K240832 · Guangdong Transtek Medical Electronics Co., Ltd. · DXN · Jun 7, 2024 · Cardiovascular
Device Facts
Record ID
K240832
Device Name
Blood Pressure Monitor (BBZ32-AA01)
Applicant
Guangdong Transtek Medical Electronics Co., Ltd.
Product Code
DXN · Cardiovascular
Decision Date
Jun 7, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
This Blood Pressure Monitor is a digital monitor intended for use in measuring blood pressure and pulse rate with arm circumference ranging from 22cm to 32cm (about 8¾"-12½") or 22cm to 42cm (about 8¾"-16½"). It is intended for adult indoor use only.
Device Story
Digital blood pressure monitor; uses oscillometric method to measure systolic/diastolic pressure and pulse rate. Input: cuff pressure oscillations detected by electronic sensor. Processing: conversion of pressure changes to electrical signals; calculation of BP and pulse rate; irregular pulse detection via pulse interval analysis. Output: BP/pulse readings on LCD; data transmission via Bluetooth to mobile app. Used in home environment by adult patients. Benefits: non-invasive, automated BP monitoring; irregular heartbeat detection. Software/App functions limited to data transfer/display (Non-device-MDDS).
Clinical Evidence
Clinical study per ISO 81060-2:2018/Amd.1:2020. Sample: 86 subjects (mean age 52±13.5), 258 datasets. Results: Criterion 1 mean differences 0.17±2.57 mmHg (SBP) / 0.79±2.09 mmHg (DBP); Criterion 2 mean differences 0.17±1.99 mmHg (SBP) / 0.79±1.68 mmHg (DBP). Device fulfilled both validation criteria.
Indicated for adult patients for non-invasive measurement of systolic/diastolic blood pressure and pulse rate in home settings. Suitable for arm circumferences 22-42cm.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Noninvasive blood pressure measurement system / HEM-7311 (K133379)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 7, 2024
Guangdong Transtek Medical Electronics Co., Ltd. Jerry Fan RA Manager Zone A. No. 105. Dongli Road. Torch Development District Zhongshan, Guangdong 528437 China
Re: K240832
Trade/Device Name: Blood Pressure Monitor (BBZ32-AA01) Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: March 26, 2024 Received: March 26, 2024
Dear Jerry Fan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Stephen C. Browning -S
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K240832
Device Name
Blood Pressure Monitor (BBZ32-AA01)
#### Indications for Use (Describe)
This Blood Pressure Monitor is a digital monitor intended for use in measuring blood pressure and pulse rate with arm circumference ranging from 22cm to 32cm (about 8%"-12½") or 22cm to 42cm (about 83/4"-161/2").
It is intended for adult indoor use only.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
{4}------------------------------------------------
# 510(k) Summary
Prepared in accordance with the requirement of 21 CFR Part 807.92
Prepared Date: 03/26/2024
1. Submitter Guangdong Transtek Medical Electronics Co., Ltd. Name: Zone A, No. 105, Dongli Road, Torch Development District, 528437 Zhongshan, Address: Guangdong, China Contact: Jerry Fan Title: RA Manager E-mail: gt-rateam@transtekcorp.com TEL: +86 15728668528
#### 2. Subject Device
| Trade/Device Name | Blood Pressure Monitor |
|-------------------|------------------------|
| Model | BBZ32-AA01 |
| Common Name | Blood Pressure Monitor |
| Classification | Class II |
| Product Code | DXN |
| Submission Type | Traditional 510(k) |
#### 3. Predicate Device
| Manufacturer: | Omron Healthcare, Inc. |
|----------------------|-------------------------------------------------------------|
| Device Name / Model: | Noninvasive blood pressure measurement system /<br>HEM-7311 |
| 510(k) Number: | K133379 |
| Classification | Class II |
| Product Code | DXN |
### 4. Device Description
Blood Pressure Monitor BBZ32-AA01 is designed to measure systolic pressure, diastolic pressure and pulse rate of adult by a non-invasive technique, with an inflatable cuff wrapped around the upper arm. The method to define systolic and diastolic pressure is similar to auscultatory method, though it uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alternations of cuff pressure into electrical signals.
Based on analysis of these signals, the systolic and diastolic blood pressure is defined, and the pulse rate is calculated. This is an extensively used technique applied in blood pressure monitors, also known as "oscillometric method".
The main components of the Blood Pressure Monitor include main unit and cuff. For the outer housing of the main unit, it's made of HIPS material. As for accompanying cuffs, two types of cuffs have been clinically validated to be matched with the device, suitable for adults with arm circumference from 22cm~42cm. The cuff is consisted of fabric and an inflatable bladder inside. For critical electronic components, there is PCB, thermistor, pressure pump, motor, release valve
{5}------------------------------------------------
and pressure sensor.
The device also enjoys a function of detecting irregular pulse rate. When measurements were performed, the monitor will record all pulse intervals and calculate the average. If two or more pulse intervals were recorded, and the difference between each interval and the average is larger than ±25% of the average; or if four or more pulse intervals were recorded, and the difference between each interval and the average is larger than ±15% of the average, the irregular pulse symbol will be displayed along with measurement results.
An embedded Bluetooth wireless connection module in the device allows it to connect with matching receiving ends. When a measurement is done, the results will be displayed on LCD, and the measured data will be transferred to the APP via Bluetooth.
With the use of software (including APP) and Bluetooth communication module, the wireless software function and hardware function are solely intended to transfer, store, convert formats, or display medical device data and results (blood pressure and pulse rate readings), without controlling or altering the functions or parameters of any connected medical devices, which is intended to transmit clinical data from Point B only, and not involved in active patient monitoring. Therefore, based on the FDA guidance titled "Medical Device Data Systems, Medical Image Storage Devices, and Medical Image Communications Devices" (issued on September 28, 2022.), this software function is belong to Non-device-MDDS, and the hardware function is belong to Device-MDDS, they are not subject to FDA laws and regulations applicable to devices.
#### 5. Indications for use
This Blood Pressure Monitor is a digital monitor intended for use in measuring blood pressure and pulse rate with arm circumference ranging from 22cm to 32cm (about 8¾ -12½´) or 22cm to 42cm (about 834"-161/2").
It is intended for adult indoor use only.
| Item | Subject Device | Predicate Device | Note |
|----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| 510(k) | Applying | K133379 | / |
| Model | BBZ32-AA01 | HEM-7311 | / |
| Applicant | Guangdong Transtek Medical<br>Electronics Co., Ltd. | OMRON HEALTHCARE, INC. | Same |
| Device Name | Blood Pressure Monitor | Blood Pressure Monitor | Same |
| Product Code | DXN | DXN | Same |
| Classification | Class II | Class II | Same |
| Regulation # | 21 CFR 870.1130 | 21 CFR 870.1130 | Same |
| Type of Use | OTC | OTC | Same |
| Intended Use /<br>Indication for<br>Use | This Blood Pressure Monitor is a<br>digital monitor intended for use in<br>measuring blood pressure and pulse<br>rate with arm circumference ranging<br>from 22cm to 32cm (about 8¾"-12½") | The device is a digital monitor<br>intended for use in measuring blood<br>pressure and pulse rate in adult patient<br>population.<br>The device detects the appearance of | Similar, refer<br>to Note 1. |
| | or 22cm to 42cm (about 8¾"-16½")<br>It is intended for adult indoor use only. | irregular heartbeats during<br>measurement and gives a warning<br>signal with readings. | |
| Arm<br>circumference | 22cm to 32cm (about 8¾"-12½") or<br>22cm to 42cm (about 8¾"-16½") | Small 17-22cm, Standard 22-32cm,<br>Large 22-42cm | Similar, refer<br>to Note 2. |
| Patient<br>Population | Adult | Adult | Same |
| Principle | Oscillometric method | Oscillometric method | Same |
| Anatomical Site | Upper Arm | Upper Arm | Same |
| Where used<br>(hospital, home,<br>ambulance, etc.) | Home | Home | Same |
| Power Supply | 4*1.5V AAA batteries | 4*AA batteries or AC adapter | Similar, refer<br>to Note 3. |
| Human Factors | Blood pressure | Blood pressure | Same |
| Measurement<br>Items | Measuring systolic and diastolic blood<br>pressure and pulse rate of intended<br>population, including irregular pulse<br>rhythm detection. | Measuring systolic and diastolic blood<br>pressure and pulse rate of intended<br>population, including irregular pulse<br>rhythm detection. | Same |
| Cuff Deflation | Automatic deflation | Automatic pressure release valve | Same |
| Blood Pressure<br>Measurement | 0mmHg ~ 299mmHg,<br>5°C - 40°C within ±3mmHg (0.4kPa) | 0mmHg ~ 299mmHg, within<br>±3mmHg or 2% of reading | Same |
| Pulse Rate<br>Measurement | 40-199 beat/minute, ±5% | 40-180 beat/minute, within ±5% of<br>reading | Similar, refer<br>to Note 2. |
| Display | LCD | LCD | Same |
| Memory | 2×99 | 2×100 | Similar, refer<br>to Note 3. |
| Operation<br>Environment | Temperature: 5°C~40°C;<br>Relative Humidity: 15%~90% RH;<br>Atmospheric: 700hPa~1060hPa | Temperature: 10°C~40°C;<br>Relative Humidity: 15%~90% RH; | Similar, refer<br>to Note 3. |
| Storage and<br>transportation<br>Environment | Temperature: -20°C~60°C;<br>Relative humidity≤93%RH,<br>non-condensing;<br>Atmospheric: 500hPa~1060hPa | Temperature: -20°C~60°C;<br>Relative humidity 10 to 95%RH,<br>non-condensing; | Similar, refer<br>to Note 3. |
| Performance | Compliance with IEC 80601-2-30 | Compliance with IEC 80601-2-30 | Same |
| Clinical | Compliance with ISO 81060-2 | Compliance with ANSI/AAMI/ISO<br>81060-2 | Same |
| Biocompatibility | All patient contact parts meet the | Compliance with ISO 10993-1 | Same |
| | requirements of ISO 10993-1/5/10/23 | | |
| Electrical Safety | Compliance with AAMI ANSI ES<br>60601-1 and IEC 60601-1-11 | Compliance with AAMI ANSI ES<br>60601-1 and IEC 60601-1-11 | Same |
| EMC | Compliance with IEC 60601-1-2 | Compliance with IEC 60601-1-2 | Same |
| Wireless | Bluetooth | Not publicly available | Similar, refer<br>to Note 3. |
#### 6. Comparison to Predicate Device
{6}------------------------------------------------
{7}------------------------------------------------
Justification for difference:
#### Note 1:
Although there are some differences between the "Intended Use / Indication for Use" of the subject device and predicate device, it is only expressed differently, and the general meaning is the same, and they both use in measuring blood pressure and pulse rate. Thus, the difference here does not raise any issues concerning safety and effectiveness.
### Note 2:
Although there are some differences between the "Arm circumference" and "Pulse Rate Measurement" of the subject device and the predicate device, both the subject device and predicate device can be applied to 22-32cm and 22-42cm arm circumference, the pulse rate range of the subject device is similar to that of the predicate device, and the subject device has also been validated according to IEC 80601-2-30 and ISO 81060-2. As demonstrated in relevant test reports, the difference here does not raise any issues concerning safety and effectiveness.
#### Note 3:
Although there are some differences between the "Power Supply", "Memory", "Operation Environment", "Storage and transportation Environment" and "Wireless" of the subject device and the predicate device, both the subject device and predicate device comply with the requirement of standard AAMI ANSI ES 60601-1, IEC 60601-1-11, IEC 60601-1-2 and IEC 80601-2-30, and the subject device also complies with the guidance "Content of Premarket Submissions for Device Software Functions". As demonstrated in relevant test reports, the difference here does not raise any issues concerning safety and effectiveness.
#### Conclusion:
Based on the comparison and analysis in this submission, it can be concluded that: the subject device is substantially equivalent to the predicate device and reference device regarding safety and effectiveness.
#### 7. Performance Data
The following performance data were provided in support of the substantial equivalence determination:
{8}------------------------------------------------
### Biocompatibility testing:
The biocompatibility evaluation for the device was conducted in accordance with the FDA Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The biocompatibility testing includes the following tests:
- Cytotoxicity
- Sensitization
- Irritation
The subject device is considered as surface contacting for a duration not exceed 24hours.
## Non-clinical data
The device has been tested according to following standards:
- IEC 60601-1: Medical electrical equipment Part 1: General requirements for basic safety ● and essential performance
- IEC 60601-1-2: Medical electrical equipment – Part 1-2: General requirement for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests
- IEC 80601-2-30: Medical electrical equipment – Particular requirements for basic safety and essential performance of automated non-invasive sphygmomanometers.
- IEC 60601-1-11: Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- FDA Guidance for Non-Automated Sphygmomanometer
## Wireless testing:
- 47 CFR Part 15, Subpart B, Subpart C
- Radio Frequency Wireless Technology in Medical Devices: Guidance for Industry and Food and Drug Administration Staff (August 14, 2013)
### Clinical data
device was tested according to ISO 81060-2:2018/Amd.1:2020 Non-invasive The sphygmomanometers - Part 2: Clinical investigation of intermittent automated measurement type AMENDMENT 1. The clinical study was performed on the device matching with cuffs ranging from 22-32cm and 22-42cm. The study population consisted of 86 qualified subjects. 258 datasets were collected from 86 subjects with aged 52±13.5. The mean differences between reference BPs and device readings were 0.17±2.57/0.79±2.09mmHg for systolic BP (SBP)/diastolic BP (DBP) of criterion 1, and 0.17 ± 1.99/0.79 ± 1.68mmHg for SBP/DBP of criterion 2. BBZ32-AA01 fulfilled both validation criteria 1 and 2 of the ISO 81060-2:2018/Amd.1:2020.
### 8. Conclusion
Based on the information presented in this 510(k) premarket notification submission, a conclusion can be drawn that the proposed subject device is considered substantially equivalent to the predicate device. The differences between the subject device and the predicate devices were
{9}------------------------------------------------
successfully tested with relevant standards and FDA guidance, and do not affect equivalent safety and effectiveness or raise new issues concerning safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.