Total hip arthroplasty is intended to replace the damaged hip joint articulation in patients at a high risk of hip dislocation, where there is evidence of sufficient sound bone to seat and support the components.
Device Story
Modular total hip prosthesis component; replaces natural hip articular surface. Device consists of UHMWPE liner locking into porous-coated titanium alloy (Ti-Al6-V4) shell. Provides >180-degree femoral head capture via titanium constraining ring; mechanically constrains femoral head. Available in lateralized neutral or face-changing orientations; compatible with 28, 32, 36mm metal femoral heads and 40, 44mm unipolar/bipolar/modular femoral heads. Used in orthopedic surgery for total hip replacement. Surgeon implants device to stabilize hip joint in high-risk patients. Benefits include reduced risk of dislocation. Current submission updates labeling for MRI compatibility and standardizes symbols/labeling.
Clinical Evidence
Bench testing only. MRI safety evaluation performed per ASTM F2052-21 (displacement force), ASTM F2213-17 (torque), ASTM F2182-19 (RF heating), and ASTM F2119-07 (image artifacts). Data demonstrates conditional safety in MR environment.
Indicated for primary or revision total hip arthroplasty patients at high risk of dislocation due to history of prior dislocation, bone loss, joint/soft tissue laxity, neuromuscular disease, or intraoperative instability, where other constrained options have been considered. For use with PINNACLE Acetabular Cup in cementless application.
Regulatory Classification
Identification
A hip joint metal/polymer constrained cemented or uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device prevents dislocation in more than one anatomic plane and has components that are linked together. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular component made of ultra-high-molecular-weight polyethylene with or without a metal shell, made of alloys, such as cobalt-chromium-molybdenum and titanium alloys. This generic type of device is intended for use with or without bone cement (§ 888.3027).
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Hip Joint Metal/Polymer Constrained Cemented or Uncemented Prosthesis.”
DePuy Hip Portfolio MRI Bundled Traditional 510(k) (K231873)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
April 4, 2024
Depuy Ireland UC Jennifer Hill Regulatory Affairs Project Leader Loughbeg, Ringaskiddy County Cork, Ireland
Re: K240639
Trade/Device Name: PINNACLE 114 Constrained Acetabular Liners Regulation Number: 21 CFR 888.3310 Regulation Name: Hip Joint Metal/Polymer Constrained Or Uncemented Prosthesis Regulatory Class: Class II Product Code: KWZ Dated: March 6, 2024 Received: March 6, 2024
Dear Jennifer Hill:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
## Limin Sun - S
Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K240639
Device Name
PINNACLETM Constrained Acetabular Liner
Indications for Use (Describe)
The PINNACLETM Constrained Acetabular Liner is indicated for use as a component of a total hip prosthesis in primary or revision patients at high risk of hip dislocation due to a history or prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease or intraoperative instability and for whom all other options of constrained acetabular components have been considered.
The PINNACLETM Constrained Acetabular Liner is indicated for use with the PINNACLETM Acetabular Cup in cementless application.
Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| Submitter Information | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | DePuy Ireland UC |
| Address | Loughbeg<br>Ringaskiddy<br>Co. Cork<br>Ireland |
| Phone number | (+44) 7834974433 |
| Fax number | N/A |
| Establishment Registration Number | 3015516266 |
| Name of contact person | Jennifer Hill |
| Date prepared | 06th March 2024 |
| Name of device | |
| Trade or proprietary name | PINNACLETM Constrained Acetabular Liners |
| Common or usual name | Acetabular Cup Prosthesis |
| Classification name | 21 CFR 888.3310: Hip joint metal/polymer constrained cemented or uncemented prosthesis. |
| Class | II |
| Classification panel | 87 Orthopedics |
| Regulation | Class II – 21 CFR 888.3310 |
| Product Code(s) | KWZ: Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer |
| Legally marketed device(s) to which<br>equivalence is claimed | K043058/K052079/K071117 - PINNACLE™ Constrained Acetabular Liners |
| Reason for 510(k) submission | DePuy Ireland UC has compiled a Special 510(k) Premarket Notification to expand the labeling for<br>the DePuy PINNACLE™ Constrained Acetabular Liners, to provide updated information regarding<br>the MRI compatibility of the devices. A program of MRI Safety Evaluation testing has been carried<br>out with full test data present and reviewed in the FDA cleared submission 'K231873 - DePuy Hip<br>Portfolio MRI Bundled Traditional 510(k)'. Appropriate amends have been made to the product<br>labels and Instructions for Use (IFU), along with the provision of new MR Safety Cards for patients. |
| | At the same time as making these labeling updates in relation to MRI compatibility, DePuy will take<br>the opportunity to standardize language and symbols used across its legacy devices for consistency,<br>including the use of new internationally recognized symbols with corresponding updated symbols<br>glossaries. |
| | Additionally, in order to harmonize with International and EU labeling requirements, the labeling<br>of subject devices which are sold Internationally will be further updated to also include a new<br>Implant Card, Implant Card Guide, and Implant Card Label. These are not US requirements and can<br>be discarded by the end user. The labeling will also be updated to include a material code with<br>corresponding material descriptions. |
| | There is no change to the indications, intended use, safety, fit, form or technological characteristics<br>of the devices. |
| Device description | The PINNACLE™ Constrained Acetabular Liner is part of a modular system designed to replace the<br>natural articular surface of the hip joint in total hip replacement. The liner is manufactured from<br>ultra-high molecular weight polyethylene (UHMWPE), which locks into a porous coated,<br>hemispherical outer shell component manufactured from titanium alloy (Ti-Al6-V4). The liner<br>component articulates with a metal femoral head of an appropriate diameter. |
| | The PINNACLE™ Constrained Acetabular Liner mechanically constrains the femoral head within<br>the inner diameters of the liner by providing greater than 180 degrees femoral head capture<br>combined with a titanium constraining ring which fits over the opening diameter of the liner. The<br>UHMWPE liner is held in the metal shell by means of a titanium locking ring.<br>The PINNACLE™ Constrained Acetabular Liners are UHMWPE acetabular cup liners that are<br>available in a lateralized neutral or lateralized face-changing orientation. The liners have inner<br>diameters compatible with standard metal femoral heads sized 28, 32, and 36mm, as well as<br>larger diameter unipolar, Self-Centering (bipolar) and Modular M femoral heads sized 40 and<br>44mm. The outer diameters (OD) are geometrically the same as other Pinnacle Acetabular Liners,<br>in a 48mm-76mm size range offering. |
| Intended use of the device | Total hip arthroplasty is intended to replace the damaged hip joint articulation in patients at a<br>high risk of hip dislocation, where there is evidence of sufficient sound bone to seat and support<br>the components. |
| Indications for use | The PINNACLE™ Constrained Acetabular Liner is indicated for use as a component of a total hip<br>prosthesis in primary or revision patients at high risk of hip dislocation due to a history of prior<br>dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease or intraoperative instability<br>and for whom all other options to constrained acetabular components have been considered.<br>The PINNACLE™ Constrained Acetabular Liner is indicated for use with the PINNACLE™<br>Acetabular Cup in cementless application. |
| Substantial Equivalence | DePuy Ireland UC has compiled a Special 510(k) Premarket Notification to expand the labeling for<br>the DePuy PINNACLE™ Constrained Acetabular Liners, to provide updated information regarding<br>the MRI compatibility of the devices. A program of MRI Safety Evaluation testing has been carried<br>out with full test data present and reviewed in the FDA cleared submission 'K231873 - DePuy Hip<br>Portfolio MRI Bundled Traditional 510(k)'. Appropriate amends have been made to the product<br>labels and Instructions for Use (IFU), along with the provision of new MR Safety Cards for patients. |
| At the same time as making these labeling updates in relation to MRI compatibility, DePuy will take the opportunity to standardize language and symbols used across its legacy devices for consistency, including the use of new internationally recognized symbols with corresponding updated symbols glossaries. | |
| Additionally, in order to harmonize with International and EU labeling requirements, the labeling of subject devices which are sold Internationally will be further updated to also include a new Implant Card, Implant Card Guide, and Implant Card Label. These are not US requirements and can be discarded by the end user. The labeling will also be updated to include a material code with corresponding material descriptions. | |
| There is no change to the indications, intended use, safety, fit, form or technological characteristics of the devices. | |
| Non-clinical testing is provided to support the conditional safety of the PINNACLETM Constrained Acetabular Liners in the MR environment, including assessment of:<br>Magnetically Induced Displacement Force (ASTM F2052-21) Magnetically Induced Displacement Torque (ASTM F2213-17) Radio Frequency (RF) Heating (ASTM F2182-19) Image Artifacts (ASTM F2119-07) | |
| The non-clinical performance data demonstrate that the Subject Devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy. | |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.