K240798 · Cntrl+, Inc. · HHW · Dec 17, 2024 · Obstetrics/Gynecology
Device Facts
Record ID
K240798
Device Name
Cntrl+ Bladder Support Pessary
Applicant
Cntrl+, Inc.
Product Code
HHW · Obstetrics/Gynecology
Decision Date
Dec 17, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.3575
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cntrl+ Bladder Support Pessary is intended for the use in adult women, over 18 years of age who experience involuntary urine loss with physical activity (stress urinary incontinence).
Device Story
Cntrl+ Bladder Support Pessary is a cone-shaped, medical-grade Santoprene device for management of stress urinary incontinence (SUI). Inserted intravaginally by the patient, the device provides mechanical support to the urethra to prevent or reduce unwanted urinary leakage during physical activities like coughing, sneezing, or exercising. The device is prescribed by a physician who determines the correct size. It is designed for daily use (up to 12 hours) and is reusable for 30-90 times. By supporting the urethra, the device reduces bladder leaks, potentially improving quality of life for women with SUI.
Clinical Evidence
Leveraged clinical data from two prospective, single-arm studies of the nearly identical Contiform device. Study 1 (N=41 completers): 24-hour pad weight test showed 72% median reduction in urine loss after 3 weeks. Study 2 (N=37 completers): 24-hour pad weight test showed 67% median reduction in urine loss after 4 weeks, with 54% of women achieving dryness (<2g). Reported adverse events included post-treatment residual urine, acute bacterial cystitis, and difficulty removing the device. No serious adverse events were reported.
Technological Characteristics
Material: Medical-grade Santoprene. Form factor: Cone-shaped vaginal pessary. Dimensions: 1.47-1.76 inches diameter. Mechanism: Mechanical urethral support. Biocompatibility: ISO 10993-1:2018 compliant. Testing: TSST-1, durability, compression force, and vaginal wall force testing.
Indications for Use
Indicated for adult women >18 years experiencing stress urinary incontinence (involuntary urine loss with physical activity).
Regulatory Classification
Identification
A vaginal pessary is a removable structure placed in the vagina to support the pelvic organs and is used to treat conditions such as uterine prolapse (falling down of uterus), uterine retroposition (backward displacement), or gynecologic hernia.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 17, 2024
Cntrl+ Inc.
% Charusheila Ramkumar
Consultant
Avania CRO Canada Inc.
250 Carlaw Ave., Suite 108
Toronto, ON M4M3L1
CANADA
Re: K240798
Trade/Device Name: Cntrl+ Bladder Support Pessary
Regulation Number: 21 CFR 884.3575
Regulation Name: Vaginal Pessary
Regulatory Class: II
Product Code: HHW
Received: November 4, 2024
Dear Charusheila Ramkumar:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K240798 - Charusheila Ramkumar
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K240798 - Charusheila Ramkumar
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica K. Nguyen -S
Jessica K. Nguyen, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K240798 | |
| Device Name Cntrl+ Bladder Support Pessary | |
| Indications for Use (Describe) The Cntrl+ Bladder Support Pessary is intended for the use in adult women, over 18 years of age who experience involuntary urine loss with physical activity (stress urinary incontinence). | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K240798
Page 1 of 4
# TRADITIONAL 510(K) SUMMARY
## 1. SUBMITTER INFORMATION
Applicant
Cntrl+ Inc.
501 Campbell St. Unit 3B
Cornwall, ON K6H6X5, Canada
Applicant Contact
Karen Brunet, President
E-mail: karen@cntrlplus.com
Phone: (613) 935-5073
Official Correspondent
Dr. Charusheila Ramkumar,
E-mail: charusheila.ramkumar@avaniaclinical.com
Avania CRO Canada Inc.
250 Carlaw Ave. Suite 108
Toronto, ON M4M3L1, Canada
Phone: (647) 773-0974
Date Prepared
December 17, 2024
## 2. DEVICE NAME
Trade Name of the Device
Cntrl+ Bladder Support Pessary
Classification Name:
Pessary, Vaginal
Classification Regulation:
21 CFR 884.3575
Device Class:
II
Product Code:
HHW
Panel:
Gastroenterology/Urology
## 3. PREDICATE DEVICE IDENTIFICATION
Uresta™ Pessary (K081385)
Predicate device was not subjected to any design related recall.
## 4. DEVICE DESCRIPTION:
The subject Pessary is designed for the management of stress urinary incontinence (SUI) in women, is a device made of medical-grade Santoprene, inserted intravaginally to support the urethra. The Cntrl+ Bladder Support Pessary reduces bladder leaks in women who are suffering from SUI by preventing or reducing unwanted urinary leakage. Use of the Cntrl+ Bladder Support Pessary is applicable for women who experience leakage when laughing, coughing, exercising, sneezing, etc. The shape of the pessary is designed to fit in the vaginal tract. The device fits most users; the correct size will be determined and prescribed by the physician. The Cntrl+ Bladder Support Pessary can be
{5}
K240798
Page 2 of 4
worn daily for up to 12 hours. The Cntrl+ Bladder Support Pessary can be reused (and cleaned) for 30-90 times.
# 5. INDICATIONS FOR USE:
The Cntrl+ Bladder Support Pessary is intended for the use in adult women, over 18 years of age who experience involuntary urine loss with physical activity (stress urinary incontinence).
# 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| Device & Predicate Device(s) | K240798 (Subject) | K081385 (Predicate) |
| --- | --- | --- |
| Device Name | Cntrl+ Bladder Support Pessary | Uresta™ Pessary |
| Indication For use | The Cntrl+ Bladder Support Pessary is intended for the use in adult women, over 18 years of age who experience involuntary urine loss with physical activity (stress urinary incontinence). | The device is for the use in adult women, over 18 years of age who experience involuntary urine loss with physical activity (stress urinary incontinence). |
| Device Type | Vaginal Pessary | Vaginal Pessary |
| Size (diameter in inch) | 1.47-1.76 | 1.73-4.77 |
| Shape | Cone-shape | Bell-shaped |
| Duration of use | Up to 12-hours in a 24-hour period Reusable for 90 days | Up to 24 hours and can be re-inserted after cleaning. The maximum duration of use is not specified. |
| Material | Bladder Support – Medical Grade Santoprene | Bladder Support – Santoprene (non-latex thermoplastic rubber) |
| Rx or OTC | RX | RX |
| Reusable or Single use | Single patient reusable | Single patient reusable |
As evidenced by the above table, both the subject and the predicate devices have the same intended use but differ in technological characteristics. Performance testing was conducted on the subject device, and it was established that the differences in technological characteristics do not raise different questions of safety or effectiveness.
# 7. SUMMARY OF NONCLINICAL TESTING:
Below is a list of the tests that were performed and successfully completed for the subject device per the specified guidance and standards:
- Biocompatibility testing according to ISO 10993-1:2018 - Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process and FDA Guidance "Use of International Standard ISO 10993-1" (2016).
- Toxic shock syndrome toxin-1 (TSST-1) testing to assess if subject device enhances the growth of Staphylococcus Aureus or alters the growth of normal vaginal microflora
{6}
K240798
Page 3 of 4
- The following bench performance testing were conducted per internal protocols designed by the sponsor:
- Durability Testing
- Compression Force Testing
- Testing to evaluate the maximum force exerted on the vaginal wall by the subject pessary
All pre-determined acceptance criteria were met.
## 8. SUMMARY OF CLINICAL EVIDENCE:
Contiform is a vaginal pessary that is not marketed in the US, but it is currently marketed in Australia and United Kingdom. The design, geometry, materials and features of the Contiform and the subject device are almost identical. The only difference between the subject device and Contiform’s design is the subject device uses a green colorant which is absent in Contiform pessary. A comparison including non-clinical testing was provided to demonstrate equivalence between Contiform and the subject device. Based upon this comparison, the available clinical data on Contiform were leveraged to support the safety and effectiveness of the Cntrl+ device, given the differences in physical dimensions and design of the subject device compared to the predicate device.
Data from two clinical studies conducted on Contiform were leveraged to support safety and effectiveness of the Cntrl+ in the intended population. The summary of those studies are provided below:
1. Morris, AR and Moore KH: The Contiform incontinence device – efficacy and patient acceptability Int Urogynecol. J (2003) 14: 412–417, DOI 10.1007/s00192-003-1094-8
**Location:**
St George Hospital, Sydney, Australia
**Study Design, Study Population, Primary Endpoints:**
This study had a median use of the subject pessary for 21 days (maximum 35 days) and was a prospective, single arm, clinical study that enrolled 59 subjects with dominant stress urinary incontinence or stress urinary incontinence (SUI). Forty one (41) subjects with mild, moderate, and severe SUI completed the study, with age range 42 - 53. After 3 weeks of using the Contiform device, the primary endpoint, urine loss reduction, was evaluated by using a 24-hours pad weight test. Urine loss was reduced by a median of 72% from baseline.
Adverse events related to post treatment residual urine (N=6), acute bacterial cystitis (N=2), difficulty removing the pessary (N =3) were reported during this study. No serious adverse events were reported.
2. Allen WA, Leek H, Izurieta A and Moore KH: Update: the “Contiform” intravaginal device in four sizes for the treatment of stress incontinence Int Urogynecol. J (2008) 19:757–761, DOI 10.1007/s00192-007-0519-1
**Location:**
St George Hospital, Sydney, Australia
**Study Design, Study Population, Primary Endpoints:**
This study was 4 weeks long and was a prospective, single arm clinical study that enrolled 73 subjects with dominant stress urinary incontinence or stress urinary incontinence (SUI). Thirty seven (37) subjects with mild, moderate, and severe SUI completed the study,
{7}
K240798
Page 4 of 4
with age range 41-54. After 4 weeks of using the Contiform device, the primary endpoint, urine loss reduction, was evaluated by using a 24-hours pad weight test. Urine loss was reduced by a median of 67% from the baseline and 20 women (54%) were dry (<2 g) during this test.
Adverse events related to post treatment residual urine (N=17) and difficulty removing the pessary (N =7) were reported during this study. No serious adverse events were reported.
The data leveraged from the Contiform clinical studies supports that the Ctrl+ Bladder Support Pessary can treat Stress Urinary Incontinence in the intended patient population.
## 9. CONCLUSIONS
Based on the information presented in this submission, it can be concluded that the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.