid base design reduces potential microbial ingression points compared to perforated-base predicates. Output: sterilized, protected medical instruments. Healthcare providers use the system to maintain sterility until point of use; tamper-evident seals provide visual confirmation of container integrity. Benefits: provides durable, reusable packaging that ensures sterility maintenance and protects instruments from damage during handling.
Clinical Evidence
Bench testing only. Testing conducted in accordance with ANSI/AAMI ST77, AAMI TIR 12, and ANSI/AAMI ST-79. Validated using the 'overkill method' with G. stearothermophilus biological indicators. Endpoints included sterility maintenance, whole package microbial challenge, biocompatibility (ISO 10993-1), material compatibility, and sterilization efficacy in V-Pro cycles. All tests passed.
Technological Characteristics
Materials: Anodized aluminum (lid/base), passivated stainless steel (hardware), platinum-cured silicone (gaskets), polypropylene (single-use filters). Design: Rigid container with gasketed lid-to-base seal, tamper-evident latching, and filter retention plates. Sterilization: Low-temperature vaporized hydrogen peroxide (V-Pro cycles) and pre-vacuum steam. Standards: ANSI/AAMI ST77, ISO 10993-1, ISO 14937.
Indications for Use
Indicated for use by healthcare providers to enclose, sterilize, transport, and store medical devices. Compatible with specific STERIS V-Pro low-temperature sterilization cycles.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
SteriTite Rigid Container Sterilization System with and MediTray products (K212711)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 31, 2024
Case Medical
Marcia Frieze
CEO
50 West Street
Bloomfield, New Jersey 07003
Re: K240898/S001
Trade/Device Name: SteriTite rigid reusable sterilization container with MediTray Products
Regulation Number: 21 CFR 880.6850
Regulation Name: Sterilization Wrap
Regulatory Class: Class II
Product Code: KCT
Dated: March 25, 2024
Received: April 1, 2024
Dear Marcia Frieze:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K240898 - Marcia Frieze
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Christopher K. Dugard -S
Christopher K. Dugard, MS
Assistant Director
Center for Devices and Radiological Health
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Enclosure
{2}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K240898
Device Name
SteriTite rigid reusable sterilization container with MediTray Products
Indications for Use (Describe)
The SteriTite container and MediTray products are a reusable sterilization container system to be used to enclose other medical devices, which are to be sterilized, transported, and stored by a health care provider. It is intended to allow sterilization of the enclosed devices and maintain sterility of the enclosed device until it is used. The solid and perforated bottom container configurations are compatible for use with the following low temperature sterilizers and cycles identified below:
- STERIS V-Pro 60: Flexible and Lumen cycle (Solid and perforated bottom container)
- STERIS V-Pro s2: Flexible and Lumen cycle (Solid and perforated bottom container)
- STERIS V-Pro maX 2: Lumen, Non-Lumen and Flexible cycle (Solid and perforated bottom container)
- STERIS V-Pro maX: Lumen, Non-Lumen and Flexible cycle (Solid and perforated bottom container)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.**
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Type of Use (Select one or both, as applicable)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}
The following tables (1-3) identify the sterilizer maximum load weight recommendations with the SteriTite containers in V-Pro sterilization cycles as well as lumen claims per cycle.
Table 1. SteriTite Container in V-Pro s2 and V-Pro 60 including Weight of Container
| Part Number | Total Load Weight in V-Pro s2/60 Non-Lumen Cycle | Total Load Weight in V-Pro s2/60 Flexible Cycle |
| --- | --- | --- |
| SC02M | 25 lbs/11.3kg | 11 lbs/4.9kg |
| SC03M | 25 lbs/11.3kg | 11 lbs/4.9kg |
| SC04M | 25 lbs/11.3kg | 11 lbs/4.9kg |
| SC02N | 25 lbs/11.3kg | 11 lbs/4.9kg |
| SC03N | 25 lbs/11.3kg | 11 lbs/4.9kg |
| SC03NL | 25 lbs/11.3kg | 11 lbs/4.9kg |
| SC04NL | 25 lbs/11.3kg | 11 lbs/4.9kg |
| SC05NL | 25 lbs/11.3kg | 11 lbs/4.9kg |
| SC04H | 25 lbs/11.3kg | 11 lbs/4.9kg |
| SC05H | 25 lbs/11.3kg | 11 lbs/4.9kg |
| SC04Q | 25 lbs/11.3kg | 11 lbs/4.9kg |
| SC05Q | 25 lbs/11.3kg | 11 lbs/4.9kg |
| SC04F | 25 lbs/11.3kg | 11 lbs/4.9kg |
Table 2. SteriTite Container Maximum Load Weight in V-Pro maX/maX 2 including Weight of Container
| Part Number | Total Load Weight in V-Pro maX/maX 2 Lumen Cycle | Total Load Weight in V-Pro maX/maX 2 Flex Cycle | Total Load Weight in V-Pro maX/maX 2 Non-Lumen Cycle |
| --- | --- | --- | --- |
| SC02M | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC03M | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC04M | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC02N | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC03N | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC03NL | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC04NL | 19.65 lbs/ 8.9kg | 24lbs/11kg | 50 lbs/22.6kg |
| SC05NL | 19.65 lbs/ 8.9kg | 24lbs/11kg | 50 lbs/22.6kg |
| SC04H | 19.65 lbs/ 8.9kg | 24lbs/11kg | 50 lbs/22.6kg |
| SC05H | 19.65 lbs/ 8.9kg | 24lbs/11kg | 50 lbs/22.6kg |
| SC06H | 19.65 lbs/ 8.9kg | 24lbs/11kg | 50 lbs/22.6kg |
| SC08H | 19.65 lbs/ 8.9kg | 24lbs/11kg | 50 lbs/22.6kg |
| SC04Q | 19.65 lbs/ 8.9kg | 24lbs/11kg | 50 lbs/22.6kg |
| SC05Q | 19.65 lbs/ 8.9kg | 24lbs/11kg | 50lbs/22.6kg |
| SC06Q | 19.65 lbs/ 8.9kg | 24lbs/11kg | 50lbs/22.6kg |
| SC08Q | 19.65 lbs/ 8.9kg | 24lbs/11kg | 50lbs/22.6kg |
| SC04F | 19.65 lbs/ 8.9kg | 24lbs/11kg | 50lbs/22.6kg |
| SC05F | 19.65 lbs/ 8.9kg | 24lbs/11kg | 50lbs/22.6kg |
| SC06F | 19.65 lbs/ 8.9kg | 24lbs/11kg | 50lbs/22.6kg |
| SC08F | 19.65 lbs/ 8.9kg | 24lbs/11kg | 50lbs/22.6kg |
| SC04L | 19.65 lbs/ 8.9kg | 24lbs/11kg | 50lbs/22.6kg |
| SC06L | 19.65 lbs/ 8.9kg | 24lbs/11kg | 50lbs/22.6kg |
| SC08L | 19.65 lbs/ 8.9kg | 24lbs/11kg | 50lbs/22.6kg |
| SC05W | 19.65 lbs/ 8.9kg | 24lbs/11kg | 50lbs/22.6kg |
{4}
Table 3. Sterilant penetration for channel lumens using solid bottom container in STERIS V-Pro sterilization
| Sterilizer | Cycle | Lumen Sterilization Claims (I.D. x Length) |
| --- | --- | --- |
| STERIS V-Pro 60 | Flexible | ≥1mm x ≤990mm: Single or Dual Lumen (Solid Bottom)* |
| STERIS V-Pro 60 | Lumen | ≥0.77mm x ≤527mm: Dual Channel (Solid Bottom)** |
| STERIS V-Pro s2 | Flexible | ≥1mm x ≤990mm: Single or Dual Lumen (Solid Bottom) |
| STERIS V-Pro s2 | Lumen | ≥0.77mm x ≤527mm: Dual Channel (Solid Bottom) |
| STERIS V-Pro maX 2 | Flexible | ≥1mm x ≤1050mm: Single Lumen (Solid Bottom) |
| STERIS V-Pro maX 2 | Lumen | ≥0.77mm x ≤527mm: Dual Channel (Solid Bottom) |
| STERIS V-Pro maX | Flexible | ≥1mm x ≤1050mm: Single Lumen (Solid Bottom) |
| STERIS V-Pro maX | Lumen | ≥0.77mm x ≤527mm: Dual Channel (Solid Bottom) |
{5}
Confidential Information
Page 1 of 8
K240898
# 510K Summary K240898
## SteriTite rigid reusable sterilization container with MediTray Products
**Date Prepared**: 03/25/2024
**Company Name**: Case Medical, Inc
50 West Street
Bloomfield NJ 07003
**Contact**: Marcia Frieze
**Email**: MFrieze@casemed.com
**Office**: 201-313-1999 x 225
**Fax**: 201-373-9090
**Trade Name**: SteriTite Container System with MediTray Products
**Common Name**: SteriTite Universal Container System
**Class of Device**: Class II
**Product Code**: KCT
**Review Panel**: General Hospital
**Establishment Registration Number**: 2248608
**Predicate Device**: K212711 -SteriTite Rigid Container Sterilization System with and MediTray products
## Description of Device
The SteriTite® Universal Container system consists of a family of rigid reusable containers and inserts that provide an effective reusable sterilization packaging system for operating room instruments. The SteriTite container, as previously cleared, is available with solid or perforated base. The container is made of anodized aluminum with passivated stainless-steel hardware and silicone gaskets, disposable single use filter, tamper evident seal, and perforated vents for sterilant penetration and air removal.
## 10.3 Indications for Use Summary
The SteriTite container and MediTray products are a reusable sterilization container system to be used to enclose other medical devices, which are to be sterilized, transported, and stored by a health care provider. It is intended to allow sterilization of the enclosed devices and maintain sterility of the enclosed device until it is used. The solid and perforated bottom container configurations are compatible for use with the following low temperature sterilizers and cycles identified below:
- STERIS V-Pro 60: Flexible and Lumen cycle (Solid and perforated bottom container)
- STERIS V-Pro s2: Flexible and Lumen cycle (Solid and perforated bottom container)
- STERIS V-Pro maX 2: Lumen, Non-Lumen and Flexible cycle (Solid and perforated bottom container)
- STERIS V-Pro maX: Lumen, Non-Lumen and Flexible cycle (Solid and perforated bottom container)
Case Medical, Inc | 50 West St. Bloomfield NJ 07003 | 201-313-1999 | www.casemed.com
{6}
Confidential Information
Page 2 of 8
K240898
The following table (Table 1) displays the SteriTite part numbers and the sterilizers they are compatible with:
Table 1. SteriTite Solid Bottom Container Compatibility with Low Temperature and Pre vacuum Steam Sterilizers.
| Part Number | Description | V-Pro maX/maX 2 | V-Pro s2/60 |
| --- | --- | --- | --- |
| SC02M | 2” High Mini-Size Solid Base | Yes | Yes |
| SCO3M | 3” High Mini-Size Solid Base | Yes | Yes |
| SCO4M | 4” High Mini-Size Solid Base | Yes | Yes |
| SC02N | 2” High Narrow-Size Solid Base | Yes | Yes |
| SC03N | 3” High Narrow-Size Solid Base | Yes | Yes |
| SC03NL | 3” High Narrow Long-Size Solid Base | Yes | Yes |
| SC04NL | 4” High Narrow Long-Size Solid Base | Yes | Yes |
| SC05NL | 5” High Narrow Long -Size Solid Base | Yes | Yes |
| SC04H | 4” High Half-Size w/ Solid Base | Yes | Yes |
| SC05H | 5” High Half-size w/ Solid Base | Yes | Yes |
| SC06H | 6” High Half-Size w/ Solid Base | Yes | N/A |
| SC08H | 8” High Half-Size w/ Solid Base | Yes | N/A |
| SC04Q | 4” High 3/4 Size w/ Solid Base | Yes | Yes |
| SC05Q | 5” High 3/4 Size w/ Solid Base | Yes | Yes |
| SC06Q | 6” High 3/4 Size w/ Solid Base | Yes | N/A |
| SC08Q | 8” High 3/4 Size w/ Solid Base | Yes | N/A |
| SC04F | 4” High Full-size w/ Solid Base | Yes | Yes |
| SC05F | 5” High Full-size w/ Solid Base | Yes | N/A |
| SC06F | 6” High Full-size w/ Solid Base | Yes | N/A |
| SC08F | 8” High Full-size w/ Solid Base | Yes | N/A |
| SC04L | 4” High Long-size w/ Solid Base | Yes | N/A |
| SC06L | 6” High Long-size w/ Solid Base | Yes | N/A |
| SC08L | 8” High Long-size w/ Solid Base | Yes | N/A |
| SC05W | 5” High Long Size w/ Solid Base | Yes | N/A |
The following table (Table 2) identifies which MediTray products are compatible with low temperature sterilizers and steam sterilization:
Table 2. MediTray Compatibility with Low Temp. Sterilizers & Steam Containerized and Wrapped
| MediTray Product | V-Pro maX/2 | V-Pro s2/60 | STERIZONE VP4 | STERRAD NX/ 100 NX | Steam |
| --- | --- | --- | --- | --- | --- |
| Baskets | Yes | Yes | Yes | Yes | Yes |
| Trays | Yes | Yes | Yes | Yes | Yes |
| Insert Boxes | Yes | Yes | Yes | Yes | Yes |
| Metal Brackets | Yes | Yes | Yes | Yes | Yes |
| Metal Partitions | Yes | Yes | Yes | Yes | Yes |
| Posts | Yes | Yes | Yes | Yes | Yes |
| Silicone Brackets | Yes | Yes | Yes | Yes | Yes |
| Racks | Yes | Yes | Yes | Yes | Yes |
| Stringers | Yes | Yes | Yes | Yes | Yes |
Case Medical, Inc | 50 West St. Bloomfield NJ 07003 | 201-313-1999 | www.casemed.com
{7}
Confidential Information
Page 3 of 8
K240898
The following table (Table 3) identifies which SteriTite accessories are compatible with low temperature sterilization:
Table 3. SteriTite Consumables for Low Temperature
| SteriTite Disposables | Steris V-Pro maX/maX2 | V-Pro s2/60 |
| --- | --- | --- |
| SCF02 Round Polypro* filter | Yes | Yes |
| SCFM02 Rectangular Polypro* filter | Yes | Yes |
| SCS01W Tamper Evident Seal White | Yes | Yes |
| SCLH2O23 Load Card Large | Yes | Yes |
| SCLH2O24 Load Card Small | Yes | Yes |
*PolyPro is a non-woven polypropylene filter material that must be used for vaporized hydrogen peroxide and gas plasma. It may be used for steam sterilization. Use round filters for standard container sizes, rectangular filters for mini, narrow, and narrow long containers.
The following tables (4-5) identify the sterilizer maximum load weight recommendations with the SteriTite containers in Low temperature sterilization:
Table 4. SteriTite Container in V-Pro s2 and V-Pro 60 including Weight of Container
| Part Number | Total Load Weight in V-Pro s2/60 Non-Lumen Cycle | Total Load Weight in V-Pro s2/60 Flexible Cycle |
| --- | --- | --- |
| SC02M | 25 lbs/11.3kg | 11 lbs/4.9kg |
| SC03M | 25 lbs/11.3kg | 11 lbs/4.9kg |
| SC04M | 25 lbs/11.3kg | 11 lbs/4.9kg |
| SC02N | 25 lbs/11.3kg | 11 lbs/4.9kg |
| SC03N | 25 lbs/11.3kg | 11 lbs/4.9kg |
| SC03NL | 25 lbs/11.3kg | 11 lbs/4.9kg |
| SC04NL | 25 lbs/11.3kg | 11 lbs/4.9kg |
| SC05NL | 25 lbs/11.3kg | 11 lbs/4.9kg |
| SC04H | 25 lbs/11.3kg | 11 lbs/4.9kg |
| SC05H | 25 lbs/11.3kg | 11 lbs/4.9kg |
| SC04Q | 25 lbs/11.3kg | 11 lbs/4.9kg |
| SC05Q | 25 lbs/11.3kg | 11 lbs/4.9kg |
| SC04F | 25 lbs/11.3kg | 11 lbs/4.9kg |
Case Medical, Inc | 50 West St. Bloomfield NJ 07003 | 201-313-1999 | www.casemed.com
{8}
Confidential Information
Page 4 of 8
K240898
Table 5. SteriTite Container Maximum Load Weight in V-Pro maX/maX 2 including Weight of Container
| Part Number | Total Load Weight in V-Pro maX/maX 2 Lumen Cycle | Total Load Weight in V-Pro maX/maX 2 Flex Cycle | Total Load Weight in V-Pro maX/maX 2 Non-Lumen Cycle |
| --- | --- | --- | --- |
| SC02M | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC03M | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC04M | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC02N | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC03N | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC03NL | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC04NL | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC05NL | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC04H | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC05H | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC06H | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC08H | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC04Q | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC05Q | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC06Q | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC08Q | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC04F | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC05F | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC06F | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC08F | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC04L | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC06L | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC08L | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
| SC05W | 19.65 lbs/ 8.9kg | 24 lbs/11kg | 50 lbs/22.6kg |
# Indications For Use Comparison
The device has no different indication for use than the predicate (K212711). The SteriTite rigid reusable sealed container system with solid base is equivalent to the SteriTite container with perforated base (predicate) previously cleared for compatibility with Steris V-Pro Sterilizers, including Steris V-Pro maX (K112904), V-Pro maX 2, V-Pro s2, V-Pro 60 (K221492, K212711). Case Medical adheres to the overkill principle for sterilization validation testing. As a worst-case scenario, half cycles are utilized, as are end of shelf-life concentrations of hydrogen peroxide for low temperature sterilization validation. The SteriTite container system has also been previously cleared for compatibility with other various low temperature sterilizers as well as steam sterilization. All functional and design characteristics are the same as the predicate previously cleared and show equivalent performance characteristics. The SteriTite container system with MediTray products presented in the current submission shares the same technological characteristics as the previously cleared SteriTite container and MediTray products. This includes anodized lid and base, passivated stainless steel hardware, gasketed seal between lid and base, platinum cured silicone gaskets, perforated vent pattern on the lid, utilizing polypropylene filters (polypropylene filters specified for H2O2 sterilizers), tamper evident latching mechanism, and one year shelf life (365 days) for sterility maintenance
Case Medical, Inc | 50 West St. Bloomfield NJ 07003 | 201-313-1999 | www.casemed.com
{9}
Confidential Information
Page 5 of 8
K240898
# Technological Comparison
Table 6. Technological Characteristics of Predicate and Subject Device
| Technological Characteristics | Predicate | Subject Device |
| --- | --- | --- |
| Anodized aluminum lid and base | Same | Same |
| Passivated stainless steel hardware | Same | Same |
| Platinum cured silicone gaskets | Same | Same |
| Perforated vent pattern on lid | Same | Same |
| Perforated vent pattern on base | Yes | Different |
| Interchangeable filter retention plate | Same | Same |
| Polypropylene single use filters | Same | Same |
| Tamper evident latching mechanism | Same | Same |
| One year shelf-life sterility maintenance | Same | Same |
| Pre-Vacuum Steam Sterilization | Same | Same |
| Low Temperature Sterilization (V-Pro) | Same | Same |
The following tables (6a-6d) list the substantial equivalence of the previously cleared perforated bottom SteriTite container with MediTray products (predicate) K212711 and the solid base SteriTite container system and MediTray products (subject device).
Table 6a Indications for Use
| INDICATIONS FOR USE: | Predicate | Subject Device |
| --- | --- | --- |
| Intended to enclose medical devices | Same | Same |
| Intended for holding instruments to be sterilized, transported, and stored | Same | Same |
| Intended to be reused | Same | Same |
| Protect instruments from damage | Same | Same |
| Maintain the sterility of enclosed medical devices | Same | Same |
| Provide a barrier to microorganisms | Same | Same |
Table 6b Device Design
| DESIGN: | Predicate | Subject Device |
| --- | --- | --- |
| Rectangular anodized aluminum container consisting of lid and base | Same | Same |
| Recessed gasket seal between lid and base | Same | Same |
| Filter system to permit entry of sterilant agent and exit | Same | Same |
| Off set retention plate | Same | Same |
| Microbial barrier migration | Same | Same |
| Interchangeable filter retention plates with gasket | Same | Same |
| Latching mechanism to secure lid and base | Same | Same |
| Tamper-proof locking mechanisms | Same | Same |
| Handles for ease of transport | Same | Same |
| Passivated stainless-steel hardware | Same | Same |
| A knife edge fit between lid and base. | Same | Same |
| Various instrument trays to correspond with container | Same | Same |
| Organizational inserts | Same | Same |
| Internal and external stacking | Same | Same |
| Aseptic presentation of contents | Same | Same |
Case Medical, Inc | 50 West St. Bloomfield NJ 07003 | 201-313-1999 | www.casemed.com
{10}
Confidential Information
Page 6 of 8
K240898
Table 6c Performance Standards/Specifications
| PERFORMANCE STANDARDS/ SPECIFICATIONS: | Predicate | Subject Device |
| --- | --- | --- |
| Sterile barrier system | Same | Same |
| Secure transport of sterilized contents to point of use | Same | Same |
| Shelf life 365 days/ 1year of sterility maintenance | Same | Same |
| Dry outcome in all methods of sterilizations | Same | Same |
| Corrosion resistant seal | Same | Same |
| Inspection Instructions included in labeling | Same | Same |
| Durable construction for long use life | Same | Same |
| User responsibilities listed in labeling | Same | Same |
| Remain sealed until opened | Same | Same |
| The filter retention plates audibly click to lock into place | Same | Same |
| Tamper evident seals on latches provide visual evidence of sterility breach | Same | Same |
| Lid and bottom are made of anodized aluminum making them lightweight | Same | Same |
Table 6d Validation
| VALIDATION: | Predicate | Subject Device |
| --- | --- | --- |
| Half Cycle Test Conditions | Same | Same |
| Maximum Load Test | Same | Same |
| Microbial Barrier | Same | Same |
| Test Organism G. Stearothermophilus | Same | Same |
| Simulated Use Test | Same | Same |
| Inoculated blades | Same | Same |
| Inoculated lumens | Same | Same |
| Occlusion / matted surfaces | Same | Same |
| Sterilization of single and dual lumen devices | Same | Same |
| Testing in lumen, non-lumen, and flexible cycles | Same | Same |
Non-Clinical Performance Testing
Case Medical, Inc. follows the “overkill method” of sterility assurance to show an elimination of a biological challenge. All SteriTite containers were tested with inner baskets or trays representative of the MediTray line of products. Testing was designed to simulate worst-case scenarios with half cycles utilized with end of shelf-life concentrations of hydrogen peroxide for low temperature sterilization. The bacterial strain G. stearothermophilus was used in all testing due to its resistance to moist heat and hydrogen peroxide. The validation testing was conducted at qualified independent laboratories in accordance with FDA guidance and available ANSI/AAMI standards such as ANSI/AAMI ST77, AAMI TIR 12 and ANSI/AMMI ST-79. Results are summarized in Table 7 below.
Table 7. Summary of Performance Testing
The solid bottom container is substantially equivalent to the perforated bottom container. However, it has greater performance features for sterility maintenance and for maintaining a microbial barrier as there are no perforations or vents in the bottom of the container.
Case Medical, Inc | 50 West St. Bloomfield NJ 07003 | 201-313-1999 | www.casemed.com
{11}
Confidential Information
Page 7 of 8
K240898
| Performance Testing | Standard | Purpose | Results |
| --- | --- | --- | --- |
| Sterility Maintenance | ANSI/AAMI ST77:2013 (R)2018 | Demonstrate the integrity of the container to maintain sterility of contents over time and with multiple handling events after sterilization. | Pass |
| Whole Package Microbial Challenge | ANSI/AAMI ST77:2013 (R)2018 | Demonstration of whole package integrity following exposure to aerosolized bacteria spores. | Pass |
| Biocompatibility | ANSI AAMI ISO10993- 1 | Acute systemic toxicity, primary irritation and sensitization of the subject device. | Pass |
| Material Compatibility / Re- Use Testing | ANSI/AAMI ST77:2013 (R)2018 | Demonstration that the subject device materials are compatible with the intended sterilization cycles. | Pass |
| Simulated Use Testing/Sterilization Efficacy | ANSI AAMI ISO 14937:2009/(R)2013 | Demonstrated sterilization efficacy of the subject device for processing medical devices using the identified sterilization processes. | Pass |
Table 8. Sterilant penetration
| Test Device | Predicate | Subject Device |
| --- | --- | --- |
| General instrumentation | Same | Same |
| Single/double channel flexible endoscopes | Same | Same |
| Single/double channel rigid endoscopes | Same | Same |
| Closed-end Cannula | N/A | Pass |
Table 9. Sterilant penetration for channel lumens using solid bottom container in low temperature sterilization and steam
| Sterilizer | Cycle | Lumen Sterilization Claims (I.D. x Length) |
| --- | --- | --- |
| STERIS V-Pro 60 | Flexible | ≥1mm x ≤990mm: Single or Dual Lumen (Solid Bottom)* |
| STERIS V-Pro 60 | Lumen | ≥0.77mm x ≤527mm: Dual Channel (Solid Bottom)** |
| STERIS V-Pro s2 | Flexible | ≥1mm x ≤990mm: Single or Dual Lumen (Solid Bottom) |
| STERIS V-Pro s2 | Lumen | ≥0.77mm x ≤527mm: Dual Channel (Solid Bottom) |
| STERIS V-Pro maX 2 | Flexible | ≥1mm x ≤1050mm: Single Lumen (Solid Bottom) |
| STERIS V-Pro maX 2 | Lumen | ≥0.77mm x ≤527mm: Dual Channel (Solid Bottom) |
| STERIS V-Pro maX | Flexible | ≥1mm x ≤1050mm: Single Lumen (Solid Bottom) |
| STERIS V-Pro maX | Lumen | ≥0.77mm x ≤527mm: Dual Channel (Solid Bottom) |
| Sterizone VP4 | Cycle 1 | ≥1.2mm x ≤1955mm: Flexible Lumens (Solid Bottom) |
| Sterizone VP4 | Cycle 1 | ≥1.45mm x ≤3500mm: Flexible Lumens (Solid Bottom) |
| Steam | Pre-Vac | ≥2mm x ≤400mm: Stainless Steel Lumen (Solid Bottom) ≥1.2mm x ≤400mm: Flexible Lumen (Solid Bottom) |
Case Medical, Inc | 50 West St. Bloomfield NJ 07003 | 201-313-1999 | www.casemed.com
{12}
Confidential Information
Page 8 of 8
K240898
*The validation study in the flexible cycle utilized lumens of 0.48 mm I.D x 100 mm L. Additional lumens in the load were 1.0mm I.D x 1050 mm L. (Attachment 013- Steris Lab Report 14463393.A) ** ≥ 0.8mm ID and ≤ 542 mm in length lumens have been shown to be equivalent to be equivalent for validation testing purposes to the predicate ≥0.77mm ID and ≤527mm in length (Attachment 011- Steris Lab report 14463391.A).
# Conclusion
The conclusion drawn from the nonclinical performance tests demonstrate that the subject device in this 510(k) submission for the SteriTite Container System and MediTray Products, is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K212711.
The solid bottom container has been previously cleared in pre-vacuum steam, EO, VP4 sterilizer and received FDA 510k clearance. This submission is for compatibility of the SteriTite solid bottom container system in the V-Pro maX, V-Pro maX2, V-Pro s2 and the V-Pro 60 sterilizer. The solid bottom container is substantially equivalent to the perforated bottom container previously cleared. However, it has greater performance features for sterility maintenance and for maintaining a microbial barrier as there are no perforations or vents in the bottom of the container. SteriTite Universal Container System and MediTray Products is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K212711.
All functional and design characteristics are the same as the predicate previously cleared in K212711 and show equivalent performance characteristics. The SteriTite container system with solid base for this submission is the same container system that has been previously cleared for steam and low temperature sterilization with perforated base. The solid base line of SteriTite container systems has only perforations on the lid compared to the predicate which has perforations in lid and base. Therefore, there are fewer inlets for microbial ingression. The solid bottom line of SteriTite container systems demonstrated the same technical and performance characteristics in the V-Pro maX, V-Pro maX 2, V-Pro s2, and V-Pro 60 as the predicate. The functional and design characteristics are the same as the predicate previously cleared and show equivalent performance characteristics.
Case Medical, Inc | 50 West St. Bloomfield NJ 07003 | 201-313-1999 | www.casemed.com
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.