The Wolff's Law Anterior Cervical Plate System is in the temporary stabilization of the cervical spine (C2-C7) during the development of solid spinal fusion in patients with instability caused by the following: • Degenerative Disc Disease (DDD) – as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies; - · Spondylolisthesis; - Spinal Stenosis: - · Trauma (including fractures); - · Tumor; - · Deformity (i.e., scoliosis, kyphosis, and/or lordosis); - · Pseudoarthrosis; and - · Failed previous fusions
Device Story
Dynamic cervical plate system for anterior cervical spine (C2-C7) stabilization; intended for use during spinal fusion. System comprises one-level plates of varying lengths and both fixed- and variable-angle screws. Incorporates nitinol compression element to provide continuous compression at each level. Used by surgeons in clinical settings to address spinal instability. Device provides mechanical support to the cervical spine; continuous compression facilitates fusion. Benefits patient by stabilizing spine during healing process. Provided sterile for single use.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical mechanical testing per ASTM F1717 (static/dynamic compression, torsion, tension), ASTM F543 (screw pullout/pushout), ASTM F2129/F3044 (corrosion), ASTM F1877 (particulate analysis), and ASTM F2004 (transformational temperature).
Technological Characteristics
Materials: Titanium alloy Ti6Al4V ELI (ASTM F136) and superelastic nitinol (ASTM F2063). Design: Dynamic cervical plate with nitinol compression element; fixed and variable-angle screws. Sterilization: Provided sterile. Single-use.
Indications for Use
Indicated for temporary stabilization of cervical spine (C2-C7) during solid spinal fusion development in patients with instability due to Degenerative Disc Disease, spondylolisthesis, spinal stenosis, trauma (fractures), tumor, deformity (scoliosis, kyphosis, lordosis), pseudoarthrosis, or failed previous fusions.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
Predicate Devices
DynaTran™ Anterior Cervical Plating (ACP) System (K083020)
SC-AcuFix® Ant-Cer Dynamic Cervical Plating System (K143626)
{0}------------------------------------------------
May 24, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters FDA in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Spinal Simplicity LLC Adam Rogers Vice President of Regulatory and Engineering 6363 College Blvd Ste 320 Overland Park. Kansas 66211
#### Re: K240592
Trade/Device Name: Wolff's Law Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: KWQ Dated: March 1, 2024 Received: March 1, 2024
Dear Adam Rogers:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Eileen
Digitally signed
by Eileen Cadel -
S
Cadel -S Date: 2024.05.24
13:52:10-04'00'
for
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
#### Indications for Use
Submission Number (if known)
K240592
Device Name
Wolff's Law Anterior Cervical Plate System
Indications for Use (Describe)
The Wolff's Law Anterior Cervical Plate System is indicated for use in the temporary stabilization of the cervical spine (C2-C7) during the development of solid spinal fusion in patients with instability caused by the following:
• Degenerative Disc Disease (DDD) – as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies;
- · Spondylolisthesis;
- Spinal Stenosis:
- · Trauma (including fractures);
- · Tumor;
- · Deformity (i.e., scoliosis, kyphosis, and/or lordosis);
- · Pseudoarthrosis; and
- · Failed previous fusions
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
ver-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
K240592 Page 1 of 2
Prepared on: 2024-05-21
| Contact Details | | 21 CFR 807.92(a)(1) |
|------------------------------------|----------------------------------------------------------------|---------------------|
| Applicant Name | Spinal Simplicity LLC | |
| Applicant Address | 6363 College Blvd Ste 320 Overland Park KS 66211 United States | |
| Applicant Contact Telephone | 913-451-4414 | |
| Applicant Contact | Mr. Adam Rogers | |
| Applicant Contact Email | arogers@spinalsimplicity.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | Wolff's Law Anterior Cervical Plate System | |
| Common Name | Spinal intervertebral body fixation orthosis | |
| Classification Name | Appliance, Fixation, Spinal Intervertebral Body | |
| Regulation Number | 888.3060 | |
| Product Code(s) | KWQ | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K083020 | DynaTran™ Anterior Cervical Plating (ACP) System | KWQ |
| K143626 | SC-AcuFix® Ant-Cer Dynamic Cervical Plating System | KWQ |
## Device Description Summary
Synthes Vectra-T System
The Spinal Simplicity Wolff's Law Anterior Cervical Plate System is a dynamic cervical plate designed for the anterior cervical spine (C2-C7). The Wolff's Law Plate System consists of one-level plates in multiple lengths with both fixed- and variable-angle screws. The plates are translatable and use a nitinol Compression Element to provide continuous compression at each level. The Wolff's Law Cervical Plate consists of components made from titanium alloy Ti6Al4V ELI per ASTM F136 and superelastic nitinol per ASTM F2063. The fixed and variable screws are made from Titanium alloy Ti6Al4V ELI per ASTM F136. The Anterior Cervical Plate System is provided to the end-user sterile and is single use.
## Intended Use/Indications for Use
21 CFR 807.92(a)(5)
KWQ
21 CFR 807.92(a)(4)
The Wolff's Law Anterior Cervical Plate System is in the temporary stabilization of the cervical spine (C2-C7) during the development of solid spinal fusion in patients with instability caused by the following:
• Degenerative Disc Disease (DDD) – as defined by neck pain with degeneration of the disc confirmed by patient history and radiographic studies;
• Spondylolisthesis;
• Spinal Stenosis;
K051665
· Trauma (including fractures);
· Tumor;
{4}------------------------------------------------
· Deformity (i.e., scoliosis, kyphosis, and/or lordosis);
• Pseudoarthrosis; and
• Failed previous fusions
## Indications for Use Comparison
The indications for use of the primary and secondary predicate devices fully encompasses the Wolff's Law Anterior Cervical Plate System indication for use statement. The types of conditions for which the Wolff's Law Anterior Cervical Plate System is used are identical to the predicate devices. The slight differences in the indications for use statements do not constitute a new intended use.
## Technological Comparison
The Wolff's Law Anterior Cervical Plate System has similar technological characteristics, surgical approach, materials, range of sizes, design and principles of operation as the predicate devices. Any differences in technological characteristics do not raise different questions of safety or effectiveness. Performance testing device has mechanical performance substantially equivalent to that of the predicate.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Non clinical mechanical tests were performed to substantial equivalence of the Wolffs Law Anterior Cervical Plate System. Custom test were also performed to support the performance and substantial equivalence of the device. FDA Guidance Document Spinal Plating Systems - Performance Criteria for Safety and Performance Based Pathway: Guidance for Industry and Food and Druq Administration Staff was used to support the performance testing and the following ASTM test were performed:
ASTM F1717 Static Compression Bending ASTM F1717 Static Torsion ASTM F1717 Static Tension Bending ASTM F1717 Dynamic Compression Bending ASTM F543 Axial Screw Pullout/Pushout ASTM F2129 & ASTM F3044 Corrosion Testing ASTM F1877 Particulate Analysis ASTM F2004 Transformational Temperature
Clinical performance testing was not performed for this submission.
Spinal Simplicity concludes that the performance testing provided in this 510(k) application demonstrates that the Wolffs Law Anterior Cervical Plate System is capable of performing as intended and is substantially equivalent to the legally marketed predicate device.
### 21 CFR 807.92(a)(5)
## 21 CFR 807.92(a)(6)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.