The iTero Lumina Pro NIRI functionality is a diagnostic aid for the detection of interproximal carious lesions above the gingiva and for monitoring the progress of such lesions.
Device Story
iTero Lumina Pro is an integrated intraoral imaging system capturing 3D optical impressions, 2D color images, and Near Infrared (NIR) images. The system uses a handheld scanning wand containing projectors, LEDs, and cameras. For NIR imaging, 850nm LEDs transmit light through teeth; healthy enamel appears dark/translucent, while dentin/caries appear bright/opaque. The device is operated by dentists in a clinical setting. The system alternates between modalities using pulsed illumination. NIR images are acquired from multiple angles and displayed to the clinician upon scan completion. Clinicians use these images to identify and monitor interproximal caries. The device uses single-use disposable wand barrier sleeves for cross-contamination prevention. It connects to a computer platform via USB for image processing, display, and storage.
Clinical Evidence
Bench testing only. Performance attributes evaluated included field of view, image sharpness, depth of field, illumination power, wavelength, spectral bandwidth, specular reflection suppression, and signal-to-noise ratio. Wand barrier sleeve testing included tensile, tear, puncture, and penetration resistance (ASTM D882, D1004, F1342, D2582, F1670) and microbial ingress/pull testing (ASTM D3330). Biocompatibility was confirmed per ISO 10993 series. Reprocessing validation complied with ANSI/AAMI ST98:2022.
Technological Characteristics
Handheld intraoral scanner; aluminum, copolyester, and glass construction. NIR imaging uses 850nm LEDs and CMOS sensors. 3D imaging uses 450nm/520nm laser dot patterns. Pulsed illumination method. USB connectivity. Complies with IEC 60601-1, IEC 80601-2-60, IEC 62471, and IEC 60825-1. Software developed per IEC 62304.
Indications for Use
Indicated for dentists to detect interproximal carious lesions above the gingiva and monitor lesion progression in patients.
Regulatory Classification
Identification
A laser fluorescence caries detection device is a laser, a fluorescence detector housed in a dental handpiece, and a control console that performs device calibration, as well as variable tone emitting and fluorescence measurement functions. The intended use of the device is to aid in the detection of tooth decay by measuring increased laser induced fluorescence.
Special Controls
*Classification.* Class II, subject to the following special controls:(1) Sale, distribution, and use of this device are restricted to prescription use in accordance with § 801.109 of this chapter;
(2) Premarket notifications must include clinical studies, or other relevant information, that demonstrates that the device aids in the detection of tooth decay by measuring increased laser induced fluorescence; and
(3) The labeling must include detailed use instructions with precautions that urge users to:
(i) Read and understand all directions before using the device,
(ii) Store probe tips under proper conditions,
(iii) Properly sterilize the emitter-detector handpick before each use, and
(iv) Properly maintain and handle the instrument in the specified manner and condition.
1. restrictions on the sale, distribution, and use of the device to prescription use in accordance with 21 CFR 801.109;
2. clinical studies, or other relevant information, that demonstrates that the device aids in the detection of tooth decay by measuring increased laser induced fluorescence; and
3. labeling that includes detailed use instructions with precautions that urge users to (i) read and understand all directions before using the device, (ii) store probe tips under proper conditions, (iii) properly sterilize the emitter-detector handpiece before each use, and (iv) properly maintain and handle the instrument in the specified manner and condition.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 16, 2024
Align Technology Ltd. % Kristin Zielinski Duggan Partner Hogan Lovells US LLP Columbia Square 555 Thirteenth Street, NW Washington, District of Columbia 20004-1109
Re: K240573
Trade/Device Name: iTero Lumina™ Pro Regulation Number: 21 CFR 872.1745 Regulation Name: Laser Fluorescence Caries Detection Device Regulatory Class: Class II Product Code: NTK, OJK Dated: July 26, 2024 Received: July 26, 2024
Dear Kristin Zielinski Duggan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
## Michael E. Adjodha -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K240573
Device Name
iTero Lumina™ Pro
Indications for Use (Describe)
The iTero Lumina Pro NRI functionality is a diagnostic aid for the detection of interproximal carious lesions above the gingiva and for monitoring the progress of such lesions.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) SUMMARY - K240573 Align Technology's iTero Lumina™ Pro System
### Submitter
Align Technology Ltd. 1 Yitzhak Rabin Rd., Petach Tikva, 4925110, lsrael
Phone: +972544622662 email: atouati@aligntech.com Contact Person: Aviva Touati Date Prepared: August 14th, 2024
Name of Device: iTero Lumina™ Pro
Common or Usual Name: Caries Detector, Laser Light, Transmission
Classification Name: Laser fluorescence caries detection device
Regulation Number: 872.1745
Regulatory Class: Class II
Product Code: NTK
Predicate Device: iTero Element 5D (K193659), Align Technology Ltd.
## Device Description
The iTero Lumina Pro is an integrated intraoral imaging system capable of capturing 3D optical impressions for CAD/CAM of dental devices, 2D color images and Near Infrared (NIR) images. Scanning cycles are employed allowing the system to quickly alternate between the three modalities, creating a virtually simultaneous capture of the 3D optical impression, 2D color and NIR images.
The principles of operation of the three modalities are briefly described hereafter.
## 1) 3D Optical Impression
During the 3D optical impression scan, 450nm (Blue) and 520nm (Green) laser light dot patterns, are operated to create the 3D model.
- 2) 2D Color Imaging
To capture the 2D color image, the intra-oral cameras are operated while the white LEDs are turned ON.
- 3) NIR Imaging
LEDs with a wavelength of 850nm transmit light through the tooth resulting in an image that translates the teeth structure (enamel and dentin) to different brightness levels. The healthy enamel appears dark and translucent, while dentin or caries lesions appear brighter and opaque.
{5}------------------------------------------------
The NIR Imaging modality is the focus of this submission.
The iTero Lumina Pro consists of the following main components:
- A. Scanning Wand used for intraoral scanning.
- B. Wand Cradle used to hold the Scanning Wand when not in use.
- C. Application Software that includes Graphical User Interface ("GUI").
- D. Computer platform
The scanning wand houses electronic, mechanical, and optical components, including projectors, LEDs and cameras which enables capturing 3D impression data as well as 2D color and NIRI images. The wand body is made of Aluminum, Copolyester and Glass. When not in use, it is stored in the Wand Cradle.
The iTero Lumina Pro features proprietary software that can be installed on the various configurations of the computer platform. The software is designed to manage the GUI, control the scanning process, display the scanned images, and connect to the stored patient data, including scanned data for display or export.
The iTero Lumina scanning wand is designed to be used with a single-use, non-sterile, disposable, wand barrier sleeve. The only patient-contacting materials are those contained within the wand barrier sleeve.
## Intended Use / Indications for Use
The iTero Lumina Pro NIRI functionality is a diagnostic aid for the detection of interproximal carious lesions above the gingiva and for monitoring the progress of such lesions.
## Summary of Technological Characteristics
NIR imaging is the technological principle for both the subject and predicate devices. For both the predicate and the subject device, the NIR LEDs create an 850nm wavelength illumination that transmits through a tooth and is captured by cameras. The NIR Image translates the teeth structure (enamel and dentin) to different brightness levels. The healthy enamel appears dark and translucent, while dentin or caries lesions appear brighter and opaque.
At a high level, the subject and predicate devices are based on the following same technological elements. The iTero Lumina Pro and the iTero Element 5D systems include similar hardware and software components designed to obtain and display NIR images for a dentist's review.
Both system's scanning wand comprise three different light sources, which are capable of capturing 3D optical impressions for CAD/CAM of dental devices, 2D color images and Near Infrared (NIR) images. In addition, the light sources operate in a substantially equivalent manner, given that they utilize periodical light source activation (i.e., alternating ON / OFF depending on the required modality: 3D optical impression, 2D color imaging and NIR imaging).
{6}------------------------------------------------
Both devices employ a similar cross contamination prevention strategy in which the user is required to apply a new disposable wand barrier sleeve before each use. The scanning process is also similar in that the user selects the scanning features, places the wand tip slightly above the subject's teeth, initiates the scan by using the control buttons and moves the scanning wand above the area which is required to be scanned.
The subject and predicate devices differ in the number of cameras and sensors and the optical pathway of the emitted lights. While the predicate device encompasses two (2) NIR LEDs and one (1) camera, the subject device has multiple NIR LEDs and cameras and thus can simultaneously acquire NIR images from several angles. In contrast, the predicate device requires more manipulation during scanning in order to capture images from several angles.
The predicate device displays the NIR images both in real time and offline, while the subject device presents the NIR images only upon completion of the scan. Despite this minor difference, both devices meet the clinical need of evaluating the NIR images to aid in diagnosis and monitoring the progress of potential interproximal caries. Clinicians do not typically evaluate the NIR images for aiding in diagnosis during scanning, but rather the evaluation occurs at the end of the scan, making this difference inconsequential.
Both the subject and predicate devices share the same indications for use.
| | iTero Lumina Pro<br>K240573<br>(Subject device) | iTero Element 5D<br>K193659<br>(Predicate device) |
|---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use/<br>Intended Use | The iTero Lumina Pro NIRI<br>functionality is a diagnostic<br>aid for the detection of<br>interproximal carious lesions<br>above the gingiva and for<br>monitoring the progress of<br>such lesions. | The iTero Element 5D is a<br>diagnostic aid for the<br>detection of interproximal<br>caries lesions above the<br>gingiva and for monitoring<br>the progress of such lesions. |
| User Population | Dentist | Dentist |
| Design | Hand-held device | Hand-held device |
| Main Device Components | Hand-held device with USB,<br>cable, software and<br>computer. | Hand-held device with USB,<br>cable, software and<br>computer. Alternatively,<br>software is installed on a<br>personal computer that<br>meets performance<br>parameters defined by the<br>company. |
| | iTero Lumina Pro<br>K240573<br>(Subject device) | iTero Element 5D<br>K193659<br>(Predicate device) |
| Functional Principle | Transillumination: Based on<br>the principle that the tooth<br>enamel is translucent to NIR<br>light hence its image appears<br>dark, while dentin and caries<br>lesions scatter NIR light<br>hence their image appears<br>bright. | Transillumination: Based on<br>the principle that the tooth<br>enamel is translucent to NIR<br>light hence its image appears<br>dark, while dentin and caries<br>lesions scatter NIR light<br>hence their image appears<br>bright. |
| NIRI Light Source | Infrared light emitting diode<br>(LED)s are used to generate<br>exactly 850nm wavelength<br>which are detected by the<br>CMOS sensor. | Infrared light emitting diode<br>(LED)s are used to generate<br>exactly 850nm wavelength<br>which are detected by the<br>CMOS sensor. |
| Installation | The computer-based<br>installation enables the<br>customer to install and<br>update the software. | The computer-based<br>installation enables the<br>customer to install and update<br>the software. |
| Power Source | The power supply of the<br>wand is USB-15V | The power supply of the wand<br>is USB-15V |
| Software | The software consists of a<br>product firmware and a<br>computer-based software,<br>which controls: - Display of<br>pictures - stores/-saves the<br>pictures - Live stream<br>controlling of camera<br>functions, such as brightness<br>and contrast. | The software consists of a<br>product firmware and a<br>computer-based software,<br>which controls: - Display of<br>pictures - Stores/saves the<br>pictures - Live stream<br>controlling of camera<br>functions, such as brightness<br>and contrast. |
| Standards with which the<br>Device Complies | IEC 60601-1 | IEC 60601-1 |
| Compatibility | USB connection | USB connection |
| Cross contamination<br>control | A single use, disposable,<br>wand barrier sleeve. | A single use, disposable,<br>wand barrier sleeve. |
| Wavelength | 850 nm | 850 nm |
| | iTero Lumina Pro<br>K240573<br>(Subject device) | iTero Element 5D<br>K193659<br>(Predicate device) |
| Illumination Type | Pulsed Illumination - method<br>of illumination that consists<br>of periodical light source<br>activation. i.e. alternating<br>ON(pulse)/OFF states. | Pulsed Illumination - method<br>of illumination that consists of<br>periodical light source<br>activation. i.e. alternating<br>ON(pulse)/OFF states. |
The table below provides a comparison of the fundamental technological characteristics of the iTero Lumina Pro System and those of its predicate device.
{7}------------------------------------------------
# align | 兴 invisalign® | iTero
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logos for Align Technology and its products. The first logo is for "align" with a blue dot above the letter "i". The second logo is for "invisalign" with a flower-like symbol to the left of the word. The third logo is for "itero".
## Performance Data
The company has performed bench testing to support the substantial equivalence of the iTero Lumina Pro to the predicate device.
For the Near Infrared Imaging (NIR) functionality, key performance attributes tested included: Field of View, Image Sharpness and Depth of Field, illumination Power, Wavelength and Spectral Bandwidth, Specular Reflection Suppression, and Signal to Noise Ratio. All testing showed similar performance to the predicate system.
Testing for the Lumina Pro wand barrier sleeve testing met requirements for dental barriers regulated under 21 C.F.R §878.4370 and demonstrated equivalent performance as the predicate device wand barrier sleeve. The testing included the following:
- Material testing of the iTero Lumina Pro wand barrier components was conducted in o accordance with ASTM D882-12 (tensile testing), ASTM D1004-13 (tear resistance), ASTM F1342/F1342M-05 and ASTM D2582-16 (puncture resistance), ASTM F1670/F1670M-17a (penetration resistance), and additional internally derived test protocols.
- Finished assembly testing of the wand barrier sleeve including Microbial Ingress o Testing, Pull testing in accordance with ASTM D3330/D3330M-04(18), and additional internally derived test protocols.
Basic Safety and Electromagnetic Compatibility (EMC) testing confirmed the iTero Lumina Pro complies with IEC 60601-1 (Medical electrical equipment - Part 1: General requirements for basic safety and essential performance), IEC 80601-2-60 (Medical electrical equipment - Part 2-60: Particular requirements for the basic safety and essential performance of dental equipment), IEC 60601-1-6 (Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability), IEC 62366-1 (Application of Usability Engineering to Medical Devices), IEC 60601-1-2 (Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and Tests), FDA Guidance on Electromagnetic Compatibility (EMC) of Medical Devices issued in June 2022 and, FDA Guidance on Radio Frequency Wireless Technology in Medical Devices issued in August 2013, along with lamps and laser safety standards including IEC 62471, Photobiological safety of lamps and lamp systems and IEC 60825-1, Safety of laser products – Part 1: Equipment classification, requirements and user's guide.
The iTero Lumina Pro software application has been designed and developed according to IEC 62304 (Medical Device Software Life-Cycle Processes) and FDA Guidance on Content of Premarket Submissions for Device Software Functions issued in June 2023.
{9}------------------------------------------------
Cybersecurity is ensured by compliance with the requirements of UL ANSI 2900-1 (Standard for Safety, Standard for Software Cybersecurity Network-Connectable Products, Part 1: General Requirements), UL ANSI 2900-2-1 (Standard for Safety Software Cybersecurity for Network-Connectable Products, Part 2-1: Particular Requirements for Network Connectable Components of Healthcare and Wellness Systems), and the FDA Guidance on Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions issued in September 2023.
Patient contacting components were shown to be biocompatible for their intended use per the recognized consensus standards in the ISO 10993-series. The device was found to be reliably cleaned through validated cleaning methods, including intermediate disinfection level validation of the wand and cradle, that comply with the requirements of ANSI/AAMI ST98:2022 standard and FDA Guidance on Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, issued in March 2015 and Appendix E which was updated in June 2017.
## Conclusions
The iTero Lumina Pro System is as safe and effective as the predicate device, the iTero Element 5D. The iTero Lumina Pro System has the same intended use and indications for use, and similar technological characteristics, and principles of operation as its predicate device. The minor differences in design do not alter the intended clinical use of the device and do not affect its safety and effectiveness when used as labeled. In addition, the minor technological differences between the iTero Lumina Pro System and its predicate devices raise no new questions of safety or effectiveness. Performance data demonstrate that the iTero Lumina Pro System is as safe and effective as the iTero Element 5D. Thus, the iTero Lumina Pro System is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.