K240472 · Dongguan Tutamen Metalwork Co., Ltd. · OAP · Oct 11, 2024 · Physical Medicine
Device Facts
Record ID
K240472
Device Name
PBM Hair Therapy Cap
Applicant
Dongguan Tutamen Metalwork Co., Ltd.
Product Code
OAP · Physical Medicine
Decision Date
Oct 11, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PBM Hair Therapy Cap is indicated to treat And promote hair growth in females who have Ludwig-Savin Scale I to II, males who have Norwood Hamilton Classifications of Ila to V patterns of hair loss, females and males with frontal patterns of hair-loss both with Fitzpatrick Skin Types I to IV.
Device Story
PBM Hair Therapy Cap; fabric cap with 84 inbuilt 660nm red LED diodes (7 strips, 12 diodes/strip); battery-powered controller; OTC use for hair growth stimulation; user-operated; 15-minute automatic timer; USB-C charging; LED indicators for power/charging status; device emits red light to scalp; intended to treat androgenetic alopecia; benefits include potential hair growth promotion.
Clinical Evidence
Bench testing only. Verification of requirements (VR0001) and usability testing (US001) performed in accordance with IEC 62366-1:2015.
Technological Characteristics
Fabric cap with 84 red LED diodes (660nm); 3.7V lithium-ion battery; USB-C charging; 15-minute automatic timer; Class 3R laser system per IEC standards.
Indications for Use
Indicated for females with Ludwig-Savin Scale I-II and males with Norwood Hamilton Classifications IIa-V androgenetic alopecia; includes frontal hair loss patterns; applicable to Fitzpatrick Skin Types I-IV.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's full name on the right. The symbol on the left is a stylized representation of human services. To the right of the symbol is a blue square containing the acronym "FDA" in white letters. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
October 11, 2024
Dongguan Tutamen Metalwork Co., Ltd Caitlyn Dzhafarov Sr. Regulatory Consultant No.3, Huangguotang Road Tangxia Town DongGuan, 523000 China
Re: K240472
Trade/Device Name: PBM Hair Therapy Cap Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: October 10, 2024 Received: February 20, 2024
Dear Caitlyn Dzhafarov:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@;tda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA TANISHA L. HITHE -S L. HITHE -S Date: 2024.10.11
2020-02-02:43 -04'00'
Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K240472
Device Name
PBM Hair Therapy Cap
#### Indications for Use (Describe)
The PBM Hair Therapy Cap is indicated to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I to II, males who have Norwood Hamilton Classifications of Ila to V patterns of hair loss, females and males with frontal patterns of hair-loss both with Fitzpatrick Skin Types I to IV.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 510(k) #: | K240472 | | 510(k) Summary | | | Prepared on: 2024-10-10 | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|------------|------------------------------------------------------------|--|-----|-------------------------|--|
| Contact Details | | | | | | 21 CFR 807.92(a)(1) | |
| Applicant Name | | | Dongguan Tutamen Metalwork Co., LTD. | | | | |
| Applicant Address | | | No.3, Huangguotang Road Tangxia Town DongGuan 523000 China | | | | |
| | Applicant Contact Telephone | | 760-402-7859 | | | | |
| Applicant Contact | | | Mr. Stephen Prior | | | | |
| Applicant Contact Email | | | stephenprior@tutamen.net | | | | |
| Correspondent Name | | | Medical Devices Pathway, LLC. | | | | |
| Correspondent Address | | | 14330 178th Ln NE Woodinville WA 98072 United States | | | | |
| Correspondent Contact Telephone | | | 3602243622 | | | | |
| Correspondent Contact | | | Ms. Caitlyn Dzhafarov | | | | |
| Correspondent Contact Email | | | cdzhafarov@meddevpath.com | | | | |
| Device Name | | | | | | 21 CFR 807.92(a)(2) | |
| Device Trade Name | | | PBM Hair Therapy Cap | | | | |
| Common Name | | | Infrared lamp | | | | |
| Classification Name | | | Laser, Comb, Hair | | | | |
| Regulation Number | | | 890.5500 | | | | |
| Product Code | | | OAP | | | | |
| Legally Marketed Predicate Devices | | | | | | 21 CFR 807.92(a)(3) | |
| Predicate Trade Name (Primary Predicate is listed first)<br>Predicate # | | | | | | Product Code | |
| K173729 | | Revian Red | | | OAP | | |
| Device Description Summary | | | | | | 21 CFR 807.92(a)(4) | |
| The PBM Hair Therapy Cap is a hair therapy device that uses red light to help stimulate hair growth in both male and females. The device<br>comes with a Cap that has LED diodes inbuilt and an attachable controller battery pack that comes with an automatic timer that limits<br>the user to 15-minute sessions. | | | | | | | |
The device uses only 660nm LED wavelength diodes on 7 strips, each strip has 12 diodes. These diodes strips are sewn into a cap and has a Type C USB connector that the controller battery pack can connect to.
The PBM Hair Therapy Cap is comprised of a fabric cap with inbuilt LED diodes that emit red light on the wavelength of 660nm. The diodes are on strips and the cap style hat has 7 strips with 12 diodes on each strip with a total of 84 LED diodes.
A battery powered controller that attaches via a C-type USB connector to the fabric hat is also provided. The controller has a built in 3.7V
{5}------------------------------------------------
ithium-ion battery, has an on/off button and in programmed to automatically turn of after 15min of use. The controller also has a 2 small indicator LED's for indicating, power on/in use, needs charging and also blinks while in charging mode. The controller charges via inbuilt C-Type USB port and charges at 500 mA.
## Intended Use/Indications for Use
The PBM Hair Therapy Cap is indicated to treat And promote hair growth in females who have Ludwig-Savin Scale I to II, males who have Norwood Hamilton Classifications of Ila to V patterns of hair loss, females and males with frontal patterns of hair-loss both with Fitzpatrick Skin Types I to IV.
## Indications for Use Comparison
The indications for use statements are substantially equivalent, just worded sightly differently. Both are indicated for the same patient populations and have the same intended use.
# Technological Comparison
Both the PBM Hair Therapy Cap Device and Revian Red utilize LLLT mechanism of action with textile hair growth on the scalp in the same indicated patient population. The technical specifications of the PBM Hair Therapy Cap and the predicate Revian Red are comparable. The noted minor differences in technological characteristics, such as battery types (lithium polymer), wave-lengths (660 nm) and number of red LEDs (84 versus 119 Red LEDs) and treatment durations (15 minutes versus 10 minutes), are considered negligible and do not affect the therapeutic value or safety profile. Both devices are classified as class 3R laser systems in accordance with IEC standards and share a substantially equivalent adverse event profile. Many of the characteristics which differ between the PBM Hair Therapy Cap and the predicate, The Time Machine Series Lasers device, are due to the PBM Hair Therapy Cap being a subset of the intended use and indications for use of the predicate device and fall within the cleared spectrum of technical characteristics of the predicate device (e.g. fall within the cleared wavelength, etc.). None of these differences raises new or different questions of safety or effectiveness.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Verification of requirements (VR0001) and usability testing (US001, in accordance with IEC 62366-1:2015 and Applying Human Factors and Usability Engineering to Medical Devices, Guidance for Industry and Food and Drug Administration Staff)
### Not Applicable
With respect to the PBM Hair Therapy Cap comparison to a predicate device, Revian Red, the PBM Hair Therapy Cap has the same intended use, similar technological characteristics, and for minor differences in technological characteristics, these have been determined to not raise and different questions of safety or effectiveness. Scientific data supporting safety and effectiveness have been provided and are similar to the data required for the legally marketed predicate device. The totality of non-clinical data provides valid scientific evidence to support the conclusion that there are no different questions of safety and effectiveness raised when compared to the predicate device. This data demonstrates the PBM Hair Therapy Cap is substantially equivalent to the predicate Revian Red based upon the totality of the evidence.
## 21 CFR 807.92(a)(5)
21 CFR 807.92(a)(5)
# 21 CFR 807.92(a)(6)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.