Single Use Reloadable Clip Applicator (HX-810LR, HX-810UR); Clip (HX-610-090, HX-610-135); Long Clip (HX-610-090L, HX-610-135L); Short Clip (HX-610-090S, HX-610-135S); Super Short Clip (HX-610-135XS)
K240617 · Olympus Medical Systems Corp. · PKL · Jun 3, 2024 · Gastroenterology, Urology
Device Facts
Record ID
K240617
Device Name
Single Use Reloadable Clip Applicator (HX-810LR, HX-810UR); Clip (HX-610-090, HX-610-135); Long Clip (HX-610-090L, HX-610-135L); Short Clip (HX-610-090S, HX-610-135S); Super Short Clip (HX-610-135XS)
Applicant
Olympus Medical Systems Corp.
Product Code
PKL · Gastroenterology, Urology
Decision Date
Jun 3, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Single Use Reloadable Clip Applicators HX-810 Series, Long Clip HX-610 Series, Short Clip HX-610 Series and Super Short Clip HX-610-135XS have been designed to be used with an Olympus endoscope for endoscopic Clip placement within the gastrointestinal (GI) tract in adult patients only for the purpose of: (1) Endoscopic marking, (2) Hemostasis for (a) Mucosal/sub-mucosal defects < 3 cm, (b) Bleeding ulcers, (c) Arteries < 2 mm, (d) Polyps < 1.5 cm in diameter, (e) Diverticula in the colon, (3) As a supplementary method, closure of G1 tract luminal perforations < 20 mm that can be treated conservatively.
Device Story
Single-use reloadable clip applicator system used with endoscopes for GI tract procedures. Device consists of applicator handle, coil sheath, and metal clips. Operator (physician) uses slider mechanism to advance clip through endoscope; pulling slider opens clip; further pulling deploys clip by breaking connector. Once deployed, clip remains in tissue until natural excretion. Used for marking, hemostasis, and perforation closure. Benefits include mechanical tissue approximation and bleeding control. No electronic components or software.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included insertability, clip opening width, rotatability, capability, withdrawal, mechanical integrity, tensile strength, and applicator repetition. Testing performed on finished devices post-simulated distribution and accelerated aging (ASTM F1980-16).
Technological Characteristics
Materials: PTFE, silicone oil, metal components. Energy: Mechanical (manual operation). Form factor: Endoscopic applicator with coil sheath and reloadable clips. Sterilization: Ethylene Oxide (applicator), Gamma-ray (clips). Connectivity: None. Software: None.
Indications for Use
Indicated for adult patients requiring endoscopic clip placement in the GI tract for marking, hemostasis (mucosal/sub-mucosal defects < 3 cm, bleeding ulcers, arteries < 2 mm, polyps < 1.5 cm, colonic diverticula), or closure of GI tract luminal perforations < 20 mm treatable conservatively.
Regulatory Classification
Identification
A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.
Special Controls
*Classification.* Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
Single Use Reloadable Clip Applicator, Clip, Long Clip, Short Clip (K183590)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 3, 2024
Olympus Medical Systems Corp. % Rebecca Walker Regulatory Program Manager Olympus Corporation of the Americas 800 West Park Drive Westborough, Massachusetts 01581
Re: K240617
Trade/Device Name: Single Use Reloadable Clip Applicator (HX-810UR); Clip (HX-610-090, HX-610-135); Long Clip (HX-610-090L, HX-610-135L); Short Clip (HX-610-090S, HX-610-135S); Super Short Clip (HX-610-135XS) Regulation Number: 21 CFR 876.4400 Regulation Name: Hemorrhoidal Ligator Regulatory Class: Class II Product Code: PKL Dated: March 5, 2024 Received: March 5, 2024
### Dear Rebecca Walker:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Sivakami Venkatachalam -S
for
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K240617
Device Name
Single Use Reloadable Clip Applicator (HX-810LR); Clip (HX-610-090, HX-610-135); Long Clip (HX-610-090L, HX-610-135L); Short Clip (HX-610-090S, HX-610-135S); Super Short Clip (HX-610-135XS)
#### Indications for Use (Describe)
The Single Use Reloadable Clip Applicators HX-810 Series, Long Clip HX-610 Series, Short Clip HX-610 Series and Super Short Clip HX-610-135XS have been designed to be used with an Olympus endoscope for endoscopic Clip placement within the gastrointestinal (GI) tract in adult patients only for the purpose of:
- (1) Endoscopic marking,
- (2) Hemostasis for
- (a) Mucosal/sub-mucosal defects < 3 cm,
- (b) Bleeding ulcers,
- (c) Arteries < 2 mm,
- (d) Polyps < 1.5 cm in diameter,
- (e) Diverticula in the colon,
(3) As a supplementary method, closure of G1 tract luminal perforations < 20 mm that can be treated conservatively.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
коло , D 13
05
| | K240617 Page 1 of | | |
|----------------|-----------------------|--|--|
| 510(k) Summary | Prepared on: 2024-03- | | |
| Contact Details | |
|-----------------|--|
|-----------------|--|
21 CFR 807.92(a)(1)
| | 21 CFR 807.92(a)(1) | | |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--------------|
| Applicant Name | Olympus Medical Systems Corp. | | |
| Applicant Address | 2951 Ishikawa-cho Hachiochi-shi Tokyo 192-8507 Japan | | |
| Applicant Contact Telephone | +81 42-642-2111 | | |
| Applicant Contact | Ms. Seiko Yunoki | | |
| Applicant Contact Email | seiko.yunoki@olympus.com | | |
| Correspondent Name | Olympus Corporation of the Americas | | |
| Correspondent Address | 800 West Park Drive Westborough MA 01581 United States | | |
| Correspondent Contact Telephone | 805-551-3023 | | |
| Correspondent Contact | Ms. Rebecca Walker | | |
| Correspondent Contact Email | Rebecca.walker@olympus.com | | |
| | 21 CFR 807.92(a)(2) | | |
| Device Trade Name | Single Use Reloadable Clip Applicator (HX-810LR, HX-810UR);<br>Clip (HX-610-090, HX-610-135);<br>Long Clip (HX-610-090L, HX-610-135L);<br>Short Clip (HX-610-090S, HX-610-135S);<br>Super Short Clip (HX-610-135XS) | | |
| Common Name | Hemorrhoidal ligator | | |
| Classification Name | Hemostatic Metal Clip For The Gi Tract | | |
| Regulation Number | 876.4400 | | |
| Product Code(s) | PKL | | |
| | 21 CFR 807.92(a)(3) | | |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | | Product Code |
| K183590 | Single Use Reloadable Clip Applicator, Clip, Long Clip, Short Clip | | PKL |
| | 21 CFR 807.92(a)(4) | | |
| Device Description Summary | | | |
{5}------------------------------------------------
The instruments (AKA EZClip) have been designed to be used with an Olympus endoscopic Clip placement within the gastrointestinal (GI) tract in adult patients only for the purpose of:
(1) Endoscopic marking,
(2) Hemostasis for
(a) Mucosal/sub-mucosal defects < 3 cm,
(b) Bleeding ulcers,
(c) Arteries < 2 mm,
(d) Polyps < 1.5 cm in diameter,
(e) Diverticula in the colon,
(3) As a supplementary method, closure of Gl tract luminal perforations < 20 mm that can be treatively.
The Single Use Reloadable Clip Applicators and Clips are single-use devices. The HX-810 series Applicators are sterilized by Ethylene Oxide Gas and the HX-610 series Clips are sterilized by Gamma-ray radiation.
The Single Use Reloadable Clip Applicators and Clips are used in a combined state. The clip connected to the hook portion of the applicator is controlled via operation wire by the slider. The clip protruded from the coil sheath opens to the maximum opening width when the slider of the applicator is pulled towards the operator. Clipping is achieved when the slider of the applicator is pulled completely and the clip connector is broken. Upon completion, the clip and the applicator are no longer connected. Once the clipping is performed, the closure force of clips remains in place until tissue necrosis occurs, and the clip is excreted naturally.
# Intended Use/Indications for Use
### 21 CFR 807.92(a)(5)
The Single Use Reloadable Clip Applicators HX-810 Series, Long Clip HX-610 Series, Short Clip HX-610 Series and Super Short Clip HX-610-135XS have been designed to be used with an Olympus endoscope for endoscopic Clip placement within the gastrointestinal (GI) tract in adult patients only for the purpose of:
(1) Endoscopic marking,
(2) Hemostasis for
(a) Mucosal/sub-mucosal defects < 3 cm,
(b) Bleeding ulcers,
(c) Arteries < 2 mm,
(d) Polyps < 1.5 cm in diameter,
(e) Diverticula in the colon,
(3) As a supplementary method, closure of Gl tract luminal perforations < 20 mm that can be treated conservatively.
## Indications for Use Comparison
### 21 CFR 807.92(a)(5)
Indications for use are the same as the predicate device.
Technological Comparison
21 CFR 807.92(a)(6)
{6}------------------------------------------------
The subject device and the predicate device have technological characteristics except for design modifications to the Single Use Reloadable Clip Applicator. The subject device has the following design modifications to the Single Use Clip Applicator:
1. Structure of operation wire has been coated with silicone oil and covered with an inner tube composed of PTFE. The outer dimensions of the operator wire and number of stranded wires have been decreased.
2. Structure inside the handle. Dimensional changes have been introduced due to the decreased outer diameter of the operator wire.
3. Patient contacting materials. Silicone oil and PTFE have been added to the operation wire assembly.
The differences between the subject and predicate devices have been validated and testing demonstrations do not raise new questions of safety or efficacy and that the subject device is substantially equivalent to the predicate device.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Product performance testing was conducted on the modified Single Use Reloadable Clip Applicator (HX-810 Series) to demonstrate equivalency and ensure the device meets its intended use and design specifications after a shelf life of three (3) years. The performance tests were conducted on finished devices, prior to and after simulated distribution and accelerated aging for the terminally sterilized medical device in accordance with ASTM F1980-16(2016). The following tests were performed:
- 1. Insertability
- 2. Clip opening width
- 3. Clip rotatability
- 4. Clip capability
- 5. Withdrawal from Endoscope
- 6. Mechanical integrity of clip assembly
- 7. Tensile Strength/coil to handle strength
- 8. Applicator repetition
Based on the risk analysis, the design modification to the Reloadable Clip Applicator (HX-810 Series) did not require additional MRI Testing, Package Integrity Testing or Human Factors Studies; therefore, testing provided under K183590 is hereby incorporated as reference to this submission. As there were no changes to the Clip (HX-610 Series), product performance testing was not repeated for the Clips.
The subject Single Use Reloadable Clip Applicators (HX-810 Series) have satisfied the pre-defined acceptance criteria for each test item described in the test plan. Therefore, it is determined that the device is as safe, as effective, and performs as well as the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.