AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K240612 · May 31, 2024
CINA-VCF
Avicenna.Ai
Retrospective clinical anonymized CT scan cases
A retrospective study of 474 clinical cases was used to validate the device's performance (sensitivity, specificity, accuracy, and ROC AUC) in identifying vertebral compression fractures in a real-world clinical setting.
Retrospective, multinational and blinded study; Retrospective, blinded performance validation
Patients aged 50 years and over undergoing non-enhanced or contrast-enhanced CT scans of the chest and/or abdomen; Sample Size: 474 clinical anonymized cases; Number of Sites: Multiple (US and OUS)
886 series collected from multiple centers in the USA and France.
—
Retrospective, multinational and blinded study of 474 clinical anonymized cases.
3 (US-board-certified expert radiologists)
Indications for Use
CINA-VCF is a radiological computer aided triage and notification software indicated for use in patients aged 50 years and over undergoing non-enhanced or contrast-enhanced CT scans which include the chest and/or abdomen. The device is intended to assist hospital networks and appropriately trained medical specialists within the standard-of-care bone health setting in workflow triage by flagging and communication of suspected positive cases of Vertebral Compression Fractures (VCF) findings. CINA-VCF uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone application in parallel to the ongoing standard of care image interpretation. The device does not alter the original medical image, and it is not intended to be used as a diagnostic device. The results of CINA-VCF are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care.
Device Story
CINA-VCF is a radiological computer-aided triage and notification software; operates on off-the-shelf server/workstation. Inputs: non-enhanced or contrast-enhanced CT scans (chest/abdomen). Processing: deep learning models detect VCF at T1-L5; additional models locate spine, identify vertebrae, and exclude prior vertebroplasty or orthopedic material. Output: notification of suspected positive cases to a worklist application; displays unannotated, compressed grayscale preview images. Used in hospital networks by trained medical specialists. Operates in parallel to standard-of-care; does not alter original images or remove cases from standard worklists. Benefits: facilitates earlier triage and prioritization of suspected VCF cases by alerting clinicians to review images sooner than standard FIFO queues.
Clinical Evidence
Retrospective, multinational, blinded study (N=474 cases; 180 positive, 294 negative). Primary endpoint: ROC AUC for VCF detection. Results: AUC 0.974 (95% CI: 0.962-0.986), exceeding 0.95 goal. Sensitivity 95.2% (95% CI: 90.7-97.9%), Specificity 92.9% (95% CI: 89.4-96.5%), Accuracy 93.7%. Mean time-to-notification for true positives was 21.7 seconds vs 117.2 seconds for predicate.
Technological Characteristics
Software-only device; runs on standard off-the-shelf server/workstation. Uses deep learning AI models for image analysis. Interoperability via DICOM protocol. Integrates with CINA Platform or compatible medical image communications devices. No hardware components; no patient-contacting materials.
Indications for Use
Indicated for patients aged 50+ undergoing non-enhanced or contrast-enhanced CT scans of the chest and/or abdomen to assist in triage/prioritization of suspected Vertebral Compression Fractures (VCF).
Regulatory Classification
Identification
Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.
Special Controls
Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Avicenna.AI % John Smith Partner Hogan Lovells, US LLP 555 Thirteenth Street, NW Washington, District of Columbia 20004
Re: K240612
May 31, 2024
Trade/Device Name: Cina-VCF Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer aided triage and notification software Regulatory Class: Class II Product Code: QFM Dated: March 5, 2024 Received: March 5, 2024
Dear John Smith:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Submission Number (if known)
#### K240612
Device Name
CINA-VCF
#### Indications for Use (Describe)
CINA-VCF is a radiological computer aided triage and notification software indicated for use in patients aged 50 years and over undergoing non-enhanced or contrast-enhanced CT scans which include the chest and/or abdomen.
The device is intended to assist hospital networks and appropriately trained medical specialists within the standard-of-care bone health setting in workflow triage by flagging and communication of suspected positive cases of Vertebral Compression Fractures (VCF) findings.
CINA-VCF uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone application in parallel to the ongoing standard of care image interpretation. The device does not alter the original medical image, and it is not intended to be used as a diagnostic device.
The results of CINA-VCF are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(K) SUMMARY
#### AVICENNA.AI's CINA-VCF
#### Submitter
# K240612
# Applicant:
AVICENNA.AI ZI Athelia, 297 Av. du Mistral Bât. A, 13600 La Ciotat France
#### Contact Person:
Stephane Berger Regulatory Manager Phone: +33 6 12 12 28 13 E-mail: stephane.berger@avicenna.ai
Date prepared: May 28, 2024
#### Device Identification
| Name of Device: | CINA-VCF |
|-----------------------|--------------------------------------------------------------------|
| Classification Name: | Radiological computer-assisted triage<br>and notification software |
| Regulation No: | 21 CFR § 892.2080 |
| Product Code: | QFM |
| Regulatory Class: | Class II |
| Classification Panel: | Radiology devices |
#### Predicate Device
The CINA-VCF device is substantially equivalent to the following predicate device with regard to indications for use, performance, and technological characteristics:
| 510(k): | K222692 |
|----------------------|-----------------------------------------------------------------|
| Trade Name: | BriefCase |
| Manufacturer: | Aidoc Medical Vision Ltd. |
| Classification Name: | Radiological computer-assisted triage and notification software |
| Regulation No: | 21 CFR § 892.2080 |
| Product Code: | QFM |
| Regulatory Class: | Class II |
#### Device Description
{4}------------------------------------------------
CINA-VCF is a radiological computer-assisted triage and notification software device.
CINA-VCF runs on a standard "off the shelf" server/workstation and consists of VCF Image Processing Application, which can be integrated, deployed and used with the CINA Platform (cleared under K200855) or other compatible medical image communications devices. CINA-VCF receives nonenhanced or contrast-enhanced CT scans (which include the chest and/or abdomen) identified by the CINA Platform or other compatible medical image communications device, processes them using algorithmic methods involving execution of multiple computational steps to identify suspected presence of Vertebral Compression Fractures (VCF) findings and generates results files to be transferred by CINA Platform or a similar medical image communications device for output to a PACS system or workstation for worklist prioritization.
DICOM images are received, recorded and filtered before processing. The series are processed chronologically by running algorithms on each series to detect suspected positive findings of a vertebral compressions fracture (VCF).
The device uses deep learning models to detect VCF at the T1-L5 level. The models were trained endto-end on a dataset of 886 series collected from multiple centers in the USA and France satisfying the device protocol and representing a large distribution of scanner models from Siemens, Philips, GE and Canon (formerly Toshiba), acquisition protocols, spine presentation and fracture location and severity. Additional models, trained on subsets of this dataset, are used to locate the spine, identify the vertebra bodies and exclude vertebra which have been subjected to vertebroplasty or contains orthopedic material.
The Worklist Application displays all incoming suspect cases, each notified case is marked with an icon. In addition, compressed, grayscale, unannotated images that are captioned "not for diagnostic use" is displayed as a preview function. This compressed preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for diagnostic use beyond notification.
Presenting the specialist with worklist prioritization facilitates earlier triage by allowing prioritization of images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.
The CINA Platform is an example of medical image communications platform for integrating and deploying the CINA-VCF image processing applications. It provides the necessary requirements for interoperability based on the standardized DICOM protocol and services to communicate with existing systems in the hospital radiology department such as CT modalities or other DICOM nodes (DICOM router or PACS for example). It is responsible for transferring, converting formats, notifying of suspected findings and displaying medical device data such as radiological data. The CINA Platform server includes the Worklist client application which receives notifications from the CINA-VCF Image Processing application.
{5}------------------------------------------------
# Intended Use / Indications for Use
CINA-VCF is a radiological computer aided triage and notification software indicated for use in patients aged 50 years and over undergoing non-enhanced or contrast-enhanced CT scans which include the chest and/or abdomen.
The device is intended to assist hospital networks and appropriately trained medical specialists within the standard-of-care bone health setting in workflow triage by flagging and communication of suspected positive cases of Vertebral Compression Fractures (VCF) findings.
CINA-VCF uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone application in parallel to the ongoing standard of care image interpretation. The device does not alter the original medical image, and it is not intended to be used as a diagnostic device.
The results of CINA-VCF are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care.
# Summary of Performance Data
The following performance data were provided in support of the substantial equivalence determination.
## Software Verification and Validation Testing
CINA-VCF complies with DICOM (Digital Imaging and Communications in Medicine) - Developed by the American College of Radiology and the National Electrical Manufacturers Association. NEMA PS 3.1 - 3.20.
Avicenna.Al conducted extensive performance validation testing and software verification and validation testing of the CINA-VCF device as standalone software. CINA-VCF is tested against its user needs and intended use by the successful execution of planned software verification and validation testing included in this submission.
Software performance, validation and verification testing demonstrated that the CINA-VCF met all design requirements and specifications associated with the intended use of the software.
#### Standalone Performance Testing
Avicenna.Al conducted a retrospective, multinational and blinded study with the CINA-VCF application with the primary endpoint to evaluate the software's performance in identifying vertebral compression fractures (VCF) on non contrast-enhanced Chest and/or Abdominal CT images performed for another clinical indications than for thoraco-lumbar vertebral compression fractures assessments, including at least three (3) consecutive measurable vertebrae in the T1-L5 portion of the spine, in 474 clinical anonymized cases.
{6}------------------------------------------------
The data was provided from multiple US (66.9%) and OUS (33.1%) clinical sites. There were 180 (37.9%) positive cases (CT with VCF) and 294 (62.1%) negative cases included in the analyses. The data was acquired by 4 different scanner makers and 38 different scanner models.
Multiple subgroups of interest were considered in the analysis. Mean age of patients included in the study was 72.1 ± [SD] 10.1 yo (MIN = 50 yo and MAX = 100 yo) and 50.8% were female. Data were acquired from different regions across the US to account for race/ethnicity in the intended US patient population. Additional scanner parameters considered were slice thickness, number of detector rows, and kVp ranges, contrast vs non-contrast, imaging protocol (chest and/or abdomen), reconstruction kernel (soft/standard). Detailed subgroup analysis were reported in the labeling.
Device Area Under the Receiver Operating Characteristic curve (ROC AUC) was computed against the ground truth established by consensus of three US-board-certified expert radiologists, as the primary endpoint, in accordance with the established required technical method under the QFM product code. A case was considered positive if at least one moderate or severe vertebral compression fracture located within the thoracic or lumbar spine was identified by the experts.
Sensitivity, Specificity and Accuracy, were also assessed as additional analyses.
The secondary endpoint was CINA-VCF time-to-notification, which was compared to the predicate device for true positive cases only, as provided by the predicate device - BriefCase (Aidoc Medical -K222692).
The ROC AUC was 0.974 [95% Cl: 0.962 - 0.986], which exceeded the 0.95 performance goal, thus, achieving the primary endpoint.
Sensitivity and Specificity were 95.2% [95% Cl: 90.7% - 97.9%] and 92.9% [95% Cl: 89.4% - 96.5%], respectively, Similarly, the overall agreement (Accuracy) was 93.7% [95% Cl: 91.1% - 95.7%], which represents very good predictions.
The CINA-VCF prioritization and triage effectiveness (time-to-notification) was evaluated by the standalone per-case processing time of the device, which corresponds to time between the end of the DICOM reception (made available for VCF image processing) and the end of processing (positive or negative identification). The results are presented below:
| Time-to-<br>Notification (seconds) | MEAN ± SD | MEDIAN | 95% Cl | MIN | MAX |
|----------------------------------------------|------------|--------|---------------|-----|------|
| CINA-VCF<br>All cases<br>(N = 474) | 23.4 ± 8.4 | 21.0 | [22.7 - 24.2] | 9.0 | 60.0 |
| CINA-VCF<br>True Positive cases<br>(N = 158) | 21.7 ± 7.5 | 20.0 | [20.5 - 22.8] | 9.0 | 45.0 |
Table 1: Time-to-Notification for CINA-VCF Image Processing Application
The mean [95% Cl] time-to-notification for all included cases (n = 474) was estimated to be 23.4 195% Cl: 22.7 - 24.2] seconds for CINA-VCF.
{7}------------------------------------------------
When taking into account only true positive cases (n = 158), the mean [95% Cl] time-to-notification was 21.7 [95% Cl: 20.5 – 22.9] seconds for CINA-VCF and 117.2 [95% Cl: 98.64 – 135.85] seconds for BriefCase, the selected predicate device.
The performance testing of the CINA-VCF device establishes that the subject device is as safe and effective as the predicate device and compatible with the same clinical use, since the performances demonstrated the clinical effectiveness of the subject device and its ability to provide effective prioritization and triage, which is substantially equivalent to that of the predicate device. This establishes that CINA-VCF device achieves its intended use and is substantially equivalent to the predicate device.
#### Substantial Equivalence
The subject CINA-VCF and the predicate BriefCase device are both intended to assist hospital networks and appropriately trained medical specialists within the standard-of-care bone health setting in workflow triage by flagging and communication of suspected positive cases of Vertebral Compression Fractures (VCFx) findings based on the analysis of medical images acquired from radiological signal acquisition systems. The CINA device provides the CINA Platform in which the subject CINA-VCF prioritization and triage application can be integrated, deployed and used. The labeling of the subject and the predicate devices clearly states that the devices are not for diagnostic use. The subject CINA-VCF device and the predicate BriefCase are software packages with substantially similar technological characteristics and principles of operation, and incorporate deep learning Al algorithms that process images, and software to send notifications and to display unannotated preview images. In both devices, the labeling instructs the user to further evaluate and diagnose based only on the original images in the local PACS.
The subject CINA-VCF and the predicate BriefCase device operate in parallel to the standard of care workflow in the sense that they do not change the original image, do not provide any marking on the output preview, and do not remove images from the standard of care FIFO (First-in-first-out) queue, thus not disturbing standard interpretation of the images by the attending specialists. The subject and predicate devices achieve time-to-notification performance improvements in a similar range of time and thus contribute similarly to effective triage and early involvement of the specialist in evaluating suspected images of vertebral compression fractures.
The standalone performance and effectiveness assessment studies demonstrated that the CINA-VCF device performs as intended is therefore substantially equivalent to the BriefCase predicate device.
The table below compares the key features of the subject and the predicate devices.
# Table 2: Comparison of key features between CINA-VCF and predicate device (Aidoc Medical Vision Ltd)
{8}------------------------------------------------
| | Subject device:<br>CINA-VCF<br>Software | Predicate device: BriefCase Software<br>(K222692) |
|-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use<br>/ Indications<br>for Use | CINA-VCF is a radiological computer aided triage and notification software indicated for use in patients aged 50 years and over undergoing non-enhanced or contrast-enhanced CT scans which include the chest and/or abdomen. The device is intended to assist hospital networks and appropriately trained medical specialists within the standard-of-care bone health setting in workflow triage by flagging and communication of suspected positive cases of Vertebral Compression Fractures (VCF) findings.<br>CINA-VCF uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone application in parallel to the ongoing standard of care image interpretation. The device does not alter the original medical image, and it is not intended to be used as a diagnostic device.<br>The results of CINA-VCF are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care. | BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of chest and abdominal CT images. The device is intended to assist hospital networks and appropriately trained medical specialists within the standard-of-care bone health setting in workflow triage by flagging and communication of suspected positive cases of Vertebral Compression Fractures (VCFx) findings.<br>BriefCase uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone application in parallel to the ongoing standard of care image interpretation. The device does not alter the original medical image and is not intended to be used as a diagnosis device.<br>The results of BriefCase are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care. |
| User<br>population | Trained medical specialists within standard-of-care bone health settings | Hospital networks and appropriately trained medical specialists within the standard of-care bone health setting |
| Anatomical<br>region of<br>interest | Chest and/or abdomen. | Chest and abdomen |
| Data<br>acquisition<br>protocol | Non-enhanced / contrast-enhanced CT scans of the chest and/or abdomen. | Chest and abdominal CT scans |
| Passive<br>notification | Yes | Yes |
| | Subject<br>device: CINA-VCF<br>Software | Predicate device: BriefCase Software<br>(K222692) |
| only, parallel<br>workflow tool | | |
| Interference<br>with standard<br>workflow | No | No |
| Algorithm | Artificial intelligence algorithm | Artificial intelligence algorithm |
| Preview<br>images | Presentation of compressed,<br>grayscale, unannotated images<br>that is marked "not for diagnostic<br>use" | Presentation of compressed, low-quality,<br>grayscale, unannotated image that is<br>captioned "not for diagnostic use" |
| Alteration of<br>original<br>image | No | No |
| Removal of<br>cases from<br>worklist<br>queue | No. The device operates in parallel<br>with the standard of care, which<br>remains the default option for all<br>cases. | No. The device operates in parallel with<br>the standard of care, which remains the<br>default option for all cases. |
| Structure | -VCF image processing application<br>- Compatibility of use with the CINA<br>Platform device (worklist and<br>Image Viewer) or other medical<br>image communications device | - AHS module (orchestrator, image<br>acquisition);<br>- ACS module (image processing);<br>- Aidoc Desktop application for workflow<br>integration (feed and non-diagnostic<br>Image Viewer). |
{9}------------------------------------------------
# Conclusion
CINA-VCF and BriefCase have the same intended use and substantially similar indications, technological characteristics, and principles of operation. The following is a summary of the substantial equivalence comparison:
- -The predicate device is legally marketed.
- CINA-VCF has the same intended use as the predicate device, i.e., Aidoc Medical Vision -Ltd. 's BriefCase, and therefore it may be found substantially equivalent.
- -CINA-VCF and the predicate device have very similar indications for use.
- -CINA-VCF has similar technological characteristics as the predicate device, i.e., the deployment of an artificial intelligence algorithm with a database of images.
{10}------------------------------------------------
- Standalone performance study demonstrates that the CINA-VCF and the predicate । BriefCase raise the same types of safety and effectiveness questions, namely, accurate detection of findings within the processed study.
Accordingly, the subject CINA-VCF device is substantially equivalent to the predicate BriefCase device
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.