K240605 · Vitrolife Sweden AB · MQL · Aug 7, 2024 · Obstetrics/Gynecology
Device Facts
Record ID
K240605
Device Name
Ultra RapidWarm™ Blast
Applicant
Vitrolife Sweden AB
Product Code
MQL · Obstetrics/Gynecology
Decision Date
Aug 7, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6180
Device Class
Class 2
Indications for Use
Ultra RapidWarm™ Blast is indicated for warming of vitrified human blastocyst stage embryos.
Device Story
Ultra RapidWarm™ Blast is a single-solution reproductive medium used for warming vitrified human blastocyst stage embryos. It consists of a MOPS buffered solution containing gentamicin, human serum albumin (HSA), and sucrose. The device simplifies the warming procedure to a single step (2 minutes) compared to traditional multi-step protocols. It is supplied in gamma-sterilized 5 mL PETG bottles. The medium is used in clinical IVF laboratory settings by embryologists. By facilitating the transition of embryos from a vitrified state to a viable state for transfer, the device supports assisted reproductive technology procedures, aiming to achieve clinical pregnancy and live birth rates comparable to standard multi-step warming methods.
Clinical Evidence
Clinical evidence includes a literature review and a prospective pilot study (86 cycles) followed by a cohort study (868 cycles) conducted at a university-based IVF center. The study compared a single-step warming protocol (using the predicate as a surrogate) against the conventional multi-step protocol. Results showed comparable embryo survival (97.8%), clinical pregnancy rates (38.2% vs 39.2%), and live birth rates (28.3% vs 29.6%) for single-step versus multi-step procedures, respectively.
Technological Characteristics
Single-solution MOPS buffered medium containing gentamicin, HSA, and sucrose. Osmolality: 547 ± 20 mOsm/kg; pH: 7.30 ± 0.10. Packaging: 5 mL PETG bottles with HDPE closures. Sterilization: Aseptic filtration. Shelf-life: 9 weeks (14 days after opening). Complies with ISO 13408-1:2008/Amd1:2013, ISO 13408-2:2018, USP <71>, USP <85>, USP <791>, USP <785>, and ASTM D4169-22.
Indications for Use
Indicated for warming of vitrified human blastocyst stage embryos.
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
August 7, 2024
Vitrolife Sweden AB Nina Arvidsson Official Applicant Gustaf Werners gata 2 Vastra Frolunda, 42132 SWEDEN
Re: K240605
> Trade/Device Name: Ultra RapidWarm™ Blast Regulation Number: 21 CFR 884.6180 Regulation Name: Reproductive Media and Supplements Regulatory Class: II Product Code: MQL Received: July 9, 2024
Dear Nina Arvidsson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}------------------------------------------------
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael T. Bailey -S
For Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240605
Device Name Ultra RapidWarm™ Blast
Indications for Use (Describe) Ultra RapidWarm™ Blast is indicated for warming of vitrified human blastocyst stage embryos.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the word "Vitrolife" in blue font, with a stylized blue wave-like graphic to the right of the word. The font is a serif font, and the wave graphic is a simple, curved line. The overall design is clean and modern. The color is a dark blue.
# 510(k) Summary - K240605
#### 1. Submitter Information
Submitted by: Vitrolife Sweden AB Gustaf Werners gata 2 SE - 421 32 Västra Frölunda Sweden
| Contact Person: | Nina Arvidsson |
|-----------------|-----------------------------|
| | Vitrolife Sweden AB |
| | Gustaf Werners gata 2 |
| | SE - 421 32 Västra Frölunda |
| | Sweden |
| Phone: | +46 31 721 80 00 |
| Fax: | +46 31 721 80 90 |
| Email: | narvidsson@vitrolife.com |
- 2. Date Prepared:
August 6, 2024
- 3. Device Identification
| Trade Name: | Ultra RapidWarm™ Blast |
|----------------------|-------------------------------------------------------------------|
| Common Name: | Warming medium for vitrified blastocysts |
| Regulatory Class: | Class II |
| Regulation Number: | 21 CFR 884.6180 |
| Regulation Name: | Reproductive media and supplements |
| Product Code: | MQL (Media, Reproductive) |
| 4. Predicate Device: | RapidWarm™ Blast (K101003) manufactured by Vitrolife<br>Sweden AB |
The predicate device has not been subject to a design related recall.
# 5. Device Description
Ultra RapidWarm Blast is intended for warming vitrified human blastocyst stage embryos. The device consists of a single solution composed of a MOPS buffered solution containing gentamicin, human serum albumin (HSA) and sucrose.
The medium is aseptically filtered into gamma sterilized 5 mL PETG bottles with HDPE closures and a tamper evident seal. The medium can be used for up to 14 days after bottle opening.
### 6. Indications for Use
Ultra RapidWarm™ Blast is indicated for warming of vitrified human blastocyst stage embryos.
{5}------------------------------------------------
# Ultra RapidWarm™ Blast
Image /page/5/Picture/2 description: The image shows the word "Vitrolife" in blue font. To the right of the word is a blue swoosh symbol. The font is a serif font and the color is a dark blue.
# 7. Comparison of intended use and technological characteristics of the subject and predicate devices
A comparison of the indications for use and technological characteristics of the subject device and the predicate device is shown in the table below:
| | Subject Device<br>(K240605) | Predicate Device<br>(K101003) | Comparison |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Ultra RapidWarmTM Blast | RapidWarm Blast | N/A |
| Indications for<br>Use | Ultra RapidWarmTM Blast<br>is indicated for warming of<br>vitrified human blastocyst<br>stage embryos. | RapidWarmTM Blast is<br>indicated for warming of<br>vitrified human blastocyst<br>stage embryos. | Same |
| Warming<br>procedure | Ultra Warm Blast (250 mM<br>sucrose): 2 minutes | Warm 1 Blast (250 mM<br>sucrose): 2 minutes<br>Warm 2 Blast (125 mM<br>sucrose): 3 minutes<br>Warm 3 Blast: 5-10<br>minutes | Different: The<br>warming procesure<br>of the subject and<br>predicate devices<br>are not the same.<br>Differences in<br>warming procedure<br>do not raise<br>different questions<br>of safety and<br>effectiveness<br>(S&E). |
| Packaging | 5 mL PETG bottle with<br>HDPE closure | 10 mL PETG bottle with<br>HDPE closure | Different: The<br>packaging of the<br>subject and<br>predicate devices<br>are not the same.<br>Differences in<br>device packgaing<br>do not raise<br>different questions<br>of S&E. |
| Storage<br>conditions | Store dark at +2 to +8 °C | Store dark at +2 to +8 °C | Same |
| pH | 7.30 ± 0.10 | 7.30 ± 0.10 | Same |
| Osmolality<br>(mOsm/kg) | 547 ± 20 | 528 ± 20 / 394±12 / 261±5 | Different: The<br>osmolality of the<br>subject and<br>predicate devices<br>are not the same.<br>Differences in<br>osmolality do not<br>raise different<br>questions of S&E. |
| Bacterial<br>endotoxins<br>(LAL assay) | < 0.5 EU/mL | < 0.5 EU/mL | Same |
| Mouse Embryo<br>Assay (MEA) | One-cell system: ≥80%<br>embryos developed to<br>expanded blastocyst at 96<br>hours after 2 minutes of<br>exposure | One-cell: ≥ 70% re-<br>expanded blastocyst 24<br>hours post-test | Different: The<br>MEA specification<br>of the subject and<br>predicate devices<br>are not the same.<br>This difference<br>does not raise a<br>different question of<br>S&E. |
| Sterilization<br>method | Aseptic filtration | Aseptic filtration | Same |
| Shelf-Life | 9 weeks<br>14 days (open bottle) | 24 weeks | Different: The<br>subject device has<br>a shorter shelf-life<br>than the predicate<br>device. Differences<br>in shelf-life do not<br>raise different<br>questions of S&E. |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Vitrolife. The word "Vitrolife" is written in a blue serif font. To the right of the word is a blue curved shape that resembles a check mark.
Ultra RapidWarm™ Blast
The subject and predicate device have the same indications for use and intended use. The subject and predicate device have different technological characteristics, including differences in warming procedure. specifications, and shelf-life. The different technological characteristics do not raise different questions of safety and effectiveness.
# 8. Summary of Non-Clinical Performance Testing
The following studies have been performed to support substantial equivalence to the predicate device:
- Aseptic filtration and filling validation testing that met the requirements of ISO 13408-. 1:2008/Amd1:2013 and ISO 13408-2:2018.
- . Shelf-life testing was conducted to support a 9-week shelf-life for the subject device through demonstration that the product specifications (shown below) were met at time 0 and after realtime aging. Testing was also included on aged samples demonstrating that medium in bottles can maintain their specifications after 14 days of simulated use conditioning after bottle opening. Testing conducted is shown below:
- Sterility testing per USP <71> (acceptance criterion: no microbial growth) -
- Bacterial endotoxins testing per USP <85> (acceptance criterion: <0.5 EU/ml) -
- pH measurements per USP <791> (acceptance criterion: 7.30±0.10) -
- -Osmolality per USP <785> (acceptance criterion: 547±20 mOsm/kq)
- Mouse embryo assay One-cell system: ≥80% embryos developed to expanded blastocyst at -96 hours after 2 minutes of exposure.
- . Transportation testing was conducted according to ASTM D4169-22 (DC-13) and cap/seal leak testing according to USP <1207.2>.
# 9. Summary of Clinical Performance Testing
A clinical evidence summary providing information to support the safety and performance of the subject device was included in this submission. The evaluation covers literature review and data from a study on a single step warming protocol identical to the subject device. The literature review summarized research
{7}------------------------------------------------
### Ultra RapidWarm™ Blast
Image /page/7/Picture/2 description: The image contains the word "Vitrolife" in blue font. To the right of the word is a blue swoosh. The font is a serif font.
conducted with single-step warming versus multi-step standard warming procedures, and the results showed comparable embryo survival rate, clinical pregnancy rate, and live birth rate between blastocysts warmed with the single-step and multi-step warming procedures, indicating that a single-step warming procedure is equivalent to multi-step warminq procedures. The clinical study used the Warm 1 Blast (K101003) as a surrogate for the subject device and compared the effectiveness of warming using a simplified warming protocol (2 minutes of warming in Warm 1 Blast) compared to standard warming protocol using RapidWarm Blast (K101003). The study was conducted in a universitybased IVF center in France. A prospective pilot study with 86 cycles (52 blastocysts in the control group and 42 blastocysts in the treatment group) was conducted, followed by a cohort study including 868 cycles (578 blastocysts in the control group and 336 blastocysts in the treatment group). The results from the cohort study showed that embryo survival rate after warming (97.8%), clinical pregnancy rate (38.2% vs. 39.2%), and live birth rate (28.3% vs. 29.6%) obtained with the single-step warming and with the conventional standard warming protocol, respectively, were comparable.
### 9. Conclusion
The results of the performance testing described above demonstrate that the subject device is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.