K240876 · Zimmer, Inc. · MBF · Jun 12, 2024 · Orthopedic
Device Facts
Record ID
K240876
Device Name
Identity Shoulder System
Applicant
Zimmer, Inc.
Product Code
MBF · Orthopedic
Decision Date
Jun 12, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3670
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Hemiarthoplasty/Conventional Total Application: · Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. - Rheumatoid arthritis. · Correction of functional deformity. · Fractures of the proximal humerus, where other methods of treatment are deemed inadequate. · Difficult clinical management problems, including cuff arthropathy, where other methods of treatment may not be suitable or may be inadequate. Optional use in revision: in some medical conditions (e.g. revision when healthy and good bone stock exists), the surgeon may opt to use primary implants in a revision procedure. Reverse Application: Zimmer Biomet Reverse Shoulder products are indicated for use in patients whose shoulder ioint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically and structurally suited to receive the implants and a functional deltoid muscle is necessary. The Zimmer Biomet Reverse Shoulder is indicated for primary, fracture, or revision total shoulder replacement for the relief of pain and significant disability due to gross rotator cuff deficiency. The assembled humeral component may be used alone for hemiarthroplasty or combined with the glenoid component or reverse components for total shoulder arthroplasty (conventional or reverse applications). The humeral components may be used cemented (biological fixation). The Titanium Humeral Head and Glenosphere components are indicated for patients with suspected cobalt alloy sensitivity. The wear properties of Titanium alloys are inferior to that of cobalt alloy. A Titanium humeral head or Glenospheres is not recommended for patients who lack suspected material sensitivity to cobalt alloy.
Device Story
Identity Shoulder System is a modular orthopedic implant for shoulder arthroplasty; provides anatomical adjustment and joint restoration. System includes humeral stems, adapters, heads, trays, and bearings; used in anatomic (hemiarthroplasty or total) or reverse configurations. Reverse configuration requires compatible glenosphere/baseplate. Current submission adds extra extended humeral trays (3 heights) and uniform thickness bearings to improve range of motion and reduce scapular notching. Used by orthopedic surgeons in clinical settings. Implants provide pain relief and restore function in patients with severe joint disease or rotator cuff deficiency. Titanium options available for patients with cobalt alloy sensitivity.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by non-clinical bench testing, including Finite Element Analysis (FEA) and range of motion analysis.
Technological Characteristics
Modular shoulder prosthesis system. Components include humeral stems, trays, bearings, and glenospheres. Materials are identical to predicate (raw and contact). Connectivity: N/A. Sterilization: Identical to predicate. Software: None.
Indications for Use
Indicated for patients with non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, proximal humerus fractures, or cuff arthropathy. Reverse application indicated for patients with grossly deficient rotator cuff, severe arthropathy, or failed prior replacement requiring functional deltoid muscle. Titanium components indicated for patients with suspected cobalt alloy sensitivity.
Regulatory Classification
Identification
A shoulder joint metal/polymer/metal nonconstrained or semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits movement in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral component made of alloys such as cobalt-chromium-molybdenum (Co-Cr-Mo) and titanium-aluminum-vanadium (Ti-6Al-4V) alloys, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, or a combination of an articulating ultra-high molecular weight bearing surface fixed in a metal shell made of alloys such as Co-Cr-Mo and Ti-6Al-4V. The humeral component and glenoid backing have a porous coating made of, in the case of Co-Cr-Mo components, beads of the same alloy or commercially pure titanium powder, and in the case of Ti-6Al-4V components, beads or fibers of commercially pure titanium or Ti-6Al-4V alloy, or commercially pure titanium powder. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance: Shoulder Joint Metal/Polymer/Metal Nonconstrained or Semi-Constrained Porous-Coated Uncemented Prosthesis.”
{0}------------------------------------------------
June 12, 2024
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. The FDA logo is on the right, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The logos are placed side by side.
Zimmer Inc. Patricia Beres Regulatory Affairs Principal 1800 West Center St. Warsaw, Indiana 46580
Re: K240876
Trade/Device Name: Identity Shoulder System Regulation Number: 21 CFR 888.3670 Regulation Name: Shoulder joint metal/polymer/metal nonconstrained or semi-constrained porouscoated uncemented prosthesis Regulatory Class: Class II Product Code: MBF, KWT, KWS, HSD, PHX, PAO Dated: May 2, 2024 Received: May 2, 2024
Dear Patricia Beres:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Joseph P. Joseph P. Russell -S Russell-s Date: 2024.06.12 15:56:26 -04'00"
for: Farzana Sharmin, PhD Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K240876
Device Name
Identity Shoulder System
#### Indications for Use (Describe)
Hemiarthoplasty/Conventional Total Application:
- · Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- · Rheumatoid arthritis.
- · Correction of functional deformity.
- · Fractures of the proximal humerus, where other methods of treatment are deemed inadequate.
- · Difficult clinical management problems, including cuff arthropathy, where other methods of treatment may not be suitable or may be inadequate.
Optional use in revision: in some medical conditions (e.g. revision when healthy and good bone stock exists), the surgeon may opt to use primary implants in a revision procedure.
Reverse Application:
Zimmer Biomet Reverse Shoulder products are indicated for use in patients whose shoulder ioint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically and structurally suited to receive the implants and a functional deltoid muscle is necessary.
The Zimmer Biomet Reverse Shoulder is indicated for primary, fracture, or revision total shoulder replacement for the relief of pain and significant disability due to gross rotator cuff deficiency. The assembled humeral component may be used alone for hemiarthroplasty or combined with the glenoid component or reverse components for total shoulder arthroplasty (conventional or reverse applications). The humeral components may be used cemented (biological fixation).
The Titanium Humeral Head and Glenosphere components are indicated for patients with suspected cobalt alloy sensitivity. The wear properties of Titanium alloys are inferior to that of cobalt alloy. A Titanium humeral head or Glenospheres is not recommended for patients who lack suspected material sensitivity to cobalt alloy.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
K240876 Page 1 of 3
| 510(k) #: K240876 | 510(k) Summary | | K240876 Page 1 of 3<br>Prepared on: 2024-05-02 | 21 CFR 807.92(a)(1) |
|-----------------------------------------------------------|----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|------------------------------------------------|---------------------|
| Contact Details | | | | |
| Applicant Name | | Zimmer Inc. | | |
| Applicant Address | | 1800 West Center St. Warsaw IN 46580 United States | | |
| Applicant Contact Telephone | | 754-376-5271 | | |
| Applicant Contact | | Ms. Patricia Beres | | |
| Applicant Contact Email | | patty.beres@zimmerbiomet.com | | |
| Device Name | | | | 21 CFR 807.92(a)(2) |
| Device Trade Name | | Identity Shoulder System | | |
| Common Name | | Shoulder joint metal/polymer/metal nonconstrained or semi-constrained<br>porous-coated uncemented prosthesis | | |
| Classification Name | | Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer, Uncemented | | |
| Regulation Number | | 888.3670 | | |
| Product Code(s) | | MBF, KWT, KWS, HSD, PHX, PAO | | |
| Legally Marketed Predicate Devices<br>21 CFR 807.92(a)(3) | | | | |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | | | Product Code |
| K213856 | Identity Shoulder System | | | MBF |
| K201315 | PERFORM Humeral System - Stem | | | PAO |
| K220914 | Tornier Perform Humeral System - Fracture | | | PHX |
| K181611 | Comprehensive Reverse Shoulder System | | | PHX |
### Device Description Summary
21 CFR 807.92(a)(4)
The Identity Shoulder System cleared previously in K213856 is a comprehensive collection of components designed with the intention of providing the modularity and adaptability necessary to facilitate individual anatomical adjustment and restoration of the glenohumeral joint during shoulder arthroplasty.
The Identity Shoulder System in the anatomic configuration is comprised of several individual components such as Humeral Stem, Fixed Angle Humeral Stem Adapter, Humeral Head. This configuration can be used as a hemi-athroplasty with the humeral head articulating against the natural glenoid bone or as an anatomic total shoulder replacement with a compatible glenoid component.
The components of the Identity Shoulder System may also be used in the reverse configuration. Individual components include a Humeral Stem, Humeral Tray and Humeral Bearing, This construct is intended to be used with a compatible glenosphere/baseplate component.
The current 510(k) is for additional sizes/styles of humeral trays and bearings for the reverse configuration. Extra Extended trays in 3 heights in a neutral configuration are being additional lateralization. A uniforn thickness
{5}------------------------------------------------
#### K240876 Page 2 of 3
bearing is also being added in an effort to improve internal/external range of motion and reduce potential of scapular notching.
# Intended Use/Indications for Use
# 21 CFR 807.92(a)(5)
Hemiarthoplasty/Conventional Total Application:
· Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- Rheumatoid arthritis.
- · Correction of functional deformity.
· Fractures of the proximal humerus, where other methods of treatment are deemed inadequate.
• Difficult clinical management problems, including cuff arthropathy, where other may not be suitable or may be inadequate.
Optional use in revision: in some medical conditions (e.g. revision when healthy and good bone stock exists), the surgeon may opt to use primary implants in a revision procedure.
Reverse Application:
Zimmer Biomet Reverse Shoulder products are in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically and structurally suited to receive the implants and a functional deltoid muscle is necessary.
The Zimmer Biomet Reverse Shoulder is indicated for primary, fracture, or revision total shoulder replacement for the relief of pain and significant disability due to gross rotator cuff deficiency.
The assembled humeral component may be used alone for hemiarthroplasty or combined with the glenoid component or reverse components for total shoulder arthroplasty (conventional or reverse applications). The humeral components may be used cemented or uncemented (biological fixation).
The Titanium Humeral Head and Glenosphere components are indicated for patients with suspected cobalt alloy sensitivity. The wear properties of Titanium and Titanium alloys are inferior to that of cobalt alloy. A Titanium humeral head or Glenospheres is not recommended for patients who lack suspected material sensitivity to cobalt alloy.
# Indications for Use Comparison
The indications for use of the subject devices are identical to the primary predicate Identity Shoulder System.
# Technological Comparison
The proposed device has similar technological characteristics as the primary predicate system.
The following items are identical to the predicate Identity Shoulder System, cleared in K213856.
• Component Materials, raw and contact
• MRI Conditional Status and MRI language
- Sterilization methods and validations
- · Packaging materials, methods and validations
- Shelf Life validations
· Biocompatibility
The rationale for substantial equivalence is based on consideration of the following characteristics:
- · Materials: Identical to primary predicate device
- · Design Features: Similar to predicate devices
- Packaging: Identical to predicate devices
- · Sterilization: Identical to primary predicate device
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Non-Clinical Tests/Justifications:
• Finite Element Analysis (FEA)
· Range of Motion Analysis
### 21 CFR 807.92(a)(5)
## 21 CFR 807.92(a)(6)
{6}------------------------------------------------
· Magnetic Resonance Imaging (MRI)
Clinical Tests None provided
Conclusion
Non-clinical testing demonstrated that the new devices do not represent new worst-case configurations than the predicate, and the subject device is substantially equivalent to the predicate device identified.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.