K240564 · Carol Cole Company Dba Nuface® · NFO · Jun 27, 2024 · Neurology
Device Facts
Record ID
K240564
Device Name
NuFACE® FIX+
Applicant
Carol Cole Company Dba Nuface®
Product Code
NFO · Neurology
Decision Date
Jun 27, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NuFACE® FIX+ Device is intended for facial stimulation.
Device Story
Portable hand-held microcurrent stimulator for facial skin stimulation; intended for over-the-counter cosmetic use. Device comprises main unit with ABS housing, chrome-plated spherical probes, and rechargeable lithium-ion battery. Operates by delivering constant biphasic square wave microcurrent via two fixed spherical electrodes. Features three adjustable microcurrent output frequency levels (8.3 Hz to 50.0 Hz) and increased output intensity compared to predicate. User applies conductivity gel (NuFACE Line Smoothing FIX Serum) before treatment. Device provides non-invasive stimulation to facial skin; intended to assist in cosmetic facial toning. Operated by the user in a home environment. Output parameters are controlled by internal microprocessor/firmware. Safety and performance verified through bench testing and compliance with electrical/mechanical safety standards (IEC 60601 series).
Clinical Evidence
No clinical or animal studies were conducted. Substantial equivalence is supported by bench testing, including output waveform and energy characteristic verification, and compliance with electrical safety and EMC standards (IEC 60601-1, IEC 60601-2-10, etc.).
Indicated for over-the-counter cosmetic facial stimulation.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 27, 2024
Carol Cole Company dba NuFACE® Robert Castro Director of Quality and Regulatory 1325 Sycamore Ave Suite A Vista, California 92081
Re: K240564
Trade/Device Name: NuFACE® FIX+ Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NFO Dated: April 15, 2024 Received: April 16, 2024
Dear Robert Castro:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Heather L. Dean -S
Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices
{2}------------------------------------------------
OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
# Indications for Use
s for Use
Submission Number (if known)
K240564
Device Name
NuFACE® FIX+
Indications for Use (Describe)
The NuFACE FIX+ Device is intended for facial stimulation and is indicated for over-the-counter cosmetic use.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 1. SUBMITTER
| Submitter: | Carol Cole Company dba NūFACE® |
|-----------------|---------------------------------------------|
| Address: | 1325 Sycamore Ave., Suite A Vista, CA 92081 |
| Phone: | 818-468-3752 |
| Fax: | 760-650-3037 |
| Email: | rcastro@nuface.com |
| Contact Person: | Robert C. Castro |
| Date Prepared: | April 15, 2024 |
# 2. DEVICE
| Name of Device: | NūFACE® FIX+ |
|--------------------------|------------------------------------------------------------|
| Common or<br>Usual Name: | FIX+ |
| Classification Name: | Transcutaneous electrical nerve stimulator for pain relief |
| Regulation Number: | 21 CFR 882.5890 |
| Regulatory Class: | Class II |
| Product Code: | NFO |
# 3. PREDICATE DEVICE
| Name of Device: | NūFACE® FIX Skin Toning Device |
|-----------------|-----------------------------------------------------------------------------------|
| 510(k) Number: | K182424<br>This predicate device has not been subject to a design-related recall. |
{5}------------------------------------------------
# 4. DEVICE DESCRIPTION
- FIX+ is a portable hand-held microcurrent stimulator for facial skin stimulation indicated for over-the-counter cosmetic use.
- -Main components: Main unit, USB power cable, and NuFACE Line Smoothing FIX Serum conductivity gel.
- Housing of the device made from Acrylonitrile Butadiene Styrene, with chrome-plated spherical probes.
- Inclusion of three microcurrent output frequency levels. -
- Integrated rechargeable lithium-ion battery -
# 5. INTENDED USE
The NuFACE® FIX+ Device is intended for facial stimulation.
# 6. INDICATIONS FOR USE
The NuFACE FIX+ Device is intended for facial stimulation and is indicated for over-the-counter cosmetic use.
# 7. DEVICE COMPONENTS
The following materials are expected to contact the user in normal use:
| Device<br>Component | Composition |
|----------------------------|----------------------------------------|
| Two (x2) Spheres | Chromium plating |
| Handheld Unit<br>Enclosure | Acrylonitrile Butadiene Styrene (ABS). |
# 8. ACCESSORIES
| Accessories | Description |
|------------------|------------------------------------------------------------------------------------------------------|
| FIX+ device | NūFACE® FIX+ |
| Power Cord | USB-to-DC Power cable (Charge only) |
| Conductivity Gel | NūFACE® Line Smoothing FIX Serum<br>Sizes: 2 fl. oz. or 5 fl. oz.<br>To be applied before treatments |
{6}------------------------------------------------
# 9. DESIGN AND TECHNOLOGICAL SPECIFICATIONS
FIX+ is based on the design of the established NūFACE® FIX Skin Toning Device (predicate device - K182424). The primary distinction from the predicate device involves a slight modification in the shape and dimensions of the device housing, the inclusion of three (3) adjustable microcurrent output frequencies, and an increased microcurrent output intensity.
#### 10. TECHNOLOGICAL CHARACTERISTICS COMPARISON
| Technological Characteristics Comparison | | |
|------------------------------------------|------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| Category | NUFACE FIX+<br>(SUBJECT DEVICE) | NUFACE FIX<br>Line Smoothing Device (PREDICATE) |
| Type of Energy<br>Output | Microcurrent | Microcurrent |
| Energy Delivery | Microcurrent is delivered via two fixed,<br>smooth spherical electrodes | Microcurrent is delivered via two fixed,<br>smooth spherical electrodes |
| Energy Flow | Microcurrent continuously alternates<br>between the positive and negative<br>electrode spheres | Microcurrent continuously alternates between<br>the positive and negative electrode spheres |
| Energy Output | Constant biphasic square<br>wave | Constant biphasic square<br>wave |
| Energy Power<br>Source | Internal rechargeable battery | Internal rechargeable<br>battery |
| Battery Type | Lithium Ion | Lithium Ion |
| Charging<br>Circuitry | Internal to Device | Internal to Device |
| Charging Method | USB Power Cable | USB Power Cable |
| Waveform | Pulsed Biphasic | Pulsed Biphasic |
| Shape | Modulated Square | Modulated Square |
| Maximum Output<br>Voltage | 28 VDC | 28 VDC |
| Software | Yes | Yes |
| Firmware | Yes | Yes |
| Technological Characteristics Comparison | | |
| Category | NuFACE FIX+<br>(SUBJECT DEVICE) | NuFACE FIX<br>Line Smoothing Device (PREDICATE) |
| Microprocessor Control | Yes | Yes |
| Output Frequency | Variable frequency<br>ranging from 8.3 Hz to<br>50.0 Hz | Approximately 8.3 Hz |
| Range of Output Frequency (Hz) | Range of 2-50 Hz | 8.3 Hz |
| Wireless Technologies | None | None |
| Wireless Capability | None | None |
| Output Intensity (μA) | 225 μA | 200 μA |
| Maximum Output Current Density | 0.480 mA/cm2 | 0.418 mA/cm2 |
{7}------------------------------------------------
# 510(k) SUMMARY NūFACE® FIX+
| Table 4: Performance Specifications Comparison | | |
|------------------------------------------------|---------------------------------|-------------------------------------------------|
| Category | NuFACE FIX+<br>(Subject Device) | NuFACE FIX Line Smoothing Device<br>(Predicate) |
| Maximum Output<br>Current | 225 $ μ $ A @ 10000Ω | 208 $ μ $ A @ 500Ω |
| Output Current when<br>not stimulating | < 1 $ μ $ A | < 1 $ μ $ A |
| Output Tolerance | +/- 10% | +/- 10% |
| Pulse Width | 10 – 60 ms | 60 ms |
{8}------------------------------------------------
# 510(k) SUMMARY NūFACE® FIX+
| Table 4: Performance Specifications Comparison | | |
|-----------------------------------------------------------------------------------------------------|---------------------------------|-------------------------------------------------|
| Category | NuFACE FIX+<br>(Subject Device) | NuFACE FIX Line Smoothing Device<br>(Predicate) |
| Frequency (Hz) | Range of 8.3 - 50 Hz | Approximately 8.3 Hz |
| Symmetrical phases | Biphasic | Biphasic |
| Phase Duration<br>(include units) (state range, if<br>applicable) (both phases,<br>if asymmetrical) | Biphasic | Biphasic |
| Net Charge<br>(µC per pulse) | 0 | 0 |
| Burst Mode (i.e., pulse trains) | | |
| a. Pulse ON time<br>(msec) | 10ms to 60ms | 60 |
| b. Pulse OFF<br>time (msec) | 10ms to 60ms | 60 |
| c. Pulses per burst | 20 | 20 |
| d. Pulses per second | 8.3 to 50 | 8.3 |
| e. Burst duration<br>(sec) | 0.4 to 2.4 | 2.4 |
| f. Duty Cycle | 50% | 50% |
# Technological Comparison Conclusion
The NūFACE® FIX+ maintains core technological similarities with the predicate NūFACE® FIX Line Smoothing Device, including microcurrent energy output and delivery. The expanded frequency range and increased microcurrent output intensity represent technological enhancements that offer additional features for the user, but do not impact the device's intended clinical use for cosmetic facial stimulation.
{9}------------------------------------------------
#### TESTING SUMMARIES 11.
### a. Safety Testing
NūFACE® FIX+ device was tested and has demonstrated compliance with applicable IEC standards concerning electrical safety and EMC.
Electrical, Mechanical, and Thermal Safety Testing were conducted according to the following standards:
| IEC 60529 | Ingress Protection |
|----------------|---------------------------------------------------------------------------------------------|
| IEC 60601-1 | Basic safety and essential performance (Electrical Equipment) |
| IEC 60601-1-11 | Basic safety and essential performance (Med Dev Home Use) |
| IEC 60601-1-2 | EMC - Basic safety and essential performance (Electromagnetic<br>Disturbances) |
| IEC 60601-1-6 | Basic safety and essential performance (Med Dev Electrical<br>Equipment) |
| IEC 60601-2-10 | Basic safety and essential performance (Med Dev Home Use –<br>nerve and muscle stimulators) |
| ISO 10993-1 | Biological Evaluation of Medical Devices |
| ISO 14971 | Risk Management (Med Dev) |
| TR 60601-4-2 | Medical Electrical Equipment |
| IEC 62366 | Usability engineering (Med Dev) |
### b. Common EM Emitters
The FIX+ was tested to all applicable clauses of the IEC 60601-1-2: Edition 4.1, IEC 60601-4-2 Edition 1.0, and IEC 55014-2:2021 standards. In addition, a Common Emitter evaluation, for some of the most common electromagnetic emitters a home user could encounter, was conducted per FDA guidance document "Electromagnetic Compatibility (EMC) of Medical Devices" (06/06/2022). A Common Emitter evaluation is provided with this submission.
### c. Bench Testing
Extensive bench testing was conducted to ensure the output waveform characteristics and the microcurrent output performance levels met the established product requirements within the designed tolerances. These tests are addressed in detail in the Performance Testing Section of this premarket notification.
Performance bench testing was conducted using production equivalent devices of the NūFACE® FIX Plus device. The testing was comprised of:
- -Output Waveform Characteristics
- -Output Energy Characteristics
{10}------------------------------------------------
Subsequent to performance bench testing. FIX+ production equivalent devices were also tested by a third-party accredited laboratory (Nemko USA) in accordance with IEC 60601-2-10 Particular requirements for the basic safety and essential performance of nerve and muscle stimulators.
Both internal performance bench testing and laboratory testing conducted concluded that the subject FIX+ device waveform and energy characteristics fall within their intended specifications and are substantially equivalent to the predicate device.
# d. Software Verification and Validation Testing
Certification and validation testing for the subject device were conducted and documented as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Details are provided in the Software Verification Section of this premarket notification.
Testing requirements were determined by the FIX+ Design Inputs & PRD and complemented with compliance testing performed at Nemko USA. All test requirements are considered as Passed and the user interface features, functions, and output of the FIX+ devices meet the specified design input criteria.
#### Animal Study and Clinical Studies e.
The provided documentation asserts that the determination of substantial equivalence for the device in question will not involve animal or clinical studies; hence, no such studies have been conducted to support this claim. Instead, a comparison of performance specifications with a predicate device will be used for the assessment.
#### Biocompatibility f.
NuFACE products classified as medical devices are subject to a Biocompatibility Evaluation per the FDA's guidance document titled "Use of International Standard ISO 10993-1, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process. The FIX+ device is categorized as a surface device which makes repeated contact with intact skin. Since this device can be used up to 2 times per treatment area and 5 days per week for the lifetime of the product, the device has longterm (>30 day) repeated contact with human tissue. Biocompatibility evaluation endpoints were assessed in accordance with the "Evaluation of Endpoints for Consideration" detailed in Attachment A of the FDA's quidance document titled "Use of International Standard ISO 10993-1, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process." The FIX+ device was evaluated for Biocompatibility as a Surface Device with contact with intact skin with a long-term contact duration of greater than 30 days.
{11}------------------------------------------------
#### 12. CONCLUSION
The NuFACE® FIX+ is found to be substantially equivalent to its predicate, the NūFACE® FIX Line Smoothing Device (K182424). This determination is based on the similarity in intended use, indications for use, and the majority of the technological characteristics between the two devices. Although there are some differences in output characteristics like pulse width and frequency range, these do not affect the devices' clinical efficacy or intended cosmetic functions. The detailed analysis provided in the Substantial Equivalence Section of the premarket notification confirms that the observed technological differences do not introduce new safety and effectiveness concerns.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.