K240760 · Steris · JOJ · Apr 10, 2024 · General Hospital
Device Facts
Record ID
K240760
Device Name
CELERITY HP Indicator Tape
Applicant
Steris
Product Code
JOJ · General Hospital
Decision Date
Apr 10, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
The Indicator Tape is intended to secure items wrapped in synthetic wrap materials until used and to distinguish between processed and unprocessed units through a color change from the start color (pink/magenta) to orange, yellow or lighter. The Indicator Tape is also intended to be used as an external process indicator when applied to synthetic wrap materials and/or Tyvek pouches. The tape may be used in the following sterilization cycles: · Lumen, Non Lumen, Fast Non Lumen, Fast, Flexible, and Specialty Cycles of the V-PRO: 1, 1 Plus, maX 2, 60, and s2 Low Temperature Sterilization Systems. · Standard and Advanced Cycles of the STERRAD® NX Sterilizer with or without ALLClear · Standard, Flex Scope, Express and DUO Cycles of the STERRAD® 100NX Sterilizer with or without ALLClear.
Device Story
CELERITY HP Indicator Tape is a 3/4" wide crepe paper tape coated with vaporized hydrogen peroxide (VHP) reactive ink. Used in clinical settings to secure sterilization wraps and provide visual confirmation of VHP exposure. Upon exposure to VHP sterilization cycles, the ink undergoes a chemical reaction, changing color from pink/magenta to orange, yellow, or lighter. The tape is applied manually by sterile processing staff. It serves as a Type 1 process indicator per ANSI/AAMI/ISO 11140-1:2014. The color change allows healthcare providers to distinguish between processed and unprocessed items, ensuring that only sterilized items are used in clinical procedures. The device is a passive chemical indicator; it does not involve electronic processing or software.
Clinical Evidence
Bench testing only. Testing included simulated use, ISO 11140-1 Type 1 performance verification, light stability, endpoint stability, and shelf-life testing. Results confirmed complete color change in full cycles, incomplete change in partial cycles, and adherence through worst-case cycles.
Technological Characteristics
Crepe paper tape with VHP-reactive ink and protective varnish coating. Dimensions: 3/4" wide, 60 yards long. Complies with ANSI/AAMI/ISO 11140-1:2014 for Type 1 process indicators. Passive chemical sensing mechanism; no energy source, software, or connectivity.
Indications for Use
Indicated for use as an external process indicator to secure synthetic wrap materials and Tyvek pouches during VHP sterilization cycles (V-PRO and STERRAD systems) and to provide visual confirmation of processing via color change.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 10, 2024
Steris Corporation Anthony Piotrkowski Director, Regulatory Affairs 5960 Heisley Rd Mentor, Ohio 44060
Re: K240760
Trade/Device Name: CELERITY HP Indicator Tape Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: JOJ Dated: March 20, 2024 Received: March 20, 2024
Dear Anthony Piotrkowski:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Stephen A. Digitally signed by Stephen A. Anisko -S Date: 2024.04.10 Anisko -S 11:42:33 -04'00'
For: Christopher Dugard Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices
{2}------------------------------------------------
OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240760
Device Name CELERITY HP Indicator Tape
Indications for Use (Describe)
The Indicator Tape is intended to secure items wrapped in synthetic wrap materials until used and to distinguish between processed and unprocessed units through a color change from the start color (pink/magenta) to orange, yellow or lighter. The Indicator Tape is also intended to be used as an external process indicator when applied to synthetic wrap materials and/or Tyvek pouches.
The tape may be used in the following sterilization cycles:
· Lumen, Non Lumen, Fast Non Lumen, Fast, Flexible, and Specialty Cycles of the V-PRO: 1, 1 Plus, maX 2, 60, and s2 Low Temperature Sterilization Systems.
· Standard and Advanced Cycles of the STERRAD® NX Sterilizer with or without ALLClear
· Standard, Flex Scope, Express and DUO Cycles of the STERRAD® 100NX Sterilizer with or without ALLClear.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K240760 510(k) Summary for CELERITY HP Indicator Tape
### Sponsor Facility
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 639-4459
Contact:
Director, Regulatory Affairs Telephone: (440) 392-7437 Fax No: (440) 357-9198 Tony piotrkowski@steris.com
Anthony Piotrkowski
Submission Date: April 10, 2024
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{5}------------------------------------------------
#### 1. Predicate Device
Trade Name: Common/Usual Name: Classification: Classification Name: 510(k) Submitter/Holder: 510(k) Number:
VERIFY VH2O2 Indicator Tape Chemical Indicator Class II Physical/chemical Sterilization Process Indicator STERIS Corporation K231488
#### 2. Device Description
The CELERITY HP Indicator Tape (Indicator Tape) is a ¾' wide crepe paper tape printed with diagonal stripes of a vaporized hydrogen peroxide (VHP) reactive ink. The ink is sealed with a varnish, the function of which is to inhibit removal of the ink via transference or by the adhesive (as the tape is dispensed). The reactive ink meets the performance specifications for a Type 1 process indicator for vaporized hydrogen peroxide as defined in ANSI/AAMI/ISO 11140-1:2014.
#### 3. Indications for Use:
The Indicator Tape is intended to secure items wrapped in synthetic wrap materials until used and to distinguish between processed and unprocessed units through a color change from the start color (pink/magenta) to orange, yellow or lighter. The Indicator Tape is also intended to be used as an external process indicator when applied to synthetic wrap materials and/or Tyvek pouches.
The tape may be used in the following sterilization cycles:
- . Lumen, Non Lumen, Fast Non Lumen, Fast, Flexible, and Specialty Cycles of the V-PRO: 1, 1 Plus, maX, maX 2, 60, and s2 Low Temperature Sterilization Systems.
- . Standard and Advanced Cycles of the STERRAD® NX Sterilizer with or without ALLClear
- . Standard, Flex Scope, Express and DUO Cycles of the STERRAD® 100NX Sterilizer with or without ALLClear.
#### 4. Technological Characteristics
The proposed and predicate devices are chemical indicators in accordance with ISO 11140-1:2014 for use in monitoring Vaporized Hydrogen Peroxide sterilization cycles. The mechanism of action and endpoint are the same and when exposed to the defined processing conditions, the proposed and predicate devices exhibit a visible color change.
{6}------------------------------------------------
| Feature | Proposed: CELERITY HP Indicator Tape | Predicate: VERIFY VH2O2 Indicator<br>Tape (K231488) | Comparison |
|-----------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use<br>/ Indications<br>for Use | The Indicator Tape is intended to secure items<br>wrapped in synthetic wrap materials until used<br>and to distinguish between processed and<br>unprocessed units through a color change<br>from the start color (pink/magenta) to orange,<br>yellow or lighter. The Indicator Tape is also<br>intended to be used as an external process<br>indicator when applied to synthetic wrap<br>materials and/or Tyvek pouches. | The Indicator Tape is intended to secure items<br>wrapped in synthetic wrap materials until used<br>and to distinguish between processed and<br>unprocessed units through a color change<br>from the start color (pink) to peach, yellow or<br>lighter. The Indicator Tape is also intended to<br>be used as an external process indicator when<br>applied to synthetic wrap materials and/or<br>Tyvek pouches. | Similar, a slight<br>change to the<br>start and end<br>color descriptor<br>to match color<br>description of<br>other Celerity<br>chemical<br>indicators and |
| | The tape may be used in the following<br>sterilization cycles:<br>• Lumen, Non Lumen, Fast Non Lumen,<br>Fast, Flexible, and Specialty Cycles of the<br>V-PRO: 1, 1 Plus, maX, maX 2, 60, and<br>s2 Low Temperature Sterilization<br>Systems.<br>• Standard and Advanced Cycles of the<br>STERRAD® NX Sterilizer with or<br>without ALLClear<br>• Standard, Flex Scope, Express and DUO<br>Cycles of the STERRAD® 100NX<br>Sterilizer with or without ALLClear | The tape may be used in the following<br>sterilization cycles:<br>• Lumen, Non Lumen, Fast Non Lumen,<br>Fast, Flexible, and Specialty Cycles of the<br>V-PRO: 1, 1 Plus, maX, maX 2, 60, and<br>s2 Low Temperature Sterilization<br>Systems.<br>• STERRAD® 100S Sterilizer (Default<br>Cycle)<br>• Standard and Advanced Cycles of the<br>STERRAD® NX Sterilizer with or<br>without ALLClear<br>• Standard, Flex Scope, Express and DUO<br>Cycles of the STERRAD® 100NX<br>Sterilizer with or without ALLClear | removal of the<br>STERRAD<br>100S Default<br>Cycle |
| Chemical<br>Indicator<br>Agent and<br>Performance<br>Specification | The indicator agent is a vaporized hydrogen<br>peroxide reactive ink that meets the<br>requirement specified in ANSI/AAMI/ISO<br>11140-1:2014 for a Type 1 Process Indicator<br>for a vaporized hydrogen peroxide process | The indicator agent is a vaporized hydrogen<br>peroxide reactive ink that meets the<br>requirement specified in ANSI/AAMI/ISO<br>11140-1:2014 for a Type 1 Process Indicator<br>for a vaporized hydrogen peroxide process | Identical |
| Device<br>Design | Tape: 3/4" wide by 60 yards long crepe<br>(masking) tape which is wound around a 3"<br>core. A hydrogen peroxide reactive ink is laid<br>down on the non-adhesive surface. The ink is<br>protected from transfer to the adhesive via a<br>coating. | Tape: 3/4" wide by 60 yards long crepe<br>(masking) tape which is wound around a 3"<br>core. A hydrogen peroxide reactive ink is laid<br>down on the non-adhesive surface. The ink is<br>protected from transfer to the adhesive via a<br>coating. | Identical |
| Shelf Life | 5 months - real-time testing is ongoing | Tape 24 months | real-time<br>testing is<br>ongoing |
| Performance<br>Limitations<br>(taken from<br>instructions<br>for use) | Do not overlap the tape onto itself as this may<br>prevent the underlying layer from being<br>appropriately exposed to hydrogen peroxide<br>resulting in a failed indicator response. | Do not overlap the tape onto itself as this may<br>prevent the underlying layer from being<br>appropriately exposed to hydrogen peroxide<br>resulting in a failed indicator response. | Identical |
# Table 1 Comparison of CI Physical Description and Technological Properties
{7}------------------------------------------------
## STERIS Traditional 510(k) PREMARKET NOTIFICATION K240760 CELERITY HP Indicator Tape
#### 5. Performance Testing
Performance testing was completed using well established methods such as those described in ISO 11140-1 and those used to verify the predicate design.
| Testing | Acceptance Criteria | Study Result |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Simulated Use<br>Testing | • Complete color change in full cycle<br>• Incomplete color change in partial cycle<br>• Remain adhered after worst-case cycle | • Complete color change in full cycle<br>• Incomplete color change in partial cycle<br>• Remains adhered after worst-case cycle |
| ISO 11140-1 Testing | Meets requirement of Type 1 VHP Indicator | Meets requirement of Type 1 VHP Indicator |
| Light Stability | Ink performs as intended when exposed to<br>light for labeled period | Ink performs as intended when exposed to<br>light for labeled period |
| Endpoint Stability | Endpoint color remains stable for labeled<br>period | Endpoint color remains stable for labeled<br>period |
| Shelf Life | Meets all performance requirements at<br>labeled shelf-life | Meets all performance requirements at labeled<br>shelf-life |
| Table 2. Verification Results Summary | |
|---------------------------------------|--|
|---------------------------------------|--|
The results of the performed testing demonstrate that the CELERITY HP Indicator Tape performs as intended.
#### 6. Conclusion
The conclusion drawn from the non-clinical testing demonstrates that the subject device, CELERITY HP Indicator Tape, submitted under K240760, is as safe, as effective, and performs as well as the legally marketed predicate device cleared under K231488.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.