K240620 · Medical Templates AG · LLZ · Mar 25, 2024 · Radiology
Device Facts
Record ID
K240620
Device Name
Cube Navigator
Applicant
Medical Templates AG
Product Code
LLZ · Radiology
Decision Date
Mar 25, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, Real-World Evidence, 3rd-Party Reviewed
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K240620 · Mar 25, 2024
Cube Navigator
Medical Templates AG
Retrospectively collected patient cases
Retrospective patient cases were used to verify the accuracy of the automated registration tool and to validate the needle planning capability by comparing software-suggested coordinates against actual physical puncture coordinates from existing clinical data.
Retrospective validation of registration and needle planning; Retrospective analysis of existing clinical datasets
Patients undergoing CT-guided percutaneous punctures; Sample Size: 5 datasets per cube type (10 total)
Not applicable for this study
Accuracy of automated registration and needle trajectory planning coordinates
Indications for Use
Cube Navigator is a non-invasive software intended for planning the needle trajectory of CT-guided, percutaneous punctures using DICOM-images and a Navigation Cube attached to a patient. It is indicated for the planning the needle trajectory of CT-guided, percutaneous punctures for therapeutic and diagnostic purposes. The software's functionality allows to automatically recognize orientation, position and type of the Navigation Cube present in the DICOM-images and to manually define virtual needle trajectories to a physician-identified target through the Navigation Cube. Based on this planning, Cube Navigator provides physician information for such purpose. Cube Navigator can only be used in combination with a Navigation Cube, specifically, either the Access Cube for instruments 10-20G (article number mt-ac-ct-20, Class 1 exempt, product code GDF) or the Puncture Cube for instruments 18-22G (article number mt-pc-ct-25, Class 1 exempt, product code GDF). Cube Navigator is a standalone software that integrates the CE-marked and FDA-cleared diagnostic Medical Imaging Viewer medDream manufactured by Softneta that provides all DICOM-related functionality like displaying and manipulating DICOM-images. The Cube Navigator is intended for interventional radiologists, who perform CT-guided interventions for therapeutic and diagnostic purposes. The intended use environment is the CT control room, which can be in a hospital or a medical office.
Device Story
Cube Navigator is standalone software for interventional radiologists to plan needle trajectories for CT-guided percutaneous punctures. Input: DICOM images containing a patient-attached Navigation Cube (Access or Puncture Cube). Operation: Software automatically detects orientation, position, and type of Navigation Cube; user manually defines virtual needle trajectory to target. Output: Visualized needle depth and entry points on the Navigation Cube interface. Context: Used in CT control rooms (hospital/office). Benefit: Simplifies needle path planning, reduces human error in manual coordinate calculation, and provides precise trajectory information. Integrates medDream viewer for DICOM manipulation.
Clinical Evidence
Bench testing only. Verification included phantom studies and retrospective patient datasets to validate automated registration accuracy of the Navigation Cube. Needle depth calculation verified by comparing software-suggested coordinates against physical needle trajectories in existing puncture data. Usability study conducted with 5 interventional radiologists confirmed safety and effectiveness. No clinical tests were submitted or relied upon.
Technological Characteristics
Standalone software integrating medDream DICOM viewer. Uses automated image processing to detect Navigation Cube (Access or Puncture variants). Provides virtual needle trajectory planning based on registered cube coordinates. Connectivity: Local or remote DICOM-connection. Software-based; no hardware components other than the separately used Navigation Cubes.
Indications for Use
Indicated for interventional radiologists performing CT-guided percutaneous punctures for therapeutic and diagnostic purposes. Used with Navigation Cube (Access Cube 10-20G or Puncture Cube 18-22G) in CT control rooms.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Medical Templates AG % Shawnnah Monterrey BeanStock Consulting 8885 Rio San Diego Dr. #237 San Diego, CA 92108 USA
Re: K240620
Trade/Device Name: Cube Navigator Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: March 5, 2024 Received: March 5, 2024
Dear Shawnnah Monterrey:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov
March 25, 2024
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Samal for
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT 8B: Division of Radiological Imaging Devices and Electronic Products OHT 8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
### Indications for Use
Submission Number (if known)
K240620
Device Name
Cube Navigator
Indications for Use (Describe)
Cube Navigator is a non-invasive software intended for planning the needle trajectory of CT-quided, percutaneous punctures using DICOM-images and a Navigation Cube attached to a patient. It is indicated for the planning the needle trajectory of CT-guided, percutaneous punctures for therapeutic and diagnostic purposes. The software's functionality allows to automatically recognize orientation, position and type of the Navigation Cube present in the DICOM-images and to manually define virtual needle trajectories to a physician-identified target through the Navigation Cube. Based on this planning, Cube Navigator provides physician information for such purpose.
Cube Navigator can only be used in combination with a Navigation Cube, specifically, either the Access Cube for instruments 10-20G (article number mt-ac-ct-20, Class 1 exempt, product code GDF) or the Puncture Cube for instruments 18-22G (article number mt-pc-ct-25, Class 1 exempt, product code GDF). Cube Navigator is a standalone software that integrates the CE-marked and FDA-cleared diagnostic Medical Imaging Viewer medDream manufactured by Softneta that provides all DICOM-related functionality like displaying and manipulating DICOM-images.
The Cube Navigator is intended for interventional radiologists, who perform CT-guided interventions for therapeutic and diagnostic purposes. The intended use environment is the CT control room, which can be in a hospital or a medical office.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{3}------------------------------------------------
| K240620 | 510(k) Summary | Prepared on: 2024-03-20 |
|---------|----------------|-------------------------|
|---------|----------------|-------------------------|
| Contact Details | 21 CFR 807.92(a)(1) |
|-----------------|---------------------|
|-----------------|---------------------|
| Applicant Name | Medical Templates AG |
|-----------------------------|--------------------------------------|
| Applicant Address | Brunnenwiese 10 Egg 8132 Switzerland |
| Applicant Contact Telephone | +41 43 50 88 59 |
| Applicant Contact | Mr. Joscha Hüttel |
| Applicant Contact Email | jh@medicaltemplates.ch |
| Device Name | 21 CFR 807.92(a)(2) |
|-------------|---------------------|
|-------------|---------------------|
| Device Trade Name | Cube Navigator |
|---------------------|------------------------------------------------|
| Common Name | Medical image management and processing system |
| Classification Name | System, Image Processing, Radiological |
| Regulation Number | 892.2050 |
| Product Code | LLZ |
| Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) |
|------------------------------------|---------------------|
|------------------------------------|---------------------|
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|-------------|----------------------------------------------------------|--------------|
| K141745 | IQQA-BODYIMAGING SOFTWARE | LLZ |
| Device Description Summary | 21 CFR 807.92(a)(4) |
|----------------------------|---------------------|
|----------------------------|---------------------|
Cube Navigator is a non-invasive medical device software used in interventional radiology for planning the needle trajectory of CTguided, percutaneous punctures. Cube Navigator provides planning functionality for needle placement through a separate device, called a Navigation Cube. Cube Navigator allows the user to register the Navigation Cube in a feasible needle traiectory using a virtual needle. Based on this planned trainctor displays needle depth and the enty points on the Navigation Cube in its user interface. The Cube Navigator software is only used for planning the needle trajectory, but requires a CT-scan with a Navigation Cube mounted on a patient for the purposes of visualization.
Navigation Cubes are sterile, disposable needle quides which are attached to the planning scan is acquired. Currently two variants exist: Puncture Cube or Access Cube, both Class 1 exempt, product code GDF. The Navigation Cubes support the use of needle sizes 10-20G (Access Cube) or 18-22G (Puncture Cube), Needle length is dependent on the trained from the displayed needle depth. Cube Navigator detects which Navigation Cube is present in the takes the average of the supported needle diameters for that Navigation Cube when calculating the thickness of the virtual needle.
Cube Navigator is a standalone software that integrates with the CE-marked and FDA-cleared diagnostic Medical Imaging Viewer medDream manufactured by Softneta (Class II, K22232), see Software Architecture Design Cube Navigator for integrated version). Gube Navigator allows the user to access the functionality of medicing images locally or through a remote DICOM-connection, by viewing and manipulating DICOM-images through the user interfaces of the Cube Navigator, as well as accessing the functionality of Cube Navigator itself. The DICOM integration is part of the base software package medDream (see the Statement recarding the DICOM-Standard - Cube Navigator for detailed information about DICOM conformity).
Cube Navigator as well as the Navigation Cubes are manufactured by Medical Templates AG. medDream is manufactured by Softneta and for the described purpose, distributed by Medical Templates AG as bundle together with Cube Navigator. When the user installs Cube Navigator, medDream is also installed, without any additional installation or configuration necessary.
{4}------------------------------------------------
Customer support for Cube Navigator is provided by Medical Templates support for medDream functions accessible through Cube Navigator. Maintenance is provided by Medical Templates AG in the form of regular updates and/or patches. These updates and patches may include updates to medDream if necessary updates for medDream are available. A link to updates or patches. along with instructions on how to install them, will be distributed via email to the user.
## Intended Use/Indications for Use
## 21 CFR 807.92(a)(5)
Cube Navigator is a non-invasive software intended for planning the needle trajectory of CT-guided, perculaneous punctures using DICOM-images and a Navigation Cube attached to a patient. It is indicated for the needle traiectory of CT-quided, perculaneous punctures for therapeutic and diagnostic purposes. The software's functionality allows to automatically recognize orientation, position and type of the Navigation Cube present in the DICON-images and to manually define virtual needle trajectories to a physician-identified target through the Navigation Cube. Based on this planning, Cube Navigator provides physician information for such purpose.
Cube Navigator can only be used in combination Cube, specifically, either the Access Cube for instruments 10-20G (article number mt-ac-ct-20, Class 1 exempt, product code GDF) or the Puncture Cube for instruments 18-22G (article number mt-pcct-25, Class 1 exempt, product code GDF). Cube Navigator is a standalone software that integrates the CE-marked and FDA-cleared diagnostic Medical Imaging Viewer manufactured by Softneta that provides all DICOM-related functionality like displaying and manipulating DICOM-images.
The Cube Navigator is interventional radiologists, who perform CT-guided interventions for therapeutic and diagnostic purposes. The intended use environment is the CT control room, which can be in a hospital of a medical office.
# Indications for Use Comparison
## 21 CFR 807.92(a)(5)
The Cube Navigator has substantially equivalent Intended Use as the predicate device, in that it is a software application that allows the user to define a virtual needle path in the displayed DICOM-images and provides information to the user for treatment planning. In comparison to the Indications For Use of the provides functionality for the planning of needle paths as one of many functions, the Cube Navigator software is limited to be used for such planning of needle paths as described in the Indication for Use Statement. The differences in the Indications for use stem from that limitation. Therefor the Indications for Use do not affect the safety and effectiveness of the device when used as labeled.
Cube Navigator is a non-invasive software intended for planning the needle trajectory of CT-guided, perculaneous punctures using DICOM-images and a Navigation Cube attached to a patient. It is indicated for the needle trajectory of CT-guided, perculaneous punctures for therapeutic and diagnostic purposes. The software's functionality allows to automatically recognize orientation, position and type of the Navigation Cube present in the DICON-images and to manually define virtual needle trajectories to a physician-identified target through the Navigation Cube. Based on this planning, Cube Navigator provides physician information for such purpose.
Cube Navigator can only be used in combination with a Navigation Cube, specifically, either the Access Cube for instruments 10-20G (article number mt-ac-ct-20, Class 1 exempt, product code GDF) or the Puncture Cube for instruments 18-22G (article number mt-pcct-25, Class 1 exempt, product code GDF). Cube Navigator is a standalone software that integrates the CE-marked and FDA-cleared diagnostic Medical Imaging Viewer meduled by Softneta that provides all DICOM-related functionality like displaying and manipulating DICOM-images.
The Cube Navigator is interventional radiologists, who perform CT-guided interventions for therapeutic and diagnostic purposes. The intended use environment is the CT control room, which can be in a hospital or a medical office.
# Technological Comparison
## 21 CFR 807.92(a)(6)
Cube Navigator software and the predicate device, are both software tools that enable the user to plan a virtual needle path in the displayed DICOM-images. While the predicate is an image analysis software application for reviewing body imaging studies that also provides functionality for that purpose, the Cube Navigator provides only functionality for that purpose. To achieve the existing DICOM-Viewer MedDream (Class II, K222320, manufactured by Softneta UAB) on which the planning functionality is based, In comparison to the predicate device functionality of the Cube Navigator is limited to the features relevant for the planning. Therefor this difference in features does not raise different questions of safety and effectiveness. As those features on t present in the Cube Navigator are possibly associated with risks regarding safety and effectiveness, this difference in features rather leaves the Cube Navigator associated with less risks overall in comparison to the predicate device.
Another difference in functional features is that the Cube Navigator provides information based on the user's defined needle path, that helps the user understand through which holes of the virtual needle is currently going and what the corresponding virtual needle is. The information provided is based on the registration performed as a first step in the process of planning and the user's input to define the virtual needle. In principle risks may arise from a faulty
{5}------------------------------------------------
registration, which is a risk that is also relevant for the automated registration for viewing and analyzing multiple timepoint volume datasets, a feature provided by the predicate device. Another source of potential risk is the virtual needle path planning, that is also a feature in the predicate device the user need to manually determine through which holes of the Navigation Cube the current virtual needle depth is. The Cube Navigator provides this information for the user and therefor simplifies this process thereby erasing potential source of error due to human error. Therefor this difference in features does not raise different questions of safety and effectiveness. Cube Navigator has the predicate device and is substantially equivalent in performance. Any technological differences do not raise any new potential safety risk. In conclusion the Cube Navigator software is substantially equivalent to the predicate device.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Verification of automated registration accuracy:Data from phantom studies and retrospectively collected patient cases are used to determine the accuracy of the registration tool to automatically detect and register Navigation of displayed needle depth:In order to verify the needle length in the sidebar a dataset is loaded, the cube is registered and the needle traiectory is planned. The calculated needle-depth is verfied by measuring the tip of the needle to the corresponding plate in the cube with the ine measurement tool provided by MedDream (FDA certified product). By comparing the two values to each other the error of the indicated needle length is determined .Side by side comparison to interactively define virtual needle path for planning The needle planning capability of the software end is verified by scanning an existing puncture, registering the cube, planning the virtual needle in the same trajectory as the physical one and comparing if the suggested coordinates for the needle plan by the software are equal to the actual coordinates used for the physical needle. This provides the advantage that it can be done retrospectively on existing data and excludes effects not related to the software such as actual puncturing by the clinician. For each cube 5 datasets from different tests or studies have been selected and the above conducted process has shown that for all cases the needle coordinates suggested by the software coincide with the actual coordinates for the physical setup. It is worth mentioning that apart from suggesting the correct coordinates for the needle plan (holes and corners) the accuracy of the physical puncture is mostly related to the physical design of the cube and its coordinate arids (holes and corners). With regards to needle diameter an average needle qauqe of 0.9 mm for Puncture Cube and 1.2 mm for Access Cube is used in the software as simulations have shown that accuracy only varies slightly for the different needle diameters. Usability study was conducted where 5 interventional radiologists from independent sites used the software to plan a intervention successfully and provided feedback along the line of the system. It was shown that the Cube Navigator is safe and effective to use.
No clinical tests have been submitted, referenced. or relied on in the premarket notification submission of substantial equivalence.
Based on the non-clinical test conducted we find Cube Navigator to be as safe, as effective and performs as well as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.