← Product Code [FMF](/productcode/FMF) · K240774

# Pen Injector (K240774)

_Wuxi Nest Biotechnology Co., Ltd. · FMF · Jun 18, 2024 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K240774

## Device Facts

- **Applicant:** Wuxi Nest Biotechnology Co., Ltd.
- **Product Code:** [FMF](/productcode/FMF.md)
- **Decision Date:** Jun 18, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5860
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The Pen Injector is a re-usable pen injector designed for single patient use by diabetics for the self-injection of a desired dose of insulin. The Pen Injector uses HUMALOG (insulin lispro) injection (U-100) available in 3 mL cartridges, and a single use detachable and disposable insulin pen needle (supplied separately, needle sizes including: 32G*4mm, 31G*4mm, 31G*5mm, and 31G*6mm).

## Device Story

Reusable mechanical pen injector for insulin delivery; accommodates 3 mL insulin cartridges and detachable needles. User manually sets dose via adjustment knob (up to 80 units, 1-unit increments); depressing button engages mechanical coupling to drive piston and expel insulin. Used in professional healthcare facilities and home environments by patients. Device provides mechanical dose delivery; no electronic or software components. Benefits include precise, repeatable insulin administration for diabetes management.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified against ISO 11608-1:2022 for dose accuracy, injection force, and injection time. Biocompatibility testing performed per ISO 10993-1, -5, -10, and -23. Shelf life verified per ASTM F1980-16; transportation testing per ASTM D4169:2022.

## Technological Characteristics

Mechanical pen injector; 3 mL cartridge capacity; 80-unit max dose; 1-unit increments. Materials evaluated for biocompatibility per ISO 10993. Complies with ISO 11608-1:2022. Non-sterile. No energy source; purely mechanical operation.

## Regulatory Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

## Predicate Devices

- NovoPen Echo® ([K182387](/device/K182387.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. Food & Drug Administration".

June 18, 2024

Wuxi NEST Biotechnology Co., Ltd. % Ryan Li, Consultant ICAS Group No.155 Pingbei Road, Minhang District Shanghai, 201109, China

Re: K240774

Trade/Device Name: Pen Injector Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: March 19, 2024 Received: March 21, 2024

Dear Ryan Li:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Shruti N. Mistry -S

Shruti Mistry Assistant Director Division of Drug Delivery and General Hospital Devices, and Human Factors Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K240774

Device Name Pen Injector

#### Indications for Use (Describe)

The Pen Injector is a re-usable pen injector designed for single patient use by diabetics for the self-injection of a desired dose of insulin. The Pen Injector uses HUMALOG (insulin lispro) injection (U-100) available in 3 mL cartridges, and a single use detachable and disposable insulin pen needle (supplied separately, needle sizes including: 32G*4mm, 31G*4mm, 31G*5mm, and 31G*6mm).

Type of Use (Select one or both, as applicable)

| <div> <span> <svg height="12" width="12"> <rect fill="white" height="12" stroke="black" stroke-width="1" width="12"></rect> <line stroke="black" stroke-width="1" x1="0" x2="12" y1="0" y2="12"></line> <line stroke="black" stroke-width="1" x1="0" x2="12" y1="12" y2="0"></line> </svg> </span>                 Prescription Use (Part 21 CFR 801 Subpart D)             </div> |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg height="12" width="12"> <rect fill="white" height="12" stroke="black" stroke-width="1" width="12"></rect> </svg> </span>                 Over-The-Counter Use (21 CFR 801 Subpart C)             </div>                                                                                                                                                          |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary - K240774

#### I. Submitter Information

Submitter Name: Wuxi NEST Biotechnology Co., Ltd. Submitter Address: No. 530, Xida Road, Meicun Industrial Park, Xinwu District, Wuxi, Jiangsu, China Contact Person: Ryan Li Position: Consultant Tel.: + 86 13701581791 Email: IryryanO211@gmail.com Date of Preparation: May 22, 2024

### II. Proposed Device

Device Trade Name: Pen Injector Common Name: Pen-injector Classification Name: Piston Syringe Regulation Number: 21 CFR 880.5860 Regulatory Class: Class II Product Code: FMF Review Panel: General Hospital

### III. Predicate Device

Primary predicate device 510(k) Number: K182387 Trade Name: NovoPen Echo®

### IV. Device Description

The Pen Injector is a reusable mechanical pen-injector capable of injecting a dose of up to 80 units of insulin, in 1 unit increments. The Pen Injector consists of a pen injector body, a push block, a injection button, a dose display window, a dose adjustment knob, a reservoir and a pen injector cap. The intended dose is mechanically set by rotating a dose adjustment knob. The insulin is injected by depressing the knob which mechanical coupling causes the piston in the insulin cartridge to move forward thereby expelling the intended dose. The Pen Injector is intended for single user and provided for non-sterile.

### V. Indication for use

The Pen Injector is a re-usable pen injector designed for single patient use by diabetics for the self-injection of a desired dose of insulin. The Pen Injector uses HUMALOG (insulin lispro) injection (U-100) available in 3 mL cartridges, and a single use detachable and disposable insulin pen needle (supplied separately, needle sizes including: 32G*4mm, 31G*4mm, 31G*5mm, and 31G*6mm).

{4}------------------------------------------------

## VI. Comparison of technological characteristics with the predicate devices

Table 1 summarizes the proposed device technological characteristics with compared to the predicate device under K182387.

| Item                      | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                           | Primary Predicate Device<br>(K182387)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Discussion                 |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Product Name              | Pen Injector                                                                                                                                                                                                                                                                                                                                                                                                                              | NovoPen Echo®                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | N/A                        |
| Product Code              | FMF                                                                                                                                                                                                                                                                                                                                                                                                                                       | FMF                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                       |
| Regulation No.            | 21 CFR 880.5860                                                                                                                                                                                                                                                                                                                                                                                                                           | 21 CFR 880.5860                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                       |
| Class                     | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                  | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                       |
| Indications for Use       | The Pen Injector is a re-usable<br>pen injector designed for single<br>patient use by diabetics for the<br>self-injection of a desired dose<br>of insulin. The Pen Injector<br>uses HUMALOG (insulin<br>lispro) injection (U-100)<br>available in 3 mL cartridges,<br>and a single use detachable and<br>disposable insulin pen needle<br>(supplied separately, needle<br>sizes including: 32G*4mm,<br>31G*4mm, 31G*5mm, and<br>31G*6mm). | The NovoPen Echo® is a<br>re-useable pen injector<br>designed for single patient use<br>by diabetics for the<br>self-injection of a desired<br>dose of insulin. The pen<br>injector uses PenFill® 3 mL<br>cartridge of NovoLog®, 100<br>units/mL (U-100) [insulin<br>aspart injection], Fiasp® 100<br>units/mL (U-100) [insulin<br>aspart injection], and a single<br>use detachable and disposable<br>pen needle (supplied<br>separately). The pen injector<br>allows the user to dial the<br>desired dose from 0.5 to 30<br>units in 0.5 unit<br>increments | Same                       |
| User<br>Environment       | Professional healthcare<br>facilities and home<br>environment                                                                                                                                                                                                                                                                                                                                                                             | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Different<br>Note 1        |
| Reusable Device           | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                       |
| Cartridge<br>Volume       | 3 mL (300 units of U-100<br>insulin)                                                                                                                                                                                                                                                                                                                                                                                                      | 3 mL (300 units of U-100<br>insulin)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same                       |
| Dose Accuracy             | Meets ISO 11608-1:2022<br>requirements                                                                                                                                                                                                                                                                                                                                                                                                    | Meets ISO 11608-1:2014<br>requirements                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                       |
| Dial Increments           | 0.01 mL per increment<br>providing one unit (1U) dose<br>increments                                                                                                                                                                                                                                                                                                                                                                       | 0.005 mL per increment<br>providing half unit (0.5U)<br>dose increments                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Different<br>Note 2        |
| Maximum<br>Delivered Dose | 80 Units                                                                                                                                                                                                                                                                                                                                                                                                                                  | 30 Units                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Different<br>Note 3        |
| Shelf Life                | 5 years                                                                                                                                                                                                                                                                                                                                                                                                                                   | 5 years                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                       |
| Service Life              | 3,000 times of reuse                                                                                                                                                                                                                                                                                                                                                                                                                      | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Different                  |
|                           |                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Note 4                     |
| Biocompatibility          | Cytotoxicity<br>Skin Sensitization<br>Irritation                                                                                                                                                                                                                                                                                                                                                                                          | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Different<br><b>Note 5</b> |
| Sterility                 | Not a sterile device                                                                                                                                                                                                                                                                                                                                                                                                                      | Not a sterile device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same                       |

{5}------------------------------------------------

## Discussion

## Note 1

The user environment for the predicate device is unknown. However, this does not raise different questions of safety and effectiveness as both device are prescription use devices.

## Note 2

Although the dial increments between the proposed device and the predicate device are different, corresponding performance tests were conducted on the proposed device to demonstrate its comply with the ISO 11608-1 standard. Therefore, this difference would not affect the substantial equivalence.

## Note 3

Although the maximum delivered dose between the proposed device and the predicate device are different, corresponding performance tests were conducted on the proposed device to demonstrate the dose accuracy performance comply with the ISO 11608-1 standard. Therefore, this difference would not affect the substantial equivalence.

## Note 4

A service life verification test was conducted to demonstrate that the proposed device is valid during its service life. Therefore, this difference would not affect the substantial equivalence.

## Note 5

The biocompatibility tests were conducted to demonstrate the proposed device is biocompatible with intended users. Therefore, this difference would not affect the substantial equivalence.

## VII. Performance Data

## Non-Clinical Performance Test Conclusion

## Biocompatibility

The proposed device, Pen Injector is categorized as skin contact with a duration of category A-limited(≤24hrs) according to FDA guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part1: Evaluation and testing within a risk management process". The Biocompatibility tests were conducted to verify that the proposed devices are not adverse to human tissue based on the following standards:

- ISO 10993-1 Biological evaluation of medical devices Part 1: Evaluation .

{6}------------------------------------------------

and testing within a risk management process

- ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Test for ● in vitro cytotoxicity
- ISO 10993-10:2021 Biological evaluation of medical devices - Part 10: Tests for skin sensitization
- ISO 10993-23:2021 Biological evaluation of medical devices Part 23: Tests ● for irritation

## Performance Testing

Device Verification Tests confirm that the Pen Injector meets the requirements of ISO 11608-1:2022 Needle-based injection systems for medical use - Requirement and test methods - Part 1: Needle-based injection systems.

The Pen Injector is intended to be used with 3mL cartridges of HUMALOG (insulin lispro) injection (U-100). Therefore, the dose accuracy, injection force and injection time performance tests of the 3mL cartridges of HUMALOG (insulin lispro) injection (U-100) with Pen Injector was performed which meets the requirements of ISO 11608-1:2022. In addition, corresponding tests are conducted between the Pen Injector and the compatible pen needles to demonstrate the compatibility.

### Shelf Life and Service Life

The Pen Injector has a five-year shelf life, as well as 3000 times of reuse service life. For shelf life, in accordance with ASTM F1980-16, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices, aging studies have been conducted to verify a five-year shelf life of the proposed device and ensure that its functionality is successfully maintained throughout the duration of this shelf life. In addition, a service life verification test was conducted on the proposed device to ensure its validity within its service life.

#### Simulated Transportation

A test is conducted according to ASTM D4169:2022 to demonstrate the proposed device could function as intended after transportation.

## Clinical Test Conclusion

No clinical study is included in this submission.

#### VIII. Conclusion

The conclusion drawn from the non-clinical tests demonstrates that the proposed device, Pen Injector is as safe, as effective, and performs as well as the legally marketed predicate device NovoPen Echo® (K182387).

---

**Source:** [https://fda-staging.innolitics.com/device/K240774](https://fda-staging.innolitics.com/device/K240774)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
