Fluent Pro Fluid Management System (FLT-200); Fluent Pro Fluid Management System Disposable Procedure Kit (6-pack) (FLT-212); Fluent Pro Fluid Management System Disposable Procedure Kit (1-pack) (FLT-212S); Fluent Pro Fluid Management System Tissue Trap Multipack (10-pack) (FLT-210); Fluent Pro Fluid Management System Waste Bag Multipack (5-pack) (FLT-205)
Fluent Pro Fluid Management System (FLT-200); Fluent Pro Fluid Management System Disposable Procedure Kit (6-pack) (FLT-212); Fluent Pro Fluid Management System Disposable Procedure Kit (1-pack) (FLT-212S); Fluent Pro Fluid Management System Tissue Trap Multipack (10-pack) (FLT-210); Fluent Pro Fluid Management System Waste Bag Multipack (5-pack) (FLT-205)
Applicant
Hologic, Inc.
Product Code
HIG · Obstetrics/Gynecology
Decision Date
Jul 25, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1700
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Fluent Pro Fluid Management System is intended to provide liquid distension of the uterus during diagnostic and operative hysteroscopy, and to monitor the volume differential between the irrigation liquid flowing into and out of the uterus while providing drive, control and suction for hysteroscopic morcellators.
Device Story
Fluent Pro Fluid Management System is a console-based device for hysteroscopic procedures; provides uterine distension via irrigation and monitors fluid deficit. System includes a console and single-use procedure kit (inflow/outflow tube sets, waste bag). Console features electro-mechanical peristaltic pumps for fluid control, pressure sensors, and a graphical user interface (GUI). Device provides drive, control, and suction for MyoSure tissue removal devices; operated by physicians in clinical settings. Input includes fluid flow data; output includes real-time pressure and fluid deficit metrics displayed on the GUI. Adjustable suction settings (Low, Medium, High) allow physician control during tissue removal. System enhances procedural safety by monitoring fluid deficit and maintaining intrauterine pressure, facilitating efficient tissue removal and visualization.
Clinical Evidence
Bench testing only. Evidence includes intrauterine pressure control and fluid deficit accuracy testing, mechanical/hardware/electrical bench testing, software unit/integration/system testing, and disposables pull force/flow rate/duty cycle testing. Biocompatibility testing per ISO 10993-1, sterilization validation per ISO 11135, and usability testing per IEC 62366-1 were performed. No clinical data was required or provided.
Technological Characteristics
Console-based hysteroscopic insufflator with electro-mechanical peristaltic pumps. Materials comply with ISO 10993-1. Connectivity includes pneumatic foot pedal and MyoSure device interface. Software-controlled GUI for pressure and deficit monitoring. Sterilization of disposable kits via Ethylene Oxide. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 62304 (software), ISO 14971 (risk management).
Indications for Use
Indicated for patients undergoing diagnostic and operative hysteroscopy requiring uterine distension and fluid volume monitoring, and for those requiring hysteroscopic morcellation.
Regulatory Classification
Identification
A hysteroscopic insufflator is a device designed to distend the uterus by filling the uterine cavity with a liquid or gas to facilitate viewing with a hysteroscope.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below that.
July 25, 2024
Hologic, Inc. Meghan Wakeford Principal Regulatory Affairs Specialist 250 Campus Drive Marlborough, Massachusetts 01752
#### Re: K240886
Trade/Device Name: Fluent Pro Fluid Management System (FLT-200); Fluent Pro Fluid Management System Disposable Procedure Kit (6-pack) (FLT-212); Fluent Pro Fluid Management System Disposable Procedure Kit (1-pack) (FLT-212S): Fluent Pro Fluid Management System Tissue Trap Multipack (10-pack) (FLT-210): Fluent Pro Fluid Management System Waste Bag Multipack (5-pack) (FLT-205) Regulation Number: 21 CFR 884.1700 Regulation Name: Hysteroscopic Insufflator Regulatory Class: II Product Code: HIG, HIH Received: June 27, 2024
Dear Meghan Wakeford:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
#### K240886 - Meghan Wakeford
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}------------------------------------------------
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jason Roberts -S
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K240886
Device Name Fluent Pro Fluid Management System
#### Indications for Use (Describe)
The Fluent Pro Fluid Management System is intended to provide liquid distension of the uterus during diagnostic and operative hysteroscopy, and to monitor the volume differential between the irrigation liquid flowing into and out of the uterus while providing drive, control and suction for hysteroscopic morcellators.
Type of Use (Select one or both, as applicable)
| <span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## Traditional 510(k) Summary
This 510(k) Summary is submitted in accordance with the requirements of 21 CFR Part 807.92
| Date Prepared: | March 29, 2024 |
|----------------------------------------------------------|-------------------------------------------------------------------------------|
| Manufacturer: | Hologic, Inc.<br>250 Campus Drive<br>Marlborough, MA 01752 USA |
| Establishment Registration #: | 1222780 |
| Contact Person: | Meghan Wakeford<br>Principal Regulatory Affairs Specialist<br>P: 508.263.6172 |
| Identification of the Device: | |
| Proprietary/Trade Name: | Fluent® Pro Fluid Management System |
| Classification Name: | Hysteroscopic Insufflator |
| Regulatory Number: | 21 CFR 884.1700 |
| Product Codes: | HIG (Hysteroscopic Insufflator), HIH (Hysteroscope and<br>accessories) |
| Device Class: | Class II |
| Review Panel: | Obstetrics/Gynecology |
| Identification of the Legally Marketed Predicate Device: | |
| Trade Name: | Fluent® Fluid Management System |
| Classification Name: | Hysteroscopic Insufflator |
| Regulatory Number: | 21 CFR 884.1700 |
| Product Codes: | HIG (Hysteroscopic Insufflator), HIH (Hysteroscope and<br>accessories) |
| Device Class: | Class II |
| Review Panel: | Obstetrics/Gynecology |
| Submitter/510(k) Holder: | Hologic, Inc. |
| Clearance: | K180825 (August 3, 2018) |
The legally marketed predicate device, the Fluid Management System (K180825) has not been subject to design-related recalls.
#### Device Description:
The Fluent Pro Fluid Management System is a hysteroscopic fluid management system and drive console for hysteroscopic morcellators. The Fluent Pro Fluid Management System consists of a console and single-use procedure kit. The single-use procedure kit consists of sterile inflow (In-FloPak) and outflow (Out-FloPak) tube sets, and a non-sterile waste bag. The FloPaks and waste bag connect for use in hysteroscopic procedures. The console includes motors that control fluid inflow and outflow for hysteroscopic insufflation, as well as a pressure sensor monitoring system and graphical user interface.
{5}------------------------------------------------
In addition, the console includes a connection for MyoSure Tissue Removal Devices and a pneumatic foot pedal to perform tissue removal procedures. The Fluent Pro Fluid Management System is compatible with currently marketed MyoSure Tissue Removal Devices.
#### Indications for Use:
The Fluent Pro Fluid Management System is intended to provide liquid distension of the uterus during diagnostic and operative hysteroscopy, and to monitor the volume differential between the irrigation liquid flowing into and out of the uterus while providing drive, control and suction for hysteroscopic morcellators.
#### Standards:
- ISO 14971:2019 Medical Devices Application of Risk Management to Medical Devices ●
- ISO 10993-1:2018 Biological Evaluation of Medical Devices Part 1: Evaluation and Testing . within a Risk Management Process
- IEC 60601-1:2005 + A1:2012 + A2:2020 – Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance
- IEC 60601-1-2:2014 + A1:2020 - Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
- IEC 62304:2006 + A1:2015 - Medical Device Software - Software Life Cycle Processes
- . IEC 62366-1:2015 + A1:2020 - Medical Devices - Part 1: Application of Usability Engineering to Medical Devices
#### FDA Guidance Documents:
- "Hysteroscopic and Laparoscopic Insufflators: Submission Guidance for a 510(k)," issued July ● 31, 1995
- "Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices Part 1: . Evaluation and Testing within a Risk Management Process"" issued September 8, 2023
- . "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile," issued January 8, 2024
- "Content of Premarket Submissions for Device Software Functions," issued June 14, 2023
- "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions," issued September 27, 2023.
- "Electromagnetic Compatibility (EMC) of Medical Devices," issued June 6, 2022
- "Applying Human Factors and Usability Engineering to Medical Devices," issued February 3, ● 2016
#### Comparison with Predicate Device:
The Hologic Fluent Pro Fluid Management System has the same intended use, method of operation, and key functional elements as the predicate device, the Hologic Fluent Fluid Management System (K180825). The proposed device presents similar technological characteristics as the predicate device in order to provide fluid distention of the uterus during diagnostic and operative hysteroscopy procedures and to monitor the volume differential between the irrigation fluid flowing into and out of the uterus, while providing drive, control, and suction for the MyoSure tissue removal devices.
{6}------------------------------------------------
#### Substantial Equivalence:
| | Fluent Pro Fluid Management System<br>Hologic, Inc.<br>Proposed Device | Fluent Fluid Management System<br>(K180825)<br>Hologic, Inc.<br>Predicate Device |
|-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | The Fluent Pro Fluid Management<br>System is intended to provide liquid<br>distension of the uterus during diagnostic<br>and operative hysteroscopy, and to<br>monitor the volume differential between<br>the irrigation liquid flowing into and out<br>of the uterus while providing drive,<br>control and suction for hysteroscopic<br>morcellators. | The Fluent Fluid Management System is<br>intended to provide liquid distension of<br>the uterus during diagnostic and<br>operative hysteroscopy, and to monitor<br>the volume differential between the<br>irrigation fluid flowing into and out of<br>the uterus while providing drive, control<br>and suction for hysteroscopic<br>morcellators. |
| Prescription<br>or OTC | Prescription Use | Prescription Use |
| Product Code | HIG, HIH | HIG, HIH |
| Regulation<br>Number | 21 CFR 884.1700 | 21 CFR 884.1700 |
| Method of<br>Use | Uterine distention to enable<br>hysteroscopic surgical removal of<br>intrauterine tissue | Uterine distention to enable<br>hysteroscopic surgical removal of<br>intrauterine tissue |
| Mechanism<br>of Action | Fluid flow with a single peristaltic pump<br>for irrigation and single peristaltic pump<br>for vacuum | Fluid flow with a single peristaltic pump<br>for irrigation and single peristaltic pump<br>for vacuum |
| Mode of<br>Operation | Provide liquid distension of the uterus<br>during diagnostic and operative<br>hysteroscopy, and to monitor the volume<br>differential between the irrigation fluid<br>flowing into and out of the uterus.<br>Vacuum is created to pull tissue into the<br>side-facing cutting window where it is<br>cut and aspirated through the cutter and<br>deposited into a collection canister. | Provide liquid distension of the uterus<br>during diagnostic and operative<br>hysteroscopy, and to monitor the volume<br>differential between the irrigation fluid<br>flowing into and out of the uterus.<br>Vacuum is created to pull tissue into the<br>side-facing cutting window where it is<br>cut and aspirated through the cutter and<br>deposited into a collection canister. |
| Inflow Pump<br>Design &<br>Mechanism | Electro-mechanical peristaltic pump | Electro-mechanical peristaltic pump |
| Pressure<br>Control<br>Range | 40-150 mmHg | 40-120 mmHg |
| Maximum<br>Intrauterine<br>Pressure | 150 mmHg | 120 mmHg |
| Maximum<br>Inflow Rate | 1000 mL/min | 650 mL/min |
| | Fluent Pro Fluid Management System<br>Hologic, Inc.<br>Proposed Device | Fluent Fluid Management System<br>(K180825)<br>Hologic, Inc.<br>Predicate Device |
| Adjustable<br>Suction<br>Settings | Low, Medium, and High | Non-adjustable |
| Intrauterine<br>Pressure<br>Display | Pressure setting and actual pressure level<br>displayed on the procedure screen | Pressure setting displayed on the<br>procedure screen |
| Deficit Limit<br>Range | 100-2500 mL in increments of 50 mL at<br>start of procedure<br><br>Deficit can be increased beyond 2500<br>mL after procedure has started based on<br>physician discretion | 100-2500 mL in increments of 50 mL at<br>start of procedure<br><br>Deficit can be increased beyond 2500<br>mL after procedure has started based on<br>physician discretion |
| Fluid Deficit<br>Tracking | Displays Deficit, Deficit Limit, and Total<br>Fluid In | Displays Deficit and Deficit Limit |
| Deficit<br>Accuracy | $\pm$ 50 mL under normal use | $\pm$ 50 mL under normal use |
| Drive<br>Control<br>Motor speed | 2200 $\pm$ 270 RPM | 2200 $\pm$ 270 RPM |
| MyoSure<br>Operating<br>Modes | System allows operation of TRD control<br>functions with or without operation of<br>the inflow and outflow pumps. | System allows operation of TRD control<br>functions during operation of the inflow<br>and outflow pumps. |
| Electrical<br>Safety &<br>EMC Testing | IEC 60601-1<br>IEC 60601-1-2 | IEC 60601-1<br>IEC 60601-1-2 |
| Disposable<br>Procedure<br>Kit<br>Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide |
| Biocompatibility<br>Testing<br>per ISO<br>10993-1 | Yes | Yes |
{7}------------------------------------------------
#### Summary of Testing:
The Fluent Pro Fluid Management System has successfully passed the following performance and simulated-use testing to demonstrate the device is substantially equivalent to the predicate devices and meets design specifications.
#### Biocompatibility Testing
Biocompatibility testing was performed on the patient contacting materials of the Fluent Pro Fluid Management System in accordance with FDA Guidance Use of International Standard ISO-10993,
{8}------------------------------------------------
"Biological evaluation of medical devices – Part 1: Evaluation and Testing within a Risk Management Process," issued on September 8, 2023. The material analysis and testing demonstrate the patient contacting materials are biocompatible and comply with the requirements of ISO 10993-1: 2018 Biological evaluation of medical devices – Part 1: Evaluation and Testing within a Risk Management Process for externally communicating, limited exposure, tissue contacting devices:
- . ISO 10993-5:2009 - Biological evaluation of medical devices - Part 5: Tests for in vitro cvtotoxicitv
- ISO 10993-10:2021 Biological evaluation of medical devices Part 10: Tests for skin ● sensitization
- ISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for systemic ● toxicity
- . ISO 10993-23:2021 - Biological evaluation of medical devices – Part 11: Tests for irritation
#### Sterilization and Shelf-Life Testing
The Fluent Pro Fluid Management System Procedure Kit is provided as a sterile, single-use device, Sterilization validation was conducted in accordance with ISO 11135:2014. The shelf-life of the disposable product and its packaging integrity were tested in accordance with ASTM F1980, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices, and has a shelf life of 36 months.
#### Electrical Safety and Electromagnetic Compatibility Testing
The Compass Revolution Colpotomy Device underwent electrical safety and electromagnetic compatibility testing in accordance with the following standards:
- IEC 60601-1:2005 + A1:2012 + A2:2020 Medical Electrical Equipment Part 1: General ● Requirements for Basic Safety and Essential Performance
- IEC 60601-1-2:2014 + A1:2020 - Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements and Tests
#### Performance Testing
The following performance testing was conducted to support the substantial equivalence of the Fluent Pro Fluid Management System to the predicate:
- Intrauterine Pressure Control and Fluid Deficit Bench Testing ●
- System Mechanical, Hardware, and Electrical Bench Testing ●
- Software Unit, Integration, and System Testing
- Disposables Pull Force. Flow Rate, and Duty Cycle Bench Testing .
#### Usability Testing
Simulated use testing was performed in accordance with IEC 62366-1:2015 + A1:2020 - Medical Devices - Part 1: Application of Usability Engineering to Medical Devices and FDA Guidance "Applying Human Factors and Usability Engineering to Medical Devices," issued February 3, 2016.
#### Conclusion:
Based on the information submitted in this premarket notification, including the intended use, technological characteristics, and operational use, the Hologic Fluent Pro Fluid Management System has been shown to be substantially equivalent to the predicate device, Fluent Fluid Management System (K180825).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.