cobas® SARS-CoV-2 Qualitative for use on the cobas® 5800/6800/8800 Systems

K240867 · Roche Molecular Systems, Inc. · QQX · Feb 11, 2025 · Microbiology

Device Facts

Record IDK240867
Device Namecobas® SARS-CoV-2 Qualitative for use on the cobas® 5800/6800/8800 Systems
ApplicantRoche Molecular Systems, Inc.
Product CodeQQX · Microbiology
Decision DateFeb 11, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3981
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K240867 · Feb 11, 2025cobas® SARS-CoV-2 Qualitative for use on the cobas® 5800/6800/8800 SystemsRoche Molecular Systems, Inc.National Football League (NFL) COVID-19 Surveillance ProgramThe NFL surveillance program data was used to assess the clinical performance (PPA/NPA) of the cobas SARS-CoV-2 Qualitative test in asymptomatic individuals using anterior nasal swab specimens.Surveillance Program; Asymptomatic; Occupational Testing

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
National Football League (NFL) COVID-19 Surveillance Program; Retrospective analysis of surveillance data; Study Period: August 2020 - January 2021Asymptomatic NFL players and staff; Sample Size: 1,776 samplesMolecular comparator test results and/or clinical adjudication performed within the NFL testing programPositive Percent Agreement (PPA) and Negative Percent Agreement (NPA)

Indications for Use

cobas® SARS-CoV-2 Qualitative for use on the cobas®5800/6800/8800 Systems is a real-time RT-PCR test intended for the qualitative detection of nucleic acids from SARS-CoV-2 in nasopharyngeal swab specimens collected from individuals with signs and symptoms of COVID-19 and in anterior nasal swab specimens collected from any individuals with or without signs and symptoms of COVID-19. Positive results are indicative of the presence of SARS-CoV-2 RNA. Positive results do not rule out bacterial infection or co-infection with other pathogens. Negative results do not preclude SARS-CoV-2 infection and should not be used as the sole basis for patient management decisions. Results are meant to be used in conjunction with clinical observations, patient history, recent exposures, epidemiological information, and laboratory data, in accordance with the guidelines provided by the relevant public health authorities.

Device Story

Fully automated real-time RT-PCR test system; processes nasopharyngeal or anterior nasal swab specimens. System performs nucleic acid extraction/purification using magnetic glass particles; target-specific amplification of ORF1 a/b and E-gene regions; detection via TaqMan FRET probes. Operated by qualified clinical laboratory personnel in clinical settings. Output: qualitative SARS-CoV-2 detection result. Results reviewed on system screen or printed report; used by clinicians alongside patient history and epidemiological data to inform management decisions. Benefits: rapid, automated diagnostic capability for symptomatic and asymptomatic populations.

Clinical Evidence

Clinical performance in asymptomatic individuals evaluated via two studies: 1) NFL surveillance program (n=1776), showing 100% PPA (95% CI: 74.1-100%) and 99.8% NPA (95% CI: 99.5-99.9%); 2) TUAH prospective longitudinal study (n=38,192), showing 94.3% PPA (95% CI: 91.4-96.8%) and 99.2% NPA (95% CI: 99.2-99.4%). Comparator algorithms utilized molecular test results and clinical adjudication.

Technological Characteristics

Real-time RT-PCR; automated nucleic acid extraction via magnetic glass particles; TaqMan FRET detection chemistry. Targets: ORF1 a/b and E-gene. Includes RNA internal control and AmpErase (uracil-N-glycosylase) for amplicon contamination control. Platform: cobas 5800/6800/8800 systems. Connectivity: integrated system software for data management.

Indications for Use

Indicated for qualitative detection of SARS-CoV-2 nucleic acids in nasopharyngeal swabs from symptomatic individuals and anterior nasal swabs from individuals with or without COVID-19 symptoms.

Regulatory Classification

Identification

A device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test is an in vitro diagnostic device intended for the detection and identification of SARS-CoV-2 and other microbial agents when in a multi-target test in human clinical respiratory specimens from patients suspected of respiratory infection who are at risk for exposure or who may have been exposed to these agents. The device is intended to aid in the diagnosis of respiratory infection in conjunction with other clinical, epidemiologic, and laboratory data or other risk factors.

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) The intended use in the labeling required under § 809.10 of this chapter must include a description of the following: Analytes and targets the device detects and identifies, the specimen types tested, the results provided to the user, the clinical indications for which the test is to be used, the specific intended population(s), the intended use locations including testing location(s) where the device is to be used (if applicable), and other conditions of use as appropriate. (2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device. (3) The labeling required under § 809.10(b) of this chapter must include: (i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens; (ii) Detailed descriptions of the performance characteristics of the device for each specimen type claimed in the intended use based on analytical studies including the following, as applicable: Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, precision, reproducibility, and clinical studies; (iii) Detailed descriptions of the test procedure(s), the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing; (iv) A warning statement that viral culture should not be attempted in cases of positive results for SARS-CoV-2 and/or any similar microbial agents unless a facility with an appropriate level of laboratory biosafety ( *e.g.,* BSL 3 and BSL 3+, etc.) is available to receive and culture specimens; and(v) A prominent statement that device performance has not been established for specimens collected from individuals not identified in the intended use population ( *e.g.,* when applicable, that device performance has not been established in individuals without signs or symptoms of respiratory infection).(vi) Limiting statements that indicate that: (A) A negative test result does not preclude the possibility of infection; (B) The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician; (C) There is a risk of incorrect results due to the presence of nucleic acid sequence variants in the targeted pathogens; (D) That positive and negative predictive values are highly dependent on prevalence; (E) Accurate results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and (F) When applicable ( *e.g.,* recommended by the Centers for Disease Control and Prevention, by current well-accepted clinical guidelines, or by published peer-reviewed literature), that the clinical performance may be affected by testing a specific clinical subpopulation or for a specific claimed specimen type.(4) Design verification and validation must include: (i) Detailed documentation, including performance results, from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, as appropriate, additional characterized clinical samples. The clinical study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained using a comparator that FDA has determined is appropriate. Detailed documentation must include the clinical study protocol (including a predefined statistical analysis plan), study report, testing results, and results of all statistical analyses. (ii) Risk analysis and documentation demonstrating how risk control measures are implemented to address device system hazards, such as Failure Modes Effects Analysis and/or Hazard Analysis. This documentation must include a detailed description of a protocol (including all procedures and methods) for the continuous monitoring, identification, and handling of genetic mutations and/or novel respiratory pathogen isolates or strains ( *e.g.,* regular review of published literature and periodic in silico analysis of target sequences to detect possible mismatches). All results of this protocol, including any findings, must be documented and must include any additional data analysis that is requested by FDA in response to any performance concerns identified under this section or identified by FDA during routine evaluation. Additionally, if requested by FDA, these evaluations must be submitted to FDA for FDA review within 48 hours of the request. Results that are reasonably interpreted to support the conclusion that novel respiratory pathogen strains or isolates impact the stated expected performance of the device must be sent to FDA immediately.(iii) A detailed description of the identity, phylogenetic relationship, and other recognized characterization of the respiratory pathogen(s) that the device is designed to detect. In addition, detailed documentation describing how to interpret the device results and other measures that might be needed for a laboratory diagnosis of respiratory infection. (iv) A detailed device description, including device components, ancillary reagents required but not provided, and a detailed explanation of the methodology, including molecular target(s) for each analyte, design of target detection reagents, rationale for target selection, limiting factors of the device ( *e.g.,* saturation level of hybridization and maximum amplification and detection cycle number, etc.), internal and external controls, and computational path from collected raw data to reported result (*e.g.,* how collected raw signals are converted into a reported signal and result), as applicable.(v) A detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review. (vi) For devices intended for the detection and identification of microbial agents for which an FDA recommended reference panel is available, design verification and validation must include the performance results of an analytical study testing the FDA recommended reference panel of characterized samples. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses. (vii) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens, the design verification and validation must include a detailed description of the identity, phylogenetic relationship, or other recognized characterization of the Influenza A and B viruses that the device is designed to detect, a description of how the device results might be used in a diagnostic algorithm and other measures that might be needed for a laboratory identification of Influenza A or B virus and of specific Influenza A virus subtypes, and a description of the clinical and epidemiological parameters that are relevant to a patient case diagnosis of Influenza A or B and of specific Influenza A virus subtypes. An evaluation of the device compared to a currently appropriate and FDA accepted comparator method. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses. (5) When applicable, performance results of the analytical study testing the FDA recommended reference panel described in paragraph (b)(4)(vi) of this section must be included in the device's labeling under § 809.10(b) of this chapter. (6) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens in addition to detection of SARS-CoV-2 and similar microbial agents, the required labeling under § 809.10(b) of this chapter must include the following: (i) Where applicable, a limiting statement that performance characteristics for Influenza A were established when Influenza A/H3 and A/H1-2009 (or other pertinent Influenza A subtypes) were the predominant Influenza A viruses in circulation. (ii) Where applicable, a warning statement that reads if infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to State or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens. (iii) Where the device results interpretation involves combining the outputs of several targets to get the final results, such as a device that both detects Influenza A and differentiates all known Influenza A subtypes that are currently circulating, the device's labeling must include a clear interpretation instruction for all valid and invalid output combinations, and recommendations for any required followup actions or retesting in the case of an unusual or unexpected device result. (iv) A limiting statement that if a specimen yields a positive result for Influenza A, but produces negative test results for all specific influenza A subtypes intended to be differentiated ( *i.e.,* H1-2009 and H3), this result requires notification of appropriate local, State, or Federal public health authorities to determine necessary measures for verification and to further determine whether the specimen represents a novel strain of Influenza A.(7) If one of the actions listed at section 564(b)(1)(A) through (D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain: (i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those influenza viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate. (ii) Within 60 days from the date that FDA notifies manufacturers that characterized influenza viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by: (A) Placing the results directly in the device's labeling required under § 809.10(b) of this chapter that accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or (B) In a section of the device's label or in other labeling that accompanies the device, prominently providing a hyperlink to the manufacturer's public website where the analytical reactivity testing data can be found. The manufacturer's website, as well as the primary part of the manufacturer's website that discusses the device, must provide a prominently placed hyperlink to the website containing this information and must allow unrestricted viewing access.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT ## I Background Information: A 510(k) Number K240867 B Applicant Roche Molecular Systems, Inc. C Proprietary and Established Names cobas SARS-CoV-2 Qualitative for use on the cobas 5800/6800/8800 Systems D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | QQX | Class II | 21 CFR 866.3981 - Device To Detect And Identify Nucleic Acid Targets In Respiratory Specimens From Microbial Agents That Cause The SARS-CoV-2 Respiratory Infection And Other Microbial Agents When In A Multi-Target Test | MI - Microbiology | ## II Submission/Device Overview: ### A Purpose for Submission: The purpose of this submission is to expand the intended use of cobas SARS-CoV-2 Qualitative for use on the cobas 5800/6800/8800 Systems, which was originally FDA-cleared under K231306 to include asymptomatic population. ### B Measurand: SARS-CoV-2 RNA ### C Type of Test: Real-time nucleic acid amplification test Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 {1} III Intended Use/Indications for Use: A Intended Use(s): See Indications for Use below. B Indication(s) for Use: cobas SARS-CoV-2 Qualitative for use on the cobas 5800/6800/8800 Systems is a real-time RT-PCR test intended for the qualitative detection of nucleic acids from SARS-CoV-2 in nasopharyngeal swab specimens collected from individuals with signs and symptoms of COVID-19 and in anterior nasal swab specimens collected from any individuals with or without signs and symptoms of COVID-19. Positive results are indicative of the presence of SARS-CoV-2 RNA. Positive results do not rule out bacterial infection or co-infection with other pathogens. Negative results do not preclude SARS-CoV-2 infection and should not be used as the sole basis for patient management decisions. Results are meant to be used in conjunction with clinical observations, patient history, recent exposures, epidemiological information, and laboratory data, in accordance with the guidelines provided by the relevant public health authorities. C Special Conditions for Use Statement(s): Rx - For Prescription Use Only IVD - For in vitro diagnostic use D Special Instrument Requirements: For use with the cobas 5800/6800/8800 Systems. IV Device/System Characteristics: A Device Description: The cobas SARS-CoV-2 Qualitative for use on the cobas 5800/6800/8800 Systems is a fully automated test system including sample preparation (nucleic acid extraction and purification) followed by PCR amplification and detection. The cobas 5800/6800/8800 Systems consists of three separate instruments, the cobas 5800 System, the cobas 6800 System and the cobas 8800 System, which allow users to perform multiple PCR-based in vitro nucleic acid amplification tests. The cobas 5800/6800/8800 Systems consist of the sample supply module, the transfer module, the processing module, and the analytic module. Automated data management is performed by the cobas 5800/6800/8800 Systems software(s), which assigns test results for all tests. Results can be reviewed directly on the system screen and printed as a report. Nucleic acid from patient samples and added internal control RNA (RNA IC) molecules are simultaneously extracted. Nucleic acid is released by addition of proteinase and lysis reagent to the sample. The released nucleic acid binds to the silica surface of the added magnetic glass particles. Unbound substances and impurities, such as denatured protein, cellular debris, and potential PCR inhibitors, are removed with subsequent wash steps and purified nucleic acid is K240867 - Page 2 of 7 {2} eluted from the magnetic glass particles with elution buffer at elevated temperature. External controls (positive and negative) are processed in the same way. Selective amplification of target nucleic acid from the sample is achieved using target specific forward and reverse primers for ORF1 a/b non-structural region that is unique to SARS-CoV-2. Additionally, a conserved region in the structural protein envelope E-gene were chosen for pan-Sarbecovirus detection. The pan-Sarbecovirus detection sets will also detect SARS-CoV-2 virus. Selective amplification of RNA IC is achieved using non-competitive sequence specific forward and reverse primers which have no homology with the coronavirus genome. A thermostable DNA polymerase enzyme is used for amplification. For more details, refer to K213804 and K231306. ## B Principle of Operation: Same as K213804 and K231306 ## V Substantial Equivalence Information: ## A Predicate Device Name(s): cobas SARS-CoV-2 Qualitative for use on the cobas 5800/6800/8800 Systems ## B Predicate 510(k) Number(s): K231306 ## C Comparison with Predicate(s): | | Submitted Device: cobas SARS-CoV-2 Qualitative (K240867) | Predicate Device: cobas SARS-CoV-2 Qualitative (K231306) | | --- | --- | --- | | Regulation Number | 21 CFR 866.3981 | Same | | Regulation Name | Device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test | Same | | Product Code | QQX | Same | | Intended Use | cobas SARS-CoV-2 Qualitative for use on the cobas 5800/6800/8800 Systems is a real-time RT-PCR test intended for the qualitative detection of nucleic acids from SARS-CoV-2 in nasopharyngeal swab specimens collected from individuals with signs and symptoms of COVID-19 and in anterior nasal swab specimens collected from any individuals with or without signs and symptoms of COVID-19. Positive results are indicative of the presence of SARS-CoV-2 RNA. Positive results do not rule out bacterial infection or co-infection with other pathogens. | cobas SARS-CoV-2 Qualitative for use on the cobas 5800/6800/8800 Systems is a real-time RT-PCR test intended for the qualitative detection of nucleic acids from SARS-CoV-2 in nasal and nasopharyngeal specimens collected from symptomatic individuals suspected of COVID-19 by their healthcare provider. Results are for the detection of SARS-CoV-2 RNA. Positive results are indicative of the presence of SARS-CoV-2 RNA; clinical correlation with patient history and other diagnostic information is necessary to determine patient infection status. Positive | K240867 - Page 3 of 7 {3} K240867 - Page 4 of 7 | | Negative results do not preclude SARS-CoV-2 infection and should not be used as the sole basis for patient management decisions. Results are meant to be used in conjunction with clinical observations, patient history, recent exposures, epidemiological information, and laboratory data, in accordance with the guidelines provided by the relevant public health authorities. | results do not rule out bacterial infection or co-infection with other pathogens. Negative results do not preclude SARS-CoV-2 infection and should not be used as the sole basis for patient management decisions. Results are meant to be used in conjunction with clinical observations, patient history, recent exposures and epidemiological information, and laboratory data, in accordance with the guidelines provided by the relevant public health authorities. cobas SARS-CoV-2 Qualitative is intended for use by qualified clinical laboratory personnel specifically instructed and trained in the techniques of real-time PCR and on the use of the cobas 5800/6800/8800 Systems. | | --- | --- | --- | | Conditions for use | For prescription use | Same | | Sample Types | Nasopharyngeal swab specimen Anterior Nasal swab specimen | Same | | Analyte Targets | SARS-CoV-2 | Same | | Sample Preparation Procedure | Automated by cobas 5800/6800/8800 Systems | Same | | Amplification Technology | Real-time PCR | Same | | Detection Chemistry | Paired reporter and quencher fluorescence labeled probes (TaqMan Technology) using fluorescence resonance energy transfer (FRET) | Same | | Controls used | Sample processing control (IC) Positive and negative control | Same | | Result Analysis | Based on PCR cycle threshold analysis | Same | VI Standards/Guidance Documents Referenced: Class II Special Controls as per 21 CFR 866.3981. VII Performance Characteristics (if/when applicable): A Analytical Performance: 1. Precision/Reproducibility: Precision/Reproducibility were previously reviewed and described in K213804. 2. Linearity: Not applicable 3. Analytical Specificity/Interference: {4} An updated in-silico analysis was performed in January 2025 using all SARS-CoV-2 sequences submitted to the GISAID database till date (as of January 15, 2025) and are reported in Table 1 below. The in-silico analysis results indicate that &gt;99.9% of sequences for SARS-CoV-2 have no changes in primer/probe binding sites at both target regions simultaneously. All sequences are predicted to be detected by at least one of the two target regions. Table 1: Updated in silico analysis of SARS-CoV-2 Qualitative Oligo Design | Target | Orf1ab | | E-gene | | Orf1ab & E-gene | | | --- | --- | --- | --- | --- | --- | --- | | Database | GISAID | | GISAID | | GISAID | | | total | 16156883 | 100.00% | 16156883 | 100.00% | 16156883 | 100.00% | | with_mismatch | 549763 | 3.40% | 87773 | 0.54% | 3560 | 0.02% | | dCp>5 or Tm<65 | 545 | 0.00% | 1175 | 0.01% | 3* | 0.00% | * The three sequences have several frameshifts, significantly long truncations and nucleotide gaps, and thus are considered submissions of lower sequencing quality. 4. Assay Reportable Range: Not Applicable 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): Same as K213804 6. Detection Limit: Same as K213804 7. Assay Cut-Off: Same as K213804 B Comparison Studies: 1. Method Comparison with Predicate Device: Not Applicable 2. Matrix Comparison: Same as K213804 C Clinical Studies: Symptomatic Population The clinical performance of cobas SARS-CoV-2 Qualitative in symptomatic population using anterior nasal swab and nasopharyngeal swab specimens was previously reviewed and described in K213804. Asymptomatic Population The clinical performance of the cobas SARS-CoV-2 Qualitative with asymptomatic subjects was assessed using real-world data (RWD) from the National Football League (NFL) COVID-19 Surveillance Program and a prospective clinical study called Test Us at Home (TUAH). K240867 - Page 5 of 7 {5} # Real-World evidence (NFL surveillance program) The clinical performance of the cobas SARS-CoV-2 Qualitative with asymptomatic subjects was assessed using real-world data collected from the 2020 National Football League (NFL) COVID-19 Surveillance Program where samples were collected and tested between August 2020-January 2021 as part of an Occupational Testing protocol. Anterior nasal swab samples were prospectively collected on a near-daily basis from NFL players and staff. The performance of cobas SARS-CoV-2 Qualitative were estimated using comparator algorithm that was based on molecular comparator test results and/or clinical adjudication performed within the NFL testing program. A total of 1776 samples were selected for analysis where the cobas SARS-CoV-2 Qualitative candidate test, and comparator test results, were evaluable to establish the COVID-19 status for each sample. The results are shown below. Table 2 Performance estimates for the cobas SARS-CoV-2 Qualitative in anterior nasal swabs in asymptomatic individuals (NFL). | Cobas SARS-CoV-2 Qualitative | Comparator algorithm | | Total | | --- | --- | --- | --- | | | Positive | Negative | | | Positive | 11 | 3 | 14 | | Negative | 0 | 1,762 | 1,762 | | Total | 11 | 1,765 | 1,776 | | PPA (n/N) (95% Confidence Interval) | 100.0 % (11/11) (95% CI: 74.1% - 100%) | | | | NPA (n/N) (95% Confidence Interval) | 99.8% (1762/1765) (95% CI: 99.5% - 99.9%) | | | Note: CI = confidence interval, PPA = positive percent agreement, NPA = negative percent agreement # Clinical Study The clinical performance of the cobas SARS-CoV-2 Qualitative with asymptomatic subjects was also assessed using data collected from the 2021 Test Us at Home (TUAH) study where samples were collected and tested for SARS-CoV-2 between October 2021 and April 2022 as part of a longitudinal study. Anterior nasal swab samples were prospectively collected every 48 hours from each participant over 15 days. The performance of cobas SARS-CoV-2 Qualitative was estimated using a comparator algorithm where two consecutive test results (molecular test) over 48 hours were used to determine comparator result. All samples (38,185) from the TUAH study that had a valid comparator algorithm and a valid candidate test result were included in the calculation of performance estimates of the cobas SARS-CoV-2 Qualitative. The results are shown in table below. Table 3. Performance estimates for the cobas SARS-CoV-2 Qualitative in anterior nasal swabs in asymptomatic individuals in (TUAH study). | Cobas SARS-CoV-2 Qualitative | Comparator algorithm | | Total | | --- | --- | --- | --- | | | Positive | Negative | | | Positive | 315 | 272 | 587 | | Negative | 19 | 37,586 | 37,605 | | Total | 334 | 37,858 | 38,192 | | PPA (n/N) (95% Confidence Interval) | 94.3% (315/334) (95% CI: 91.4% - 96.8%) * | | | K240867 - Page 6 of 7 {6} NPA (n/N) (95% Confidence Interval) 99.2% (37,586/37,858) (95% CI: 99.2% - 99.4%) * * Confidence intervals were estimated using a bootstrapping method. Note: CI = confidence interval, PPA = positive percent agreement, NPA = negative percent agreement D Clinical Cut-Off: Not applicable E Expected Values/Reference Range: The clinical performance of cobas SARS-CoV-2 Qualitative in asymptomatic population was evaluated in a prospective clinical study between October 2021 and April 2022. The positivity rate for the detection of SARS-CoV-2 by the cobas SARS-CoV-2 Qualitative was 1.8% for asymptomatic anterior nasal swab specimens. VIII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. K240867 - Page 7 of 7
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...