K240517 · Epic Medical Pte. , Ltd. · ONB · Dec 26, 2024 · General Hospital
Device Facts
Record ID
K240517
Device Name
ProSeal™ Vented Universal Vial Adaptor
Applicant
Epic Medical Pte. , Ltd.
Product Code
ONB · General Hospital
Decision Date
Dec 26, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The ProSeal™ Closed System drug Transfer Device (CSTD) mechanically prohibits environmental contaminants from entering the system and the escape of drug or vapor concentrations from thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills. The ProSeal™ system also prevents the introduction of microbial contaminations into the drug or fluid path for up to 7 days when used as intended.
Device Story
ProSeal™ Vented Universal Vial Adaptor is a component of the ProSeal™ CSTD system; facilitates closed-system fluid transfer between standard drug vials and ProSeal™ Injectors (syringe adaptors). Device features a dual-lumen spike and double-membrane septum design; self-sealing elastomeric membranes prevent drug/vapor escape and microbial ingress during connection/disconnection. Integrated 0.1-micron PTFE air-vent enables pressure equalization to prevent vial collapse or spray-back. Used by trained healthcare professionals/pharmacists in clinical settings. Output is a closed fluid path for hazardous drug reconstitution/transfer; minimizes exposure to aerosols/spills and maintains sterility for up to 7 days.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench performance testing, including positive/negative air pressure leakage, water leak integrity, vial perforator penetration force, fragmentation, vapor containment (per NIOSH CSTD 2016 draft protocol), and microbial ingress testing. Biocompatibility testing performed per ISO 10993-1 for externally communicating, blood path indirect, prolonged contact devices.
Technological Characteristics
Vial adaptor with integrated spike and claws; materials include polypropylene (P705JM) and polyisoprene membrane. Features 0.1-micron PTFE air-vent. Dimensions: 54.4 x 34.0 mm. Sterile barrier packaging (medical paper/plastic film). Sterilization: Ethylene Oxide (EO), SAL 10^-6. Connectivity: None (mechanical).
Indications for Use
Indicated for use by adequately trained healthcare professionals or pharmacists in clinical settings for the closed system transfer of antineoplastic and hazardous drugs.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
Arisure® Closed System Drug Transfer Device (CSTD) (K201422)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the full name of the agency on the right. The symbol on the left is a stylized representation of a human figure, while the FDA acronym is in blue, followed by the full name of the agency, "U.S. Food & Drug Administration," in a smaller font size.
December 26, 2024
Epic Medical Pte. Ltd. Freddie Lee CEO/md 105 Cecil Street #20-04, The Octagon Singapore, SG 069534 Singapore
Re: K240517
Trade/Device Name: ProSeal™ Vented Universal Vial Adaptor Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: ONB Dated: November 25, 2024 Received: November 25, 2024
Dear Freddie Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Porsche Bennett
Porsche Bennett For David Wolloscheck, Ph.D Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K240517
Device Name ProSeal™ Vented Universal Vial Adaptor
#### Indications for Use (Describe)
The ProSeal™ Closed System drug Transfer Device (CSTD) mechanically prohibits environmental contaminants from entering the system and the escape of drug or vapor concentrations from thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills. The ProSeal™ system also prevents the introduction of microbial contaminations into the drug or fluid path for up to 7 days when used as intended.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|--|
| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | |
| <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 으며 미
## K240517 – 510(k) Summary
#### I. Submitter
Epic Medical Pte. Ltd. 105 Cecil Street #20-04, The Octagon, Singapore 069534. Phone: +65 9635 2618 / +66 81 761 5292
Contact Person: Mr. Freddie LEE, Chief Executive Officer/ Managing Director Date Prepared: December 26, 2024 Content and Format: Prepared in accordance with 21 CFR 807.92 Type of Submission: Traditional
#### II. Subject Device
| 510(k) Number: | K240517 |
|---------------------|--------------------------------------------------------------------------------|
| Trade/ Device Name: | ProSeal™ Vented Universal Vial Adaptor |
| Common/ Usual Name: | Closed Antineoplastic and Hazardous Drug Reconstitution and<br>Transfer Device |
| Regulation Number: | 21 CFR 880.5440 |
| Regulation Name: | Intravascular administration set |
| Regulatory Class: | Class: II |
| Product Code: | ONB |
#### Predicate III.
| 510(k) Number: | K201422 |
|---------------------|--------------------------------------------------------------------------------|
| Trade/ Device Name: | Arisure® Closed System Drug Transfer Device (CSTD) |
| Common/ Usual Name: | Closed Antineoplastic and Hazardous Drug Reconstitution and<br>Transfer Device |
| Regulation Number: | 21 CFR 880.5440 |
| Regulation Name: | Intravascular administration set |
| Regulatory Class: | Class: II |
| Product Code: | ONB |
{5}------------------------------------------------
# 다이다
#### IV. Device Description
This FDA Submission is to include an authorization to legally market the ProSeal" Vented Universal Vial Adaptor to the existing ProSeal™ CSTD devices system of eleven (11) devices that were cleared for sales in the US, the most recent being the ToxiSeal™ Vial Adaptor (K241823).
The ProSeal™ Vented Universal Vial Adaptor is a component of the ProSeal™ CSTD system intended for connection with the interface membranes between any standard vials and ProSeal™ CSTD component devices for close system fluid transfer. When connected to a standard vial and engaged with a ProSeal™ Injector (Syringe Adaptor), fluid can be transferred to and from the connecting component device in a closed system. The ProSeal™ Vented Universal Vial Adaptor and its corresponding interface membranes exhibit a dry connection with the communicating surfaces in a fluid transfer. The use of this component and its appropriate ProSeal™ CSTD component device reduces the risk of microbial ingress for up to 168 hours (or 7 days). when used as intended.
The closed transfer of liquid that takes place with the use of the ProSeal™ CSTD system as follows:
- A double membrane septum design utilizing self-sealing elastomeric membranes tightly fits . together when the system components engage. A cannula within the ProSeal™ Injector housing perforates the double membrane for the transfer of liquid. When the cannula is retracted, the membranes seal off the transfer of environmental contaminants into the system and/or escape of drug or vapor concentrations outside the system, thereby minimizing the individual and environmental exposure to drug vapor, aerosols, and also minimizing the risk of microbial contamination.
#### v. Indications for Use Statement
The ProSeal™ Closed System drug Transfer Device (CSTD) mechanically prohibits environmental contaminants from entering the system and the escape of drug or vapor concentrations from the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills. The ProSeal™ system also prevents the introduction of microbial contaminations into the drug or fluid path for up to 7 days when used as intended.
#### Comparison of Technological Characteristics VI.
The ProSeal™ Vented Universal Vial Adaptor, a system component of the ProSeal™ CSTD and the Predicate, the Arisure® CSTD's Vial Adaptor, share the following key characteristics:
- Role of the CSTD system component devices ●
- Residual fluid
- Close system technology/ mechanism of action ●
- Biocompatibility ●
- Sterile barrier packaging top web and base web ●
- Pvrogenicitv .
- Validated shelf-life .
An overview table summarizing the comparison of key characteristics between the Subject ProSeal™ Vented Universal Vial Adaptor and the Predicate Arisure® CSTD's Vial Adaptor is provided hereunder.
{6}------------------------------------------------
| | Characteristics | Predicate Device (K201422) | Subject Device (K240517) | Comments,<br>discussions |
|-----------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| | compared | Arisure® CSTD - Vial Adaptor | ProSeal™ Vented Universal Vial Adaptor | |
| | Intended use and<br>Indications for Use<br>statement | Arisure® is a Closed System Drug Transfer<br>Device (CSTD) that mechanically prohibits<br>the release of drugs in vapor, aerosol or<br>liquid form during preparation and<br>administration, and prevents the introduction<br>of microbial and airborne contaminants into<br>the drug or fluid path, allowing the system to<br>minimize exposure of individuals, healthcare<br>personnel, and the environment to hazardous<br>drugs. | The ProSeal™ Closed System drug Transfer<br>Device (CSTD) mechanically prohibits<br>environmental contaminants from entering<br>the system and the escape of drug or vapor<br>concentrations from the system, thereby<br>minimizing individual and environmental<br>exposure to drug vapor, aerosols, and spills.<br>The ProSeal™ system also prevents the<br>introduction of microbial contaminations<br>into the drug or fluid path for up to 7 days,<br>when used as intended. | Same |
| | Intended user<br>population | Adequately trained health care professionals<br>or pharmacists | Adequately trained health care professionals<br>or pharmacists | Same |
| | Intended use<br>environment | Clinical setting | Clinical setting | Same |
| | Intended drug type | Antineoplastic and hazardous drug | Antineoplastic and hazardous drug | Same |
| | Intended Injector | Syringe adapter – a component of the<br>Arisure® CSTD | ProSeal™ Injector – a component of the<br>ProSeal™ CSTD (cleared K192075) | Different<br>See Comment #1 |
| | Single use or reusable | Single use only | Single use only | Same |
| | Role of the CSTD system<br>component devices | Connector (Fluid transfer interface to standard<br>external vials) self-sealing elastomeric<br>membrane on all system components | Connector (Fluid transfer interface to standard<br>external vials), self-sealing elastomeric<br>membrane on all system components | Same |
| Device subcomponents and materials | Injection site<br>protective seal<br>tab | N/A (injection site not present) | Protective seal tab for injection site (PP<br>348N) | Different<br>See Comment #1 |
| | Septum/seal | Flat, split septum of elastomeric membrane –<br>TPE Medalist 145 | Injection site & its polyisoprene membrane<br>(cleared K241823)(fluid path) | Different<br>See Comment #2 |
| | Vial adaptor<br>claws | Acrylonitrile butadiene styrene (ABS),<br>Terlux 2802 | Polypropylene (PP, P705JM) integrated<br>whole subcomponent with vial adaptor<br>spike | Different |
| | Vial adaptor<br>spike | Polycarbonate | Polypropylene (PP, P705JM)(fluid path)<br>integrated whole subcomponent with vial<br>adaptor claws | See Comment #3 |
| | Filter | PTFE membrane, 0.2 micron | PTFE membrane, 0.1 micron | Different<br>See Comment #4 |
| Characteristics<br>compared | Predicate Device (K201422)<br>Arisure® CSTD - Vial Adaptor | Subject Device (K240517)<br>ProSeal™ Vented Universal Vial Adaptor | Comments,<br>discussions | |
| Device dimensions | External dimensions (L×W) mm (inch):<br>· Vial adapter: 63.50 × 57.15 (2.5 × 2.25)<br>· Vial spike: 101.40 × 12.70 (4.0 × 0.5) | External dimensions (L×W) mm (inch):<br>· Vial adaptor claws: 34.4 × 34.0 (1.4 × 1.3)<br>· Total L×W: 54.4 × 34.0 (2.1 × 1.3) | Different<br>See Comment #1 | |
| Residual fluid | <0.05 mL for 13 mm & 20 mm<br><1 mL for 28 mm | <0.05 mL | Same | |
| Drug vial interface | Arisure™ Vial Adaptor snap lock and<br>elastomeric membrane - TPE Medalist 145 | ProSeal™ Vented Universal Vial Adaptor: snap<br>lock and elastomeric membrane -<br>polyisoprene | Different<br>See Comment #2 | |
| Closed system<br>technology/<br>mechanism of action | Physical barrier to prevent all mass from<br>crossing the system boundary | Physical barrier to prevent all mass from<br>crossing the system boundary | Same | |
| Design configuration and<br>function of drug vial<br>perforator | Arisure™ Vial Adaptor: plastics, ABS Terlux<br>2802 vial adaptor claws and<br>polycarbonate vial adapter spike<br>Predicate vial adaptor incorporates the<br>Universal vial spike - a dual lumen spike -<br>that makes for smooth fluid transfers while it<br>minimizes the drug residue left in the vial.<br>When used with the Arisure™ Syringe<br>Adaptor, the system provides transfer of<br>compounds safely and efficiently while<br>maintaining a closed system | ProSeal™ Vented Universal Vial Adaptor:<br>plastic, polypropylene integrated vial<br>adaptor claws and vial adaptor spike<br>Subject vial adaptor incorporates the<br>Universal vial spike - a dual lumen spike -<br>that makes for smooth fluid transfers while it<br>minimizes the drug residue left in the vial.<br>When used with the ProSeal™ Injector, the<br>system provides transfer of compounds<br>safely and efficiently while maintaining a<br>closed system | Different<br>See Comment #3 | |
| Pressure equalization in<br>drug vial | Integrated PTFE air-vent for air equalization<br>into and out of the drug vial<br>Predicate vented adaptor incorporates a<br>venting mechanism incorporating a 0.2<br>micron air filter that minimizes aerosols and<br>surface contamination and allows for the<br>release of pressure, thus neutralizing vial<br>pressure during drug transfer<br>The venting mechanism enables pressure<br>equalization within the vial, reducing the risk<br>of vial collapse or spray-back during drug<br>withdrawal, while maintaining a closed<br>system | Integrated PTFE air-vent for air equalization<br>into and out of the drug vial<br>Subject vented adaptor incorporates a<br>venting mechanism incorporating a 0.1<br>micron air filter that minimizes aerosols and<br>surface contamination and allows for the<br>release of pressure, thus neutralizing vial<br>pressure during drug transfer<br>The venting mechanism enables pressure<br>equalization within the vial, reducing the risk<br>of vial collapse or spray-back during drug<br>withdrawal, while maintaining a closed<br>system | Different<br>See Comment #4 | |
| Characteristics<br>compared | Predicate Device (K201422)<br>Arisure® CSTD - Vial Adaptor | Subject Device (K240517)<br>ProSeal™ Vented Universal Vial Adaptor | Comments,<br>discussions | |
| Biocompatibility | Externally Communicating Device, Blood<br>Path Indirect, Prolonged Contact (>24hr to<br>30d) | Externally Communicating Device, Blood<br>Path Indirect, Prolonged Contact (>24hr to<br>30d) | Same | |
| Sterile barrier packaging<br>top web and base web | Medical grade paper and medical plastic film,<br>heat sealed. Form, fill, seal packaging with top<br>web material sealed to bottom web material | Medical grade paper and medical plastic film,<br>heat sealed. Form, fill, seal packaging with top<br>web material sealed to bottom web material | Same | |
| Sterilization process | Gamma Irradiation, SAL 10-6 | Ethylene Oxide (EO), SAL 10-6 | Different<br>See Comment #5 | |
| Pyrogenicity | Non-pyrogenic | Non-pyrogenic | Same | |
| Prescription use or over-<br>the-counter use | R only | R only | Same | |
| Shelf-life validation | 36 months | 36 months | Same | |
{7}------------------------------------------------
{8}------------------------------------------------
## Discussion of differences in technological characteristics
## Comment #1
Differences in subcomponents (noted in Comments #1 to #4) related to their subcomponent types, dimensions and materials, as well as the different intended injectors, are considered, noting that all these aspects achieve the same intended use. Key differences in materials are addressed individually in Comment #2 and Comment #3, and difference in air-vent size, in Comment #4. Biocompatibility evaluations and functional performance testing demonstrate that all the differences do not raise new questions regarding safety and effectiveness.
## Comment #2
Polyisoprene vs. TPE: The Subject device's elastomeric membrane is made of the same polyissoprene as cleared K241823 ToxiSeal" Vial Adaptor subcomponent membrane. Biocompatibility and performance testing demonstrate that the difference does not raise new questions regarding safety and effectiveness.
## Comment #3
Polypropylene vs. ABS Terlux 2012 & polycarbonate: The Subject device's vial spike material is made of the same polypropylene as cleared K241823 ToxiSeal" Vial Adaptor's subcompatibility evaluation and performance testing demonstrate that the difference does not raise new questions regarding safety and effectiveness.
## Comment #4
0.1 micron air-vent vs 0.2 micron air-vent The difference in 0.2 micron air-vent in Subject device is vice is noted. The cleared ToxiSeal" Vial Adaptor incorporates the same air-vent (0.1 micron) (K241823). Performance testing demonstrates that this difference does not raise new questions regarding safety and effectiveness.
### Comment #5
EO sterilization vs. gamma sterilization: Use of gamma versus EO processes does not raise any new questions regarding safety and effectiveness. Both are assured to the same SAL as demonstrated by adequate sterility testing.
{9}------------------------------------------------
#### VII. Performance Data Supporting Substantial Equivalence
### A. Functional Performance
The ProSeal™ Vented Universal Vial Adaptor described in this Summary were tested and demonstrated to be in conformance with the following ISO and FDA recognized standards:
- ISO 8871-5:2016, Elastomeric parts for parenterals and for devices for pharmaceutical use Part ● 5: Functional requirements and testing
- ISO 8536-2:2010, Infusion equipment for medical use Part 2: Closures for infusion bottles ●
- ISO 8536-4:2019, Infusion equipment for medical use Part 4: Infusion sets for single use, ● gravity feed
- . ISO 22413:2010, Transfer sets for pharmaceutical preparations – Requirements and test methods
- Intravascular-Administration-Sets-Premarket-Notification-Submissions-[510(k)]---● Guidance-for-Industry-and-FDA-Staff
- NIOSH 2016, Performance Test (draft) Protocol for Closed System Transfer Devices ● Used During Pharmacy Compounding and Administration of Hazardous Drugs
Bench performance verifications and validations performed on the Subject device and cleared device:
- Positive and negative air pressure leakage test to ISO 8536-4:2019
- Positive pressure water leak integrity test to ISO 8536-4:2019 .
- Penetration force test of vial perforator to ISO 22413:2021 ●
- Fragmentation test to ISO 22413:2021 and to ISO 8871-5:2016 ●
- Vapor containment test per NIOSH CSTD 2016 draft test protocol (functional) from testing data . on devices cleared under K241823
- Microbial ingress test per FDA guidance, ANSI AAMI CN27:2021 and ISO 80369-7:2021 . (functional) from testing data on device cleared under K222929
- B. Biocompatibility
In accordance with ISO 10993-1:2018, the ProSeal™ Vented Universal Vial Adaptor is classified as: Externally Communicating Device, Blood Path Indirect, Prolonged Contact (>24hr to 30d). The following testing are referenced from testing on cleared devices of ProSeal™ CSTD:
- Cytotoxicity to ISO 10993-5 under K192075 ●
- Sensitization to ISO 10993-10 under K192075 ●
- Intracutaneous Reactivity to ISO 10993-10 under K192075 .
- Acute Systemic Toxicity to ISO 10993-11 under K192075
- . 14-day Subacute/ Subchronic Acute Systemic Toxicity to ISO 10993-11 under K222929
- In-vitro Hemolysis Assessment to ISO 10993-4 under K192075 and K192075/S001 ●
- . Material Mediated Pyrogenicity to ISO 10993-11 under K192075
- . Chemical Characterization and Toxicological Risk Management to ISO 10993-18 and ISO 10993-17 under K192075/S001 and K222929
Particulate matter testing was conducted on the Subject device, in accordance with ISO 8536-4:2019, Infusion equipment for medical use - Part 4: Infusion sets for single use, gravity feed, Annex A.2 and USP <788> Particulate Matter in Injections.
{10}------------------------------------------------
# 6010
- C. Sterility, Shipping, and Shelf-Life
The Subject device complies with sterilization requirements of ISO 11135:2014, Sterilization of Health Care Products - Ethylene Oxide - Part 1: Requirements for Development, Validation and Routine Control of a Sterilization Process for Medical Devices and the following testing/ evaluations:
- Simulated shipping testing per ASTM D4169-16, Standard Practice for Performance Testing of Shipping Containers and System under K192075
- -Package Integrity Tests per ASTM F1980-16, Standard guide for accelerated aging of sterile barrier systems for medical devices and Sterile Barrier Packaging Testing performed on the proposed device: Seal strength – ASTM F88/F88M-21, Standard test method for seal strength of flexible barrier materials, ASTM F1929-15, Standard test method for detecting seal leaks in porous medical device packaging by dye penetration, EN 868-5:2018, Packaging materials and systems for medical devices which are to be sterilized – Part 5: Heat and self-sealable pouches and reels of paper and plastic film construction – Requirements and test methods
- । Pyrogen Tests per ANSI/AAMI ST72/2019, Bacterial endotoxins – Test methods, routing monitoring, and alternatives to batch testing, USP 40<151>, Pyrogen test (USP rabbit test), USP-NF <161>, Medical Devices-Bacterial Endotoxin and Pyrogen Tests, USP-NF <85>, Bacterial Endotoxins Test and testing will be conducted on every lot
- Shelf-life of 3 years has been validated using the FDA recognized standard, ASTM 1980-16, -Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
## VIII. Clinical Tests
Not applicable.
#### IX. Conclusion
The differences between the Predicate and the Subject device do not raise any new or different questions of safety or effectiveness. The ProSeal" Vented Universal Vial Adaptor is substantially equivalent to the Predicate, the Arisure "CSTD - Vial Adaptor, cleared under K201422 with respect to the indications for use, principles of operation and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.