IntelliBlate Microwave Ablation System Console (IB-CON); IntelliBlate Microwave Ablation System Mobile Cart (IB-CART); IntelliBlate Ximitry Probe Assembly, 13G x 15cm (IB-XPA-1315); IntelliBlate Ximitry Probe Assembly, 13G x 20cm (IB-XPA-1320); IntelliBlate Ximitry Probe Assembly, 13G x 27cm (IB-XPA-1327)
K240480 · Varian Medical Systems, Inc. · NEY · Jul 19, 2024 · General, Plastic Surgery
Device Facts
Record ID
K240480
Device Name
IntelliBlate Microwave Ablation System Console (IB-CON); IntelliBlate Microwave Ablation System Mobile Cart (IB-CART); IntelliBlate Ximitry Probe Assembly, 13G x 15cm (IB-XPA-1315); IntelliBlate Ximitry Probe Assembly, 13G x 20cm (IB-XPA-1320); IntelliBlate Ximitry Probe Assembly, 13G x 27cm (IB-XPA-1327)
Applicant
Varian Medical Systems, Inc.
Product Code
NEY · General, Plastic Surgery
Decision Date
Jul 19, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K240480 · Jul 19, 2024
IntelliBlate Microwave Ablation System Console (IB-CON); IntelliBlate Microwave Ablation System Mobile Cart (IB-CART); IntelliBlate Ximitry Probe Assembly, 13G x 15cm (IB-XPA-1315); IntelliBlate Ximitry Probe Assembly, 13G x 20cm (IB-XPA-1320); IntelliBlate Ximitry Probe Assembly, 13G x 27cm (IB-XPA-1327)
Varian Medical Systems, Inc.
Published clinical literature (systematic literature review)
A systematic literature review was conducted to support the indications for use by evaluating clinical outcomes and complication rates from 23 publications involving microwave ablation systems in routine clinical settings.
Systematic Literature Review; Clinical Outcomes; Soft Tissue Ablation
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Systematic Literature Review of Microwave Ablation; Systematic literature review and meta-analysis
Radio-frequency ablation, cryoablation, or surgical resection
Safety and effectiveness of microwave ablation for soft tissue coagulation
Indications for Use
The IntelliBlate Microwave Ablation System is intended for coagulation (ablation) of soft tissue. The IntelliBlate Microwave Ablation System is not intended for use in cardiac procedures. The IntelliBlate Ximitry Probe Assembly, used with the IntelliBlate Microwave Ablation System is intended for coagulation (ablation) of soft tissue. The IntelliBlate Ximitry Probe Assembly is not intended for use in cardiac procedures.
Device Story
Microwave ablation system for soft tissue coagulation; utilizes 2.45 GHz microwave energy to induce cellular death via coagulation necrosis. System comprises tabletop console with integrated peristaltic pump for coolant delivery and sterile, single-use disposable Ximitry Probe Assemblies. Operated by clinicians in facility settings. Console features graphical user interface (GUI) and touchscreen for monitoring and control; supports two independent ablation channels. Probe assembly integrates power cable and coolant tubing (supply/return) into single bundle; includes ceramic trocar tip and multiple thermocouples for temperature monitoring. System provides ablation, track ablation, and needle lock modes. Output power up to 170W (single) or 2x 150W (double). Safety features include automatic shutdown upon temperature limit or high reflected power. Output affects clinical decision-making by enabling targeted tissue necrosis; benefits patients by providing minimally invasive soft tissue ablation.
Clinical Evidence
No clinical data. Evidence consists of non-clinical bench testing (functional, ablation performance in ex-vivo porcine liver/kidney/muscle, reliability, electrical safety, EMC), GLP animal studies (porcine soft tissue), and systematic literature review of predicate device performance (1,628 patients, 2,438 tumors).
Indicated for coagulation (ablation) of soft tissue. Not for use in cardiac procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Emprint Ablation System with Thermosphere Technology (K200796)
Neuwave Certus 140 2.45 GHz Ablation System and Accessories (K100744)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the full name of the agency on the right. The symbol on the left is the Department of Health & Human Services logo. The FDA acronym is in a blue square, and the full name of the agency, "U.S. Food & Drug Administration," is in blue text to the right of the square.
July 19, 2024
Varian Medical Systems, Inc. Lynn Allman Sr. Director Regulatory Affairs 3100 Hansen Way Palo Alto, California 94304
# Re: K240480
Trade/Device Name: IntelliBlate Microwave Ablation System Console (IB-CON); IntelliBlate Microwave Ablation System Mobile Cart (IB-CART); IntelliBlate Ximitry Probe Assembly, 13G x 15cm (IB-XPA-1315); IntelliBlate Ximitry Probe Assembly, 13G x 20cm (IB-XPA-1320); IntelliBlate Ximitry Probe Assembly, 13G x 27cm (IB-XPA-1327) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: NEY Dated: June 17, 2024 Received: June 18, 2024
Dear Lynn Allman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Colin K. Colin K. Chen -S Date: 2024.07.19 Chen -S 13:35:14 -04'00'
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices
for
{2}------------------------------------------------
OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240480
Device Name
IntelliBlate Microwave Ablation System Console (IB-CON); IntelliBlate Microwave Ablation System Mobile Cart (IB-CART); IntelliBlate Ximitry Probe Assembly, 13G x 15cm (1B-XPA-1315); IntelliBlate Ximitry Probe Assembly, 13G x 20cm (1B-XPA-1320); IntelliBlate Ximitry Probe Assembly, 13G x 27cm (IB-XPA-1327).
Indications for Use (Describe) IntelliBlate Microwave Ablation System:
The IntelliBlate Microwave Ablation System is intended for coagulation (ablation) of soft tissue. The IntelliBlate Microwave Ablation System is not intended for use in cardiac procedures.
IntelliBlate Ximitry Probe Assembly:
The IntelliBlate Ximitry Probe Assembly, used with the IntelliBlate Microwave Ablation System is intended for coagulation (ablation) of soft tissue.
The IntelliBlate Ximitry Probe Assembly is not intended for use in cardiac procedures.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows three blue circles in a row. The circles are evenly spaced and are all the same size. The background is white. The circles are a bright, vibrant blue.
Varian Medical Systems, Inc 3100 Hansen Way Palo Alto, CA 94304-1038
Telephone: 1.650.493.4000 Toll Free: 800.544.4636 www.varian.com
# PREMARKET NOTIFICATION
K240480 510(k) Summary IntelliBlate Microwave Ablation System and Accessories The following information is provided as required by 21 CFR 807.92
#### Submitter's Information: l.
| Name and Address: | Varian Medical Systems Inc.<br>9825 Spectrum Drive, Building 2 Austin, TX 78717 |
|-------------------|---------------------------------------------------------------------------------|
| Contact Name: | Lynn Allman, Senior Director Regulatory Affairs |
| E-mail: | submissions.support@varian.com |
| Date Prepared: | July 17, 2024 |
#### ll. Device Information:
| Proprietary Name: | IntelliBlate Microwave Ablation System and Accessories |
|-----------------------|----------------------------------------------------------------|
| Common/ Usual Name: | System, Ablation, Microwave and Accessories |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories |
| Classification Panel: | General & Plastic Surgery |
| Regulation Number: | 21 CFR 878.4400 |
| Product Code | NEY |
#### III. Predicate Devices:
Emprint Ablation System with Thermosphere Technology (K200796) Neuwave Certus 140 2.45 GHz Ablation System and Accessories (K100744)
#### IV. Subject Device Description:
The IntelliBlate Microwave Ablation System is a software-controlled microwave therapy delivery system with graphical user interface and hardware to control and monitor the therapy delivered to the target tissues.
The system consists of a 2.45 GHz microwave generator (console) with an integrated peristaltic pump assembly for coolant fluid delivery, a coolant Fluid Mount and an optional mobile cart for the console. Accessories for the system include IntelliBlate Ximitry Probe Assembly - a sterile, single-use disposable microwave ablation assembly with ablation probes and a cassette for coolant fluid tubing (closed cooling circuit).
{5}------------------------------------------------
#### V. Intended Use/ Indications of Use Statement:
### IntelliBlate Microwave Ablation System
The IntelliBlate Microwave Ablation System is intended for coagulation (ablation) of soft tissue.
The IntelliBlate Microwave Ablation System is not intended for use in cardiac procedures.
### IntelliBlate Ximitry Probe Assembly
The IntelliBlate Ximitry Probe Assembly, used with the IntelliBlate Microwave Ablation System is intended for coagulation (ablation) of soft tissue.
The IntelliBlate Ximitry Probe Assembly is not intended for use in cardiac procedures.
#### VI. Substantial Equivalence Discussion:
The following table compares the subject device to the predicate devices with respect to indications for use, principles of operation, technological characteristics, materials, and performance, and forms the basis for the determination of substantial equivalence.
### Table 1: Comparison of Technological Characteristics between subject and predicate devices
{6}------------------------------------------------
| Features/ Characteristics | Subject Device<br>IntelliBlate Microwave Ablation System<br>and Accessories | Predicate Device<br>Emprint Ablation System with<br>Thermosphere Technology (K200796) | Predicate Device<br>Neuwave Certus 140 2.45 GHz<br>Ablation System and Accessories<br>(K100744) | Comparison |
|-------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product code, FDA<br>regulation and device<br>class | NEY, 21 CFR 878.4400 and Class II | NEY, 21 CFR 878.4400 and Class II | NEY, 21 CFR 878.4400 and Class II | Same |
| 510(k) clearance(s) | No prior 510(k) clearances | K200796 | K100744 | Subject device is a new device<br>with no prior regulatory<br>clearances. |
| Manufacturer | Varian Medical Systems | Covidien | Neuwave Medical Inc. | Different manufacturer |
| Prescription (Rx only) | Yes | Yes | Yes | Same |
| Intended Use | Intended for coagulation (ablation) of<br>soft tissue in laparoscopic,<br>intraoperative, and percutaneous<br>ablation procedures | Intended for coagulation (ablation) of<br>soft tissue in laparoscopic,<br>intraoperative, and percutaneous<br>ablation procedures. | The Neuwave Medical Certus 140<br>2.45 GHz Ablation System and<br>Accessories are intended for the<br>ablation (coagulation) of soft tissue.<br>The Certus 140™ 2.45 GHz Ablation<br>System is not intended for use in<br>cardiac procedures.<br>NeuWave recommends against the<br>use of the Certus 140 2.45 GHz<br>Ablation System in the following<br>situations:<br>* Pregnant patients - potential risks<br>to patient and/or fetus have not<br>been established<br>* Patients with implantable<br>pacemakers or other electronic<br>implants. Implanted electronic<br>devices may be adversely affected by<br>microwave power<br>The system is designed for facility use<br>and should only be used under the<br>orders of a clinician. | Same indication of use -subject<br>and predicate devices are<br>indicated for coagulation<br>(ablation) of soft tissue.<br>Subject device does not include<br>percutaneous, laparoscopic,<br>endoscopic, and intraoperative<br>coagulation (ablation) of soft<br>tissue, including partial or<br>complete ablation of<br>nonresectable liver tumors. |
| Features/ Characteristics | Subject Device<br>IntelliBlate Microwave Ablation System<br>and Accessories | Predicate Device<br>Emprint Ablation System with<br>Thermosphere Technology (K200796) | Predicate Device<br>Neuwave Certus 140 2.45 GHz<br>Ablation System and Accessories<br>(K100744) | Comparison |
| Indications for Use<br>Statement for IntelliBlate<br>Microwave Ablation<br>System | The IntelliBlate Microwave Ablation<br>System is intended for coagulation<br>(ablation) of soft tissue.<br><br>The IntelliBlate Microwave Ablation<br>System is not intended for use in<br>cardiac procedures. | The Emprint™ Ablation System is<br>intended for use in percutaneous,<br>laparoscopic, endoscopic, and<br>intraoperative coagulation (ablation) of<br>soft tissue, including partial or<br>complete ablation of nonresectable<br>liver tumors.<br><br>The Emprint™ Ablation System is not<br>intended for use in cardiac procedures. | The Neuwave Medical Certus 140<br>2.45 GHz Ablation System and<br>Accessories are intended for the<br>ablation (coagulation) of soft tissue.<br><br>The Certus 140™ 2.45 GHz Ablation<br>System is not intended for use in<br>cardiac procedures.<br><br>NeuWave recommends against the<br>use of the Certus 140 2.45 GHz<br>Ablation System in the following<br>situations:<br>* Pregnant patients - potential risks<br>to patient and/or fetus have not<br>been established<br>* Patients with implantable<br>pacemakers or other electronic<br>implants. Implanted electronic<br>devices may be adversely affected by<br>microwave power<br>The system is designed for facility use<br>and should only be used under the<br>orders of a clinician. | Same indication of use –subject<br>and predicate devices are<br>indicated for coagulation<br>(ablation) of soft tissue.<br><br>Subject device does not include<br>percutaneous, laparoscopic,<br>endoscopic, and intraoperative<br>coagulation (ablation) of soft<br>tissue, including partial or<br>complete ablation of<br>nonresectable liver tumors. |
| Indications for Use<br>Statement for IntelliBlate<br>Ximitry Probe Assembly | The IntelliBlate Ximitry Probe Assembly,<br>used with the IntelliBlate Microwave<br>Ablation System is intended for<br>coagulation (ablation) of soft tissue.<br><br>The IntelliBlate Ximitry Probe Assembly<br>is not intended for use in cardiac<br>procedures. | The Covidien Emprint™ Ablation System<br>is intended for use in percutaneous,<br>laparoscopic, endoscopic, and<br>intraoperative coagulation (ablation) of<br>soft tissue, including partial or<br>complete ablation of nonresectable<br>liver tumors.<br><br>The Covidien Emprint™ Ablation<br>System is not intended for use in<br>cardiac procedures. | The Neuwave Medical Certus 140<br>2.45 GHz Ablation System and<br>Accessories are intended for the<br>ablation (coagulation) of soft tissue.<br><br>The Certus 140™ 2.45 GHz Ablation<br>System is not intended for use in<br>cardiac procedures.<br><br>NeuWave recommends against the<br>use of the Certus 140 2.45 GHz<br>Ablation System in the following<br>situations. | Same indication of use -subject<br>and predicate devices are<br>indicated for coagulation<br>(ablation) of soft tissue.<br><br>Subject device does not include<br>percutaneous, laparoscopic,<br>endoscopic, and intraoperative<br>coagulation (ablation) of soft<br>tissue, including partial or<br>complete ablation of<br>nonresectable liver tumors. |
| Features/ Characteristics | Subject Device | Predicate Device | Predicate Device | Comparison |
| | IntelliBlate Microwave Ablation System<br>and Accessories | Emprint Ablation System with<br>Thermosphere Technology (K200796) | Neuwave Certus 140 2.45 GHz<br>Ablation System and Accessories<br>(K100744) | |
| | | | * Pregnant patients - potential risks<br>to patient and/or fetus have not<br>been established<br>* Patients with implantable<br>pacemakers or other electronic<br>implants. Implanted electronic<br>devices may be adversely affected by<br>microwave power<br>The system is designed for facility use<br>and should only be used under the<br>orders of a clinician. | |
| Operating Principle for<br>microwave ablation | Microwave ablation is heating of the<br>target tissue based on agitation of<br>water molecules inducing cellular death<br>via coagulation (ablation) necrosis | Microwave ablation is heating of the<br>target tissue based on agitation of<br>water molecules inducing cellular death<br>via coagulation (ablation) necrosis | Microwave ablation is heating of the<br>target tissue based on agitation of<br>water molecules inducing cellular<br>death via coagulation (ablation)<br>necrosis | Same |
| System Configuration | | | | |
| Generator | Tabletop Generator (console) | Cart mounted Generator | Cart mounted Generator | Same |
| No of independent<br>ablation channels | 2 | 1 | 3 | Subject device supports 2<br>independent ablation channels |
| Cooling System | Peristaltic Pump Assembly integrated<br>into the generator for each ablation<br>channel | Peristaltic Pump Assembly separate and<br>external to the generator | CO2 based cooling system | Subject and predicate device<br>(K200796) has peristaltic pump<br>assembly for cooling of the<br>ablation probes. |
| Touch Screen/ Graphical<br>User Interface | Yes | No (Physical Buttons and digital display) | Yes | GUI/ Touch Screen |
| Accessories | | | | |
| Ablation Probe Assembly | Includes ablation probe, coolant tubing,<br>coolant spike, integrated cassette for<br>coolant delivery and power cable. | Includes ablation probe, coolant tubing,<br>coolant spike and power cable. Coolant<br>delivery is external via manual<br>connection to the external pump. | Includes a probe handle, a 1.4M<br>cable, CO2 gas coolant tubing and a<br>connector assembly. An intermediate<br>junction box or Power Distribution | Subject device does not need an<br>additional reusable cable to<br>connect probe to the generator. |
| Features/ Characteristics | Subject Device<br>IntelliBlate Microwave Ablation System<br>and Accessories | Predicate Device<br>Emprint Ablation System with<br>Thermosphere Technology (K200796) | Predicate Device<br>Neuwave Certus 140 2.45 GHz<br>Ablation System and Accessories<br>(K100744) | Comparison |
| | Optional accessories – laser alignment disc and depth guides for ablation probes | | Module (PDM) is used between the probe and console. | |
| Optional Accessories | Temperature sensor probe, mobile cart | Remote Temperature Probe, mobile cart, footswitch | Footswitch | Similar optional accessories |
| Ablation | | | | |
| Ablation features | Ablation, Track Ablation and Needle Lock | Ablation.<br>Device can be used for track ablation as a use workflow. | Ablation, Cauterize (track ablation), Tissue Loc, Planar Coagulation | Additional pre-configured ablation features in subject device |
| Target Ablation time | Up to 10 minutes. | Up to 10 minutes | Up to 10 minutes | Same |
| Generator | | | | |
| Device Usage | Multi-use | Multi-use | Multi-use | Same |
| Compatible Disposable/<br>Single Use Devices | Microwave ablation probe (up to 2)<br>Optional Temperature Sensor Probe (up to 2) | Microwave ablation probe (1)<br>Optional Remote Temperature Probe (1) | Microwave ablation probe (up to 3) | Subject device can connect up to 2 probes |
| Generator Output<br>Frequency | 2.45GHz ± 50MHz | 2.45GHz ± 50MHz | 2.45GHz ± 25MHz | Same |
| Maximum RF power for<br>ablation (measured at<br>output port): | 170W (single needle ablation)<br>2x 150W (double needle ablation) | 150W (single needle ablation) | 140W (single needle ablation)<br>2x 95W (double needle ablation)<br>3x 65W (triple needle ablation) | Increased output power to compensate for cable losses. |
| Power Settings | Adjusts microwave power setpoint in 5-watt increments | Adjusts microwave power setpoint in 5-watt increments. | Adjusts microwave power setpoint in 5-watt increments | Same |
| Temperature Monitoring | Yes | Yes | Yes | Same |
| Safety Feature -<br>Automatic system<br>Shutdown when ablation<br>probe temperature | Yes | Yes | Yes | Same |
| Features/ Characteristics | Subject Device | Predicate Device | Predicate Device | Comparison |
| | IntelliBlate Microwave Ablation System<br>and Accessories | Emprint Ablation System with<br>Thermosphere Technology (K200796) | Neuwave Certus 140 2.45 GHz<br>Ablation System and Accessories<br>(K100744) | |
| exceed limits and due to<br>high reflected power | | | | |
| Safety Standard<br>Compliance | IEC 60601-1, 60601-1-6, 60601-2-6,<br>60601-1-2 | IEC 60601-1, 60601-1-6, 60601-2-6,<br>60601-1-2…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.