The non-contact forehead thermometer is an electronic thermometer using an infrared detector (thermopile detector) to detect body temperature from forehead center in people of all ages.
Device Story
Non-contact infrared thermometer (models TH48FE, TH09F, THD2FE) measures infrared energy radiated from forehead; IR sensor collects energy via lens; application-specific integrated circuitry converts signal to oral temperature value. Device intended for lay person use in home or clinical settings. User initiates measurement via buttons; results displayed on LCD. Provides non-invasive temperature monitoring; aids in clinical decision-making regarding fever management. Powered by AAA or CR2032 batteries.
Clinical Evidence
Clinical accuracy validation conducted per ISO 80601-2-56 on 113 subjects across age groups: A1 (1-3 months, n=15), A2 (3 months-1 year, n=21), B (1-5 years, n=38), and C (>5 years, n=39). Results confirmed compliance with ISO 80601-2-56 requirements.
Technological Characteristics
Thermopile IR sensor; ABS plastic housing; LCD display; powered by 2x1.5V AAA or 1x3.0V CR2032 batteries. Complies with ISO 10993-1/5/10/23 (biocompatibility), ASTM 1965-98, ISO 80601-2-56 (performance), and IEC 60601-1/1-2/1-11 (safety/EMC).
Indications for Use
Indicated for body temperature measurement from the forehead center in people of all ages.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
RII, Non-Contact Clinical Thermometer, Model THB0F (K121428)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. Food & Drug Administration".
June 5, 2024
Radiant Innovation Inc. Lynn Chen OA Manager 1F, No.3 Industrial E. 9th Rd., Science-Based Industrial Park HsinChu, 300 Taiwan
Re: K240648
Trade/Device Name: Non-contact Forehead Thermomete (TH48FE, TH09F, THD2FE) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: February 26, 2024 Received: March 7, 2024
Dear Lynn Chen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
David Walloschek
David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn,
{2}------------------------------------------------
General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K240648
Device Name
Non-contact Forehead Thermometer (TH48FE, TH09F, THD2FE)
Indications for Use (Describe)
The non-contact forehead thermometer is an electronic thermometer using an infrared detector (thermopile detector) to detect body temperature from forehead center in people of all ages.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains a logo for Radiant Innovation. The logo is in blue and white, with the word "RADIANT" in large, bold letters. Above the word "RADIANT" is the word "INNOVATION" in smaller letters. Below the logo is the text "RADIANT INNOVATION INC." and the website address "http://www.radiantek.com.tw".
# K240648 – 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
## 1. Submitter's Identification:
Radiant Innovation Inc., 1F, No.3 Industrial E. 9th Rd., Science-Based Industrial Park, HsinChu, Taiwan Name of contact person: Ms. Lynn Chen TEL: +886 3 6111666 Ext. 8123 FAX: +886 3 5670089 E-mail: lynnchen@radiantek.com.tw Date Summary Prepared: Jun/05/2024
### 2. Device:
Device Trade Name: Non-contact Forehead Thermometer (TH48FE, TH09F, THD2FE) Classification Name: Thermometer, Electronic, Clinical Device Classification: Class II Regulation Number: 21CFR 880.2910 Device Panel: General Hospital Product Code(s): FLL
### 3. Predicate Device :
510(k) #K121428 Device Trade Name: RII, Non-Contact Clinical Thermometer, Model THB0F Classification Name: Thermometer, Electronic, Clinical Device Classification: Class II Regulation Number: 21CFR 880.2910 Device Panel: General Hospital Product Code(s): FLL
### 4. Device Description:
The thermometer (Mode: TH48FE, TH09F, THD2FE) are electronic thermometer using an IR sensor to measure infrared energy radiated from the forehead. This energy is collected through the lens and converted to an oral temperature value.
The thermometer (Mode: TH48FE, TH09F, THD2FE), consists of the following parts:
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains a logo for Radiant Innovation Inc. The logo features the word "RADIANT" in white letters on a blue rectangle, with the word "INNOVATION" in smaller letters above it. Below the rectangle, the words "RADIANT INNOVATION INC." are printed in black, followed by the URL "http://www.radiantek.com.tw".
- (1) IR sensor with a built-in ambient temperature sensor (Function: Receives infrared signals radiated by temperature)
- (2) Application-Specific Integrated Circuitry include software (Function: Calculate the infrared signal from the forehead and convert it into oral temperature)
- (3) LCD display (Function: Display measurement results)
- (4) 2 buttons (Star button, Scan button), 1 button for TH09F (Function: Start booting and measuring actions)
- (5) 2x1.5V AAA dry batteries, 1 x 3.0V CR2032 battery for TH09F) (Function: Provide product power)
### 5. Indications for Use:
The non-contact forehead thermometer is an electronic thermometer using an infrared detector (thermopile detector) to detect body temperature from forehead center in people of all ages.
| Characteristics | Subject device<br>(TH48FE/TH09F/THD2FE) | Predicate device (THB0F) | Comment |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| 510(k)# | K240648 | K121428 | / |
| Manufacturer | Radiant Innovation Inc. | Radiant Innovation Inc. | Same |
| Indications for use | The non-contact forehead<br>thermometer is an electronic<br>thermometer using an infrared<br>detector (thermopile detector)<br>to detect body temperature<br>from forehead center in<br>people of all ages. | The Non-contact Clinical<br>Thermometer, Model THB0F<br>is an infrared thermometer<br>intended for the intermittent<br>measurement of human body<br>temperature in people of all<br>ages. | Same |
| Intended user | Lay person | Lay person | Same |
| Measuring Range | Forehead mode:<br>82.4~109.4°F(28~43°C) for<br>TH48F/TH09F<br>93.2~108°F(34~42.2°C) for<br>THD2Fe<br>Surface mode:<br>-7.6~176°F(-22~80°C) for<br>TH48FE/THD2FE | Forehead mode:<br>93.2~108°F(34~42.2°C)<br>Surface mode:<br>-7.6~176°F(-22~80°C) | Different 1 |
| Accuracy | Forehead mode:<br>$\pm$ 0.4 °F (0.2°C) within<br>95~107.6°F (35~42°C), $\pm$ 0.5<br>°F (0.3°C) for other range.<br>Surface mode (for TH48FE/<br>THD2FE):<br>0.5°F ( $\pm$ 0.3°C) within | Forehead mode:<br>$\pm$ 0.4°F(0.2°C) within<br>96.8~102°F(36~39°C), $\pm$ 0.5<br>°F(0.3°C) for other range<br>Surface mode:<br>$\pm$ 4% of reading or $\pm$ 4°F(2°C)<br>whichever is greater | |
#### 6. Comparison of technological characteristics with the predicate device:
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image features the word "RADIANT" in large, bold, white letters against a blue background. Above the word "RADIANT" is the word "INNOVATION" in smaller, white letters. The overall design is simple and modern, with a focus on the company name and its association with innovation.
RADIANT INNOVATION INC.
| | 93.2~109.4°F (34~43°C),<br>others ±4% or ±4°F (2°C)<br>whichever is greater. | | |
|-------------------------------|------------------------------------------------------------------------------|------------------------------------------------|-------------|
| Display Resolution | 0.1°F(0.1°C) | 0.1°F(0.1°C) | Same |
| Memory | 25 sets for TH48FE/THD2FE<br>1 set for TH09F | 60 sets | Different 2 |
| | | | |
| Scale Selection | °F/°C | °F/°C | Same |
| Display Type | LCD | LCD | Same |
| Sensor Type | Thermopile | Thermopile | Same |
| Case | ABS | ABS | Same |
| Power Source | AAA(1.5V)*2 for<br>TH48FE/THD2FE<br>CR2032(3V)*1 for TH09F | AAA(1.5V)*2 | Different 3 |
| | | | |
| | | | |
| Operating<br>Condition | 50~104°F(10~40°C) | 50~104°F(10~40°C) | Same |
| Biocompatibility | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | Same |
| | | | |
| | | | |
| Performance | ASTM 1965-98<br>ISO 80601-2-56 | ASTM 1965-98<br>ISO 80601-2-56 | Same |
| | | | |
| Electric Safety<br>and<br>EMC | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11 | Same |
| | | | |
| | | | |
#### Justification for the differences:
- 1) Different Measurement range/ Accuracy:
The measurement range and accuracy of the device under evaluation is different from the marked equivalent device. While the device under evaluation have been tested and validated according to standard ISO 80601-2-56 including accuracy and the measuring range of the device under evaluation meet the accuracy and the minimum rated output range of clinical thermometer requirement, therefore, such difference will not affect the safety and performance of the device under evaluation.
- 2) Different of Memory:
The memory function of subject devices has been validated in accordance IEC 62304 and FDA Guidance for the Content of Premarket Submissions for Software Contained. The difference will not raise safety or effectiveness issues on the subject devices.
- 3) Different of Power Supply:
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Radiant Innovation Inc. The logo is in blue and white, with the word "RADIANT" in large, bold letters. Above the word "RADIANT" is the word "INNOVATION" in smaller letters. Below the logo is the text "RADIANT INNOVATION INC." and the website address "http://www.radiantek.com.tw".
The power supply of TH09F is 1 x CR2032 battery that different from other models, all devices have been proven to be safe and effective based on the IEC 60601-1 and IEC 60601-1-2 testing.
The Non-contact Forehead Thermometer, Model TH48FE, TH09F, THD2FE have the similar characteristics as the predicate device. Moreover, the subject devices demonstrate product safety by successful completion of testing to the IEC 60601-1 standard and electromagnetic standard IEC 60601-1-2. The performance test demonstrates the TH48FE, TH09F, THD2FE meet the ISO 80601-2-56 standard and concludes that any difference in their characteristics do not raise any safety and effectiveness issues. Thus, the Non-contact Forehead Thermometer, Model TH48FE, TH09F, THD2FE are substantially equivalent to the predicate device.
## 7. Summary of non-clinical testing (Performance testing-bench):
The following performance data were provided to verify that the subject device met all design specifications and provided support of the substantial equivalence determination.
- Risk Analysis developed in according with ISO 14971:2019 -
- The recognized consensus standards for safety of medical electrical equipment: IEC 60601--1, IEC 60601-1-11 for safety and IEC 60601-1-2 for electromagnetic compatibility are complied.
- Biocompatibility Evaluation for patient contacting components: ISO 10993-5, ISO 10993--10 and ISO 10993-23 are complied.
- Bench testing was conducted on the thermometer device. The system complies with the ISO -80601-2-56.
- -Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions". The software for this device was considered as a "Basic" documentation level.
### 8. Summary of clinical Accuracy Validation Test:
Clinical testing is conducted per ISO 80601-2-56 Clause 201.102 Clinical Accuracy Validation. The Clinical tests were evaluated on 113 subjects and thermometer was evaluated in three age groups including subgroup A1 and A2: A1-one month up to three months and A2-three months to one year, B-older than one years and younger than five years and C-older than five years old. Subjects selected for each model in this clinical trial are: A1) group 15, A2) group 21, B) group 38 and C) group 39. Total number of subjects is 113.
The test report showed the clinical performance of the subject devices complied with the
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for Radiant Innovation Inc. The logo features the word "RADIANT" in white letters on a blue background, with the word "INNOVATION" above it in smaller letters. Below the logo is the text "RADIANT INNOVATION INC." and the website address "http://www.radiantek.com.tw".
requirement of ISO 80601-2-56.
## 9. Conclusions:
Based on comparison of the Indications for use and technological characteristics applicable safety standards, verification and validation testing, the differences between Non-contact forehead thermometer (TH48FE, TH09F, THD2FE) and predicate device do not raise new or different questions of safety and effectiveness. Thus, the Non-contact Forehead Thermometer, Model TH48FE, TH09F, THD2FE are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.