Ti-Base & Master Fix are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations. All digitally-designed Ti-Base and Master Fix are intended to be sent to an ARUM DENTISTRY validated milling center for manufacture.
Device Story
Two-piece dental abutment system comprising titanium base (Ti-Base or Master Fix) and CAD/CAM-designed zirconia superstructure; titanium base connects to endosseous dental implant via abutment screw; zirconia superstructure milled at validated center and cemented to titanium base; used by dental professionals in clinical settings; provides support for prosthetic restorations to restore chewing function; design features include anti-rotation mechanisms and cement adhesion slots; titanium components fabricated from Ti-6Al-4V Eli; zirconia components from Y-TZP blocks; non-sterile; single-use.
Clinical Evidence
No clinical data included. Evidence consists of bench testing, including biocompatibility (ISO 10993-5, ISO 10993-12), moist heat sterilization validation (ISO 17665-1, ISO 17665-2), and non-clinical MRI safety assessment based on scientific rationale and published literature.
Technological Characteristics
Two-piece abutment system; Ti-6Al-4V Eli (ASTM F136) titanium base; Y-TZP zirconia superstructure; cement-retained; non-sterile; end-user sterilization; dimensions: Ti-Base Ø3.35mm, Master Fix Ø4.5-6.5mm; anti-rotation design; compatible with specific ARUM implant systems.
Indications for Use
Indicated for patients requiring prosthetic restorations supported by endosseous dental implants in the maxillary or mandibular arch.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
DESS Dental Smart Solutions by Terrats Medical SL (K191986)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
ARUM DENTISTRY Co., Ltd. Won-Yi Choi Official Correspondent 23, Gukjegwahak 11-ro, Yuseong-gu Daeieon, 34002 REPUBLIC OF KOREA
June 5, 2024
Re: K240603
Trade/Device Name: Ti-Base & Master Fix Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: March 4, 2024 Received: March 4, 2024
### Dear Won-Yi Choi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240603
Device Name
Ti-Base & Master Fix
#### Indications for Use (Describe)
The Ti-Base & Master Fix are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations. All digitally-designed Ti-Base and Master Fix are intended to be sent to an ARUM DENTISTRY validated milling center for manufacture.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:10px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Arum Dentistry. The logo consists of a geometric shape on the left and the words "ARUM DENTISTRY" on the right. The geometric shape is made up of a red triangle on top and a smaller yellow triangle on the bottom left. The words "ARUM DENTISTRY" are in a bold, sans-serif font.
#### K240603
## 7. 510(K) Summary
#### Submitter
ARUM DENTISTRY Co., Ltd. Won-Yi Choi 23, Gukjeqwahak 11-ro, Yuseong-qu Daejeon, 34002 Republic of Korea Email: arum_ra@arumdentistry.com Tel. +82-42-935-3644 Fax. +82-42-935-3633
#### Device Information
- . Trade Name: Ti-Base & Master Fix
- . Common Name: Endosseous Dental Implant Abutment
- . Classification Name: Abutment, Implant, Dental, Endosseous
- Product Code: NHA ●
- . Panel: Dental
- . Regulation Number: 21 CFR 872.3630
- . Device Class: Class II
- Date Prepared: 06/04/2024 ●
#### Predicate Devices
The subject device is substantially equivalent to the following predicate devices:
#### Primary Predicate
- . K191986, DESS Dental Smart Solutions by Terrats Medical SL
#### Reference Device
- K190112, Non-Sterile Zirconia Block (Model: Finebase, Montblanc, Trione HT, Trione C, . Trione HT+) by Fine Advanced Compound Co., Ltd
- K193260, U-Cem Premium & MAZIC Cem by Vericom Co., Ltd. .
- . K213506, NB 1 SA Implant System by ARUM DENTISTRY Co., Ltd.
- K230725, NB Implant System by ARUM DENTISTRY Co., Ltd. ●
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo for "Arum Dentistry". The logo consists of a geometric shape on the left and the text "ARUM DENTISTRY" on the right. The geometric shape is made up of two triangles, one red and one light orange. The text "ARUM" is in a larger font than "DENTISTRY", and both are in a dark brown color.
### General Description
#### Device Components
- 1) Abutment
- Ti-Base
- Master Fix
- Abutment Screw
The Ti-Base and Master Fix consist of a two-piece abutment, where the titanium base is premanufactured abutment that will be used to support a CAD/CAM designed superstructure (the second part of the two-piece abutment) that compose the final abutment.
| No. | Device Name | Dimension |
|-----|----------------------------------------|---------------------------------------------------------------------------------------|
| 1 | Ti-Base | Ø3.35 (D) x 4.0 mm (Post Height) |
| 2 | Master Fix | Ø4.5, 5.5, 6.5 (D) x 3.5, 4.0, 5.0, mm (Post Height) |
| 3 | Abutment Screw<br>(Cleared in K213506) | Ø 2.35 (D) x 8.4 mm(L) |
| 4 | Master Fix Screw | Ø2.25 (D) x 10.3, 11.3, 11.8, 12.3, 12.8, 13.3, 13.8, 14.3,<br>14.8, 15.3, 15.8mm (L) |
The dimension ranges of the subject device are below:
Below are the abutment's features:
| Name | Uses | Surface |
|----------------------------------------|-------------------------------------------------------------------------------------------------|---------|
| Ti-Base | The Ti-Base is used as a support of prosthesis to<br>restore the patient's chewing function. | N/A |
| Master Fix | The Master Fix is used as a support of prosthesis to<br>restore the patient's chewing function. | N/A |
| Abutment Screw<br>(Cleared in K213506) | The Abutment Screw is used for connect fixture and<br>abutment. | N/A |
| Master Fix Screw | The Master Fix Screw is used for connect fixture and<br>abutment. | N/A |
Ti-Base, Master Fix, Abutment Screw and Master Fix Screw are provided non-sterilized.
Ti-Base is enclosed with Abutment Screw in a packing. Master Fix is enclosed with Master Fix Screw in a packing. These devices are intended for single use only. All digitally designed custom abutments for use with Ti-Base or Master Fix are to be sent to ARUM DENTISTRY validated milling center for manufacture. All superstructures are to be manufactured from zirconia (cleared
{6}------------------------------------------------
ARUM
**DENTISTRY**
K240603
K190112). Digitally designed CAD/CAM abutments must have a 0.5 mm minimum gingival height dimension.
The Titanium Base abutment is composed of two-piece abutment that is a titanium base at the bottom and a zirconia superstructure (CAD/CAM patient specific superstructure) at the top. The zirconia superstructure is straight only and is not to be designed to provide an angle or divergence correction.
For the Ti-Base the design parameters for the CAD/CAM zirconia superstructure are:
Minimum wall thickness - 0.5 mm;
Minimum post height* for single-unit restorations - 4.5 mm;
Maximum gingival height - 5.0 mm;
Minimum gingival height - 0.5 mm;
Angulation - 0°
and All zirconia superstructures are for straight abutments only.
*Post Height is measured above the gingival height of the final patient-matched design
For the Master Fix, the design parameters for the CAD/CAM zirconia superstructure are:
Minimum wall thickness - 0.5 mm;
Minimum post height* for single-unit restorations - 4.0 mm;
Maximum qinqival height - 5.0 mm:
Minimum gingival height - 0.5 mm;
Angulation - 0°
and All zirconia superstructures are for straight abutments only.
*Post Height is measured above the gingival height of the final patient-matched design
The Ti-Base and Master Fix is compatible with the following implant systems.
| Manufacturer | 510(k) No. | Implant System<br>Compatibility | Dimension |
|--------------------------------|------------|---------------------------------|-------------------------------------|
| ARUM<br>DENTISTRY<br>Co., Ltd. | K213506 | NB 1 SA Implant System | Ø 3.8, 4.0, 4.5, 5.0, 5.5, 6.0, 6.5 |
| ARUM<br>DENTISTRY<br>Co., Ltd. | K230725 | NB Implant System | Ø 3.8, 4.0, 4.5, 5.0, 5.5, 6.0, 6.5 |
The titanium bases are used as part of a two-piece abutment, where the base is premanufactured from titanium alloy (Ti-6Al-4V Eli) and the top half is a CAD/CAM zirconia superstructure, milled at a validated milling center. These pieces are cemented together to form the final abutment.
Raw material Zirconia Block
#### K190112, Non-Sterile Zirconia Block by FINE ADVANCED COMPOUND Co., Ltd.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Arum Dentistry. The logo consists of a geometric shape on the left and the words "ARUM DENTISTRY" on the right. The geometric shape is made up of two triangles, one red and one light orange. The words "ARUM DENTISTRY" are in a bold, sans-serif font.
## Dental Cement
K193260, U-Cem Premium & MAZIC Cem by Vericom Co., Ltd.
### Zirconia structure Details
| Specification | Zirconia Block |
|------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade name | Non-Sterile Zirconia Block |
| Common name | Dental zirconia blanks/blocks |
| Manufacturer | FINE ADVANCED COMPOUND Co., Ltd |
| 510(k) No. | K190112 |
| Product code | EIH |
| Regulatory class | Class II |
| Intended use | Non-Sterile Zirconia Block (Model name: Finebase, Montblanc, Trione HT,<br>Trione C, Trione HT+) are indicated for the production of artificial teeth in<br>fixed or removable dentures, or for jacket crowns, facings, and veneers. |
## Indication for Use
Ti-Base & Master Fix are intended to be used in conjunction with endosseous dental implants in
the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally-designed Ti-Base and Master Fix are intended to be sent to an ARUM DENTISTRY validated milling center for manufacture.
### Materials
Pre-Manufactured titanium components of Abutment Screws are fabricated from Ti-6Al-4V Eli (Conforming to ASTM F136).
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains a logo for "ARUM DENTISTRY". The logo consists of a red triangle and a smaller, light orange triangle on the left side. To the right of the triangles, the word "ARUM" is written in bold, dark brown letters, with the word "DENTISTRY" written in the same font and color directly below it.
# Summaries of Technology Characteristics
# 1) Ti-Base
| | Subject Device | Primary Predicate |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | ARUM DENTISTRY Co., Ltd. | Terrats Medical SL |
| Device Name | Ti-Base & Master Fix | DESS Dental Smart Solutions |
| 510(k) No. | N/A | K191986 |
| Intended Use/<br>Indications for<br>use | Ti-Base & Master Fix are intended to be used in<br>conjunction with endosseous dental implants in the<br>maxillary or mandibular arch to provide support for<br>prosthetic restorations.<br>All digitally-designed Ti-Base and Master Fix are<br>intended to be sent to an ARUM DENTISTRY<br>validated milling center for manufacture. | DESS Dental Smart Solutions abutments are<br>intended to be used in conjunction with endosseous<br>dental implants in the maxillary or mandibular arch to<br>provide support for prosthetic restorations.<br>All digitally designed custom abutments for use with<br>Ti Base abutments or Pre-milled (Blank) abutments<br>are to be sent to a Terrats Medical validated milling<br>center for manufacture. |
| Diameter (⌀) | 3.35 | 4.5 - 6.5 |
| Material | Ti-6Al-4V Eli | Ti-6Al-4V Eli |
| Superstructure<br>Material | Zirconia (Y-TZP)<br>Non-Sterile Zirconia Block, cleared in K190112 | Zirconia (Y-TZP)<br>(Not stated in 510(k) Summary) |
| Cement Material | Cement<br>U-Cem Premium & MAZIC Cem, cleared in K193260 | Multilink Hybrid Abutment Cement<br>(Ivoclar Vivadent AG, Cleared in K130436) |
| Sterilization | End User Sterilization | End User Sterilization |
| Design<br>Parameter | Minimum wall thickness – 0.5 mm;<br>Minimum post height for single-unit restorations – 4.5<br>mm;<br>Maximum gingival height – 5.0 mm<br>Minimum gingival height – 0.5 mm<br>Angulation - 0° | Minimum wall thickness - 0.4 mm<br>Minimum post height - 4.2 mm<br>Maximum gingival height - 6.0 mm |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the logo for Arum Dentistry. The logo consists of a red triangle and a yellow triangle on the left side of the image. To the right of the triangles, the words "ARUM" and "DENTISTRY" are written in a bold, dark brown font.
Г
| | <b>1. Similarities</b><br>The Ti-Base has the same intended use for, technological characteristics to the K191986. The Ti-Base have<br>same device characteristics with the Primary Predicate such as material, sterilization, intended use, functions,<br>structure. |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Substantial<br>Equivalent<br>Discussion | <b>2. Differences</b><br>The minor differences with the primary predicated K191986 are slightly different dimension and insignificant<br>shapes. However, the subject diameters are in the range of diameters of primary predicate and this dimensional<br>difference doesn't affect device safety and effectiveness. The shape is only for the manufacture's properties<br>such as design and compatible to their own dental implant design. These minor differences do not impact safety<br>and effectiveness because these differences are related to the compatible implant design and are mitigated by<br>the mechanical performing testing. |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image contains the logo for Arum Dentistry. The logo features a geometric design on the left, consisting of a red triangle above a smaller, light orange triangle. To the right of the geometric design is the text "ARUM" in a bold, dark brown font, with the word "DENTISTRY" directly below it in the same font and color. The overall design is clean and modern, conveying a sense of professionalism.
2) Master Fix
| | Subject Device | Primary Predicate |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | ARUM DENTISTRY Co., Ltd. | Terrats Medical SL |
| Device Name | Ti-Base & Master Fix | DESS Dental Smart Solutions |
| 510(k) No. | N/A | K191986 |
| Intended Use/<br>Indications for<br>use | Ti-Base & Master Fix are intended to be used in<br>conjunction with endosseous dental implants in the<br>maxillary or mandibular arch to provide support for<br>prosthetic restorations.<br>All digitally designed abutments for use with Ti-Base<br>and Master Fix are intended to be manufactured at an<br>ARUM DENTISTRY validated milling center. | DESS Dental Smart Solutions abutments are<br>intended to be used in conjunction with endosseous<br>dental implants in the maxillary or mandibular arch to<br>provide support for prosthetic restorations.<br>All digitally designed custom abutments for use with<br>Ti Base abutments or Pre-milled (Blank) abutments<br>are to be sent to a Terrats Medical validated milling<br>center for manufacture. |
| Diameter (ø) | 4.5, 5.5, 6.5 | 4.5 - 6.5 |
| Material | Ti-6Al-4V Eli | Ti-6Al-4V Eli |
| Superstructure<br>Material | Zirconia (Y-TZP)<br>Non-Sterile Zirconia Block, cleared in K190112 | Zirconia (Y-TZP)<br>(Not stated in 510(k) Summary) |
| Cement Material | Cement<br>U-Cem Premium & MAZIC Cem, cleared in K193260 | Multilink Hybrid Abutment Cement<br>(Ivoclar Vivadent AG, Cleared in K130436) |
| Sterilization | End User Sterilization | End User Sterilization |
| Design<br>Parameter | Minimum wall thickness - 0.5 mm;<br>Minimum post height for single-unit restorations - 4.5 mm;<br>Maximum gingival height – 5.0 mm<br>Minimum gingival height - 0.5 mm<br>Angulation - 0° | Minimum wall thickness – 0.4 mm<br>Minimum post height - 4.2 mm<br>Maximum gingival height - 6.0 mm |
| Substantial<br>Equivalent<br>Discussion | <b>1. Similarities</b><br>The Master Fix has the same intended use for, technological characteristics to the K191986. The Master Fix<br>have same device characteristics with the Primary Predicate such as material, sterilization, intended use,<br>functions, structure. | |
| | <b>2. Differences</b><br>The predicate device has a notch for anti-rotation, but the subject device features an enhanced anti-<br>rotation design with two flat surfaces to prevent rotation between the zirconia superstructure and the<br>abutment, allowing for precise positioning and orientation of the superstructure.<br>The subject device features a slot in the post design, providing a larger area for improved cement adhesion,<br>with differences in shape and design compared to the predicate device. These differences have been validated<br>through mechanical performance testing and do not affect safety or effectiveness. Therefore, the design<br>changes aim to improve the stability and retention of the prosthesis. | |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image contains the logo for Arum Dentistry. The logo consists of a red triangle and a yellow triangle on the left side of the image. To the right of the triangles, the words "ARUM" and "DENTISTRY" are written in a bold, dark brown font.
Г
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image contains the logo for Arum Dentistry. The logo consists of a geometric shape on the left and the text "ARUM DENTISTRY" on the right. The geometric shape is made up of two triangles, one red and one light orange. The text "ARUM" is above the text "DENTISTRY", and both are in a dark brown color.
## 3) Abutment Screw
| | Subject Device | Reference Device |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| Manufacturer | ARUM DENTISTRY Co., Ltd. | ARUM DENTISTRY Co., Ltd. |
| Device Name | Ti-Base & Master Fix | NB 1 SA Implant System |
| 510(k) No. | N/A | K213506 |
| Material | Ti-6Al-4V Eli | Ti-6Al-4V Eli |
| Range of Diameters (ø) | 2.2 | 2.35 |
| Length (mm) | 10.8 | 8.4 |
| Sterilization | End User Sterilization | End User Sterilization |
| Substantial Equivalent<br>Discussion | 1. Similarities<br>The Abutment Screw has same indication for use, principle of operation, functions, and material to the<br>predicate K213506.<br>2. Differences<br>The difference between the subject and reference device is the dimensions of the device. However, it<br>does not affect device's fundamental functions and safety; therefore, it is substantial equivalent. | |
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for Arum Dentistry. The logo consists of a geometric shape on the left and the words "ARUM DENTISTRY" on the right. The geometric shape is made up of two triangles, one red and one light orange. The words "ARUM DENTISTRY" are in a bold, sans-serif font, with "ARUM" on top and "DENTISTRY" on the bottom.
# 4) Master Fix Screw
| | Subject Device | Reference Device |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| Manufacturer | ARUM DENTISTRY Co., Ltd. | ARUM DENTISTRY Co., Ltd. |
| Device Name | Ti-Base & Master Fix | NB 1 SA Implant System |
| 510(k) No. | N/A | K213506 |
| Material | Ti-6Al-4V Eli | Ti-6Al-4V Eli |
| Range of Diameters (ø) | 2.25 | 2.35 |
| Length (mm) | 10.3, 11.3, 11.8, 12.3, 12.8, 13.3, 13.8, 14.3,<br>14.8, 15.3, 15.8 | 8.4 |
| Sterilization | End User Sterilization | End User Sterilization |
| Substantial Equivalent<br>Discussion | 1. Similarities<br>The Abutment Screw has same indication for use, principle of operation, functions, to secure a dental<br>abutment to a dental implant in the upper or lower jaw, and material to the predicate K213506.<br><br>2. Differences<br>The screw's design features a larger head to increase retention in screw-retained prostheses. The<br>extended head fills the abutment's empty space, enhancing stability and preventing screw fractures.<br>Additionally, the design facilitates easier access for the screw driver.<br>However, it does not affect device's fundamental functions and safety; therefore, it is substantial<br>equivalent. | |
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image contains the logo for Arum Dentistry. The logo consists of a red triangle and a yellow triangle on the left side. To the right of the triangles is the text "ARUM" in bold, followed by "DENTISTRY" in bold on the line below.
#### Performance Data
Non-clinical data submitted to demonstrate substantial equivalence include: biocompatibility testing according to ISO 10993-5 and ISO 10993-12 and moist heat sterilization validation according to ISO 17665 -1 and ISO 17665-2.
No clinical data were included in this submission.
#### MR Environment Condition
Non-Clinical worst-case MRI review was performed to evaluated the NB Mini Implant System devices in the MRI environment using scientific rationale and published literature (e.g., Terry O. Woods, Jana Delfino, & Sunder Rajan. (2019). Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices. Journal of Testing and Evaluation 49.2, 783-795), based on the entire system including all variations (all compatible implant bodies, dental abutments, and fixation screws) and material composition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque.
#### Conclusion
The Subject device and the Primary predicate device have the same intended use, have the same technological characteristics, and are made of the same materials. The Subject device and the Primary predicate device encompass the same range of physical dimensions and are to be sterilized using the same methods. Although there is slightly difference design of the anti-rotation feature, this difference is the same purpose. This minor difference do not impact safety and effectiveness because this difference are related to increase the adhesion of cement to the prosthesis. The data included in this premarket notification demonstrate substantial equivalence to the primary predicate device listed above. Therefore, the subject device and primary predicate device are substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.