VitaVitro Single Lumen Oocyte Retrieval Needle is intended for ultrasound guided transvaginal aspiration of oocytes from ovarian follicles for patients undergoing Assisted Reproductive procedures. VitaVitro Double Lumen Oocyte Retrieval Needle is intended for ultrasound guided transvaginal aspiration and flushing of oocytes from ovarian follicles for patients undergoing Assisted Reproductive procedures.
Device Story
Device consists of stainless steel needle with echogenic tip, suction catheter, vacuum catheter, and (for double lumen) flushing catheter; used for transvaginal oocyte retrieval under ultrasound guidance. Physician operates device to aspirate or flush ovarian follicles; output is collected oocytes for assisted reproduction. Echogenic tip enhances visualization during ultrasound-guided procedure. Device is disposable, sterile, and single-use.
Clinical Evidence
Bench testing only. Includes biocompatibility (cytotoxicity, sensitization, irritation), Mouse Embryo Assay (MEA >80% blastocyst development), endotoxin testing (<5EU/device), sterility (SAL 10^-6), package integrity (ASTM F88, F1929), and mechanical performance (ISO 7864, ISO 20695, ISO 9626) including ultrasound detectability, flow rate, and tensile strength.
Technological Characteristics
Materials: Stainless steel, TPU, silicone, PE. Features: Echogenic needle tip, 17G/18G gauges, 33-35cm lengths. Sterilization: Ethylene oxide (ISO 11135). Standards: ISO 10993 (biocompatibility), ISO 7864 (needle/appearance), ISO 20695 (mechanical), ISO 9626 (puncture/rigidity), ASTM D 4169 (shipping), ASTM F88/F1929 (packaging).
Indications for Use
Indicated for patients undergoing Assisted Reproductive procedures requiring ultrasound-guided transvaginal aspiration (single lumen) or aspiration and flushing (double lumen) of oocytes from ovarian follicles.
Regulatory Classification
Identification
Assisted reproduction needles are devices used in in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures to obtain gametes from the body or introduce gametes, zygote(s), preembryo(s) and/or embryo(s) into the body. This generic type of device may include a single or double lumen needle and component parts, including needle guides, such as those used with ultrasound.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services USA logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 15, 2024
Donghai Pan Official Applicant R601, Building B, Hai Ke Xing Tech Park Baoshan Road No. 16 Shenzhen, Guangdong 518118 CHINA
Re: K240523
> Trade/Device Name: VitaVitro® Single Lumen Oocyte Retrieval Needle (ORNS-17G, ORNS-18G): VitaVitro® Double Lumen Oocyte Retrieval Needle (ORND-17G) Regulation Number: 21 CFR 884.6100 Regulation Name: Assisted Reproduction Needles Regulatory Class: II Product Code: MQE Received: October 15, 2024
Dear Donghai Pan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page
2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Monica D. Garcia -S
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K240523
Device Name
VitaVitro® Single Lumen Oocyte Retrieval Needle (ORNS-17G, ORNS-18G) VitaVitro® Double Lumen Oocyte Retrieval Needle (ORND-17G)
Indications for Use (Describe)
VitaVitro Single Lumen Oocyte Retrieval Needle is intended for ultrasound guided transvaginal aspiration of oocytes from ovarian follicles for patients undergoing Assisted Reproductive procedures.
VitaVitro Double Lumen Oocyte Retrieval Needle is intended for ultrasound guided transvaginal aspiration and flushing of oocytes from ovarian follicles for patients undergoing Assisted Reproductive procedures.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY - K240523
#### 1. Submission Sponsor
Shenzhen VitaVitro Biotech Co., Ltd. R601, Building B, Hai Ke Xing Tech Park Baoshan Road No. 16 Shenzhen 518118 Guangdong, China Contact Person: Mr. Donghai Pan Title: International Regulatory Specialist Tel: 86-755-84511813 Fax: 86-755-85235226
Email: pandh(@vitavitro.com
#### 2. Date Prepared
November 14, 2024
#### 3. Device Identification
| Trade/Device Name: | VitaVitro® Single Lumen Oocyte Retrieval Needle (ORNS-<br>17G, ORNS-18G); VitaVitro® Double Lumen Oocyte<br>Retrieval Needle (ORND-17G) |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Oocyte Retrieval Needle |
| Regulation Number: | 21 CFR § 884.6100 |
| Regulation Name: | Assisted Reproduction Needles |
| Device Classification: | Class II |
| Product Code: | MQE (Needle, Assisted Reproduction) |
# 4. Predicate Device
| Predicate<br>Device | Trade Name | Ovum Pick-Up Aspiration Needle |
|---------------------|------------------|--------------------------------|
| | Manufacturer | Cook OB/GYN |
| | 510(k) No. | K983593 |
| | Regulatory Class | Class II |
| | Product Code | MQE |
The predicate device has not been subject to any design-related recalls.
# 5. Device Description
The Oocyte Retrieval Needle is composed of a stainless steel needle protective sleeve, needle handle, suction catheter, silicone plug, vacuum catheter, flushing catheter and taper joint. The device includes both dual lumen (ORND) and single-lumen needles (ORNS). Three models are available: ORNS-17G, ORNS-18G and ORND-17G. The oocyte retrieval needles are available in 17 or 18 gauge sizes and two lengths: 33 and 35 cm. The needles are provided with a suction catheter of 100 cm length and a vacuum
{5}------------------------------------------------
catheter with a length of 35 cm, and the ORND-17G also includes a flushing catheter of 100 cm length. Needles have an echogenic tip that enhances the visualization of the needle tip under ultrasound.
This product is a disposable, sterile medical device, sterilized by ethylene oxide.
## 6. Indications For Use
VitaVitro Single Lumen Oocyte Retrieval Needle is intended for ultrasound guided transvaginal aspiration of oocytes from ovarian follicles for patients undergoing Assisted Reproductive procedures.
VitaVitro Double Lumen Oocyte Retrieval Needle is intended for ultrasound guided transvaginal aspiration and flushing of oocytes from ovarian follicles for patients undergoing Assisted Reproductive procedures.
# 7. Substantial Equivalence Discussion
The tables below list the comparisons of the indications for use and technological characteristics of the subject and predicate devices.
{6}------------------------------------------------
| Parameter | Subject Device | Predicate Device | Comparison |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | VitaVitro Single Lumen<br>Oocyte Retrieval Needle is<br>intended for ultrasound<br>guided transvaginal aspiration<br>of oocytes from ovarian<br>follicles for patients<br>undergoing Assisted<br>Reproductive procedures.<br>VitaVitro Double Lumen<br>Oocyte Retrieval Needle is<br>intended for ultrasound<br>guided transvaginal aspiration<br>and flushing of oocytes from<br>ovarian follicles for patients<br>undergoing Assisted<br>Reproductive procedures. | The Ovum Pick-Up Aspiration<br>Needles used for aspiration and<br>flushing of oocytes from ovarian<br>follicles. | Different: The<br>indications for use for<br>the subject device and<br>predicate device are not<br>identical; however, they<br>have the same intended<br>use (i.e. aspiration<br>or/and flushing of<br>oocytes from ovarian<br>follicles) |
| Design<br>Features | needle, needle protective<br>sleeve, needle handle,<br>aspiration catheter, silicone<br>plug, vacuum catheter,<br>flushing catheter and taper<br>joint. | needle, aspiration<br>line, plug, with or without<br>vacuum line, flushing line and<br>taper joint. | Different |
| Dimension | ORNS-17G, ORNS-18G:<br>Needle: 33 cm<br>Aspiration catheter:100 cm<br>Vacuum catheter:35 cm<br>Needle Gauge: 17G or 18G<br>ORND-17G<br>Needle: 35 cm<br>Aspiration catheter:100 cm<br>Vacuum catheter:35 cm<br>Flushing catheter:100 cm<br>Needle Gauge: 17G | G44360, G55248, G55252:<br>Needle length:30 cm or 35 cm<br>Aspiration line: 60 cm,90 cm<br>Needle gauge:16G or 17G<br>Vacuum line: Not available<br>G55796, G55798:<br>Needle Gauge: 16G or 17G<br>Needle Length: 35 cm<br>Aspiration Tubing: 75 cm<br>Flushing Tubing: 70 or 100<br>cm Vacuum line: Not<br>available | Different |
| Parameter | Subject Device | Predicate Device | Comparison |
| Device<br>Materials | Stainless steel, TPU, silicone<br>PE | Stainless steel,<br>Tetrafluoroethylene<br>(NRT[TFE]) | Different |
| Needle type | Single Lumen Ovum<br>Aspiration Needles and<br>Double Lumen Ovum<br>Aspiration Needles | Single Lumen Ovum<br>Aspiration Needles and<br>Double Lumen Ovum<br>Aspiration Needles | Same |
| Endotoxin | < 5EU/device | Not publicly available | Different |
| MEA | ≥80% embryos developed to<br>expanded blastocyst at 96h | Not publicly available | Different |
| Sterile | No microbial growth | No microbial growth | Same |
| Single Use | Yes | Yes | Same |
| Shelf life | 2 years | Not publicly available | Different |
{7}------------------------------------------------
The subject and predicate device have the same intended use - to transvaginally aspirate oocytes from ovarian follicles with the aid of ultrasound imaging. In addition, as shown in the table above, there are technological differences between the subject and predicate device, including the device components, device dimensions, materials used, performance testing, and shelf-life. However, these differences do not raise different questions of safety and effectiveness.
# 8. Summary of Non-Clinical Performance
The following studies have been performed for the device to support the substantial equivalence to the predicate device in accordance with the 2023 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process. "
- . Biocompatibility evaluation
- | Cytotoxicity (ISO 10993-5:2009)
- Sensitization (ISO 10993-10:2021)
- Irritation (ISO 10993-23:2021)
Testing results demonstrate that the subject devices are non-cytotoxic, non-irritating, and non-
{8}------------------------------------------------
sensitizing.
- 1-cell MEA per the 2021 FDA guidance document Mouse Embryo Assay for Assisted Reproduction Technology Devices: >80% embryos developed to expanded blastocyst at 96h
- . Endotoxin testing: < 5EU/device, LAL testing by Gel Clot assay per USP<85>
- . Sterility testing
- ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Sterile, ethylene oxide, SAL of 10-6 (ISO 11135:2014)
- Residual testing (ISO 10993-7:2008) |
- . Shipping and package integrity testing
- Simulated shipping per ASTM D 4169:2016 DC13
- ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Package integrity per ASTM F88/F88M:2021 (seal peel strength), ASTM F1929:2015 (dye penetration)
- . Mechanical performance testing
- ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Appearance per ISO 7864:2016
- Dimensional verification |
- Hydraulic resistance per ISO 20695:2020, Annex D
- -Vacuum leakage
- -Tensile strength test per ISO 20695:2020, Annex C
- -Flow rate per ISO 20695:2020, Annex E
- -Puncture needle test per ISO 9626:2016, Annex B-D: Rigidity test, Toughness test, Corrosion resistance
- Needle penetration force per ISO 7864:2016, Annex D ---
- ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Ultrasound detectability test
- Collection tube compatibility test
- Shelf-life testing was conducted to ensure that product specifications for endotoxin, MEA, ●
sterility, and package integrity were met over the proposed shelf-life period.
# 9. Conclusion
The results of the performance testing described above demonstrate that the subject devices are as safe and effective as the predicate device and support a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.