Amydi-med Disposable Non-invasive EEG Sensor is applied directly to an adults/pediatrics patient's skin to record physiological signals (e.g., the electroencephalogram, EEG) in the healthcare facility / hospital.
Device Story
Disposable, non-invasive EEG sensor; applied to patient skin (forehead, temporal, above eyebrow, mastoid process) to record physiological signals (EEG). Device consists of flexible PCB with 3-6 electrodes, polyethylene base pad with pressure-sensitive adhesive, nylon tine disks, polyurethane foam disks, and aqueous gel. Used in healthcare facilities/hospitals by neurologists or clinicians. Provides interface between patient skin and EEG monitoring equipment. Benefits include ease of use and accurate electrode placement for signal acquisition. Single-patient use; non-sterile; maximum 24-hour duration.
Clinical Evidence
Bench testing only. Biocompatibility testing per ISO 10993-1, -5, -10 (cytotoxicity, irritation, sensitization). Electrical performance testing per ANSI/AAMI EC12:2000 (R2015). Adhesive performance testing per IEC 60601-2-2 (pull, conformability, fluid tolerance). Shelf life validated via accelerated aging (60°C, 46 days) and real-time testing to support 1-year duration.
Technological Characteristics
Flexible PCB with Ag/AgCl electrodes; polyethylene base pad with medical-grade pressure-sensitive adhesive; nylon tine disks; polyurethane foam disks; aqueous gel (water, NaCl, gums, glycerin, preservatives); polycarbonate cable connector. Non-sterile. 3-6 electrodes. Complies with ANSI/AAMI EC12:2000 and IEC 60601-2-2.
Indications for Use
Indicated for adult and pediatric patients in healthcare facilities/hospitals for recording physiological signals, specifically EEG.
Regulatory Classification
Identification
A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 19, 2024
Shenzhen Amydi-med Electrics Tech Co., Ltd. Dave Yungvirt CEO Third Party Review Group, LLC 25 Independence Blvd Warren, New Jersey 07059
Re: K240668
Trade/Device Name: Amydi-med Disposable Non-invasive EEG electrodes Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous electrode Regulatory Class: Class II Product Code: GXY Dated: August 21, 2023 Received: March 8, 2024
Dear Dave Yungvirt:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Tushar Bansal -S
for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Rehabilitation Devices
{2}------------------------------------------------
OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240668
#### Device Name
Amydi-med Disposable Non-invasive EEG Sensor
(Models: AMD-DE-AF0001-1, AMD-DE-AF0002-1, AMD-DE-AF0004-1, AMD-DE-AF0004-1, AMD-DE-AF0005-1, AMD-DE-AF0006-1,AMD-DE-AF0007-1, AMD-DE-PF0001-1, AMD-DE-PF0004-1)
### Indications for Use (Describe)
Amydi-med Disposable Non-invasive EEG Sensor is applied directly to an adults/pediatrics patient's skin to record physiological signals (e.g., the electroencephalogram, EEG) in the healthcare facility / hospital.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary K240668
The 510(k) summary is submitted in accordance with the requirements of 21 CFR 807.92.
### 1. SUBMITTER
Shenzhen Amydi-med Electrics Tech Co., Ltd.
601, No.15 Building, Xianan No. 3 Industrial Zone, Shangcun, Gongming Street, Guangming District,518106 Shenzhen, Guangdong, PEOPLE'S REPUBLIC OF CHINA
| Contact Person: | Cuifang Yuan (Vice General Manager) |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Email | 2810076565@qq.com |
| Tel: | +86-755-23408983 |
| Primary Correspondent: | Maple Chen (RA Consultant) |
| Email: | maple@gvmmedical.com |
| Tel: | 610-451-7988 |
| Date Prepared: | March 14, 2024 |
| 2. DEVICE | |
| Name of Device: | Amydi-med Disposable Non-invasive EEG Sensor<br>(Models: AMD-DE-AF0001-1, AMD-DE-AF0002-1, AMD-DE-AF0003-1, AMD-DE-AF0004-1, AMD-DE-AF0005-1, AMD-DE-AF0006-1, AMD-DE-AF0007-1, AMD-DE-PF0001-1, AMD-DE-PF0004-1) |
| Common Name: | EEG Sensor |
| Classification Name: | Cutaneous electrode (21 CFR 882.1320) |
| Regulatory Class: | II |
| Product Code: | GXY |
| 3. PREDICATE DEVICE | |
Name of Device: Disposable Non-invasive EEG Sensor
{5}------------------------------------------------
| (Models: B-BIS-6A, B-BIS-6A-02, B-BIS-6A-<br>01, B-BIS-6A-03, B-BIS-5A, B-BIS-5A-01, B-<br>BIS-4A, B-BIS-4A-01, B-BIS-4P, B-BIS-4P-01,<br>B-BIS-3A, B-BIS-3A-02, B-BIS-3A-01, B-BIS-<br>3A-03, B-BIS3AL, B-BIS-3AL-01, B-BIS-3AR,<br>B-BIS-3AR-01) | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| 510(k) Number: | K220448 |
| Submitter of Predicate Device: | Shenzhen Med-link Electronics Tech Co., Ltd. |
### 4. DEVICE DESCRIPTION
Amydi-med Disposable Non-invasive EEG Sensor is a low impedance, single patient use, disposable, non-sterile provided sensor that is designed for application to the sites including forehead, temporal, above eyebrow, and mastoid process. It is designed to provide ease of use and accurate electrode placement.
The main components of this sensor are the flexible PCB with electrodes, polyethylene base pad with medical grade pressure sensitive adhesive on its double sides, nylon plastic tine disks, polyurethane foam disks, aqueous gel ( , and cable connector. Each sensor has three (3) to six (6) electrodes on the flexible PCB for different application sites.
## 5. INDICATIONS FOR USE
Amydi-med Disposable Non-invasive EEG Sensor is applied directly to an adults/pediatrics patient's skin to record physiological signals (e.g., the electroencephalogram, EEG) in the healthcare facility/hospital.
# 6. COMPARISON OF THE SUBJECT DEVICE WITH THE PREDICATE DEVICE
A comparison is performed between the subject device and predicate device. The result is summarized in Table 1.
Table 1 Comparison of the subject device and predicate device
| Items | Subject Device | Predicate Device | Discussion |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|
| 510(k) Number | K240668 | K220448 | N/A |
| 510(K)<br>Submitter | Shenzhen Amydi-med<br>Electrics Tech Co., Ltd. | Shenzhen Med-link<br>Electronics Tech Co.,<br>Ltd. | N/A |
| Classification /<br>Regulation | Class II, 21 CFR 882.1320 | Class II, 21 CFR<br>882.1320 | Same |
| Product Code | GXY | GXY | Same |
| Common<br>Name | EEG Sensor | EEG sensor | Same |
| Indications<br>for Use | Amydi-med Disposable<br>Non-invasive EEG Sensor<br>is applied directly to an<br>Adults/pediatrics patient's<br>skin to record physiological<br>signals (e.g., the<br>electroencephalogram,<br>EEG) in the healthcare<br>facility/hospital. | Disposable Non-<br>invasive EEG Sensor is<br>applied directly to the<br>patient's skin to enable<br>recordings of<br>electrophysiological<br>(such as EEG) signals. | Similar, Note 1 |
| Environment<br>of Use | Prescription | Prescription | Same |
| Target User | Neurologists | N/A | Similar, Note 1 |
| Target Patient | Adults and pediatric | Adult and pediatric | Similar, Note 1 |
| Application<br>Sites | Forehead/temporal/above<br>eyebrow/mastoid process | Forehead/temporal/abo<br>ve eyebrow/mastoid<br>process | Same |
| Material | Flexible printed circuit<br>board, electrodes screen-<br>printed with cured<br>Ag/AgCl and dielectric ink,<br>polyethylene base pad with<br>medical grade pressure<br>sensitive adhesive, nylon<br>plastic tine disks,<br>polyurethane foam disks,<br>aqueous gel (Water,<br>Sodium Chloride, Gum<br>Acacia, Guar Gum,<br>Xanthan Gum, Potassium<br>Bitartrate, Glycerin,<br>Methylparaben and<br>Propylparaben), and cable<br>connector (polycarbonate). | Flexible printed circuit<br>board, 3M Foam Tape<br>(Silicone coated paper<br>liner, Foam backing,<br>Acrylate Adhesive),<br>AgCl electrode, and<br>conductive gel (a<br>mixture of the<br>following components:<br>Water, Sodium<br>Chloride, Gum Acacia,<br>Guar Gum, and<br>Xanthan Gum,<br>Potassium Bitartrate,<br>Glycerin,<br>Methylparaben and<br>Propylparaben.) | Note 2 |
| Number of<br>Electrodes | 3/4/6 | 3/4/5/6 | Note 3 |
| Duration of<br>Use | Maximum 24 hours | Maximum 24 hours | Same |
| Usage | Disposable | Disposable | Same |
| How to Be<br>Supplied | Non-Sterile | Non-Sterile | Same |
| Biocompatibi<br>lity<br>Performance | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | Same |
| Electrical<br>Safety | Not Tested | Complies with<br>ANSI/AAMI | Note 4 |
| | | ES60601-1 | |
| Electrical<br>Performance | Comply with the<br>ANSI/AAMI EC12: 2000<br>(R2015) Disposable ECG<br>Electrodes. | Complies with<br>ANSI/AAMI EC 12 | Same |
| Adhesive<br>Performance | IEC 60601-2-2 Ed. 6.0<br>b:2017 Medical Electrical<br>Equipment – Part 2-2:<br>Particular Requirements for<br>The Basic Safety and<br>Essential Performance of<br>High Frequency Surgical<br>Equipment and High<br>Frequency Surgical<br>Accessories, (a) Pull Test,<br>(b) Conformability Test, (c)<br>Fluid Tolerance Test. | Not publicly available | Note 5 |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
Note 1. The expression is different, but the facts are the same.
Note 2. The materials and construction are the same. The composition of the materials is the same, though the brand of materials may be different. It will not raise new safety or effective issues. However, a Biocompatibility test was conducted and demonstrated the subject device's safety.
Note 3. The electrodes number of the subject device is covered by the predicate scope. No risks or safety issues will be raised.
Note 4. The requirement is not applied to Disposable Non-invasive EEG Electrodes as per the guidance document titled "Cutaneous Electrodes for Recording Purposes – Performance Criteria for Safety and Performance Based Pathway" (issued on August 14, 2020).
Note 5. The subject device passed one more test than the predicate. No safety or risks issue raised.
{9}------------------------------------------------
Summary: From the comparison table, the subject device has the same intended use, constructions, materials, and technological characteristics as the predicate device.
### 7. SAFETY AND PERFORMANCE DATA
According to the Non-invasive EEG Sensor specific guidance document titled "Cutaneous Electrodes for Recording Purposes - Performance Criteria for Safety and Performance Based Pathway" (issued on August 14, 2020), we conducted the following safety and electrical performance assessment using the specified FDA recognized standards (including methods and acceptable criteria).
#### Biocompatibility Performance i.
The biocompatibility evaluation for the subject device was conducted in accordance with ANSI/AAMI/ISO 10993-1: 2018 Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process. The subject device is considered to be surface medical device contacting with intact skin and the contact duration is less than 24 hours. The biocompatibility evaluation endpoints are:
#### • Cytotoxicity
#### Irritation and skin sensitization •
The subject device successfully passed the above tests.
#### ii. Electrical Performance
Electrical testing was conducted on the subject device. The subject device complies with the ANSI/AAMI EC12: 2000 (R2015) Disposable ECG Electrodes.
#### iii. Adhesive Performance
The subject device has been designed and validated to satisfy the adhesive requirements per IEC 60601-2-2 Ed. 6.0 b:2017 Medical Electrical Equipment - Part 2-2: Particular Requirements for The Basic Safety and Essential Performance of High Frequency Surgical Equipment and High Frequency Surgical Accessories, (a) Pull Test, (b) Conformability Test, (c) Fluid Tolerance Test.
{10}------------------------------------------------
#### iv. Shelf Life
Shelf life testing were completed on the subject device following Accelerated aging test (60°C, 46 days) to support the proposed shelf life (1 year). The real time stability testing also demonstrates the shelf life of 1 year.
Summary: The subject device complies with the FDA recognized consensus standards for safety and performance criteria.
## 8. CONCLUSION
The subject device has the same intended use, constructions, materials, and technological characteristics as the predicate device. The subject device complies with the FDA recognized consensus standards for safety and performance criteria as required in the Non-invasive EEG Sensor specific guidance document titled "Cutaneous Electrodes for Recording Purposes – Performance Criteria for Safety and Performance Based Pathway" (issued on August 14, 2020).
Based on the safety and performance-based pathway, it can be concluded that the subject device is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.