← Product Code [HQD](/productcode/HQD) · K240618

# Acuity 200™ (fluoroxyfocon A) Rigid Gas Permeable Contact Lens; Acuity 100™ (hexafocon A) Rigid Gas Permeable Contact Lens; Acuity 200™ with Tangible® Hydra-PEG® (fluoroxyfocon A) Rigid Gas Permeable Contact Lens; Acuity 100™ with Tangible® Hydra-PEG® (hexafocon A) Rigid Gas Permeable Contact Lens (K240618)

_Acuity Polymers, Inc. · HQD · Apr 3, 2024 · Ophthalmic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K240618

## Device Facts

- **Applicant:** Acuity Polymers, Inc.
- **Product Code:** [HQD](/productcode/HQD.md)
- **Decision Date:** Apr 3, 2024
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 886.5916
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic
- **Attributes:** Therapeutic

## Indications for Use

The Acuity 200™ (fluoroxyfocon A), Acuity 100™ (hexafocon A), Acuity 200™ with Tangible® Hydra-PEG® (fluoroxyfocon A), and Acuity 100™ with Tangible® Hydra-PEG® (hexafocon A) Rigid Gas Permeable Contact Lenses are indicated for daily wear for the correction of refractive error (myopia, presbyopia and/or astigmatism) in aphakic and non-aphakic persons with non-diseased eyes. The lenses may be prescribed for daily wear in otherwise nondiseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g., LASIK) surgery. The Acuity 200™ (fluoroxyfocon A), Acuity 100™ (hexafocon A), Acuity 200™ with Tangible® Hydra-PEG® (fluoroxyfocon A), and Acuity 100™ with Tangible® Hydra-PEG® (hexafocon A) Rigid Gas Permeable Contact Lenses are also indicated for therapeutic use in eyes with ocular surface disease (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome, dry eve syndrome and Filamentary Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservor and protection against an adverse environment. When prescribed for therapeutic use for a distorted cornea or ocular surface disease, the lens may concurrently provide correction of refractive error. The Acuity 200™ (fluoroxyfocon A), Acuity 100™ (hexafocon A), Acuity 200™ with Tangible® Hydra-PEG® (fluoroxyfocon A), and Acuity 100™ with Tangible® Hydra-PEG® (hexafocon A) Rigid Gas Permeable Contact Lenses may be cleaned and disinfected using a chemical (not heat) lens care system.

## Device Story

Rigid gas permeable (RGP) contact lenses manufactured from siloxanyl fluoromethacrylate copolymers; available in spherical, toric, multifocal, scleral, and aspheric designs. Some models treated with Tangible® Hydra-PEG® polymer to enhance surface wetting without altering mechanical/optical properties. Lenses are lathe-cut; tinted for visibility; optional UV absorbers. Used by eye care practitioners to manage refractive errors and therapeutic conditions (irregular corneas, ocular surface disease). Lenses provide expanded tear reservoir and environmental protection for diseased eyes. Patients wear lenses daily; cleaned/disinfected via chemical systems. Output is corrected vision and ocular surface protection.

## Clinical Evidence

No new clinical data presented. Clinical performance for the ocular surface disease indication was previously addressed in 510(k) K233325 for fluoroxyfocon A and hexafocon A scleral contact lenses.

## Technological Characteristics

Materials: siloxanyl fluoromethacrylate copolymers (fluoroxyfocon A, hexafocon A). FDA Material Group #3. Lathe-cut production. Surface treatment: Tangible® Hydra-PEG® (PEG polymer). Oxygen permeability (Dk): 200 (Acuity 200) and 111 (Acuity 100). Refractive index: 1.430 (Acuity 200) and 1.415 (Acuity 100). Hardness: 78-80 Shore D. Modulus: 1194-1496 MPa. Connectivity: N/A. Sterilization: Chemical disinfection system.

## Regulatory Identification

A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.

## Predicate Devices

- Visionary Optics Scleral Contact Lens ([K233325](/device/K233325.md))
- Acuity 200™ (fluoroxyfocon A) Rigid Gas Permeable Contact Lens ([K201194](/device/K201194.md))
- Acuity 100™ (hexafocon A) Rigid Gas Permeable Contact Lens ([K162005](/device/K162005.md))
- Acuity 200 with Tangible® Hydra-PEG® (fluoroxyfocon A) Rigid Gas Permeable Contact Lens ([K213538](/device/K213538.md))
- Acuity 100™ (hexafocon A) with Tangible® Hydra-PEG®, TYRO™-97 (hofocon A) with Tangible® Hydra-PEG® Rigid Gas Permeable Contact Lens ([K230965](/device/K230965.md))
- OPTIMUM GP (roflufocon D, roflurocon E) Daily Wear Contact Lens, HEXA 100 (hexafocon A) Daily Wear Contact Lens ([K180616](/device/K180616.md))

## Reference Devices

- Menicon Unique pH Multi-purpose Solution ([K130805](/device/K130805.md))
- Boston Simplus® Multi-Action Solution ([K024289](/device/K024289.md))
- Deep Well Lens Case ([K171539](/device/K171539.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 3, 2024

Bret Andre, Principal Consultant Andre Vision and Device Research 6119 Canter Lane West Linn, OR 97068

## Re: K240618

Trade/Device Name: Acuity 200™ (fluoroxyfocon A) Rigid Gas Permeable Contact Lens; Acuity 100™ (hexafocon A) Rigid Gas Permeable Contact Lens; Acuity 200™ with Tangible® Hydra-PEG® (fluoroxyfocon A) Rigid Gas Permeable Contact Lens; Acuity 100™ with Tangible® Hydra-PEG® (hexafocon A) Rigid Gas Permeable Contact Lens Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid Gas Permeable Contact Lens Regulatory Class: Class II Product Code: HQD Dated: February 26, 2024 Received: March 6, 2024

Dear Bret Andre:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}------------------------------------------------

Page 2

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products): good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# J Angelo Green -S

J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K240618

### Device Name

Acuity 200™ (fluoroxyfocon A) Rigid Gas Permeable Contact Lens, Acuity 100™ (hexatocon A) Rigid Gas Permeable Contact Lens, Acuity 200™ with Tangible® Hydra-PEG® (fluoroxyfocon A) Rigid Gas Permeable Contact Lens, and Acuity 100™ with Tangible® Hydra-PEG® (hexafocon A) Rigid Gas Permeable Contact Lens

## Indications for Use (Describe)

The Acuity 200™ (fluoroxyfocon A), Acuity 100™ (hexafocon A), Acuity 200™ with Tangible® Hydra-PEG® (fluoroxyfocon A), and Acuity 100™ with Tangible® Hydra-PEG® (hexafocon A) Rigid Gas Permeable Contact Lenses are indicated for daily wear for the correction of refractive error (myopia, presbyopia and/or astigmatism) in aphakic and non-aphakic persons with non-diseased eyes. The lenses may be prescribed for daily wear in otherwise nondiseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g., LASIK) surgery.

The Acuity 200™ (fluoroxyfocon A), Acuity 100™ (hexafocon A), Acuity 200™ with Tangible® Hydra-PEG® (fluoroxyfocon A), and Acuity 100™ with Tangible® Hydra-PEG® (hexafocon A) Rigid Gas Permeable Contact Lenses are also indicated for therapeutic use in eyes with ocular surface disease (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome, dry eve syndrome and Filamentary Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservor and protection against an adverse environment. When prescribed for therapeutic use for a distorted cornea or ocular surface disease, the lens may concurrently provide correction of refractive error.

The Acuity 200™ (fluoroxyfocon A), Acuity 100™ (hexafocon A), Acuity 200™ with Tangible® Hydra-PEG® (fluoroxyfocon A), and Acuity 100™ with Tangible® Hydra-PEG® (hexafocon A) Rigid Gas Permeable Contact Lenses may be cleaned and disinfected using a chemical (not heat) lens care system.

| Type of Use (Select one or both, as applicable)                                                            |                                                                                            |
|------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| <div> <span> <b>                Prescription Use (Part 21 CFR 801 Subpart D)           </b> </span> </div> | <div> <span>            Over-The-Counter Use (21 CFR 801 Subpart C)         </span> </div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{4}------------------------------------------------

# SPECIAL 510 (k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

# The assigned 510(k) number is: K240618

#### I. SUBMITTER

| Date Prepared:  | March 30, 2024                                                                           |
|-----------------|------------------------------------------------------------------------------------------|
| Name:           | Acuity Polymers, Inc.                                                                    |
| Address:        | 1667 Lake Avenue, Suite 354<br>Rochester, NY 14615<br>United States                      |
| Contact Person: | James A. Bonafini, Jr<br>President                                                       |
| Phone number:   | (585) 458-8409                                                                           |
| Consultant:     | Bret Andre<br>Andre Vision and Device Research<br>6119 Canter Ln.<br>West Linn, OR 97068 |
| Phone number:   | (503) 372-5226                                                                           |

#### II. DEVICE

| Trade Name:             | Acuity 200 TM (fluoroxyfocon A) Rigid Gas Permeable Contact Lens;<br>Acuity 100 TM (hexafocon A) Rigid Gas Permeable Contact Lens;<br>Acuity 200 TM with Tangible® Hydra-PEG® (fluoroxyfocon A) Rigid Gas<br>Permeable Contact Lens;<br>Acuity 100 TM with Tangible® Hydra-PEG® (hexafocon A) Rigid Gas<br>Permeable Contact Lens |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common<br>Name:         | Daily wear rigid gas permeable contact lens                                                                                                                                                                                                                                                                                       |
| Classification<br>Name: | Rigid gas permeable contact lens. (21 CFR 886.5916)                                                                                                                                                                                                                                                                               |
| Regulatory<br>Class:    | Class II                                                                                                                                                                                                                                                                                                                          |
| Product Code:           | HQD                                                                                                                                                                                                                                                                                                                               |
| ~Reason for Submission~ | Indications for Use Change                                                                                                                                                                                                                                                                                                        |

{5}------------------------------------------------

#### III. PREDICATE DEVICE

The Acuity 200™ (fluoroxyfocon A), Acuity 100™ (hexafocon A), Acuity 200™ with Tangible® Hydra-PEG® (fluoroxyfocon A), and Acuity 100™ with Tangible® Hydra-PEG® Rigid Gas Permeable Contact Lenses for daily wear are substantially equivalent to the following predicate devices:

- . "Visionary Optics Scleral Contact Lens (roflufocon D, roflufocon E, hexafocon A, hexafocon B, tisilfocon A, fluoroxyfocon A)" By Visionary Optics LLC 510(k) number: K233325 Product Code: HQD
- l "Acuity 200TM (fluoroxyfocon A) Rigid Gas Permeable Contact Lens" By Acuity Polymers, Inc. 510(k) number; K201194 Product Code: HQD
- . "Acuity 100™ (hexafocon A) Rigid Gas Permeable Contact Lens" By Acuity Polymers, Inc. 510(k) number; K162005 Product Code: HQD
- I Acuity 200 with Tangible® Hydra-PEG® (fluoroxyfocon A) Rigid Gas Permeable Contact Lens By Acuity Polymers, Inc. 510(k) number; K213538
	- Product Code: HQD
- "Acuity 100™ (hexafocon A) with Tangible® Hydra-PEG®, TYRO™-97 (hofocon A) . with Tangible® Hydra-PEG® Rigid Gas Permeable Contact Lens" By Acuity Polymers, Inc. 510(k) number; K230965 Product Code: HQD
- . "OPTIMUM GP (roflufocon D, roflurocon E) Daily Wear Contact Lens, HEXA 100 (hexafocon A) Daily Wear Contact Lens" By Contamac Ltd. 510(k) number; K180616 Product Code: HQD

#### DEVICE DESCRIPTION IV.

The Acuity 200™ (fluoroxyfocon A), Acuity 100™ (hexafocon A), Acuity 200™ with Tangible® Hydra-PEG® (fluoroxyfocon A), and Acuity 100™ with Tangible® Hydra-PEG® Rigid Gas Permeable Contact Lenses are manufactured from machine latheable rigid gas permeable materials composed of siloxanyl fluoromethacrylate copolymers that are tinted for visibility and available with or without an ultraviolet (UV) light absorbers.

Acuity 200™ with Tangible® Hydra-PEG® (fluoroxyfocon A), and Acuity 100™ with Tangible® Hydra-PEG® Rigid Gas Permeable Contact Lenses are treated to incorporate Hydra-PEG Technology developed by Tangible Sciences. This PEG polymer is permanently attached to the surface and is designed to enhance surface (wetting) properties while not affecting the

{6}------------------------------------------------

mechanical or optical properties of the underlying material.

The Acuity 200™ (fluoroxyfocon A), Acuity 100™ (hexafocon A), Acuity 200™ with Tangible® Hydra-PEG® (fluoroxyfocon A), and Acuity 100™ with Tangible® Hydra-PEG® Rigid Gas Permeable Contact Lenses for daily wear are made available in spherical, toric, multifocal, scleral and aspheric designs in the following lens parameters:

| Parameter          | Range              | Tolerance                                                                                                                            |
|--------------------|--------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Base Curve         | 4.00mm to 11.50mm  | $\pm$ 0.05 mm                                                                                                                        |
| Center Thickness   | 0.08mm to 0.75mm   | $\pm$ 0.02 mm                                                                                                                        |
| Diameter           | 7.0mm to 21.0mm    | $\pm$ 0.10mm                                                                                                                         |
| Spherical Power    | -20.00D to +20.00D | $\pm$ 0.12 (0 to = 5D)<br $\pm$ 0.18 (5 to = 10.0D)<br $\pm$ 0.25 (10 to = 15D)<br $\pm$ 0.37 (15 to = 20D)<br $\pm$ 0.50 (over 20D) |
| Cylindrical Power  | Up to 9.00D        | $\pm$ 0.25 (0 to = 2D)<br $\pm$ 0.37 (2 to = 4D)<br $\pm$ 0.50 (over 4D)                                                             |
| Multifocal Power   | +1.00D to 4.00D    | $\pm$ 0.25D                                                                                                                          |
| Surface Appearance |                    | Lenses should be clear with no surface<br>defect                                                                                     |

The following table depicts the physical properties of the Acuity 200™ (fluoroxyfocon A) and Acuity 200TM with Tangible® Hydra-PEG® Rigid Gas Permeable Contact Lens:

|                             | Fluoroxyfocon A Rigid Gas Permeable Contact Lens                               |
|-----------------------------|--------------------------------------------------------------------------------|
| Refractive Index            | 1.430                                                                          |
| Modulus (MPa)               | 1194                                                                           |
| Hardness (Shore D)          | 78                                                                             |
| Specific Gravity            | 1.18                                                                           |
| Oxygen Permeability<br>(Dk) | 200 x 10-11 (cm²/sec) (ml O2/ml x mm Hg @ 35°C)                                |
| Color Additives             | Visibility Tints – D&C Green #6, D&C Violet #2, Solvent Yellow 18, D&C Red #17 |

The following table depicts the physical properties of the Acuity 100™ and Acuity 100™ with Tangible® Hydra-PEG® (hexafocon A) Rigid Gas Permeable Contact Lens:

|                             | Hexafocon A Rigid Gas Permeable Contact Lens                      |
|-----------------------------|-------------------------------------------------------------------|
| Refractive Index            | 1.415                                                             |
| Modulus (MPa)               | 1496                                                              |
| Hardness (Shore D)          | 80                                                                |
| Specific Gravity            | 1.27                                                              |
| Oxygen Permeability<br>(Dk) | $111 \times 10^{-11}$ (cm²/sec) (ml O2/ml x mm Hg @ 35°C)         |
| Color Additives             | Visibility Tints — D&C Green #6, D&C Violet #2, Solvent Yellow 18 |

{7}------------------------------------------------

| MODEL            | Acuity 200<br>(fluoroxyfocon A) | Acuity 100 (hexafocon A)      | Acuity 200 (fluoroxyfocon A) with Tangible Hydra-PEG                                                    | Acuity 100 (hexafocon A) with Tangible Hydra-PEG                                                        |
|------------------|---------------------------------|-------------------------------|---------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| Ship Method      | Dry                             | Dry                           | Dry or Wet*                                                                                             | Dry or Wet*                                                                                             |
| Storage Solution | N/A                             | N/A                           | Menicon Unique pH Multi-purpose Solution (K130805) &<br>Boston Simplus® Multi-Action Solution (K024289) | Menicon Unique pH Multi-purpose Solution (K130805) &<br>Boston Simplus® Multi-Action Solution (K024289) |
| Shipping Case    | Deep Well Lens Case (K171539)   | Deep Well Lens Case (K171539) | Deep Well Lens Case (K171539)                                                                           | Deep Well Lens Case (K171539)                                                                           |

The following table depicts the shipping method, storage solution for wet shipping, and shipping case used for each lens model:

* Lenses coated with Tangible Hydra-PEG are stored wet.

#### INDICATIONS FOR USE V.

The Acuity 200™ (fluoroxyfocon A), Acuity 100™ (hexafocon A), Acuity 200™ with Tangible® Hydra-PEG® (fluoroxyfocon A), and Acuity 100™ with Tangible® Hydra-PEG® Rigid Gas Permeable Contact Lenses are indicated for daily wear for the correction of refractive error (myopia, hyperopia, presbyopia and/or astigmatism) in aphakic and non-aphakic persons with non-diseased eves. The lenses may be prescribed for daily wear in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g., LASIK) surgery.

The Acuity 200™ (fluoroxyfocon A), Acuity 100™ (hexafocon A), Acuity 200™ with Tangible® Hydra-PEG® (fluoroxyfocon A), and Acuity 100™ with Tangible® Hydra-PEG® Rigid Gas Permeable Contact Lenses are also indicated for therapeutic use in eyes with ocular surface disease (e.g. ocular Graft-versus-Host disease. Siögren's syndrome, dry eve syndrome and Filamentary Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for a distorted cornea or ocular surface disease, the lens may concurrently provide correction of refractive error.

The Acuity 200™ (fluoroxyfocon A), Acuity 100™ (hexafocon A), Acuity 200™ with Tangible® Hydra-PEG® (fluoroxyfocon A), and Acuity 100™ with Tangible® Hydra-PEG® Rigid Gas Permeable Contact Lenses may be cleaned and disinfected using a chemical (not heat) lens care system.

{8}------------------------------------------------

### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICE

The Acuity 200™ (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses and Acuity 100™ (hexafocon A) Rigid Gas Permeable Contact Lenses are substantially equivalent to the predicate device (cleared under K233325) in terms of the following:

- Intended use - daily wear contact lenses
- 트 Indications for use - Therapeutic (irregular cornea and ocular surface disease)
- 트 Classification - Lenses, Rigid Gas Permeable, Daily Wear (21 CFR 886.5916)
- 트 FDA material group - group # 3 fluoro silicone acrylate
- USAN materials (fluoroxyfocon A and hexafocon A)
- Production method - lathe cut

The Acuity 200™ (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses are substantially equivalent to the predicate device (cleared under K201194) in terms of the following:

- I Intended use - daily wear contact lenses
- 트 Classification - Lenses, Rigid Gas Permeable, Daily Wear (21 CFR 886.5916)
- 트 FDA material group - group # 3 fluoro silicone acrylate
- USAN material (fluoroxyfocon A)
- Lens designs
- Production method - lathe cut (same facility, processes, and quality control procedures)
- 트 Final packaging and shipping

The Acuity 100™ (hexafocon A) Rigid Gas Permeable Contact Lenses are substantially equivalent to the predicate device (cleared under K162005) in terms of the following:

- Intended use daily wear contact lenses ■
- 트 Classification - Lenses, Rigid Gas Permeable, Daily Wear (21 CFR 886.5916)
- FDA material group - group # 3 fluoro silicone acrylate
- USAN material (hexafocon A)
- Lens designs
- Production method lathe cut (same facility, processes, and quality control procedures) 트
- I Final packaging and shipping

The Acuity 200™ with Tangible® Hydra-PEG® (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses are substantially equivalent to the predicate device (cleared under K213538) in terms of the following:

- Intended use - daily wear contact lenses
- 트 Classification - Lenses, Rigid Gas Permeable, Daily Wear (21 CFR 886.5916)
- FDA material group – group # 3 fluoro silicone acrylate
- 트 USAN material (fluoroxyfocon A)
- 트 Available with Tangible® Hydra-PEG®
- Lens designs

{9}------------------------------------------------

- ' Production method - lathe cut (same facility, processes, and quality control procedures)
- 트 Final packaging and shipping

The Acuity 100™ with Tangible® Hydra-PEG® (hexafocon A) Rigid Gas Permeable Contact Lenses are substantially equivalent to the predicate device (cleared under K230965) in terms of the following:

- 트 Intended use - daily wear contact lenses
- 트 Classification - Lenses, Rigid Gas Permeable, Daily Wear (21 CFR 886.5916)
- FDA material group group # 3 fluoro silicone acrylate ■
- 트 USAN material (hexafocon A)
- Available with Tangible® Hydra-PEG®
- 트 Lens designs
- 트 Production method - lathe cut (same facility, processes, and quality control procedures)
- I Final packaging and shipping

The Acuity 200™ with Tangible® Hydra-PEG® (fluoroxyfocon A) Acuity 100™ with Tangible® Hydra-PEG® (hexafocon A) Rigid Gas Permeable Contact Lenses are substantially equivalent to the predicate device (cleared under K180616) in terms of the following:

- Intended use - daily wear contact lenses
- 트 Classification - Lenses, Rigid Gas Permeable, Daily Wear (21 CFR 886.5916)
- FDA material group – group # 3 fluoro silicone acrylate
- 트 USAN material (hexafocon A)
- 트 Available with Tangible® Hydra-PEG®
- 트 Indications for use - Therapeutic (irregular cornea and ocular surface disease)

{10}------------------------------------------------

The following matrix illustrates the production method, lens function and materistics of the Acuity 200™ (fluoroxyfocon A), Acuity 100™ (hexafocon A), Acuity 200™ with Tangible® Hydra-PEG® (fluoroxyfocon A), and Acuity 100™ with Tangible® Hydra-PEG® Rigid Gas Permeable Contact Lenses, as well as the predicate devices.

|                                 | Acuity 200™<br>Acuity 100™<br>Rigid Gas<br>Permeable<br>Contact Lens         | Visionary Optics<br>Scleral Contact<br>Lens                                                        | Acuity 200™<br>Rigid Gas<br>Permeable<br>Contact Lens                        | Acuity 100™<br>Rigid Gas<br>Permeable<br>Contact Lens                           | Acuity 200™<br>with<br>Tangible®<br>Hydra-PEG®                                  | Acuity 100™<br>TYRO™ 97<br>with<br>Tangible®<br>Hydra-PEG®                         | OPTIMUM<br>GP<br>HEXA100<br>Daily Wear<br>Contact<br>Lens                          |                                                                                                                                                                                                                                             | Indications for Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |  |
|---------------------------------|------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
|                                 | Subject Device                                                               | Predicate Device<br>(K233325)                                                                      | Predicate Device<br>(K201194)                                                | Predicate<br>Device<br>(K162005)                                                | Predicate<br>Device<br>(K213538)                                                | Predicate<br>Device<br>(K230965)                                                   | Predicate<br>Device<br>(K180616)                                                   | Acuity 200™<br>(fluoroxyfocon<br>A), Acuity 100™<br>(hexafocon A),<br>Acuity 200™<br>with Tangible®<br>Hydra-PEG®<br>(fluoroxyfocon<br>A), and Acuity<br>100™ with<br>Tangible®<br>Hydra-PEG®<br>Rigid Gas<br>Permeable<br>Contact Lenses s | The Acuity 200™ (fluoroxyfocon A), Acuity 100™ (hexafocon A), Acuity 200™ with Tangible®<br>Hydra-PEG® (fluoroxyfocon A), and Acuity 100™ with Tangible® Hydra-PEG® Rigid Gas Permeable<br>Contact Lenses are indicated for daily wear for the correction of refractive error (myopia, hyperopia,<br>presbyopia and/or astigmatism) in aphakic and non-aphakic persons with non-diseased eyes. The lenses<br>may be prescribed for daily wear in otherwise non-diseased eyes that require a rigid contact lens for the<br>management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or<br>following penetrating keratoplasty or refractive (e.g., LASIK) surgery.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |  |
| Classification                  | Class II<br>Lenses, Rigid Gas<br>Permeable, Daily<br>Wear<br>21 CFR 886.5916 | Class II<br>Lenses, Rigid Gas<br>Permeable, Daily<br>Wear<br>21 CFR 886.5916                       | Class II<br>Lenses, Rigid<br>Gas Permeable,<br>Daily Wear<br>21 CFR 886.5916 | Class II<br>Lenses, Rigid<br>Gas Permeable,<br>Daily Wear<br>21 CFR<br>886.5916 | Class II<br>Lenses, Rigid<br>Gas Permeable,<br>Daily Wear<br>21 CFR<br>886.5916 | Class II<br>Lenses, Rigid<br>Gas<br>Permeable,<br>Daily Wear<br>21 CFR<br>886.5916 | Class II<br>Lenses,<br>Rigid Gas<br>Permeable,<br>Daily Wear<br>21 CFR<br>886.5916 | with Tangible®<br>Hydra-PEG®<br>(fluoroxyfocon<br>A), and Acuity<br>100™ with<br>Tangible®<br>Hydra-PEG®<br>Rigid Gas<br>Permeable<br>Contact Lenses s                                                                                      | The Acuity 200™ (fluoroxyfocon A), Acuity 100™ (hexafocon A), Acuity 200™ with Tangible®<br>Hydra-PEG® (fluoroxyfocon A), and Acuity 100™ with Tangible® Hydra-PEG® Rigid Gas Permeable<br>Contact Lenses are also indicated for therapeutic use in eyes with ocular surface disease (e.g. ocular<br>Graft-versus-Host disease, Sjögren's syndrome, dry eye syndrome and Filamentary Keratitis), limbal<br>stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders<br>of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster,<br>Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the<br>presence of an expanded tear reservoir and protection against an adverse environment. When prescribed<br>for therapeutic use for a distorted cornea or ocular surface disease, the lens may concurrently provide<br>correction of refractive error.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |  |
| Product Code                    | HQD                                                                          | HQD                                                                                                | HQD                                                                          | HQD                                                                             | HQD                                                                             | HQD                                                                                | HQD                                                                                | (Subject Device)                                                                                                                                                                                                                            | The Acuity 200™ (fluoroxyfocon A), Acuity 100™ (hexafocon A), Acuity 200™ with Tangible®<br>Hydra-PEG® (fluoroxyfocon A), and Acuity 100™ with Tangible® Hydra-PEG® Rigid Gas Permeable<br>Contact Lenses may be cleaned and disinfected using a chemical (not heat) lens care system.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |  |
| FDA Group #                     | Group # 3 Fluoro<br>Silicone Acrylate                                        | Group # 3 Fluoro<br>Silicone Acrylate                                                              | Group # 3 Fluoro<br>Silicone Acrylate                                        | Group # 3<br>Fluoro Silicone<br>Acrylate                                        | Group # 3<br>Fluoro Silicone<br>Acrylate                                        | Group # 3<br>Fluoro<br>Silicone<br>Acrylate                                        | Group # 3<br>Fluoro<br>Silicone<br>Acrylate                                        | Visionary Optics<br>Scleral Contact<br>Lens (roflufocon<br>D, roflufocon E,<br>hexafocon A,<br>hexafocon B,<br>tisilfocon A,<br>fluoroxyfocon A)<br>(K233325)                                                                               | The Visionary Optics Scleral Contact Lens for daily wear is indicated for use for the management of<br>multiple ocular conditions, such as, degenerations that lead to an irregular corneal shape (e.g.<br>keratoconus, keratoglobus, pellucid marginal degeneration, Salzmann's Nodular Degeneration),<br>dystrophies (e.g. Cogan's dystrophy, Granular Corneal Dystrophy, Lattice Corneal Dystrophy,), post-<br>surgery (e.g. corneal transplant, LASIK, radial keratotomy), and corneal scarring from infection or<br>trauma.<br>The Visionary Optics Scleral Contact Lens for daily wear is also indicated for therapeutic management<br>of ocular surface disease including dry eye (e.g. ocular manifestations of Graft-versus-Host disease,<br>Sjogren's syndrome, dry eye syndrome), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome,<br>chemical and thermal burns, radiation, filamentary keratitis), epidermal ocular disorders, disorders of the<br>skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster,<br>Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the<br>presence of an expanded tear reservoir and protection against an adverse environment. When prescribed<br>for therapeutic use for distorted cornea or ocular surface disease, the Visionary Optics Scleral Contact<br>Lens may incidentally provide correction of refractive error in persons with myopia, hyperopia,<br>astigmatism or presbyopia.<br>Eye care practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning,<br>disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the<br>lens may be cleaned and disinfected using a chemical (not heat) lens care system. |  |
| Material (USAN)                 | fluoroxyfocon A,<br>hexafocon A                                              | roflufocon D.<br>roflufocon E.<br>hexafocon A.<br>hexafocon B.<br>tisilfocon A,<br>fluoroxyfocon A | fluoroxyfocon A                                                              | hexafocon A                                                                     | fluoroxyfocon<br>A                                                              | hexafocon A;<br>hofocon A                                                          | hexafocon<br>A;<br>roflufocon<br>D;<br>roflufocon E                                | Acuity 200™<br>(fluoroxyfocon<br>A) Rigid Gas<br>Permeable<br>Contact Lens<br>(K201194)                                                                                                                                                     | The Acuity 200™ (fluoroxyfocon A) Rigid Gas Permeable Contact Lenses are indicated for daily wear<br>for the correction of refractive error (myopia, hyperopia, presbyopia and/or astigmatism) in aphakic and<br>non-aphakic persons with non-diseased eyes. The lens may be prescribed in spherical and aspheric<br>powers ranging from -20.00 D to +20.00 D for daily wear.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |  |
| Production Method               | Lathe-Cut                                                                    | Lathe-Cut                                                                                          | Lathe-Cut                                                                    | Lathe-Cut                                                                       | Lathe-Cut                                                                       | Lathe-Cut                                                                          | Lathe-Cut                                                                          | Acuity 100™<br>(hexafocon A)<br>Rigid Gas<br>Permeable<br>Contact Lens<br>(K162005)                                                                                                                                                         | The Acuity 100™ (hexafocon A) Rigid Gas Permeable Contact Lenses are indicated for daily<br>wear for the correction of refractive error (myopia, hyperopia, presbyopia and/or astigmatism) in<br>aphakic and non-aphakic persons with non-diseased eyes. The lens may be prescribed in spherical and<br>aspheric powers ranging from -20.00 D to +20.00 D for daily wear. The lenses may be prescribed for<br>daily wear in otherwise non-diseased eyes that require a rigid contact lens for the management of<br>irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following<br>penetrating keratoplasty or refractive (e.g., LASIK) surgery. The lens may be disinfected using a<br>chemical disinfection system only.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |  |
| Intended Use                    | Daily Wear                                                                   | Daily Wear                                                                                         | Daily Wear                                                                   | Daily Wear                                                                      | Daily Wear                                                                      | Daily Wear                                                                         | Daily Wear                                                                         | Acuity 200™<br>with Tangible®                                                                                                                                                                                                               | Acuity 200™ with Tangible® Hydra PEG® (fluoroxyfocon A) Rigid Gas Permeable Contact Lens is<br>indicated for daily wear for the correction of refractive error (myopia, hyperopia, presbyopia and/or                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |  |
| Water Content (%)               | <1%                                                                          | <1%                                                                                                | <1%                                                                          | <1%                                                                             | <1%                                                                             | <1%                                                                                | <1%                                                                                |                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |  |
| UV Absorber<br>Available        | Yes                                                                          | Yes                                                                                                | Yes                                                                          | Yes                                                                             | Yes                                                                             | Yes                                                                                | Yes                                                                                |                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |  |
| Hydra-PEG®<br>Coating Available | Yes                                                                          | No                                                                                                 | No                                                                           | No                                                                              | Yes                                                                             | Yes                                                                                | Yes                                                                                |                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |  |

{11}------------------------------------------------

{12}------------------------------------------------

| Hydra-PEG®<br>Rigid Gas<br>Permeable<br>Contact Lens<br>(K213538)                                                                                   | astigmatism) in aphakic and non-aphakic persons with non-diseased eyes. The lenses may be prescribed<br>for daily wear in otherwise non-diseased eyes that require a rigid contact lens for the management of<br>irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following<br>penetrating keratoplasty or refractive (e.g., LASIK) surgery. The Acuity™ 200 with Tangible® Hydra-<br>PEG® (fluoroxyfocon A) Rigid Gas Permeable Contact Lens may be cleaned and disinfected using a<br>chemical (not heat) lens care system.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|-----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Acuity 100™ and<br>TYRO™ 97 with<br>Tangible®<br>Hydra-PEG®<br>Rigid Gas<br>Permeable<br>Contact Lenses<br>(K230965)                                | Acuity™ 100 (hexafocon A) with Tangible™ Hydra-PEG® Rigid Gas Permeable Contact Lenses are<br>indicated for daily wear for the correction of refractive error (myopia, hyperopia, presbyopia and/or<br>astigmatism) in aphakic and non-aphakic persons with non-diseased eyes. The lenses may be prescribed<br>for daily wear in otherwise non-diseased eyes that require a rigid contact lens for the management of<br>irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following<br>penetrating keratoplasty or refractive (e.g., LASIK) surgery. The lens may be disinfected<br>using a chemical disinfection system only.<br>TYRO™-97 (hofocon A) with Tangible™ Hydra-PEG® Rigid Gas Permeable Contact Lenses are<br>indicated for daily wear for the correction of visual acuity in non-aphakic persons with non-diseased<br>eyes that are myopic, hyperopic or presbyopic and which may exhibit corneal astigmatism. Also, the<br>lenses may be prescribed in otherwise non-diseased eyes that require a gas permeable contact lens for<br>the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or<br>following penetrating keratoplasty or refractive (e.g., LASIK) surgery. The lenses may be disinfected<br>only by using chemical disinfection.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| OPTIMUM GP<br>(roflufocon D,<br>roflurocon E)<br>Daily Wear<br>Contact Lens,<br>HEXA100<br>(hexafocon A)<br>Daily Wear<br>Contact Lens<br>(K180616) | OPTIMUM GP (roflufocon D, roflufocon E) and HEXA100 (hexafocon A) Daily Wear Contact<br>Lenses are indicated for the correction of refractive ametropia (myopia, hyperopia, astigmatism and<br>presbyopia) in aphakic and non aphakic persons with non-diseased eyes. Also, the lenses may be<br>prescribed in otherwise non-diseased eyes that require a rigid contact lens for the management of<br>irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following<br>penetrating keratoplasty or refractive surgery.<br>Furthermore, eyes suffering from certain ocular surface disorders may benefit from the physical<br>protection, aqueous hydrated environment and the saline bath provided by scleral lens designs.<br>OPTIMUM GP (roflufocon D, roflufocon E) and HEXA100 (hexafocon A) Daily Wear Contact<br>Lenses are indicated for therapeutic use for the management of irregular and distorted corneal<br>surfaces where the subject:<br>1. cannot be adequately corrected with spectacle lenses<br>2. requires a rigid gas permeable contact lens surface to improve vision<br>3. is unable to wear a corneal rigid gas permeable lens due to corneal distortion or surface<br>irregularities<br>Common causes of corneal distortion include but are not limited to corneal infections, trauma,<br>tractions as a result of scar formation secondary to refractive surgery (e.g. LASIK or radial<br>keratotomy) or corneal transplantation. Causes may also include corneal degeneration (e.g.<br>keratoconus, keratoglobus, pellucid marginal degeneration, Salzmann's nodular degeneration) and<br>corneal dystrophy (e.g., lattice dystrophy, granular corneal dystrophy, Reis-Bucklers<br>dystrophy, Cogan's dystrophy). The OPTIMUM GP (roflufocon D, roflufocon E) and HEXA100<br>(hexafocon A) Daily Wear Contact Lenses are also indicated for therapeutic use in eyes with<br>ocular surface disease (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome, dry eye syndrome<br>and Filamentary Keratitis), limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical<br>radiation and thermal burns), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic<br>keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g.<br>anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and<br>protection against an adverse environment. When prescribed for therapeutic use for a distorted<br>cornea or ocular surface disease, the OPTIMUM GP (roflufocon D, roflufocon E) and HEXA100<br>(hexafocon A) Daily Wear Contact Lenses may concurrently provide correction of refractive error. |

{13}------------------------------------------------

#### PERFORMANCE DATA VII.

## ~ Non-Clinical Studies ~

Non-clinical testing to demonstrate the safety and effectiveness of the Acuity 200™ (fluoroxyfocon A), Acuity 100™ (hexafocon A), Acuity 200™ with Tangible® Hydra-PEG® (fluoroxyfocon A), and Acuity 100™ with Tangible® Hydra-PEG® Rigid Gas Permeable Contact Lenses have been addressed through previously cleared 510(k) premarket notifications as described in the following table:

|                                                       | Acuity 200TM<br>(fluoroxyfocon A)<br>Rigid Gas<br>Permeable<br>Contact Lens | Acuity 100TM<br>(hexafocon A)<br>Rigid Gas<br>Permeable<br>Contact Lens | Acuity 200™ with<br>Tangible® Hydra-<br>PEG® (fluoroxyfocon<br>A) Rigid Gas Permeable<br>Contact Lens | Acuity 100™ with<br>Tangible® Hydra-<br>PEG® (hexafocon A)<br>Rigid Gas Permeable<br>Contact Lens |
|-------------------------------------------------------|-----------------------------------------------------------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Manufacturing<br>Design Verification                  | K201194                                                                     | K162005                                                                 | K213538                                                                                               | K230965                                                                                           |
| Physicochemical<br>and Mechanical<br>Property Testing | K201194                                                                     | K162005                                                                 | K213538                                                                                               | K230965                                                                                           |
| Extraction Studies                                    | K201194                                                                     | K162005                                                                 | K213538                                                                                               | K230965                                                                                           |
| Bioburden Testing                                     | K201194                                                                     | K162005                                                                 | K213538                                                                                               | K230965                                                                                           |
| Solution<br>Compatibility                             | K201194                                                                     | K162005                                                                 | K213538                                                                                               | K230965                                                                                           |
| Shelf-Life Studies                                    | K201194                                                                     | K162005                                                                 | K213538                                                                                               | K230965                                                                                           |
| Biocompatibility                                      | K201194                                                                     | K162005                                                                 | K213538                                                                                               | K230965                                                                                           |

## ~ Clinical Studies ~

Clinical performance of the ocular surface disease indication was addressed for fluoroxyfocon A and hexafocon A scleral contact lenses in 510(k) K233325.

{14}------------------------------------------------

# VIII. CONCLUSIONS

## Substantial Equivalence

Information presented in this premarket notification establishes that the Acuity 200™ (fluoroxyfocon A), Acuity 100™ (hexafocon A), Acuity 200™ with Tangible® Hydra-PEG® (fluoroxyfocon A), and Acuity 100™ with Tangible® Hydra-PEG® Rigid Gas Permeable Contact Lenses for daily wear are as safe and effective as the predicate devices when used in accordance with the labeled directions for use and for the proposed indications.

## Risks and Benefits

The risks of the Acuity 200™ (fluoroxyfocon A), Acuity 100™ (hexafocon A), Acuity 200™ with Tangible® Hydra-PEG® (fluoroxyfocon A), and Acuity 100™ with Tangible® Hydra-PEG® Rigid Gas Permeable Contact Lenses are the same as those normally attributed to the wearing of rigid gas permeable (RGP) daily wear contact lenses. The benefits to the patient are the same as those for other RGP contact lenses.

---

**Source:** [https://fda-staging.innolitics.com/device/K240618](https://fda-staging.innolitics.com/device/K240618)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).