← Product Code [NLH](/productcode/NLH) · K240826

# Reprocessed DECANAV™ Electrophysiology Catheter; Reprocessed WEBSTER® Duo-Decapolar Electrophysiology Catheter (K240826)

_Sterilmed, Inc. · NLH · Oct 11, 2024 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K240826

## Device Facts

- **Applicant:** Sterilmed, Inc.
- **Product Code:** [NLH](/productcode/NLH.md)
- **Decision Date:** Oct 11, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1220
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Reprocessed DECANAV™ Electrophysiology Catheter is indicated for electrophysiological mapping of cardiac structures i.e., recording and stimulation, in the Coronary Sinus. In addition, the Reprocessed DECANAV™ Catheter is used with compatible Carto® 3 EP Navigation Systems to provide catheter tip location information. The Reprocessed WEBSTER® Duo-Decapolar Electrophysiology Catheter is indicated for electrophysiological mapping of cardiac structures, i.e., stimulation and recording only. In addition, the Reprocessed WEBSTER® Duo-Decapolar Catheter is designed to facilitate electrogram mapping in the atrial region of the heart and coronary sinus.

## Device Story

Sterile, single-patient-use reprocessed electrophysiology catheters; used for cardiac mapping, stimulation, and recording. DECANAV model integrates with CARTO 3 EP Navigation System for tip location; Duo-Decapolar model used for electrogram mapping. Operated by physicians in clinical settings. High-torque shaft with deflectable tip containing platinum/iridium (DECANAV) or platinum (Duo-Decapolar) electrodes. Tip deflection controlled via proximal handpiece with sliding piston. Interfaces with standard recording equipment via cables. Reprocessed maximum one time; includes cleaning, sterilization, and functional testing. Benefits include diagnostic mapping of cardiac electrical signals to guide clinical decision-making.

## Clinical Evidence

Bench testing only. No clinical data. Performance validated through simulated use, visual inspection, and fatigue testing (joint bond strength, torsional resiliency, tip buckle, fluid integrity, deflection/flexation cycling, electrical continuity/leakage, connector cycling, cable retention). Biocompatibility testing per ISO 10993-1 for external communicating device, short duration contact (<24 hours).

## Technological Characteristics

7F diameter, 110-115cm length. Platinum/iridium or platinum electrodes. High-torque shaft with deflectable tip. Manual piston-actuated deflection. Interfaces with CARTO 3 EP Navigation System (magnetic field location) and standard recording equipment. Sterilized via ethylene oxide. Biocompatibility per ISO 10993-1. Packaging validated per ASTM D4169, F88, F2096. Shelf-life validated per ASTM 1980-07.

## Regulatory Identification

An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.

## Predicate Devices

- Biosense Webster CS RefStar™ (DECANAV) Catheter ([K080425](/device/K080425.md))
- Biosense Webster CS RefStar™ (DECANAV) Catheter ([K231312](/device/K231312.md))
- Biosense Webster Duo-Decapolar Catheter ([K101991](/device/K101991.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 11, 2024

Sterilmed, Inc. Rachel Poltilove Regulatory Affairs 5010 Cheshire Pkwy N, Suite 2 Plymouth, Minnesota 55446

Re: K240826

Trade/Device Name: Reprocessed DECANAV™ Electrophysiology Catheter; Reprocessed WEBSTER® Duo-Decapolar Electrophysiology Catheter Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter Or Electrode Recording Probe Regulatory Class: Class II Product Code: NLH Dated: September 10, 2024 Received: September 11, 2024

Dear Rachel Poltilove:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Aneesh S. Deoras -S

Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# K240826 - Rachel Poltilove

# The following devices are included in the scope of this 510(k) premarket notification:

| Device Name                                                                | Biosense Webster<br>Model Numbers | Sterilmed<br>Model Numbers | Description                                                                                                  |
|----------------------------------------------------------------------------|-----------------------------------|----------------------------|--------------------------------------------------------------------------------------------------------------|
| Reprocessed<br>DECANAV™<br>Electrophysiology<br>Catheter                   | R7D282CT                          | RR7D282CT                  | D Curve, 7F, 115cm, Electrodes: 11, SP: 2-8-2mm,<br>Tip Length: 2.4mm, Pin Connector: 34                     |
| Reprocessed<br>WEBSTER® Duo-<br>Decapolar<br>Electrophysiology<br>Catheter | R7F282CT                          | RR7F282CT                  | F Curve, 7F, 115cm, Electrodes: 11, SP: 2-8-2mm,<br>Tip Length: 2.4mm, Pin Connector: 34                     |
| Reprocessed<br>WEBSTER® Duo-<br>Decapolar<br>Electrophysiology<br>Catheter | D728260RT                         | RD728260RT                 | Large Curve 270, 7F, 110 cm, Electrodes: 20, SP: 2-9-3...61...3-9-3mm, Tip Length: 2mm, Pin<br>Connector: 10 |

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known) K240826

#### Device Name

Reprocessed DECANAVTM Electrophysiology Catheter Reprocessed WEBSTER® Duo-Decapolar Electrophysiology Catheter

### Indications for Use (Describe)

The Reprocessed DECANAV™ Electrophysiology Catheter is indicated for electrophysiological mapping of cardiac structures i.e., recording and stimulation, in the Coronary Sinus. In addition, the Reprocessed DECANAV™ Catheter is used with compatible Carto® 3 EP Navigation Systems to provide catheter tip location information.

The Reprocessed WEBSTER® Duo-Decapolar Electrophysiology Catheter is indicated for electrophysiological mapping of cardiac structures, i.e., stimulation and recording only. In addition, the Reprocessed WEBSTER® Duo-Decapolar Catheter is designed to facilitate electrogram mapping in the atrial region of the heart and coronary sinus.

| Type of Use (Select one or both, as applicable)                               |                                                                              |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span>    Prescription Use (Part 21 CFR 801 Subpart D)   </div> | <div> <span> </span>    Over-The-Counter Use (21 CFR 801 Subpart C)   </div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

## 510(k) Summary

This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.

| Date Prepared:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | October 11, 2024                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter and<br>Manufacturer:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Sterilmed, Inc.<br>5010 Cheshire Parkway N, Suite 2<br>Plymouth, MN 55446<br>www.sterilmed.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Manufacturing<br>Facility Address:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 11400 73rd Avenue<br>North Maple Grove, MN 55369                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Official<br>Correspondent:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Dr. Rachel Poltilove<br>Regulatory Consultant<br>Sterilmed, Inc.<br>Tel: 301-213-7224<br>Email: rpoltilo@its.jnj.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Trade Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Reprocessed DECANAV™ Electrophysiology Catheter<br>Reprocessed WEBSTER® Duo-Decapolar Electrophysiology<br>Catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Classification<br>Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Catheter, Recording, Electrode, Reprocessed                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Common Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Diagnostic Electrophysiology Catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Device<br>Classification:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Class II, 21 CFR 870.1220                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Product Code                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | NLH                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Primary Predicate<br>Device:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Biosense Webster CS RefStar™ (DECANAV) Catheter (K080425)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Secondary<br>Predicate Devices:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Biosense Webster CS RefStar™ (DECANAV) Catheter (K231312)<br>Biosense Webster Duo-Decapolar Catheter (K101991)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Device Description:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | The Reprocessed DECANAV™ Catheter is a sterile, single patient<br>use device designed to be used with the CARTO® 3 EP Navigation<br>System (a magnetic field location technology) to facilitate<br>electrophysiological mapping of the heart. The catheter has a high<br>torque shaft with a deflectable tip section containing an array of<br>platinum/iridium electrodes that can be used for stimulation and<br>recording of cardiac electrical signals. The Reprocessed<br>DECANAV™ Catheter has a single proximal electrode that can be<br>used for unipolar recording signals. The Reprocessed DECANAV™<br>Catheter tip deflection is controlled by a proximal hand piece that<br>features a thumb operated sliding piston and is offered in curve                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | types D and F. Pushing the thumb knob forward causes the<br>catheter tip to bend (curve); when the knob is pulled back, the tip<br>straightens. The plane of the curved tip can be rotated during use.<br>The Reprocessed DECANAV™ Catheter interfaces with standard<br>recording equipment and CARTO® 3 EP Navigation System<br>equipment via interface cables with the appropriate connectors.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | The Reprocessed WEBSTER® Duo-Decapolar Catheter (DDP) is a<br>sterile, single patient use device designed to facilitate<br>electrophysiological mapping of the heart. The catheter has a<br>high-torque shaft with a deflectable tip section containing an<br>array of platinum electrodes that can be used for stimulation and<br>recording. Tip deflection is controlled at the proximal end by a<br>tubular handpiece in which a piston slides. When the piston is<br>pushed forward with the thumb knob, the tip is deflected<br>(curved). When the piston is pulled back, the tip straightens. The<br>high torque shaft allows the plane of the curved tip to be rotated<br>to facilitate accurate positioning of the catheter tip at the desired<br>site. The catheter interfaces with standard recording equipment<br>via interface cables with the appropriate connectors. |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | RR7D282CT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Reprocessed DECANAV™ Electrophysiology Catheter,<br>D Curve, 7F, 115cm, Electrodes: 11, SP: 2-8-2mm, Tip<br>Length: 2.4mm, Pin Connector: 34                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Model Numbers                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | RR7F282CT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Reprocessed DECANAV™ Electrophysiology Catheter,<br>F Curve, 7F, 115cm, Electrodes: 11, SP: 2-8-2mm, Tip<br>Length: 2.4mm, Pin Connector: 34                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | RD728260RT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Reprocessed WEBSTER® Duo-Decapolar<br>Electrophysiology Catheter, Large Curve 270, 7F, 110<br>cm, Electrodes: 20, SP: 2-9-3613-9-3mm, Tip<br>Length: 2mm, Pin Connector: 10                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| tip location information.<br>Indications For<br>Use:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | The Reprocessed DECANAV™ Electrophysiology Catheter is<br>indicated for electrophysiological mapping of cardiac structures,<br>i.e., recording and stimulation, including in the Coronary Sinus. In<br>addition, the Reprocessed DECANAV™ Catheter is used with<br>compatible Carto® 3 EP Navigation Systems to provide catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | and coronary sinus.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | The Reprocessed WEBSTER® Duo-Decapolar Electrophysiology<br>Catheter is indicated for electrophysiological mapping of cardiac<br>structures, i.e., stimulation and recording only. In addition, the<br>Reprocessed WEBSTER® Duo-Decapolar Catheter is designed to<br>facilitate electrogram mapping in the atrial region of the heart                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Technological<br>Characteristics:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | The design, materials, function, and intended use of the<br>Reprocessed DECANAV™ and Reprocessed WEBSTER® Duo-<br>Decapolar Electrophysiology Catheters are the same as that of the<br>predicate devices. There are no changes to the claims, clinical<br>applications, patient population, performance specifications, or<br>method of operation.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Functional and<br>Safety Testing:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Representative samples of reprocessed devices were tested to<br>demonstrate appropriate functional characteristics. Process<br>validation testing was performed to validate cleaning and<br>sterilization as well as device packaging. In addition, the<br>manufacturing process includes visual and validated functional<br>testing of all products produced.<br>The Reprocessed DECANAV™ and Reprocessed WEBSTER® Duo-<br>Decapolar Electrophysiology Catheters are reprocessed no more<br>than one (1) time. The catheters are marked and tracked through<br>the reprocessing cycle. After the devices have reached the<br>maximum number of reprocessing cycles, the devices are rejected<br>from further reprocessing. It is Sterilmed's policy to restrict its<br>reprocessing to exclude devices previously reprocessed by other<br>reprocessors. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Summary of<br>Non-Clinical Tests<br>Conducted:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Specific non-clinical tests performed included: cleaning<br>validation, sterilization verification, ethylene oxide residual<br>testing (ISO 10993-7), packaging validation (ASTM D4169, ASTM<br>F88, ASTM F2096), and shelf-life validation (ASTM 1980-07). In<br>addition, validation of functional performance (bench testing)<br>was performed through simulated use, visual inspection, and<br>fatigue testing. Testing performed:<br>• Joint Bond Strength<br>• Torsional Resiliency<br>• Tip Buckle<br>• Fluid Integrity<br>• Deflection Cycling<br>• Flexation Cycling<br>• Shaft Stiffness<br>• Electrical Continuity<br>• Electrical Leakage<br>• Electrical Connector Cycling<br>• Cable to Handle Retention Strength<br>• CARTO® 3 System Compatibility of Duo-Decapolar                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| CARTO® 3 System Recognition of DECANAV Electrical Resistance and Isolation Coronary Sinus Handling              Performance testing shows the Reprocessed DECANAV™ and<br>Reprocessed WEBSTER® Duo-Decapolar Electrophysiology<br>Catheters perform as originally intended.<br>In addition, the devices were tested for biocompatibility per<br>ISO 10993-1 for external communicating device, short<br>duration contact with circulating blood (<24 hours).<br>Biocompatibility testing included:  Cytotoxicity Sensitization Irritation/Intracutaneous Reactivity Acute Systemic Toxicity Pyrogenicity Hemocompatibility including Thrombogenicity, Hemolysis,<br>and Complement Activation |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Conclusion:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Sterilmed conducted performance testing for the Reprocessed<br>DECANAV™ and Reprocessed WEBSTER® Duo-Decapolar<br>Electrophysiology Catheters against the OEM predicate devices,<br>Biosense Webster CS RefStar™ (DECANAV) Catheter (K080425 &<br>K231312) and Biosense Webster Duo-Decapolar Catheter<br>(K101991). Results demonstrated substantial equivalence to the<br>predicate devices.<br>Based on testing results, the Reprocessed DECANAV™ and<br>Reprocessed WEBSTER® Duo-Decapolar Electrophysiology<br>Catheters are substantially equivalent to the predicate devices,<br>the Biosense Webster CS RefStar™ (DECANAV™) Catheter and the<br>Biosense Webster Duo-Decapolar Catheter                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

The following devices are included in the scope of this 510(k) submission:

| Device Name                                                                | Biosense Webster<br>Model Numbers | Sterilmed<br>Model Numbers | Description                                                                                                  |
|----------------------------------------------------------------------------|-----------------------------------|----------------------------|--------------------------------------------------------------------------------------------------------------|
| Reprocessed<br>DECANAV™<br>Electrophysiology<br>Catheter                   | R7D282CT                          | RR7D282CT                  | D Curve, 7F, 115cm, Electrodes: 11, SP: 2-8-2mm,<br>Tip Length: 2.4mm, Pin Connector: 34                     |
| Reprocessed<br>WEBSTER® Duo-<br>Decapolar<br>Electrophysiology<br>Catheter | R7F282CT                          | RR7F282CT                  | F Curve, 7F, 115cm, Electrodes: 11, SP: 2-8-2mm,<br>Tip Length: 2.4mm, Pin Connector: 34                     |
| Reprocessed<br>WEBSTER® Duo-<br>Decapolar<br>Electrophysiology<br>Catheter | D728260RT                         | RD728260RT                 | Large Curve 270, 7F, 110 cm, Electrodes: 20, SP: 2-9-3...61...3-9-3mm, Tip Length: 2mm, Pin<br>Connector: 10 |

---

**Source:** [https://fda-staging.innolitics.com/device/K240826](https://fda-staging.innolitics.com/device/K240826)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
