K240847 · Ide Vision, Ltd. · FCX · Oct 30, 2024 · Gastroenterology, Urology
Device Facts
Record ID
K240847
Device Name
MedicCO2LON (MedicCO2LON)
Applicant
Ide Vision, Ltd.
Product Code
FCX · Gastroenterology, Urology
Decision Date
Oct 30, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
The MedicCO2LON device is intended to be used for colonic distension with CO2 gas for CT Colonography and conventional Colonoscopy procedures. MedicCO2LON is designed to function with specific colonic insufflation administration sets that allow passage of gas from the device to the colorectal cavity. MedicCO2LON should only be used with administration sets specifically designed for this purpose. The device should only be used by personnel trained in the practice of undertaking CT Colonography and conventional Colonoscopy examinations and the use of an automatic insufflation device. The device should only be used for CT Colonography or conventional Colonoscopy procedures and should not be used for any other patient examination procedure.
Device Story
MedicCO2LON is an automated, table/trolley-mounted insufflation device for colonic distension using CO2 gas. It connects to an external CO2 supply and delivers gas to the colorectal cavity via a standard administration set. The device regulates gas flow and pressure (up to 30mmHg) based on user-defined settings via a graphic touch screen. It displays real-time output pressure, gas flow rate, delivered volume, and supply level. The device is operated by trained clinicians during CT Colonography or conventional colonoscopy. It incorporates redundant mechanical and electronic safety pressure relief mechanisms to prevent over-pressurization. By providing controlled, automated insufflation, it assists in visualizing the colon, potentially improving diagnostic accuracy and patient comfort compared to manual methods.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing verified pressure accuracy (+/- 1mmHg), flow accuracy (+/- 10% at 3LPM), volume accuracy (+/- 20%), and the functionality of redundant mechanical and electronic overpressure relief systems.
Technological Characteristics
Electro-mechanical pneumatic insufflator. Construction: PC/ABS external shell, metal internal frame. Power: Internal mains-connected power supply (3.3V, 12V, 24V DC). Interface: LED backlit color LCD touch screen. Connectivity: IEC mains inlet, low-pressure CO2 input, controlled gas output. Safety: Independent redundant mechanical pressure relief valve (set 75-78mmHg) and electronic pressure relief (50mmHg for 5 seconds).
Indications for Use
Indicated for colonic distension with CO2 gas during CT Colonography and conventional Colonoscopy procedures. For use by trained clinical personnel.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
Predicate Devices
BRACCO PROTO 2 L TOUCH Colon Insufflator (K132192)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 30, 2024
Nicholas See IDE Vision, Ltd. Lodge Way Portskewett Caldicot. NP265PS United Kingdom
Re: K240847
Trade/Device Name: MedicCO2LON (MedicCO2LON) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: FCX Dated: October 3, 2024 Received: October 3, 2024
Dear Nicholas See:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Sivakami Venkatachalam -S
for
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K240847
Device Name
MedicCO2LON (MedicCO2LON)
Indications for Use (Describe)
The MedicCO2LON device is intended to be used for colonic distension with CO2 gas for CT Colonography and conventional Colonoscopy procedures.
MedicCO2LON is designed to function with specific colonic insufflation administration sets that allow passage of gas from the device to the colorectal cavity. MedicCO2LON should only be used with administration sets specifically designed for this purpose.
The device should only be used by personnel trained in the practice of undertaking CT Colonography and conventional Colonoscopy examinations and the use of an automatic insufflation device.
The device should only be used for CT Colonography or conventional Colonoscopy procedures and should not be used for any other patient examination procedure.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a logo for iDEVision. The logo is white text on a black background. The text is in a sans-serif font and is stylized.
# 510K Summary
Date prepared 9th May 2024, Updated 23rd October 2023
## General Information:
| Contact Person: | Nicholas Adrian See |
|-----------------|-------------------------------------------------------------------------------------------------------|
| Address: | IDE Vision Ltd<br>Lodge Way<br>Portskewett<br>Caldicot<br>NP26 5PS<br>Monmouthshire<br>United Kingdom |
| Telephone: | +(44) 1291 425425 |
| Email: | nick@doppler.co.uk |
## Device Identification:
| Trade Name: | MedicCO2LON (MedicCO2LON) |
|----------------------|---------------------------|
| Common Name: | Insufflator, Colonic |
| Classification Name: | Endoscope and accessories |
| | 21 CFR 876.1500 |
| Product Code: | FCX |
## Predicate Device Information:
| Predicate Device: | BRACCO PROTO <span style="vertical-align: sub;">2</span> L TOUCH Colon Insufflator |
|-------------------|------------------------------------------------------------------------------------|
| | 510(k) K132192 |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for iDEVision. The logo is white text on a black background. The text is sans-serif and the "i" in iDEVision is lowercase.
- IDE Vision Ltd Lodge Way, Portskewett, Caldicot NP26 5PS United Kingdom
## Device Description:
MedicCO2LON is an automated insufflation device designed for administering and requlating colonic distension by insufflation with carbon dioxide qas, in preparation for and during CT Colonography (CTC or Virtual Colonoscopy) and conventional Colonoscopy.
#### Construction:
The MedicCO2LON colonic insufflator is constructed using a plastic (PC/ABS) external shell covering a metal internal frame, internal components are low voltage (3.3V, 12V and 24V DC) powered from an internal mains connected power supply.
None of the parts of the MedicCO2LON contact the patient directly and the user only has transient contact with the unit, with it being table or trolley mounted.
#### Use period:
The units are for short term use (normally no more than 30 min) and provides. to the patient, CO2 via an external pressure regulator to the device to insufflate the patient using a standard administration set at up to 30mmHg
### Intended Use:
The MedicCO2LON device is intended to be used for colonic distension with CO.gas for CT Colonography and conventional Colonoscopy procedures.
MedicCO2LON is designed to function with specific colonic insufflation administration sets that allow passage of gas from the device to the colorectal cavity. MedicCO2LON should only be used with administration sets specifically designed for this purpose.
The device should only be used by personnel trained in the practice of undertaking CT Colonography and conventional Colonoscopy examinations and the use of an automatic insufflation device.
The device should only be used for CT Colonography or conventional Colonoscopy procedures and should not be used for any other patient examination procedure.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for iDEVision. The logo is white text on a black background. The text is stylized and modern.
# Product Comparison:
| Parameter | MedicCO2LON | PROTOC02L TOUCH |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) | | K132192 |
| Indication for Use | Administer and regulate CO2 as a<br>distention media for the colon<br>during CT Colonography (CTC or<br>Virtual Colonoscopy) and<br>conventional Colonoscopy. | Administer and regulate CO2 as a<br>distention media for the colon<br>during CT Colonography (CTC or<br>Virtual Colonoscopy) and<br>conventional Colonoscopy. |
| Applied Gas | CO2 | CO2 |
| Design | Electro-Mechanical Pneumatic<br>System to regulate both Flow and<br>Pressure of CO2 | Electro-Mechanical Pneumatic<br>System to regulate both Flow and<br>Pressure of CO2 |
| Anatomical Sites | Rectal administration of CO2 via<br>enema tip or catheter lumen | Rectal administration of CO2 via<br>enema tip or catheter lumen |
| Radiation | No ionizing radiation emitted | No ionizing radiation emitted |
| Thermal | No thermal energy introduced into<br>patient. | No thermal energy introduced into<br>patient. |
| Power supply | Switch Mode Power supply housed<br>within unit | Switch Mode Power supply<br>housed within unit |
| User Connections | IEC Mains inlet,<br>Low pressure CO2 input<br>Controlled Gas output | IEC Mains inlet,<br>Low pressure CO2 input<br>Controlled Gas output |
| Display type | LED Back lit colour LCD,<br>Display Area 152 x 91mm | Back lit colour LCD, |
| Displayed data | Current output pressure<br>Gas Flow Rate<br>Delivered gas volume<br>Gas supply level | Current output pressure<br>Gas Flow Rate<br>Delivered gas volume<br>Gas supply level |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for iDEVision. The logo is white text on a black background. The text is stylized, with the "D" in "iDEVision" having a unique design.
| Controls type | Graphic touch screen controls | Graphic touch screen controls |
|--------------------------|---------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Controls data | Start / Stop / Resume Flow | Start / Stop Flow |
| | Reset Delivered Volume | Reset Delivered Volume |
| | Set Target pressure | Set Target Pressure |
| | Set Maximum Flow | |
| Gas Control | | |
| Maximum total gas volume | 12.0 L | No maximum limit |
| Gas Volume Profile | Flow stops at: 4,6,8,10,12L | Flow stops at 3L to 10L |
| Default target flow rate | 3.0 LPM | 3.0LPM |
| Gas Flow Start Profile | 0.5lpm per 0.5l delivered volume to<br>target flow rate | 1.0lpm for 0 to 0.5l delivered<br>volume<br>2.0lpm for 0.5 to 1.0l delivered<br>volume<br>3.0lpm for greater than 1.0l<br>delivered volume |
| Target pressure | 0 to 30mmhg<br>Default 20 mmHg | 0 to 35mmHg |
| Volume Accuracy | +/-20% of current reading display | Unknown |
|-------------------------------------------------|-------------------------------------------------|--------------------------------------------|
| Flow Adjustment | 0.5LPM | 1.0LPM |
| Flow Accuracy | +/-10% at 3LPM | +20% @ 3LPM |
| Pressure Accuracy | +/- 1mmHg of current reading | +/- 10% |
| Safely overpressure Relief | Independent Redundant<br>Mechanical Relief | Independent Redundant<br>Mechanical Relief |
| | Safety Pressure Relief<br>pressure (mechanical) | 1.5psi (78mmHg) nominal |
| Safety Pressure Relief<br>Accuracy (mechanical) | +/- 20% of trip pressure | Unknown |
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image is a logo for iDEVision. The logo is white text on a black background. The text is stylized, with the "i" having a dot above it, the "D" being a blocky shape, and the "Vision" being in a more traditional font. The logo is simple and modern.
| Safety Pressure Relief<br>pressure (Electronic) | 50mmHg for 5 seconds | 50mmHg for 5 seconds |
|-------------------------------------------------|----------------------|----------------------|
|-------------------------------------------------|----------------------|----------------------|
In all cases the insufflation gas is CO2 and this is externally connected to the units which then deliver it via an external Administration Kit. The control interfaces and displays vary slightly with 1 unit using touch screen LCD and the ProtoCO₂L using an LED display and mechanical controls.
The data displayed is the same in style but the MedicCO2Lon adds a display of the delivery gas flow rate to provide the doctor with an indication as to the rate the gas is entering the patient.
Both units deliver a volume of gas of at least 12L with the MedicCO2LON delivering 12L and the ProtoCO2L having no limit. The normal delivered volume for insufflation is between 2 – 4L, allowing for leaks 12L will be enough for all patients. Should more gas be required it implies a serious leak and this should be investigated rather than allowing limitless gas to enter the patient.
Pressure on the MedicCO2LON has a target of 30mmHg with the ProtoCO2L having a 25mmHg maximum. In all cases these values can be set to different levels by the doctor and the MedicCO2LON has pre-sets that can be set to any level the doctors require up to 30mmHg and these values are recalled for each patient.
The flow rate on both units is the same with a maximum of 3LPM, the MedicCO2LON allows an increase in 0.5LPM steps while the ProtoCO2L uses 1LPM steps. The smaller the steps the less stress applied to the patient.
Both units use mechanical safety pressure discharge valves set at 75 – 78mmHg.
## Conclusion:
The MedicCO2LON device for colonic distension with CO.gas for CT Colonography and conventional Colonoscopy procedures is considered to be substantially equivalent to the predicate device. It does not incorporate any significant changes in intended use, method of operation, material, or design that could affect the safety or effectiveness of the device.
#### Clinical:
The units are used for the same clinical condition and purpose.
They are used at the same site in the body;
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image is a logo for iDEVision. The logo is white text on a black background. The text is stylized and modern.
They all use the same Administration Kit to supply the gas into the patient.
The units have similar critical performance according to expected clinical effect its intended use.
#### Technical:
The MedicCO2LON and the other unit are used under similar conditions of use.
The MedicCO2LON has similar specifications and properties.
The MedicCO2LON is of similar design being a standalone mains connected insufflator supplied externally with CO2.
The MedicCO2LON has similar principles of operation, Pressure, Flow, Volume and connection to the Patient.
#### Biological:
The MedicCO2LON uses the same materials in contact with the same human tissues in that the Administration Set (not manufactured my Ultrasound Technologies Ltd) is the same or similar to all others on the market. The Kits are ALL interchangeable and depend on the hospitals purchasing arrangements.
#### Tested Predicate Comparison.
All characteristics have been tested and are detailed in the documents listed:
A summery is as follows:
| Function | MedicCO2LON<br>Performance | Test report document<br>Identification | Predicate device<br>Performance |
|------------------------------|----------------------------|----------------------------------------|---------------------------------|
| Displayed Data | Current Input Pressure | MCODOC121-102<br>Appendix Image | Current output pressure |
| | Gas Flow Rate | MCODOC121-102<br>Appendix Image | Not defined |
| | Delivered Gas Volume | MCODOC121-102<br>Appendix Image | Delivered Gas Volume |
| | Gas Supply level | MCODOC121-102<br>Appendix Image | Gas Supply level |
| Touch Screen Control<br>Data | Start / Stop / Resume flow | MCODOC121-102<br>Appendix Image | Start / Stop flow |
| | Reset Delivered Volume | MCODOC121-102<br>Appendix Image | Reset Delivered Volume |
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for iDEVision. The logo is white text on a black background. The text is stylized, with the "i" having a dot above it and the "D" being a geometric shape.
| | Set Target Pressure | MCODOC121-102<br>Appendix Image | Set Target Pressure |
|-------------------------------------------------|-------------------------------------------------|--------------------------------------------|------------------------------------------|
| | Set Maximum Flow | MCODOC121-102<br>Appendix Image | Not defined |
| | Maximum total gas<br>volume | 12.0L | MCODOC166-101 Page 7<br>section 3.3 |
| Gas Volume Profile | | Stops at 4,6,8,10,12L | MCODOC121-102 Volume<br>Accuracy test |
| | Default target flow rate | 3.0LPM | Power up default on<br>screen |
| Gas Flow Start Profile | | 0.5LPM per 0.5L Volume. | MCODOC121-102 Flow<br>rate Accuracy test |
| | Target Pressure | 0 to 30mmHg | MCODOC121-102<br>Pressure Accuracy test |
| Volume Accuracy | | +/- 20% at current<br>reading. | MCODOC121-102 Volume<br>Accuracy test |
| | Flow Adjustment | 0.5LPM | MCODOC121-102 Flow<br>rate Accuracy test |
| Flow Accuracy | | +/- 10% at 3LPM | MCODOC121-102 Flow<br>rate Accuracy test |
| | Pressure Accuracy | +/- 1mmHg of current<br>reading | MCODOC121-102<br>Pressure Accuracy test |
| Safety overpressure relief | | Independent redundant<br>mechanical relief | MCODOC166-101 Page 6<br>section 3.2 |
| | Safety pressure relief<br>(mechanical) | 1.5psi (78mmHg) | MCODOC166-101 Page 6<br>section 3.2 |
| Safety pressure relief<br>accuracy (mechanical) | | +/- 20mmHg of trip<br>pressure | MCODOC166-101 Page 6<br>section 3.2 |
| | Safety pressure relief<br>accuracy (electronic) | 50mmHg for 5 seconds | MCODOC166-101 Page 6<br>section 3.2 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.