K240499 · Airos Medical, Inc. · JOW · Oct 11, 2024 · Cardiovascular
Device Facts
Record ID
K240499
Device Name
ARTAIRA Arterial Compression Device (AACD01)
Applicant
Airos Medical, Inc.
Product Code
JOW · Cardiovascular
Decision Date
Oct 11, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ARTAIRA Arterial Compression Device's is an intermittent pneumatic compression device intended for treatment of patients with the following conditions: - Intermittent claudication - Rest pain - Diabetic Foot - Ischemic neuritis - Arterial ulcers - Gangrene - Poor runoff.
Device Story
Intermittent pneumatic compression device; increases lower extremity blood flow/circulation. System comprises hard casing with air pump, valves, and reservoir; connects via tubing to compression garments (left/right leg) featuring foot/ankle and calf air chambers. Device software controls pneumatic pressure delivery in timed, cyclical bursts. Intended for home use per physician prescription. User interface includes color LCD displaying pressure and countdown timer; LED indicators for device state and errors. Benefits patient by providing mechanical compression therapy to manage arterial conditions.
Clinical Evidence
Bench testing only. Performance testing included user interface/error indicator testing, noise testing, compression sequence timing accuracy, pressure accuracy, therapy time accuracy, garment integrity, and air leakage testing.
Technological Characteristics
Intermittent pneumatic compression system; hard casing with air pump, valves, and reservoir. Biocompatible compression garments with dual air chambers (foot/ankle and calf). Color LCD display; LED status/error indicators. Standards: ISO 14971:2019, ISO 10993-1:2018, ISTA 3A:2018, ES 60601-1, IEC 60601-1-2, IEC 61000-3-2, IEC 61000-3-3, IEC 60601-1-6, IEC 60601-1-11.
Indications for Use
Indicated for patients with intermittent claudication, rest pain, diabetic foot, ischemic neuritis, arterial ulcers, gangrene, or poor runoff.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
October 11, 2024
AIROS Medical, Inc. Darren Behuniak Vice President, Operations and Marketing 2501 Monroe Blvd., Suite 1200 Audubon, Pennsylvania 19403
Re: K240499
Trade/Device Name: ARTAIRA Arterial Compression Device (AACD01) Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: September 16, 2024 Received: September 17, 2024
Dear Darren Behuniak:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Nicole M. Gillette -S
Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural, and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240499
#### Device Name
ARTAIRA Arterial Compression Device
Indications for Use (Describe)
The ARTAIRA Arterial Compression Device's is an intermittent pneumatic compression device intended for treatment of patients with the following conditions:
- · Intermittent claudication
- · Rest pain
- Diabetic Foot
- · Ischemic neuritis
- · Arterial ulcers
- · Gangrene
- Poor runoff.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for AIROS Medical. The logo consists of a hexagon with a green "A" inside, followed by the text "AIROS" in a simple, sans-serif font. Below "AIROS" is the word "MEDICAL" in a smaller font size.
### 510(k) Summary
September 9, 2024
### Submitter:
AIROS Medical, Inc. 2501 Monroe Blvd, Suite 1200 Audubon, PA 19403
## Contact Person:
Darren Behuniak, VP Operations & Marketing Email: dbehuniak(@airosmedical.com Phone: 866-991-6956
#### Common Classification & Proprietary Names
Trade Name: ARTAIRA Arterial Compression Device
#### Classification
| Classification Name | 21 CFR Regulation | Product Code | Class |
|-----------------------------|-------------------|--------------|-------|
| Compressible Limb<br>Sleeve | 870.5800 | JOW | II |
#### Predicate Device
The ARTAIRA Arterial Compression Device is substantially equivalent to the predicate device, the ArtAssist Device.
| | Predicate ArtAssist Device |
|--------------------------|----------------------------|
| 510(k) | K131146 |
| 21 CFR Regulation Number | 870.5800 |
| Product Code | JOW |
| Classification | II |
#### Device Description
The AIROS ARTAIRA Arterial Compression Device is an intermittent pneumatic compression device intended to increase blood flow and increase circulation in the lower extremities by providing timed compression to the foot/ankle and calf. The device consists of a hard casing with an air pump, a set of air valves, an air reservoir controlled by the device software to deliver a rapid burst of pneumatic pressure to the treatment area. The device connects through air tubing to a pair of compression garments (left and right leg), each with two air chambers: one on the underfoot (Foot/Ankle chamber) and one on the calf area (Calf Chamber). The device can either be used with one garment or two garments. The system is intended to be used primarily at home settings according to a physician-prescribed therapy.
www.airosmedical.com 1.866.991.6956
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for AIROS Medical. The logo consists of a hexagon with the letter A inside of it on the left, and the word "AIROS" above the word "MEDICAL" on the right. The hexagon is gray, the letter A is green, and the words "AIROS MEDICAL" are gray.
# Indication for Use
The ARTAIRA Arterial Compression Device is an intermittent pneumatic compression device intended for treatment of patients with the following conditions:
- Intermittent claudication
- Rest pain ●
- Diabetic Foot
- Ischemic neuritis ●
- Arterial ulcers
- Gangrene ●
- Poor runoff. ●
### Technological Characteristics
The manufacturer believes that the technological characteristics of the ARTAIRA are substantially equivalent to those of the predicate device. The ARTAIRA has the same pressure setting, as well as the same compression and non-compression timing sequences. Both devices are used with associated compression garments that feature air chambers in both the foot/ankle area and the underfoot. The garments can be used on one limb or two and the material is biocompatible. The device and grarment system provides cyclical air compression sequences intended to increase blood flow to the lower extremities.
The ARTAIRA also features a large color LCD screen that can be easily viewed by the user. The LCD screen displays pressure as well as a countdown timer so the user knows how much therapy time is left in the treatment session. Additional device features include LED lights to indicate device states and multiple error indicators that are triggered in the event of an error with the device.
#### Functional Performance Testing
Testing was performed and to ensure that the system meets its specifications. The functional performance testing includes the following tests:
- User Interface & Error Indicator Testing
- Noise Testing
- Compression Sequence Timing Accuracy Testing ●
- Pressure Accuracy Testing ●
- Therapy Time Accuracy Testing ●
- Garment Integrity Testing
- Air Leakage Testing
#### Standards
www.airosmedical.com 1.866.991.6956
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for AIROS Medical. The logo consists of a hexagon with a green letter A inside, followed by the text "AIROS" in gray, with the word "MEDICAL" underneath in a smaller font. The logo is clean and modern, with a focus on the company name.
- ISO 14971:2019 Medical Devices - Application of Risk Management to Medical Devices
- ISO 10993-1 2018 Biological evaluation of medical devices Part 1: Evaluation and ● testing within a risk management process
- ISTA 3A:2018 Packaged Products for Parcel Delivery System Shipment ●
- ES 60601-1: 2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 -● Medical electrical equipment- Part 1: General requirements for basic safety and essential performance
- . IEC 60601-1-2: 2014 + AMD1:2020 CSV - Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral Standard: Electromagnetic disturbances- Requirements and tests
- IEC 61000-3-2: 2018 A1: 2020 EMC Part 3-2: Limits Limits for harmonic current . emissions (Equipment input current 16A per phase)
- IEC 61000-3-3: 2013 A1: 2017 EMC - Part 3-3: Limits - Limitation of voltage changes, voltage fluctuations and flicker in public low-voltage supply systems for equipment with rated current ≤ 16 A per phase and not subject to conditional connections
- IEC 60601-1-6: 2010/AMD1: 2013 Medical electrical equipment- Part 1-6: General requirements for basic safety and essential performance- Collateral Standard: Usability
- IEC 60601-1-11: 2015 Medical electrical equipment- Part 1-11: General requirements . for basic safety and essential performance- Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
# Statement of Substantial Equivalence
The ARTAIRA is substantially equivalent in technology, function, operating parameters, and indications to the ArtAssist predicate device There are no new risks introduced with the ARTAIRA.
## Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act and 21 CFR Part 807, and based on the information provided in this pre-market notification, AIROS Medical, Inc., believes that the ARTAIRA is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.