C2 CryoBalloon Ablation System (C2 CryoBalloon Ablation System)
K240457 · Pentax of America, Inc. · GEH · Jul 24, 2024 · General, Plastic Surgery
Device Facts
Record ID
K240457
Device Name
C2 CryoBalloon Ablation System (C2 CryoBalloon Ablation System)
Applicant
Pentax of America, Inc.
Product Code
GEH · General, Plastic Surgery
Decision Date
Jul 24, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K240457 · Jul 24, 2024
C2 CryoBalloon Ablation System (C2 CryoBalloon Ablation System)
Pentax of America, Inc.
Retrospective medical record/chart review; Endoscopic report findings; Published clinical literature
A retrospective study of 28 patients was used to support the safety and effectiveness of the device for the treatment and management of GAVE, comparing outcomes to a published literature cohort of 33 patients treated with alternative modalities (APC/RFA).
Multi-center retrospective study of C2 CryoBalloon for GAVE; Retrospective cohort study; Follow-up/Duration: 6 months
28 subjects with GAVE previously treated with APC, RFA, LNC, or treatment-naïve; Sample Size: 28; Number of Sites: Multi-center
Published study of 33 GAVE patients treated with APC or RFA
Mean number of PRBC transfusions, transfusion independence, endoscopic GAVE eradication
Indications for Use
The C2 CryoBalloon™ Ablation System is intended to be used as a cryosurgical tool in the field of general surgery, specifically for endoscopic applications, to include ablation of Barrett's Esophagus with dysplasia and treatment and management of Gastric Antral Vascular Ectasia (GAVE).
Device Story
Cryosurgical system for endoscopic tissue ablation; utilizes nitrous oxide to cool a balloon at the distal end of a catheter. System components: sterile single-use catheter, reusable controller, reusable foot pedal, single-use cartridge, and NitroClip accessory. Physician inserts catheter through endoscope (min 3.7mm working channel) to treatment site; balloon inflated with nitrous oxide; tissue visualized through balloon; cryogen spray cools balloon to ablate tissue; exhaust managed via controller. Used in clinical settings by physicians. Output is controlled tissue ablation; clinical decision-making guided by endoscopic visualization and controller settings. Benefits include effective management of GAVE and Barrett's Esophagus with dysplasia.
Clinical Evidence
Clinical evidence includes a multi-center retrospective study (n=28) for GAVE treatment, showing 100% technical success, 71% transfusion independence at 6 months, and 88% GAVE eradication. Two studies (n=25 each) evaluated 180° catheter BE treatment at 1.0mm/s and 1.2mm/s doses; results showed 90-95% BE regression with minimal adverse events. No animal data provided.
Technological Characteristics
Cryosurgical unit; nitrous oxide-based cooling; compliant balloon diffuser; catheter compatible with endoscopes (min 3.7mm ID, max 105cm length). Software-activated controller; user-controlled via foot pedal. RFID-enabled catheter parameters. Sterilization method not specified.
Indications for Use
Indicated for patients requiring endoscopic cryosurgical ablation of Barrett's Esophagus with dysplasia or treatment and management of Gastric Antral Vascular Ectasia (GAVE).
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 24, 2024
PENTAX of America, Inc. Gurvinder Singh Nanda Senior Director, Regulatory and Quality 3 Paragon Drive Montvale, New Jersey 07645
Re: K240457
Trade/Device Name: C2 CryoBalloon Ablation System (C2 CryoBalloon Ablation System) Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit And Accessories Regulatory Class: Class II Product Code: GEH Dated: June 21, 2024 Received: June 21, 2024
Dear Gurvinder Singh Nanda:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long H. Chen -Si Digitally signed by Long H.Chen -S
Date: 2024.07.24 08:20 Date: 2024.07.24 08:36:02 -04'00'
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
#### Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240457
Device Name C2 CryoBalloon Ablation System
Indications for Use (Describe)
The C2 CryoBalloon™ Ablation System is intended to be used as a cryosurgical tool in the field of general surgery. specifically for endoscopic applications, to include ablation of Barrett's Esophagus with dysplasia and treatment and management of Gastric Antral Vascular Ectasia (GAVE).
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
#### I. SUBMITTER
PENTAX of America 3 Paragon Drive Montvale, NJ 07645 Phone: 650-722-4189
Contact Name: Gurvinder Singh Nanda, PhD. Date Prepared: June 19, 2024
II. DEVICE
Name of Device: C2 CryoBalloon™ Ablation System Common or Usual Name: Cryosurgical Unit and Accessories Classification Name: Cryosurgical Unit and Accessories Regulatory Class: 21 CFR §878.4350 Product Code: GEH
#### III. PREDICATE DEVICE
C2 CryoBalloon™ Ablation System, K212814
IV. REFERENCE DEVICE
BARRX Medical HALO® System, K093008
#### V. DEVICE DESCRIPTION
The subject device is a cryosurgical unit with a nitrous oxide cooled balloon that is compatible with commercially available endoscopes with a minimum working chamel inner diameter of 3.7 mm and maximum length of 105 cm. The subject device is a cryosurgical system comprised of four components including a Catheter (sterile, single use), Controller (non-sterile, reusable), Foot Pedal (non-sterile, reusable), and Cartridge (non-sterile, single use). The system also includes NitroClip as an accessory that is intended to reduce the nitrous oxide exposure in the room.
The subject device is used to ablate unwanted tissue by application of extreme cold. The balloon at the distal end of the Catheter comes in contact with tissue and is inflated with nitrous oxide. Tissue is visualized through the pre-inflated balloon, and the treatment site is selected by adjusting the endoscope and Controller. The nitrous oxide spray cools the balloon to ablate the unwanted tissue, and the nitrous oxide exhausts through the Controller.
{4}------------------------------------------------
#### VI. INDICATIONS FOR USE
The C2 CryoBalloon™ Ablation System is intended to be used as a cryosurgical tool in the field of general surgery, specifically for endoscopic applications, to include ablation of Barrett's Esophagus with dysplasia and treatment and management of Gastric Antral Vascular Ectasia (GAVE).
#### VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Cryoablation is the fundamental technological principle for both the subject C2 CryoBalloon Ablation System and the predicate C2 CryoBalloon Ablation System (K212824). There have been no significant changes to the technological characteristics of the C2 CryoBalloon Ablation System since of the predicate device. The subject and the legally marketed predicate device C2 CryoBalloon Ablation System have the same design, materials, energy source, and principles of operation.
The subject and the predicate devices are based on the following same technological elements:
- · Inserted through an endoscope to access the treatment site
- · Application of cryogen to ablate (freeze) the unwanted tissue
- · Use of a compliant balloon to position the treatment diffuser and to contain and exhaust the cryogen
- · User-controlled (trigger/foot pedal) to release cryogen
- Software activated Controller
The only difference is the indication of use of the subject device now includes treatment and management of Gastric Antral Vascular Ectasia (GAVE).
The inclusion in the labeling of a lower dose to treat patients with short segment Barrett's Esophagus (BE) using the subject C2 CryoBalloon 180 catheter is within the device's current intended use and indications for use. The change of the catheter's RFID value of Maximum Treatment Speed from 1.0mm/s to 1.2mm/s for the lower dose falls under existing design verification. The addition of this lower treatment rate is also supported by bench and clinical data demonstrating its safety and effectiveness.
Two caution statements have been removed from the Instructions for Use. These modifications are supported by bench and / or clinical data regarding the safety and effectiveness of its use, which do not raise different questions of safety and effectiveness.
## VIII.PREFORMANCE DATA
#### Bench Testing A.
Bench testing was performed to verify the results of the temperature profiling data of the 180 Standard catheter at 1.0 mm/s and 1.2 mm/s speeds. The tests passed all acceptance criteria. Testing results verified
{5}------------------------------------------------
the spray radius and spray uniformity of the Standard 180 Catheter, confirming that the 180 Catheter can be used at speeds of 1.0 and 1.2mm/s.
Test method validation was performed for the C2 CryoBalloon Catheters that use the 180-degree diffuser. Temperature distribution tests were performed to verify temperature spray profiles for catheters, where the profile determines what percentage of the spray pattern reaches optimal temperatures for treatment. The results of the testing demonstrated that the temperature distribution can be produced consistently and accurately.
No animal data was used to support this 510(k) premarket notification.
- B. Clinical Testing
- 1. Gastric Antral Vascular Ectasia (GAVE)
A multi-center retrospective study evaluated the C2 CryoBalloon system focal for the treatment and management of GAVE. 28 subjects were included based on endoscopic report findings and history of clinical evidence of GAVE. All patients in the study had been previously treated to manage their GAVE disease with argon plasma coagulation (APC), radiofrequency ablation (RFA), liquid nitrogen cryotherapy (LNC) or were treatment naïve. The primary outcome was the mean number of packed red blood cells (PRBC) transfused before treatment versus after CryoBalloon ablation (CBA). Outcomes were compared to a published study of 33 GAVE patients that were treated with APC or RFA from three tertiary referral centers in the United States.
With the CryoBalloon ablation system treatment, technical success was 100% with no device malfunctions and no serious adverse events. Efficacy at 6 months was 71% for transfusion independence. Endoscopic success at 6 months was 88% GAVE eradication with 2.4 mean CryoBalloon ablation treatments.
Literature review shows clinical success for CryoBalloon ablation treatment (62-83%) is comparable to other modalities (50-100%). Blood hemoglobin increased 4.6 g/dl with CryoBalloon ablation treatment versus 2.55 g/dl with APC/RFA. Monthly packed red blood cell transfusions decreased by 0.77 units with CryoBalloon ablation treatment versus 1.45 units with APC/RFA.
Overall, the study shows CryoBalloon ablation treatment is effective for long-term GAVE treatment, with 6-month outcomes similar or better than other modalities and supported by literature review.
- 2. 180° Catheter Treatment dose
Two studies evaluated the safety and efficacy of the 180° CryoBalloon ablation catheter for treatment of Barrett's esophagus (BE) related. The first study enrolled 25 patients at a 1.0mm/s dose in a dose-finding phase to establish the effective dose that resulted in BE regression at 8-10 weeks of ≥60 %. The study's results demonstrated a median BE regression as evaluated of 90 % (95 % CI 70 %-90%) after a single 1.0mm/s treatment. Technical success was 96% per protocol. One patient (1/23; 4 %), developed a stricture that was resolved with two dilations. No severe bleeding, perforation, or other SAEs occurred. Five patients had adverse events. Three patients had moderate strictures not requiring dilation, one patient had dark
{6}------------------------------------------------
colored stools (no melena), which resolved spontaneously, and one patient experienced heavy chest pain on the day of treatment and additional analgesics were prescribed.
The second study was a multicenter study treating 25 patients with short segment BE using a lower 180° CryoBalloon Ablation catheter dose of 1.2mm/s. Average BE surface regression was 95% at a 3-month follow-up visit. One serious adverse event (post-procedure bleeding from a patient with double-dosed anticoagulation therapy due to a pre-existing cardiac condition.) occurred which resolved. No strictures requiring dilation occurred. Two mild adverse events occurred post procedure in which no treatments were required.
In summary, both studies demonstrate the 180° CryoBalloon ablation system catheter is safe and effective for BE treatment using doses of 1.0mm/s and 1.2mm/s. Excellent technical success and BE regression were achieved with minimal adverse events. The data supports the lower dose labeling for short segment BE treatment.
### IX. CONCLUSIONS
C2 CryoBalloon Ablation System meets all substantial equivalence criteria for use as a cryosurgical tool in the field of general surgery, specifically for endoscopic applications, to include ablation of Barrett's Esophagus with dysplasia and treatment and management of Gastric Antral Vascular Ectasia(GAVE) based on comparable use, technology, data, and minor differences not affecting safety/effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.