The SKOUT® system is a software device designed to detect potential colorectal polyps in real time during colonoscopy examinations. It is indicated as a computer-aided detection tool providing colorectal polyps location information to assist qualified and trained gastroenterologists in identifying potential colorectal polyps during colonoscopy examinations in adult patients undergoing colorectal cancer screening or surveillance. The SKOUT® system is only intended to assist the gastroenterologist in identifying suspected colorectal polyps and the gastroenterologist is responsible for reviewing SKOUT® suspected polyp areas and confirming the presence or absence of a polyp based on their own medical judgment. SKOUT® is not intended to replace a full patient evaluation, nor is it intended to be relied upon to make a primary interpretation of endoscopic procedures, medical diagnosis, or recommendations of treatment/course of action for patients. SKOUT® is indicated for white light colonoscopy only.
Device Story
SKOUT® system is a software-based computer-aided detection (CADe) tool for real-time analysis of high-definition endoscopic video during colonoscopy. It processes video input to identify potential colorectal polyps; outputs visual markers (blue rectangular outlines) superimposed on the endoscopic monitor. System includes a device status indicator (green 'S' for active, gray 'S' for tool presence, red 'X' for error). Operates in clinical settings; used by gastroenterologists. Automatically pauses detection when surgical tools are detected or lighting is inadequate to prevent obstruction. Includes a mode selection button for manual bypass to original video feed in case of software error. Does not replace clinical judgment; intended for white light colonoscopy only. Benefits include assisting clinicians in polyp identification during screening or surveillance.
Clinical Evidence
No new clinical data provided. The device relies on clinical performance data previously submitted in K213686. Non-clinical performance testing included software verification and validation, and bench testing for true/false positives, pixel degradation, and video delays, all of which met design requirements.
Technological Characteristics
Software-based CADe system; processes high-definition endoscopic video via Serial Digital Interface (SDI). Features include real-time AI-based polyp detection, GUI status indicators, and a hardware bypass switch. Compatible with Olympus EVIS EXERA III and FUJI EC760 series endoscopes. Powered by hospital mains. Software performs de-interlacing on video input. Designed for zero-latency marker annotation.
Indications for Use
Indicated for adult patients undergoing colorectal cancer screening or surveillance colonoscopy. Used as a computer-aided detection tool to assist gastroenterologists in identifying potential colorectal polyps during white light colonoscopy.
Regulatory Classification
Identification
A gastrointestinal lesion software detection system is a computer-assisted detection device used in conjunction with endoscopy for the detection of abnormal lesions in the gastrointestinal tract. This device with advanced software algorithms brings attention to images to aid in the detection of lesions. The device may contain hardware to support interfacing with an endoscope.
Special Controls
In combination with the general controls of the FD&C Act, the gastrointestinal lesion software detection system is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including detection of gastrointestinal lesions and evaluation of all adverse events.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Testing must include:
(i) Standalone algorithm performance testing;
(ii) Pixel-level comparison of degradation of image quality due to the device;
(iii) Assessment of video delay due to marker annotation; and
(iv) Assessment of real-time endoscopic video delay due to the device.
(3) Usability assessment must demonstrate that the intended user(s) can safely and correctly use the device.
(4) Performance data must demonstrate electromagnetic compatibility and electrical safety, mechanical safety, and thermal safety testing for any hardware components of the device.
(5) Software verification, validation, and hazard analysis must be provided. Software description must include a detailed, technical description including the impact of any software and hardware on the device's functions, the associated capabilities and limitations of each part, the associated inputs and outputs, mapping of the software architecture, and a description of the video signal pipeline.
(6) Labeling must include:
(i) Instructions for use, including a detailed description of the device and compatibility information;
(ii) Warnings to avoid overreliance on the device, that the device is not intended to be used for diagnosis or characterization of lesions, and that the device does not replace clinical decision making;
(iii) A summary of the clinical performance testing conducted with the device, including detailed definitions of the study endpoints and statistical confidence intervals; and
(iv) A summary of the standalone performance testing and associated statistical analysis.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" on the bottom.
April 19, 2024
Iterative Scopes, Inc. % Janice Hogan Partner Hogan Lovells US LLP 1735 Market St. Floor 23 Philadelphia, Pennsylvania 19103
Re: K240781
Trade/Device Name: SKOUT® system Regulation Number: 21 CFR 876.1520 Regulation Name: Gastrointestinal Lesion Software Detection System Regulatory Class: Class II Product Code: ONP Dated: March 21, 2024 Received: March 21, 2024
Dear Janice Hogan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Shanil P. Haugen -S
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K240781
Device Name
SKOUT® system
#### Indications for Use (Describe)
The SKOUT® system is a software device designed to detect potential colorectal polyps in real time during colonoscopy examinations. It is indicated as a computer-aided detection tool providing colorectal polyps location information to assist qualified and trained gastroenterologists in identifying potential colorectal polyps during colonoscopy examinations in adult patients undergoing colorectal cancer screening or surveillance.
The SKOUT® system is only intended to assist the gastroenterologist in identifying suspected colorectal polyps and the gastroenterologist is responsible for reviewing SKOUT® suspected polyp areas and confirming the presence or absence of a polyp based on their own medical judgment. SKOUT® is not intended to replace a full patient evaluation, nor is it intended to be relied upon to make a primary interpretation of endoscopic procedures, medical diagnosis, or recommendations of treatment/course of action for patients. SKOUT® is indicated for white light colonoscopy only.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(K) SUMMARY
#### SKOUT® system
#### K240781
#### Submitter Contact Information:
Address: Iterative Health, Inc. (Iterative Scopes) 14 Arrow Street Floor 3 Cambridge, MA 02138
Phone: (617) 209-9773
Contact: Caitlyn Seidl, VP, Clinical, Regulatory and Medical Affairs
Date Prepared: March 21, 2024
Name of Device: SKOUT® system
Classification Name: Gastrointestinal lesion software detection system (21 CFR 876.1520, Product Code QNP)
Predicate Device: SKOUT® System, Iterative Scopes, K230658
#### Device Description
The SKOUT® system is a software-based computer aided detection (CADe) system for the analysis of high-definition endoscopic video during colonoscopy procedures. The SKOUT system is intended to aid gastroenterologists with the detection of potential colorectal polyps during colonoscopy by providing an informational visual aid on the endoscopic monitor using trained software that processes the endoscopic video in real time.
Users will primarily interact with the SKOUT system by observing the software display, including the polyp detection box and device status indicator signal.
#### Polyp Detection Notification
The SKOUT system has a main graphical user interface (GUI) feature of the polyp detection notification. The polyp detection is a two-dimensional blue rectangular outline generated around any suspected polyps on the endoscopic video feed. If there is no polyp detected, the bounding box does not appear. SKOUT system pauses polyp detection when an endoscopic tools are detected in the video feed to ensure that the bounding box does not hinder any surgical procedure, biopsy, or resection or this may also occur when lighting conditions are deemed to be inadequate.
The polyp detection notification enables users to:
- . Detect potential colorectal polyps during colonoscopy examinations in adult patients undergoing a colorectal cancer screening or surveillance procedure.
- Utilize a tool that provides additional information for endoscopic observation.
{5}------------------------------------------------
## Device Status Indicator
The SKOUT system has an additional GUI feature that notifies users of the current device status (active or error):
- a two-dimensional green box with letter (S) when the device is powered on and actively ● processing video.
- . a two-dimensional gray box with letter (S) when a surgical tool is present.
- a red (X) with an error message; when there is an error with the video processing function of the SKOUT system, the green box will be replaced with a red X and error message to indicate an error has occurred.
## Intended Use / Indications for Use
The SKOUT system is a software device designed to detect potential colorectal polyps in real time during colonoscopy examinations. It is indicated as a computer-aided detection tool providing colorectal polyps location information to assist qualified and trained gastroenterologists in identifying potential colorectal polyps during colonoscopy examinations in adult patients undergoing colorectal cancer screening or surveillance.
The SKOUT system is only intended to assist the gastroenterologist in identifying suspected colorectal polyps and the gastroenterologist is responsible for reviewing SKOUT suspected polyp areas and confirming the presence or absence of a polyp based on their own medical judgment. SKOUT is not intended to replace a full patient evaluation, nor is it intended to be relied upon to make a primary interpretation of endoscopic procedures, medical diagnosis, or recommendations of treatment/course of action for patients. SKOUT is indicated for white light colonoscopy only.
#### Summary of Technological Characteristics
The subject device is a modified version of the predicate device, which was FDA-cleared on April 7, 2023 (K230658). The key characteristics of the subject SKOUT system and the predicate SKOUT system are compared in the following table.
| | Subject Device:<br>SKOUT System (K240781) | Predicate Device:<br>SKOUT System (K230658) | Comparison |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Product Code | QNP | QNP | Same |
| Regulation | 21 CFR 876.1520 | 21 CFR 876.1520 | Same |
| Device Class | Class II | Class II | Same |
| Indications for<br>Use | The SKOUT system is a<br>software device designed to<br>detect potential colorectal polyps<br>in real time during colonoscopy<br>examinations. It is indicated as a<br>computer-aided detection tool<br>providing colorectal polyps<br>location information to assist<br>qualified and trained | The SKOUT system is a<br>software device designed to<br>detect potential colorectal polyps<br>in real time during colonoscopy<br>examinations. It is indicated as a<br>computer-aided detection tool<br>providing colorectal polyps<br>location information to assist<br>qualified and trained | Same |
| | Subject Device:<br>SKOUT System (K240781) | Predicate Device:<br>SKOUT System (K230658) | Comparison |
| | gastroenterologists in identifying<br>potential colorectal polyps during<br>colonoscopy examinations in<br>adult patients undergoing<br>colorectal cancer screening or<br>surveillance. | gastroenterologists in identifying<br>potential colorectal polyps during<br>colonoscopy examinations in<br>adult patients undergoing<br>colorectal cancer screening or<br>surveillance. | |
| | The SKOUT system is only<br>intended to assist the<br>gastroenterologist in identifying<br>suspected colorectal polyps and<br>the gastroenterologist is<br>responsible for reviewing<br>SKOUT suspected polyp areas<br>and confirming the presence or<br>absence of a polyp based on<br>their own medical judgment.<br>SKOUT is not intended to<br>replace a full patient evaluation,<br>nor is it intended to be relied<br>upon to make a primary<br>interpretation of endoscopic<br>procedures, medical diagnosis,<br>or recommendations of<br>treatment/course of action for<br>patients. SKOUT is indicated for<br>white light colonoscopy only. | The SKOUT system is only<br>intended to assist the<br>gastroenterologist in identifying<br>suspected colorectal polyps and<br>the gastroenterologist is<br>responsible for reviewing<br>SKOUT suspected polyp areas<br>and confirming the presence or<br>absence of a polyp based on<br>their own medical judgment.<br>SKOUT is not intended to<br>replace a full patient evaluation,<br>nor is it intended to be relied<br>upon to make a primary<br>interpretation of endoscopic<br>procedures, medical diagnosis,<br>or recommendations of<br>treatment/course of action for<br>patients. SKOUT is indicated for<br>white light colonoscopy only. | |
| User<br>Population | Adult patients undergoing<br>colorectal cancer screening or<br>surveillance colonoscopy | Adult patients undergoing<br>colorectal cancer screening or<br>surveillance colonoscopy | Same |
| Technological<br>Characteristics | The SKOUT system is<br>composed of a single piece<br>hardware design and software<br>designed to highlight portions of<br>the colon where the device<br>detects potential colorectal<br>polyps. | The SKOUT system is<br>composed of a single piece<br>hardware design and software<br>designed to highlight portions of<br>the colon where the device<br>detects potential colorectal<br>polyps. | Same |
| Software<br>Algorithm | The SKOUT system utilizes an<br>artificial intelligence-based<br>algorithm to perform the polyp<br>detection function. | The SKOUT system utilizes an<br>artificial intelligence-based<br>algorithm to perform the polyp<br>detection function. | Same |
| Power Source | Hospital mains power | Hospital mains power | Same |
| | Subject Device:<br>SKOUT System (K240781) | Predicate Device:<br>SKOUT System (K230658) | Comparison |
| Safety<br>Features | The Mode Selection Button<br>allows for instantaneous toggling<br>between the SKOUT video feed<br>and the bypass video feed in the<br>event of software error that<br>affects video quality.<br><br>The polyp detection marker is<br>disabled if a biopsy tool enters<br>the field of view to prevent<br>obstruction of the area of interest<br>during intervention.<br><br>SKOUT system GUI also has a<br>device status indicator that<br>notifies users of the current<br>device status (active or error):<br>a two-dimensional green<br>box with letter (S) when<br>the device is powered on<br>and actively processing<br>video.a two-dimensional gray<br>box with letter (S) when a<br>surgical tool is present.a red (X) with an error<br>message; when there is<br>an error with the video<br>processing function of<br>the SKOUT system, the<br>green box will be<br>replaced with a red X<br>and error message to<br>indicate an error has<br>occurred. | The Mode Selection Button<br>allows for instantaneous toggling<br>between the SKOUT video feed<br>and the bypass video feed in the<br>event of software error that<br>affects video quality.<br><br>The polyp detection marker is<br>disabled if a biopsy tool enters<br>the field of view to prevent<br>obstruction of the area of interest<br>during intervention.<br><br>SKOUT system GUI also has a<br>device status indicator that<br>notifies users of the current<br>device status (active or error):<br>a two-dimensional green<br>box with letter (S) when<br>the device is powered on<br>and actively processing<br>video.a two-dimensional gray<br>box with letter (S) when a<br>surgical tool is present.a red (X) with an error<br>message; when there is<br>an error with the video<br>processing function of<br>the SKOUT system, the<br>green box will be<br>replaced with a red X<br>and error message to<br>indicate an error has<br>occurred. | Same |
| Video Input | Serial Digital Interface (SDI); de-<br>interlacing applied during pre-<br>processing | Serial Digital Interface (SDI); no<br>de-interlacing applied during pre-<br>processing | Similar |
| Device Output | SKOUT system generates<br>markers in the form of blue<br>rectangles superimposed on the<br>endoscopic video when potential<br>colorectal polyps are identified.<br>SKOUT markers are not<br>accompanied by a sound. | SKOUT system generates<br>markers in the form of blue<br>rectangles superimposed on the<br>endoscopic video when potential<br>colorectal polyps are identified.<br>SKOUT markers are not<br>accompanied by a sound. | Same |
| | Subject Device:<br>SKOUT System (K240781) | Predicate Device:<br>SKOUT System (K230658) | Comparison |
| | The polyp detection marker is<br>disabled if a biopsy tool enters<br>the field of view to prevent<br>obstruction of the area of interest<br>during intervention. | The polyp detection marker is<br>disabled if a biopsy tool enters<br>the field of view to prevent<br>obstruction of the area of interest<br>during intervention. | |
| Compatible<br>Endoscopes | Olympus EVIS EXERA III and<br>FUJI EC760 series | Olympus EVIS EXERA III and<br>FUJI EC760 series | Same |
| Video Delay | Video delay due to marker<br>annotation = 0ms (no SE,<br>minimum resolution 1.1ms) | Video delay due to marker<br>annotation = 0.0ms (No SE,<br>minimum resolution 1.1ms) | Same |
| | Video delay due to device = 0ms<br>(no SE, minimum resolution<br>1.1ms) | Video delay due to device =<br>0.0ms (No SE, minimum<br>resolution 1.1ms) | |
| Pixel Level<br>Degradation | No pixel level degradation is<br>introduced by SKOUT to the<br>Endoscopic System. | No pixel level degradation is<br>introduced by SKOUT to the<br>Endoscopic System. | Same |
## Table 1: Comparison of Key Characteristics
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
## Performance Testing
Non-clinical performance testing was conducted to demonstrate that the SKOUT system is as safe and effective as the predicate device.
- Software verification and validation testing was conducted to confirm the SKOUT system software meets design requirements for its intended use, per the device's design documentation in line with recommendations outlined in General Principles of Software Validation, Guidance for Industry and FDA Staff.
- Additional bench software testing was performed to confirm the device meets the special . controls in 21 CFR 876.1520 for true and false positives, pixel degradation and video delays.
SKOUT system demonstrated passing results in all applicable testing.
#### Conclusions
The SKOUT system has the same intended use, indications for use, technological characteristics, and principles of operation as its predicate device. The minor differences in software do not affect its safety and effectiveness when used as labeled. The inference algorithms have remained the same, therefore clinical performance remains unchanged from the clinical performance submitted in K213686. Performance data demonstrates that the SKOUT system is as safe and effective as the predicate device. Thus, the SKOUT system is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.